← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K180874

# 3M High Performance Surgical Mask, 3M High Performance Surgical Mask with Face Shield (K180874)

_3M Healthcare · FXX · Jul 31, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K180874

## Device Facts

- **Applicant:** 3M Healthcare
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Jul 31, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The 3M™ High Performance Surgical Mask is intended to be worn to protect both patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face mask(s) are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device(s) and provided non-sterile.

## Device Story

3M High Performance Surgical Mask (1838R) and version with face shield (1838FSG) are duckbill-style surgical masks. Device consists of three layers: polypropylene spunbond inner/outer cover webs and polypropylene melt blown filter web. Features include tie strings for securement and malleable polyethylene-coated steel nosepiece for fit. Used in clinical settings by healthcare personnel to provide barrier protection against blood, body fluids, and particulates. Device is single-use, disposable, and non-sterile. Provides physical barrier to reduce cross-contamination risk between patient and clinician.

## Clinical Evidence

Bench testing only. Performance validated against ASTM F2100 Level 2 requirements, including ASTM F1862 (synthetic blood penetration), ASTM F2299 (particulate filtration), ASTM F2101 (bacterial filtration), MIL-M36954C (differential pressure), and 16 CFR 1610 (flammability). Biocompatibility testing performed per ISO 10993-1, ISO 10993-5, and ISO 10993-10 confirmed device is non-cytotoxic, non-irritating, and non-sensitizing.

## Technological Characteristics

Duckbill design; 3-layer construction (polypropylene spunbond cover webs, polypropylene melt blown filter). Polyethylene-coated steel nosepiece. Meets ASTM F2100 Level 2. Biocompatible per ISO 10993-1/5/10. Non-sterile, single-use. No electronic components or software.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Fluidshield* Surgical Mask with Expanded Chamber ([K143287](/device/K143287.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, which features a stylized caduceus. To the right is the logo for the U.S. Food & Drug Administration (FDA), with the FDA acronym in a blue square and the full name "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 11, 2019

3M Health Care Kristin Totushek Advanced Regulatory Affairs Specialist 3M Health Care 2510 Conway Ave. Building 275-5W-06 St. Paul, MN 55144

Re: K180874

Trade/Device Name: 3M High Performance Surgical Mask, 3M High Performance Surgical Mask with Face Shield

Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: June 14, 2018 Received: June 20, 2018

Dear Kristin Totushek:

This letter corrects our substantially equivalent letter of July 31, 2018.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## Clarence W. Murray Iii III -S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K180874

Device Name

3MTM High Performance Surgical Mask

Indications for Use (Describe)

The 3M™ High Performance Surgical Mask is intended to be worn to protect both patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face mask(s) are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device(s) and provided non-sterile.

3MTM High Performance Surgical Mask - 1838R 3M™ High Performance Surgical Mask with Face Shield - 1838FSG

| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 3M™ High Performance Surgical Mask
K180874

| Sponsor Information:   | 3M Health Care<br>2510 Conway Ave<br>3M Center, Building 275-5W-06<br>St. Paul, MN 55144                                                                                                                                                                                                                                                                                                                     |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:        | Kristin Totushek, RAC<br>Title: Advanced Regulatory Affairs Specialist<br>(651) 736-6117<br>ktotushek@mmm.com                                                                                                                                                                                                                                                                                                |
| Date of Summary:       | July 30, 2018                                                                                                                                                                                                                                                                                                                                                                                                |
| Proprietary Name:      | 3MTM High Performance Surgical Mask                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name:           | Surgical Mask                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification Name:   | Surgical Apparel                                                                                                                                                                                                                                                                                                                                                                                             |
| Review Panel:          | General and Plastic Surgery                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code:          | FXX                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Classification: | Class II per (21 CFR §878.4040)                                                                                                                                                                                                                                                                                                                                                                              |
| Predicate Device:      | Fluidshield* Surgical Mask with Expanded Chamber<br>(K143287)                                                                                                                                                                                                                                                                                                                                                |
| Intended Use:          | The 3M™ High Performance Surgical Mask is intended to be<br>worn to protect both patient and healthcare personnel from<br>transfer of microorganisms, body fluids, and particulate material.<br>These face mask(s) are intended for use in infection control<br>practices to reduce potential exposure to blood and body fluids.<br>This is a single use, disposable device(s) and provided non-<br>sterile. |

{4}------------------------------------------------

#### Model Numbers: 3MTM High Performance Surgical Mask 1838R 1838FSG 3MTM High Performance Surgical Mask with Face Shield

### Device Description:

The 3M™ High Performance Surgical Mask has a duckbill design consisting of three layers that are comprised of an (polypropylene spunbond) inner and outer cover web and (polypropylene melt blown) filter web. The mask has ties to secure the mask over the users' mouth and face and includes a malleable nosepiece to provide a firm fit over the nose. They may be produced with or without a face shield. This is a single use, disposable device, provided non-sterile.

