← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K173062

# Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) (K173062)

_V&Q Manufacturing Corporation · FXX · Jan 25, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K173062

## Device Facts

- **Applicant:** V&Q Manufacturing Corporation
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Jan 25, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.

## Device Story

Single-use, multi-layer surgical face mask; protects against microorganisms, blood, body fluids, and particulates. Construction: spunbond polypropylene inner/outer layers; meltblown polypropylene filter layer; pliable aluminum nosepiece; urethane elastic earloops or spunbond polypropylene ties. Used in operating rooms and clinical settings by healthcare personnel. Provides physical barrier; reduces transfer of contaminants between wearer and environment. Benefits: infection control; protection against fluid/particulate exposure.

## Clinical Evidence

Bench testing only. Performance testing included: ASTM F2299 (Particle Filtration Efficiency: 99.74% at 0.1μm), ASTM F1862 (Fluid resistance: pass at 120mm Hg), ASTM F2101-14 (Bacterial Filtration Efficiency: 99.4%), 16 CFR Part 1610 (Flammability Class 1), and MIL-M-36954C. Biocompatibility testing per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization) confirmed the device is non-cytotoxic, non-sensitizing, and non-irritating.

## Technological Characteristics

Multi-layer mask; spunbond polypropylene (inner/outer), meltblown polypropylene (filter). Nosepiece: aluminum strip with PP covering. Headband: urethane elastic fiber or spunbond polypropylene. Dimensions: 175mm x 95mm. Flat-pleated design. Not made with natural rubber latex. Biocompatibility per ISO 10993-5 and ISO 10993-10. Performance standards: ASTM F2299, ASTM F1862, ASTM F2101-14, 16 CFR 1610, MIL-M-36954C.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Face Mask, Surgical Mask, Surgical Face Mask ([K122717](/device/K122717.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 25, 2018

V&Q Manufacturing Corporation % Tracv Che Registered engineer Feiying Drug & Medical Consulting Technical Service Group B-3F 3005, Bldg.1, Southward Ruifeng Business Center, No 22 Guimiao Road Shenzhen, 518000 CN

Re: K173062

Trade/Device Name: Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 3. 2018 Received: January 5, 2018

Dear Tracy Che:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173062

Device Name

Non Woven Face Mask (Model: VQN0185W(earloop) and VQN0185B(ties))

Indications for Use (Describe)

Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) is intended for single use by operating room personnel and other general healthcare workers and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510 (k) Summary

## K173062

### (1) Applicant information

| 510 (k) owner's name: | V&Q Manufacturing Corporation                          |
|-----------------------|--------------------------------------------------------|
| Address:              | #B1614 optical valley time square, Wuhan, Hubei, CHINA |

| Contact person:           | Jacky Jin            |
|---------------------------|----------------------|
| Phone number:             | 86-27-87680925       |
| Fax number:               | 86-27-87680926       |
| Email:                    | jackyj@chinavqmc.com |
| Date of summary prepared: | 2018.01.24           |

# (2) Proprietary name of the device

| Trade name:        | Non Woven Face mask (Models: VQN0185W (earloop) and<br>VQN0185B (ties)) |
|--------------------|-------------------------------------------------------------------------|
| Regulation name:   | Surgical apparel                                                        |
| Regulation number: | 21 CFR 878.4040                                                         |
| Product code       | FXX                                                                     |
| Review panel:      | General & Plastic Surgery                                               |
| Regulation class:  | Class II                                                                |

| Sponsor           | Tiger Medical Products Ltd.                  |  |
|-------------------|----------------------------------------------|--|
| Device Name       | Face Mask, Surgical Mask, Surgical Face Mask |  |
| 510(k) Number     | K122717                                      |  |
| Product Code      | FXX                                          |  |
| Regulation Number | 21 CFR 878.4040                              |  |
| Regulation Class  | II                                           |  |

#### (3) Predicate device

# (4) Description/ Design of device

Non Woven Face Mask is a single use multi-layer mask with outer layer and inner layer (spunbond polypropylene) that sandwich a meltblown polypropylene filter material. There are two options for the surgical mask to be secured on users via earloops or ties. Earloops are of urethane elastic fiber and not made with natural rubber latex; and ties are of spunbond polypropylene and also not made with natural rubber latex. The nose piece is a pliable white aluminum strip, covered by polypropylene covering. All of the materials used in the construction of the surgical mask are being 

{4}------------------------------------------------

used in currently marketed devices. Non Woven Face Mask has two models which are VQN0185W and VQN0185B. They are basically the same, the only difference is VQN0185W adopts earloops and VQN0185B adopts ties to secure the mask on user.

## (5) Indications for use

Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.

| Component of Device<br>Requiring<br>Biocompatibility | Material of<br>Component                                                                                                                             | Body Contact Category<br>(ISO 10993-1) | Contact Duration<br>(ISO 10993-1) |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|-----------------------------------|
| Non Woven Face Mask                                  | Spunbond<br>polypropylene,<br>meltblown<br>polypropylene,<br>urethane<br>elastic fiber,<br>white<br>aluminum<br>strip, blue<br>color master<br>batch | Surface-contacting<br>device: skin     | < 24hours                         |

## (6) Materials

The body-contacting material used in the Non Woven Face Mask have all passed biocompatibility test. Details can be seen in "Biocompatibility Discussion".

