MODEL 2082 TEMP PLUS III THERMOMETER/MODEL 2087 TEMP PLUS III WALL MOUNT THERMOMETER
Device Facts
| Record ID | K955846 |
|---|---|
| Device Name | MODEL 2082 TEMP PLUS III THERMOMETER/MODEL 2087 TEMP PLUS III WALL MOUNT THERMOMETER |
| Applicant | Y |
| Product Code | FLL · General Hospital |
| Decision Date | Oct 1, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.2910 |
| Device Class | Class 2 |
Indications for Use
The intended use for the IVAC Models 2082/2087 Thermometers is to measure oral, rectal and axillary temperatures in clinical practice.
Device Story
Thermistor-based electronic thermometer; measures oral, rectal, axillary temperatures. Probe tip thermistor contacts tissue; microprocessor measures circuit resistance; converts resistance to temperature; displays result. Uses NASA-developed predictive algorithm to accelerate sensor response time; reduces display time to ~10 seconds. Used in clinical settings by healthcare professionals. Output provides rapid temperature readings to assist clinical decision-making and patient monitoring.
Clinical Evidence
Bench testing only; no clinical data provided. Equivalence based on technological comparison and performance validation of the predictive algorithm.
Technological Characteristics
Thermistor-based electronic thermometer; microprocessor-controlled; predictive algorithm for accelerated response. Dimensions/form factor consistent with previous models. No specific material standards cited.
Indications for Use
Indicated for clinical measurement of oral, rectal, and axillary body temperature in patients requiring temperature monitoring.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
- IVAC Models 2080/2085 Temp Plus II Thermometers
- Diatek SureTemp 4 Thermometer
Related Devices
- K042202 — TEMPTELLER - HIGH SPEED DIGITAL THERMOMETERS, MODELS DT-302, DT-312, DT-412, DT-502 · Oriental System Technology, Inc. · Sep 10, 2004
- K180131 — Vicks RapidRead Digital Thermometer · Kaz USA, Inc., A Helen of Troy Company · Aug 14, 2018
- K021851 — DIGITAL CLINICAL THERMOMETER, MODEL KD-193 · K-Jump Health Co., Ltd. · Jul 17, 2002
- K962497 — Q-TIPS, FAICHNEY, PRIVATE LABEL · Faichney Medical Co. · Oct 8, 1996
- K041646 — THERMOTEK Q5 · Saat , Ltd. · Jul 22, 2004