K112116 · Amperor-Electronic(Sz) Co., Ltd. · FLL · Oct 4, 2011 · General Hospital
Device Facts
Record ID
K112116
Device Name
DIGITAL CLINICAL THERMOMETER MODEL V901US
Applicant
Amperor-Electronic(Sz) Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Oct 4, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Digital Thermometer Model V901US is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm. This thermometer is intended for the adult and pediatric population. The Digital Thermometer Model V901US is a hand- held, non-sterile, reusable clinical thermometer intended for the determination of human temperature in either predictive or Instant mode ( 10-seconds in oral, rectal mode and under the arm mode-predicative temperature ), or standard mode (actual determination of temperature).
Device Story
Hand-held, reusable digital clinical thermometer; measures human body temperature via oral, rectal, or axillary sites. Operates in predictive (instant) mode or standard (actual) mode. User-operated by clinicians or patients (OTC). Device processes thermal sensor input to calculate temperature; provides visual display of results and audible alerts (10 beeps for measurement completion, fever alert). Features include auto-power down after 10 minutes and a password-protected factory mode for calibration verification. Output assists healthcare providers in assessing patient febrile status; aids in clinical decision-making regarding patient health and fever management.
Clinical Evidence
Bench testing only. Verification activities confirmed performance specifications including measurement accuracy in water-bath simulations, battery functionality, auto-power down timing, and audible alert verification. No clinical data presented.
Technological Characteristics
Hand-held electronic thermometer; powered by 1.5V LR41 battery. Sensing via thermistor. Features LCD display with 'Lo'/'Hi' indicators. Connectivity: None. Sterilization: Non-sterile. Software: Embedded firmware for temperature calculation, predictive algorithm, and user interface management (beeps, auto-off, fever alert).
Indications for Use
Indicated for intermittent measurement and monitoring of human body temperature in adult and pediatric populations via oral, rectal, or axillary routes.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
K021851 — DIGITAL CLINICAL THERMOMETER, MODEL KD-193 · K-Jump Health Co., Ltd. · Jul 17, 2002
K201032 — Phoenix Digital Thermometer · Phoenix Innovative Healthcare Manufacturers Pvt. , Ltd. · Oct 29, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Digital Clinical Thermometer Model V901US
# KII2116
### 510(K) SUMMARY
#### OCT 4 2011
#### AS REQUIRED BY SECTION 21 CFR 807.92(c)
Amperor Electronics (SZ) Co., Ltd. Digital Clinical Thermometer Model V901US
#### 1. SUBMITTED BY, CONTACT PERSON AND DATE OF THIS SUMMARY:
A. Company Name: Amperor Electronics (SZ) Co., Ltd.
Address: 8, Sarpotau Building, Ku-Su Village, Shi-Shian Town, Baoan County,
Shenzhen, Guang dong P.R. China
B. Contact Person
Amperor Electronics : Steel Chen, Director, Safety and Regulatory Kaz USA, Inc: Raj Kasbekar, Global Vice-President, Regulatory Affairs
C. Date of Preparation of this Summary: September 12, 2011
#### 2. DEVICE NAME:
#### Table 1: Device Names
| | Device |
|---------------------|---------------------------------|
| Proprietary Name | V901US |
| Common/Usual Name | Digital Clinical Thermometer |
| Classification Name | Clinical electronic thermometer |
#### 3. DEVICE CLASSIFICATION:
Digital Clinical Thermometer V901US (21CFR 880.2910, Product Code FLL) has been classified under section 510 of the Act as Class II (Clinical Electronic Thermometer).
2
{1}------------------------------------------------
#### 4. INTENDED USE / INDICATIONS FOR USE:
The Digital Thermometer Model V901US is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm. The device is intended for the adult and pediatric population.
The Digital Thermometer Model V901US is a hand- held, non-sterile, reusable clinical thermometer intended for the determination of human temperature in either predictive ( Instant ) mode ( 10-seconds in oral, rectal mode and under the arm mode-predicative temperature, or standard mode (actual determination of temperature).
#### 5. SUBSTANIAL EQUIVALENCE:
The digital thermometer model V901US is substantially equivalent to the predicate device V9XX (K082266) made by Amperor electronics (SZ) Co., Ltd., for the following reasons:
- 1) It has the same intended use/indication for use as the predicate.
- 2) It has the same operating principle and technological characteristics as the predicate.
- 3) The digital thermometer model V901US has some minor changes from the predicate device that include speed of temperature determination and display, change in battery type, display graphics, audible beep intensity and power down time. These changes have been verified and validated ( as part of performance testing) and are included as part of this submission. A summary of these verification and validation activities is attached. These changes raise no new issues of safety and effectiveness.
The changes (minor) to the new device include:
- 1. The speed of temperature determination and display has increased.
- 2. Change in battery type from CR1225(3V) to LR41(1.5V).
- 3. The power down time has been changed from 90s to 10min.
- 4. The intensity of the three color backlight is reduced.
- 5. The indication icon for less than 32C has been changed from "L" to"Lo".
- 6. The indication icon for more than 42.9C has been changed from "H" to"Hi".