# (7) Technological characteristics and substantial equivalence

| Item                      | Subject device                                                                | Predicate device                                | Remark    |
|---------------------------|-------------------------------------------------------------------------------|-------------------------------------------------|-----------|
| Trade name                | Non Woven Face Mask<br>(Models: VQN0185W<br>(earloop) and VQN0185B<br>(ties)) | Face Mask, Surgical Mask,<br>Surgical Face Mask | /         |
| 510 (k) number            | K173062                                                                       | K122717                                         | /         |
| Regulation<br>number      | 21 CFR 878.4040                                                               | 21 CFR 878.4040                                 | Identical |
| Regulation<br>description | Surgical apparel                                                              | Surgical apparel                                | Identical |

{5}------------------------------------------------

|           | Product code                      | FXX                                                                                                                                                                                                                                                                                                    | FXX                                                                                                                                                                                                                                                                                                                        | Identical         |
|-----------|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
|           | Class                             | II                                                                                                                                                                                                                                                                                                     | II                                                                                                                                                                                                                                                                                                                         | Identical         |
|           | Indications for use/ Intended use | Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials. | Surgical mask (with different trade names: Face Mask, Surgical Mask, Surgical Face Mask) is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and airborne particulates. | Similar           |
|           | Inner layer                       | Spun-bond polypropylene                                                                                                                                                                                                                                                                                | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                    | Similar           |
|           | Middle layer                      | Meltblown polypropylene                                                                                                                                                                                                                                                                                | Meltblown polypropylene                                                                                                                                                                                                                                                                                                    |                   |
|           | Outer layer                       | Spun-bond polypropylene                                                                                                                                                                                                                                                                                | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                    |                   |
| Materials | Nosepiece                         | White aluminum strip covered by PP covering                                                                                                                                                                                                                                                            | White aluminum strip with PP covering                                                                                                                                                                                                                                                                                      |                   |
|           | Headband                          | Urethane elastic fiber or spun-bond polypropylene                                                                                                                                                                                                                                                      | Urethane elastic fiber or spun-bond polypropylene                                                                                                                                                                                                                                                                          |                   |
|           | Mask style                        | Flat pleated                                                                                                                                                                                                                                                                                           | Flat pleated                                                                                                                                                                                                                                                                                                               | Identical         |
|           | Design feature                    | Earloop or tie-on                                                                                                                                                                                                                                                                                      | Earloop or tie-on                                                                                                                                                                                                                                                                                                          | Identical         |
|           | Dimensions                        | 175mm×95mm                                                                                                                                                                                                                                                                                             | Approx 170mm×90 mm                                                                                                                                                                                                                                                                                                         | Similar           |
| Latex     |                                   | Not made with natural rubber latex                                                                                                                                                                                                                                                                     | Latex Free                                                                                                                                                                                                                                                                                                                 | Identical         |
|           | Performance test result           |                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                            |                   |
|           | Fluid resistance                  | Pass at 120mm Hg                                                                                                                                                                                                                                                                                       | Fluid resistant                                                                                                                                                                                                                                                                                                            | Similar           |
|           | Particle Filtration Efficiency    | Average 99.74% at 0.1μ m                                                                                                                                                                                                                                                                               | Average 99.54% at 0.1 micron                                                                                                                                                                                                                                                                                               | Similar           |
|           | Bacterial Filtration Efficiency   | Average 99.4%                                                                                                                                                                                                                                                                                          | >99.9%                                                                                                                                                                                                                                                                                                                     | Similar           |
|           | Flammability Class                | 1                                                                                                                                                                                                                                                                                                      | 1                                                                                                                                                                                                                                                                                                                          | Identical         |
|           | Delta – P                         | Average 2.7 mmH2O/cm2                                                                                                                                                                                                                                                                                  | Average 3.38 mmH2O/cm2                                                                                                                                                                                                                                                                                                     | Difference Note 1 |
|           | Biocompatibility                  | ISO10993-5 and ISO10993-10:                                                                                                                                                                                                                                                                            | ISO10993-5 and ISO10993-10:                                                                                                                                                                                                                                                                                                | Identical         |

{6}------------------------------------------------

| Under the conditions of   | Under the conditions of the |  |
|---------------------------|-----------------------------|--|
| the studies employed, the | studies employed, the       |  |
| device is non-cytotoxic,  | device is non-cytotoxic,    |  |
| non-sensitizing, and      | non-sensitizing, and        |  |
| non-irritating.           | non-irritating.             |  |

- A Note 1:
The Delta-P of the subject device is smaller than that of the predicate device which means user may feel cooler wearing the subject device, since a lower Delta-P translates to increased breathability.

### (8) Non-clinical studies and tests performed

The performance tests of Non Woven Face Mask were conducted.

- ASTM F2299 Standard Test Method for Determining the Initial Efficiency of Materials Used A in Medical Face Masks to Penetration by Particulates Using Latex Spheres
- A ASTM F1862 Standard test method for resistance of medical face masks to penetration by synthetic blood (Horizontal projection of fixed volume at a known velocity)
- A ASTM F 2101-14 Standard Test Method For Evaluating The Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus.
- > MIL-M-36954C Military Specification - Mask, Surgical, Disposable
- 16 CFR Part 1610 STANDARD FOR THE FLAMMABILITY OF CLOTHING TEXTILES A

During use, the Non Woven Face Mask will directly contact with user's skin, so we have it tested to demonstrate conformance to the following standards.

- > ISO 10993-5, Biological Evaluation Of Medical Devices -- Part 5: Tests For InVitro Cytotoxicity
- A ISO 10993-10, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization

#### (9) Conclusion

Based on the non-clinical tests performed, the subject device is as safe, as effective, and performs as well as the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K173062](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K173062)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
