← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K201852

# Surgical Mask-Model Number: Ear Loop, Flat-Pleated (K201852)

_Huizhou Foryou Medical Devices Co., Ltd. · FXX · Mar 2, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K201852

## Device Facts

- **Applicant:** Huizhou Foryou Medical Devices Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Mar 2, 2021
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

## Device Story

Surgical mask; flat-pleated design; three-layer construction (spunbonded polypropylene outer/inner layers; melt-blown polypropylene filter middle layer). Features elastic ear loops and malleable polypropylene/metal wire nosepiece for fit. Intended for single-use, non-sterile, disposable application in clinical settings. Protects patient and personnel from transfer of microorganisms, body fluids, and particulates. Reduces exposure to blood and body fluids during infection control procedures.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-19, ASTM F1862/F1862M-17 (fluid resistance), ASTM F2299/F2299M-03 (particulate filtration efficiency 99.79%), ASTM F2101-19 (bacterial filtration efficiency 99.70%), MIL-M-36954C (differential pressure 3.813 mm H2O/cm²), and 16 CFR 1610 (flammability Class 1). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

Three-layer flat-pleated mask; spunbonded polypropylene outer/inner layers; melt-blown polypropylene filter. Malleable polypropylene/metal wire nosepiece; spandex ear loops. Non-sterile; single-use. Meets ASTM F2100-19 Level 2 performance requirements.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K182515](/device/K182515.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

March 2, 2021

Huizhou Foryou Medical Devices Co., Ltd Guosheng Tan Development Engineer No. 1 Shangxia North Road, Dongjiang Hi-tech Industry Park Huizhou, Guangdong 516005 China

Re: K201852

Trade/Device Name: Surgical Mask-Model Number: Ear Loop, Flat-Pleated Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 27, 2021 Received: February 1, 2021

Dear Guosheng Tan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: 0MB No. 0910-0120 Expiration Date: 0613012020 See PRA Statement below.

510(k) Number (if known) K201852

Device Name Surgical Mask

#### Indications for Use (Describe)

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions , search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."

FORM FDA 3881 (7117)

Page 1 of 1

PSC Publishing Scrdccs (301)443-67-IO EF

{3}------------------------------------------------

# 510(k) Summary

This 510(k) Summary information is being submitted in accordance with the requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K201852

# 1. Submitter:

Huizhou Foryou Medical Devices Co., Ltd.

| Address:        | No.1 Shangxia North Road, Dongjiang Hi-tech Industry<br>Park, Huizhou, Guangdong, China |
|-----------------|-----------------------------------------------------------------------------------------|
| Phone:          | +86-0752-5302185                                                                        |
| Fax:            | +86-0752-5302020                                                                        |
| Contact Person: | Guosheng Tan                                                                            |
| Date Prepared:  | March 2, 2021                                                                           |

2. Subject Device:

| Trade Name:          | Surgical Mask    |
|----------------------|------------------|
| Common Name:         | Surgical Mask    |
| Classification Name: | Mask, Surgical   |
| Review Panel:        | General Hospital |
| Product Code:        | FXX              |
| Regulation Number:   | 21 CFR §878.4040 |
| Device Class:        | Class II         |

- 3. Predicate Device:
K182515

Surgical Face Mask

Wuhan Dymex Healthcare Co., Ltd

Page 1 of 4

{4}------------------------------------------------

### 4. Device Description:

The Surgical Mask is a single use, Flat Pleated mask with ear loops and nose piece. The Surgical Mask is consisting of three layers, the inner and outer layers are made of spunbonded polypropylene, and the middle layer is made of melt blown polypropylene filter. The Surgical Mask uses different colors to distinguish the inner and outer layers, the inner layer is white and the outer layer is blue. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of polypropylene and metal wire. The Surgical Mask is sold non-sterile and is intended to be single use, disposable devices.

# 5. Indications for Use:

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

# 6. Technological Characteristics Comparison

Table 5.1 Comparison of technological characteristics between subject device

| Item         | Subject Device<br>(K201852)                                                                                                                        | Predicate Device<br>(K182515)                                                                                                                             | Comparison |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended Use | The Surgical Mask is<br>intended to be worn to<br>protect both the patient<br>and healthcare personnel<br>from transfer of<br>microorganisms, body | The Surgical Face Masks<br>are intended to be worn to<br>protect both the patient<br>and healthcare personnel<br>from transfer of<br>microorganisms, body | Same       |

