K960441 · Steritec Products Mfg Co, Inc. · JOJ · May 8, 1997 · General Hospital
Device Facts
Record ID
K960441
Device Name
STERITEC INTEGRAPH MODEL CI-101
Applicant
Steritec Products Mfg Co, Inc.
Product Code
JOJ · General Hospital
Decision Date
May 8, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The Steritec Integraph is designed to provide an integrated response to steam sterilization at 132 C for sterilizers operating for 3 ½ minutes or longer.
Device Story
Chemical indicator/integrator for steam sterilization monitoring; used in sterilizer chambers. Device consists of indicator strip that changes color from purple to green upon exposure to specified steam sterilization conditions (132 C for ≥3.5 minutes). Provides integrated response to sterilization parameters; serves as visual confirmation of cycle completion. Used by healthcare personnel to verify sterilization efficacy; provides margin of safety by ensuring no 'PASS' result occurs if biological spore survivors are present. Validated against biological indicators.
Clinical Evidence
Bench testing only. Performance validated against biological spore strips across multiple sterilization temperatures and times. Stability testing conducted under high temperature, high humidity, and long-term storage (up to 2.5 years). Results showed no false 'PASS' indications; device consistently correlated with biological indicator results.
Technological Characteristics
Chemical sterilization indicator/integrator. Color-changing chemistry (purple to green) responsive to steam sterilization parameters (132 C, ≥3.5 minutes). Form factor: indicator strip. Stable under high heat and humidity.
Indications for Use
Indicated for use as a chemical sterilization indicator/integrator to monitor steam sterilization cycles at 132 C for 3.5 minutes or longer.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
K080136 — INSTANT READOUT CHEMICAL INTEGRATOR CHALLENGE PACK, MODEL LT- 105 · Steritec Products, Inc. · Aug 15, 2008
Submission Summary (Full Text)
{0}
K960441
SteriTec Products Mfg. Co., Inc.
Steritec Integraph Integrators
FDA 510(k) Application-K960441
Original Submission-January 24, 1996
Current Date-May 2, 1997
510(k) SUMMARY
MAY - 8 1997
# 510(k) PREMARKET NOTIFICATION SUMMARY
510(k) Premarket Notification Summary
## SUBMITTER:
- STERITEC PRODUCTS MFG. CO., INC.
680 Atchison Way - Suite 600
Castle Rock, CO 80104
(303) 660-4201
- Establishment Registration Number: 2028456
- Date: May 2, 1997
- **TOM ROLL**
Printed name of person required to submit 510(k)
[Handwritten signature]
- Signature of person required to submit 510(k)
- **PRESIDENT**
Title of person submitting 510(k)
Proprietary Name: Integraph
Common/Usual Name: Steam Sterilization Indicator/Integrator
Classification Name: Sterilization Indicator
Classification:
FDA has classified Physical/Chemical Sterilization Indicators in Class II under Classification Number 80JOJ, Regulation Number 880.2800.
{1}
FROM : SteriTec Prod.
PHONE NO. : 303 660 4213
P03
SteriTec Products Mfg. Co., Inc.
Steritec Integraph Integrators
FDA 510(k) Application-K960441
Original Submission-January 24, 1996
Current Date-May 2, 1997
510(k) SUMMARY
## Identification of Predicate devices
The Steritec Integraph CI 101 is a chemical indicator/integrator equivalent in construction and operation to the ATI Steam Clox (predicate device). Since the Steritec Integraph is equivalent in construction to the predicate device, it will perform equivalently.
## Description of 510(k) submission device:
### Intended use:
The Steritec Integraph is designed to provide an integrated response to steam sterilization at 132 C for sterilizers operating for 3 ½ minutes or longer.
### Comparison to Predicate device:
Compared to the predicate device our performance data shows the Integraph Integrator to be substantially equivalent.
## Performance Testing:
To demonstrate the performance of the SteriTec Integraph the product was tested:
1. At multiple sterilization temperatures and times against biological indicators.
2. For stability after storage in high temperature conditions.
3. For stability after storage in high humidity conditions.
4. For stability after storage for two and one half years.
5. For stability after being sterilized and then stored for two years.
## No false "PASS" Indications
In all cases they indicated properly whether the sterilization conditions had been met or not in the sterilizer chamber as verified by biological spore strips. There were NO instances in which the SteriTec strips indicated that sterilization conditions were met when they were not (NO FALSE PASS RESULTS).
## Margin of Safety
Test results show the SteriTec strips provide a "margin of safety". During these tests, No SteriTec strip tested provided a "PASS" test result in tests when there were any biological spore survivors.
{2}
SteriTec Products Mfg. Co., Inc.
Steritec Integraph Integrators
FDA 510(k) Application-K960441
Original Submission-January 24, 1996
Current Date-May 2, 1997
510(k) SUMMARY
were any biological spore survivors.
## Stability Testing:
Independent laboratory testing showed the SteriTec Integraph is stable after being subjected to extremes of heat and high humidity. In addition, under room temperature storage of both exposed and unexposed strips for 2 years the strips also performed satisfactorily. All strips maintained their purple color after storage testing. After exposure to sterilization conditions, the indicators properly changed color from purple to green.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.