STERIS STEAM PI

K112256 · STERIS Corporation · JOJ · Jan 11, 2012 · General Hospital

Device Facts

Record IDK112256
Device NameSTERIS STEAM PI
ApplicantSTERIS Corporation
Product CodeJOJ · General Hospital
Decision DateJan 11, 2012
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The STERIS STEAM n is process indicator that undergoes a visual color change from a pink to copper color when exposed to the following steam sterilization cycle parameters: 121℃ (250°F), 30 min, steam sterilization gravity cycle . 132°C (270°F), 15 min, steam sterilization gravity cycle 135℃ (275°F), 10 min, steam sterilization gravity cycle ● 132°C (270°F), 4 min, steam sterilization dynamic air removal cycle . 135°C (275°F), 3 min, steam sterilization dynamic air removal cycle .

Device Story

STERIS STEAM π is a single-use chemical process indicator; consists of indicator ink printed on polyethylene terephthalate substrate. Device functions by undergoing visual color change from pink to copper when exposed to specific steam sterilization parameters (121°C-135°C). Used in clinical or healthcare settings to confirm sterilization cycle exposure. Provides immediate visual feedback to healthcare personnel regarding processing status of sterilized items. Does not require complex operation or interpretation; serves as a simple pass/fail indicator for sterilization cycle verification.

Clinical Evidence

Bench testing only. Device verified against ANSI/AAMI ISO 11140-1:2005 requirements for Class 1 process indicators using a resistometer per ANSI/AAMI ISO 18472.

Technological Characteristics

Single-use chemical indicator; ink printed on polyethylene terephthalate substrate. Class 1 process indicator per ANSI/AAMI ISO 11140-1:2005. Sensing principle: chemical color change reaction triggered by steam sterilization temperature/time parameters.

Indications for Use

Indicated for use as a process indicator to verify exposure to steam sterilization cycles (gravity and dynamic air removal) at specified temperatures and durations. No specific patient population or contraindications defined.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 1 1 2012 Image /page/0/Picture/2 description: The image shows the word "STERIS" in a bold, sans-serif font. To the right of the word STERIS is a registered trademark symbol. Below the word STERIS is a graphic of several horizontal lines that are curved in the middle. # 510(k) Summary For STERIS STEAM π STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198 Contact: Bill Brodbeck Senior Manager, Regulatory Affairs Telephone: (440) 392-7690 Fax No: (440) 357-9198 Summary Date: January 4, 2012 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {1}------------------------------------------------ #### 1. Device Name | Indicators Models: | STERIS STEAM π | |----------------------|-----------------------------------------------------------------------------------------------| | Common Name: | Chemical Indicator | | Classification Name: | Physical/chemical sterilization process indicators<br>(21 CFR 880.2800 (b), Product Code JOJ) | #### 2. Predicate Device - Verify STEAM Value Indicators (K060103) . #### 3. Device Description The proposed STERIS STEAM n is composed of a single ink printed on polyethylene terephthalate. The indicator inks change from pink to copper color when exposed to temperature ranges of 121℃ to 135℃ (250ºF to 275ºF). #### 4. Intended Use The STERIS STEAM n is process indicator that undergoes a visual color change from a pink to copper color when exposed to the following steam sterilization cycle parameters: - 121℃ (250°F), 30 min, steam sterilization gravity cycle ● - . 132°C (270°F), 15 min, steam sterilization gravity cycle - 135℃ (275°F), 10 min, steam sterilization gravity cycle ● - 132°C (270°F), 4 min, steam sterilization dynamic air removal cycle . - 135°C (275°F), 3 min, steam sterilization dynamic air removal cycle . #### Description of Safety and Substantial Equivalence ട. The proposed and predicate devices are both single use indicators for use in steam sterilization cycles. The differences between the proposed STERIS STEAM n Indicators and the predicate device are limited to differences in general design of the ink and material. However, these differences do not affect the intended use, performance characteristics, nor do the differences raise any new issues of safety and efficacy. {2}------------------------------------------------ #### 6. Performance Testing Performance testing was conducted to verify that the proposed STERIS STEAM π meets the requirements for process [Class 1] indicators as defined in ANSVAAMI ISO 11140-1:2005 using a resistometer to ANSVAAMI ISO 18472. The data provided by this testing demonstrates that the proposed STERIS STEAM π device is substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular emblem with the department's name encircling a stylized image of an eagle. The eagle is depicted with three lines forming its body and wings, and two lines representing its legs and feet. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Robert Sullivan Senior Director, FDA Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060 JAN 1 1 2012 Re: K112256 Trade/Device Name: STERIS STEAM π Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: January 4, 2012 Received: January 5, 2012 Dear Mr. Sullivan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2 - Mr. Sullivan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Anthony D. Nutter Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K112256 STERIS STEAM T Device Name: Indications for Use: The STERIS STEAM n is a process indicator that undergoes a visual color change from a pink to brown color when exposed to the following steam sterilization cycle parameters: - 121°C (250°F), 30 min, steam sterilization gravity cycle . - 132°C (270°F), 15 min, steam sterilization gravity cycle . - 135℃ (275℃F), 10 min, steam sterilization gravity cycle . - 132°C (270°F), 4 min, steam sterilization dynamic air removal cycle ● - 135℃ (275℃), 3 min, steam sterilization dynamic air removal cycle . Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ X (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Elishh P. Currie-Wells (Division Sign-Off) (Division Sign-on) Division of Anesthesiology, General Hospital Division of Control, Dental Devices 510(k) Number: K112256
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...