### and predicate device

{5}------------------------------------------------

|  |                  | fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices<br>to reduce the potential<br>exposure to blood and body<br>fluids. This is a single use,<br>disposable device(s),<br>provided non-sterile. | fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices<br>to reduce the potential<br>exposure to blood and<br>body fluids. This is a single<br>use, disposable device(s),<br>provided non-sterile. |                                 |      |
|--|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|------|
|  | Model            | Ear Loops, Flat Pleated, 3 layers                                                                                                                                                                                                                          | Ear Loops, Flat Pleated, 3 layers                                                                                                                                                                                                                          | Same                            |      |
|  | Material         | Outer layer                                                                                                                                                                                                                                                | Spun-bond polypropylene                                                                                                                                                                                                                                    | Spun-bond polypropylene         | Same |
|  |                  | Middle layer                                                                                                                                                                                                                                               | Melt blown polypropylene filter                                                                                                                                                                                                                            | Melt blown polypropylene filter | Same |
|  |                  | Inner layer                                                                                                                                                                                                                                                | Spun-bond polypropylene                                                                                                                                                                                                                                    | Spun-bond polypropylene         | Same |
|  | Nosepiece        | Malleable polypropylene<br>and metal wire                                                                                                                                                                                                                  | Malleable polyethylene<br>wire                                                                                                                                                                                                                             | Similar                         |      |
|  | Ear loops        | Spandex                                                                                                                                                                                                                                                    | Spandex                                                                                                                                                                                                                                                    | Same                            |      |
|  | Color            | Blue/White                                                                                                                                                                                                                                                 | Yellow                                                                                                                                                                                                                                                     | Different*                      |      |
|  | Dimension        | Length                                                                                                                                                                                                                                                     | $17.5cm \pm 1cm$                                                                                                                                                                                                                                           | $17.5cm \pm 0.2cm$              | Same |
|  |                  | Width                                                                                                                                                                                                                                                      | $9.5cm \pm 1cm$                                                                                                                                                                                                                                            | $9.5cm \pm 0.2cm$               | Same |
|  | Sterility        |                                                                                                                                                                                                                                                            | Non-Sterile                                                                                                                                                                                                                                                | Non-Sterile                     | Same |
|  | Use              |                                                                                                                                                                                                                                                            | Single Use, Disposable                                                                                                                                                                                                                                     | Single Use, Disposable          | Same |
|  | ASTM F2100 Level |                                                                                                                                                                                                                                                            | Level 2                                                                                                                                                                                                                                                    | Level 2                         | Same |

*The difference in the color does not raise additional questions for safety and effectiveness. Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all color additives.

# 7. Non-Clinical Performance Testing

Non-clinical tests were conducted to verify that the subject device met all design specifications as the predicate device. The test results demonstrate that the subject device conforms to the recognized standards ASTM F2100-19, ISO 10993-5 and ISO 10993-10 in addition to the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks-Premarket Notification [510(k)] Submission issued on March 5, 2004. Detailed test results are shown in the table 5.2.

{6}------------------------------------------------

| Performance<br>Characteristics          | Test Method          | Acceptance Criteria                                                    | Test Result                       |
|-----------------------------------------|----------------------|------------------------------------------------------------------------|-----------------------------------|
| Fluid Resistance<br>Performance         | ASTM F1862/F1862M-17 | 29 out of 32 passes at<br>120mmHg                                      | 32 out of 32 passes at<br>120mmHg |
| Particulate<br>Filtration<br>Efficiency | ASTM F2299/F2299M-03 | ≥98%                                                                   | 99.79%                            |
| Bacterial Filtration<br>Efficiency      | ASTM F2101-19        | ≥98%                                                                   | 99.70%                            |
| Differential<br>Pressure (ΔΡ)           | MIL-M-36954C         | <6.0mmH2O/cm²                                                          | 3.813 mm H2O/cm²                  |
| Flammability                            | 16 CFR 1610          | Class 1                                                                | Class 1                           |
| Cytotoxicity                            | ISO 10993-5:2009     | Under the conditions of<br>the study, the device is<br>non-cytotoxic.  | Non-cytotoxic                     |
| Irritation                              | ISO 10993-10:2010    | Under the conditions of<br>the study, the device is<br>non-irritating. | Non-irritating                    |
| Sensitization                           | ISO 10993-10:2010    | Under the conditions of<br>the study, the device is<br>non-sensitizing | Non-sensitizing                   |

Table 5.2 Summary of Non-clinical Performance Testing

# 8. Clinical Performance Test

No clinical study is included in this submission.

# 9. Conclusion

The conclusion drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated K182515, Surgical Face Mask, Wuhan Dymex Healthcare Co., Ltd.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K201852](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K201852)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