This device is not made with Natural Rubber Latex.

## Technological Characteristics:

The 3M™ High Performance Surgical Mask is substantially equivalent to the current Fluidshield* Surgical Mask with Expanded Chamber (K143287) as these products conform with ASTM F2100 standard and meet Level 2 performance specifications. Both the submission and predicate products address, "Guidance for Industry and FDA Staff: Surgical Masks -Premarket Notification 510(K) submissions. Issued March 5, 2004".

### Comparison to Predicate Device:

The 3M™ High Performance Surgical Mask, the subject of this submission, is substantially equivalent to the Fluidshield* Surgical Mask with Expanded Chamber cleared under K143287. The submission and predicate devices claim Level 2 in accordance with ASTM F2100 product standard.

{5}------------------------------------------------

| Feature                       | K180874<br>3M™ High Performance<br>Surgical Mask<br>1838R                                                                                                                                                                                                                                                                                                                                                                         | K180874<br>3M™ High Performance<br>Surgical Mask with Face<br>Shield<br>1838FSG                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>Halyard Heath, Inc.<br>Fluidshield* Surgical<br>Mask with Expanded<br>Chamber<br>K143287                                                                                                                                                                                                                                                                                                                             |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use               | 3M™ High Performance<br>Surgical Mask is intended<br>to be worn to protect the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices<br>to reduce the potential<br>exposure to blood and<br>body fluids. This is a<br>single use, disposable<br>device(s), provided non-<br>sterile. | 3M™ High Performance<br>Surgical Mask is intended<br>to be worn to protect the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices<br>to reduce the potential<br>exposure to blood and<br>body fluids. This is a<br>single use, disposable<br>device(s), provided non-<br>sterile. | The Expanded Chamber<br>Surgical Face Mask is<br>intended to be worn to<br>protect the patient and<br>healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face<br>masks are intended for<br>use in infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single use,<br>disposable device(s),<br>provided non-sterile. |
| Materials                     |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Outer Cover<br>Web            | Polypropylene Spunbond,<br>(w Print) – Upper Top Half<br>Polypropylene Spunbond,<br>White – Lower Bottom<br>Half                                                                                                                                                                                                                                                                                                                  | Polypropylene Spunbond,<br>(w Print) – Upper Top Half<br>Polypropylene Spunbond,<br>White – Lower Bottom<br>Half                                                                                                                                                                                                                                                                                                                  | Polypropylene Spunbond<br>(w Print) – Upper Top<br>Half<br>Polypropylene<br>Spunbond, White –<br>Lower Bottom Half                                                                                                                                                                                                                                                                                                                       |
| Middle<br>Web/Filter          | Polypropylene Melt Blown                                                                                                                                                                                                                                                                                                                                                                                                          | Polypropylene Melt Blown                                                                                                                                                                                                                                                                                                                                                                                                          | Polypropylene Spunbond<br>and Polypropylene Melt<br>Blown                                                                                                                                                                                                                                                                                                                                                                                |
| Inner Cover<br>Web            | Polypropylene Spunbond                                                                                                                                                                                                                                                                                                                                                                                                            | Polypropylene Spunbond                                                                                                                                                                                                                                                                                                                                                                                                            | Polypropylene/Polyester                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Tie Strings                   | Polypropylene Spunbond<br>or<br>PolyethyleneTerephthalate                                                                                                                                                                                                                                                                                                                                                                         | Polypropylene Spunbond<br>or<br>PolyethyleneTerephthalate                                                                                                                                                                                                                                                                                                                                                                         | Polyester Spunbond or<br>Polypropylene Spunbond                                                                                                                                                                                                                                                                                                                                                                                          |
| Top and<br>Bottom<br>Binding  | Polypropylene Spunbond                                                                                                                                                                                                                                                                                                                                                                                                            | Polypropylene Spunbond                                                                                                                                                                                                                                                                                                                                                                                                            | Polypropylene Spunbond                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Nose Piece                    | Polyethylene coated steel<br>wire                                                                                                                                                                                                                                                                                                                                                                                                 | Polyethylene coated steel<br>wire                                                                                                                                                                                                                                                                                                                                                                                                 | Wire (Material<br>Unknown)                                                                                                                                                                                                                                                                                                                                                                                                               |
| Feature                       | K180874<br>3M™ High Performance<br>Surgical Mask<br>1838R                                                                                                                                                                                                                                                                                                                                                                         | K180874<br>3M™ High Performance<br>Surgical Mask with Face<br>Shield<br>1838FSG                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>Halyard Heath, Inc.<br>Fluidshield* Surgical<br>Mask with Expanded<br>Chamber<br>K143287                                                                                                                                                                                                                                                                                                                             |
| Design Features               |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Color                         | Blue - Outer Cover Web<br>(Upper) White - Outer<br>Cover Web (Lower)                                                                                                                                                                                                                                                                                                                                                              | Blue - Outer Cover Web<br>(Upper) White - Outer<br>Cover Web (Lower)                                                                                                                                                                                                                                                                                                                                                              | Variety                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Face Shield                   | No                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Mask Design                   |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Style                         | Duckbill                                                                                                                                                                                                                                                                                                                                                                                                                          | Duckbill                                                                                                                                                                                                                                                                                                                                                                                                                          | Duckbill                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Single Use                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterility                     |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Sterile                       | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                       | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                       | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Specifications and Dimensions |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Length                        | $187 mm \pm 4 mm$                                                                                                                                                                                                                                                                                                                                                                                                                 | $187 mm \pm 4 mm$                                                                                                                                                                                                                                                                                                                                                                                                                 | $8.3" \pm 0.4"$                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Width                         | $107 mm \pm 6 mm$                                                                                                                                                                                                                                                                                                                                                                                                                 | $107 mm \pm 6 mm$                                                                                                                                                                                                                                                                                                                                                                                                                 | $7.5" \pm 0.11"$                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Technological Characteristics |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                               | Product Performance Specifications Per ASTM F2100 - Meets ASTM Level 2                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Fluid                         | ASTM F1862                                                                                                                                                                                                                                                                                                                                                                                                                        | ASTM F1862                                                                                                                                                                                                                                                                                                                                                                                                                        | ASTM F1862                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Particulate                   | ASTM F2299                                                                                                                                                                                                                                                                                                                                                                                                                        | ASTM F2299                                                                                                                                                                                                                                                                                                                                                                                                                        | ASTM F2299                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Bacterial                     | ASTM F2101                                                                                                                                                                                                                                                                                                                                                                                                                        | ASTM F2101                                                                                                                                                                                                                                                                                                                                                                                                                        | ASTM F2101                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Differential                  | MIL-M36954C                                                                                                                                                                                                                                                                                                                                                                                                                       | MIL-M36954C                                                                                                                                                                                                                                                                                                                                                                                                                       | MIL-M36954C                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Flammability                  | 16 CFR 1610                                                                                                                                                                                                                                                                                                                                                                                                                       | 16 CFR 1610                                                                                                                                                                                                                                                                                                                                                                                                                       | 16 CFR 1610                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Biocompatibility              |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Results                       | Biocompatible, Non-<br>cytotoxic, Non- irritating,<br>Non-Sensitizing                                                                                                                                                                                                                                                                                                                                                             | Biocompatible, Non-<br>cytotoxic, Non-irritating,<br>Non-Sensitizing                                                                                                                                                                                                                                                                                                                                                              | Biocompatible, Non-<br>cytotoxic, Non-<br>irritating, Non-<br>Sensitizing                                                                                                                                                                                                                                                                                                                                                                |

