BASIS FLEXIBLE TIP DIGITAL THERMOMETER

K971045 · Polymedica Industries, Inc. · FLL · May 2, 1997 · General Hospital

Device Facts

Record IDK971045
Device NameBASIS FLEXIBLE TIP DIGITAL THERMOMETER
ApplicantPolymedica Industries, Inc.
Product CodeFLL · General Hospital
Decision DateMay 2, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Basis® Flexible Tip Digital Thermometer is a clinical thermometer intended for the determination of oral, rectal and axillary body temperature determination in humans.

Device Story

Digital thermometer with flexible tip; measures human body temperature via oral, rectal, or axillary sites. Device utilizes electronic sensing technology to determine temperature; provides digital readout for clinician or patient use. Intended for intermittent temperature monitoring; aids in clinical assessment of patient health status.

Clinical Evidence

Bench testing only; device conforms to ASTM E1112 Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature.

Technological Characteristics

Electronic thermometer; flexible tip design; conforms to ASTM E1112 standards. Digital display output.

Indications for Use

Indicated for oral, rectal, and axillary body temperature measurement in humans of all ages.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K971045 510(k) Basis® Flexible Tip Digital Thermometer PolyMedica Industries, Inc. MAY - 2 1997 510(k) Summary Proprietary Name: Basis® Flexible Tip Digital Thermometer Common Name: Digital Thermometer with Flexible Tip Classification: Class II Submitter Details: PolyMedica Industries, Inc. 581 Conference Place Golden, CO 80401 Tel: (303)271-0300 Fax: (303)271-0163 Contact: Ms. Susan Collins The Basis® Flexible Tip Digital Thermometer is a clinical thermometer intended for the determination of oral, rectal and axillary body temperature determination in humans. In terms of physical requirements and operating parameters, the thermometer conforms to ASTM E1112, "Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature". Basis® Flexible Tip Digital Thermometer is substantially equivalent to PolyMedica Industries, Inc's Basis® Digital Thermometer and the Becton Dickinson Flexible™ Digital Thermometer. 53
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...