{6}------------------------------------------------

{7}------------------------------------------------

| ASTM F2100          | Standard Specification for Performance of Materials Used in Medical Face Masks                                                                                     |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM F1862          | Standard Test Method for Resistance of Medical Face Masks to Penetration by<br>Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)         |
| ASTM F2299          | Standard Test Method for Determining the Initial Efficiency of Materials Used in<br>Medical Face Masks to Penetration by Particulates Using Latex Spheres          |
| ASTM F2101          | Standard Method for Evaluating the Bacterial Filtration Efficiency (BFE) of<br>Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus<br>aureus |
| MIL-M-<br>36954C    | MIL-M36954C - Military Specification, Surgical Mask, Disposable - Differential<br>Pressure (Delta P)                                                               |
| 16 CFR Part<br>1610 | Standard for the Flammability of Clothing Textiles                                                                                                                 |
| ISO10993-1          | Biological evaluation of medical devices - Part 1: Evaluation and testing within<br>a risk management process                                                      |
| ISO10993-5          | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity                                                                                 |
| ISO10993-10         | Biological evaluation of medical devices - Part 10: Tests for irritation and skin<br>sensitization                                                                 |

The following standards have been met for the 3M™ High Performance Surgical Mask.

## Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the 3M™ High Performance Surgical Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device K143287.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K180874](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K180874)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