{2}------------------------------------------------
- 7. "Valid complete" beeps have been changed from 5 Beeps to 10 Beeps.
- 8. Fever alert point has been changed from 38.3C to 37.8C.
- 9. Changed the appearance of housing.
{3}------------------------------------------------
Digital Clinical Thermometer Model V901US
| Design Control Activities Summary | | | | |
|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Device Modification | Change in Risk | Verification<br>Activity | Acceptance<br>Criteria | Results of<br>Verification |
| | | | | |
| Device Related Changes | | | | |
| Changes to device hardware, software or sensors. | | | | |
| | | Changes to device hardware. | | |
| Battery change from<br>type:CR1225(unrechargeable)3V<br>to<br>type:LR41(unrechargeable)1.5V | No change in risk<br>profile | The verification<br>activity for the<br>change in battery<br>includes<br>confirmation that<br>the unit can go<br>into the operating<br>mode and factory<br>mode and carry<br>out its normal<br>functions. Start<br>the unit and hold<br>the start button<br>for 3 seconds. The<br>unit should enter<br>the factory mode.<br>Take a few<br>readings in this<br>mode. Shut the<br>unit off and start<br>it again. This will<br>put the unit in the<br>operating mode.<br>Take a few<br>readings in this<br>mode. | The Unit should<br>go into the<br>factory mode and<br>operating mode<br>and function as<br>expected. | PASS. Traceable to<br>V901US Design<br>Verification Tests<br>1 thru 28. |
| Backlight change from three<br>color: green,orange and red to<br>do not have backlight. | No change in risk<br>profile | Start the unit.<br>Ensure that the<br>unit does not have<br>backlight in<br>normal and<br>elevated<br>temperature | Verify by<br>observation that<br>unit does not<br>have backlight. | PASS. Traceable to<br>V901US Design<br>Verification Test<br>11. |
| The indication icon for less than<br>32°C has been<br>changed from "L" to"Lo" | No change in risk<br>profile | For the normal<br>operating mode,<br>the software shall<br>signal the user<br>with no language-<br>specific words<br>other than Err, HI<br>and Lo. These<br>messages are<br>widely<br>understood and<br>have precedent<br>for successful<br>acceptance in<br>other<br>internationally<br>distributed<br>products. | Verify by<br>observation that<br>the "Lo" indicator<br>comes up for<br>temperature<br>below 32.0°C or<br>90°F | PASS. Traceable to<br>V901US Design<br>Verification Test<br>12. |
| The indication icon for more<br>than 42.9°C has been changed<br>from "H" to"Hi" | No change in risk<br>profile | For the normal<br>operating mode,<br>the software shall<br>signal the user<br>with no language-<br>specific words<br>other than Err, HI<br>and Lo.These<br>messages are<br>widely<br>understood and<br>have precedent<br>for successful<br>acceptance in<br>other<br>internationally<br>distributed<br>products. | Verify by<br>observation that<br>the "Hi" indicator<br>comes up for<br>temperature<br>above 42.9°C or<br>109.9°F | PASS. Traceable to<br>V901US Design<br>Verification Test<br>12. |
| Changes to device software. | | | | |
| Measurement time change from<br>8 seconds to 10 seconds. | No change in risk<br>profile | Start the unit.<br>Take a reading<br>using a simulated<br>waterbath. Record<br>the time required<br>to take a reading.<br>The average time<br>shall be less than<br>10 seconds. | The average<br>measurement<br>time shall be less<br>than 10 seconds. | PASS. Traceable to<br>V901US Design<br>Verification Test 1. |
| The amount of beeps after a<br>measurement is complete has<br>changed from 5 beeps to 10<br>beeps. | No change in risk<br>profile | Start the unit.<br>Take a reading<br>using a simulated<br>water-bath. Verify<br>that the number<br>of beeps to<br>indicate end of a<br>measurement is<br>10. | Verify by<br>observation that<br>the number of<br>beeps after a<br>measurement is<br>complete is 10. | Pass. Traceable to<br>V901US Design<br>Verification Tests<br>2 and 26. |
| The Fever alert point has been<br>changed from 38.3°C/100.9 to<br>37.8°C/100.0. | No change in risk<br>profile | Start the unit. Set<br>the water bath<br>temperature to<br>above 100 deg F.<br>Take a reading in<br>the water-bath.<br>Verify that the<br>fever alert beeps<br>sound. | Verify by<br>observation that<br>the number of<br>beeps for fever is<br>activated. | PASS. Traceable to<br>V901US Design<br>Verification Test<br>26. |
| The power down time has been<br>changed from 90s to 10min. | No change in risk<br>profile | Start the unit.<br>Keep it in the on<br>state for over 10<br>minutes.<br>Software control<br>unit should turns<br>itself off<br>automatically if<br>there is no action<br>taken for 10<br>minutes (± 1<br>minute). | Verify that Auto<br>power turns the<br>unit off in 10 ± 1<br>minutes when no<br>action is taken. | Pass. Traceable to<br>V901US Design<br>Verification Test 3. |
| A separate Factory mode was<br>added that can be triggered<br>only by inputting a password. | In the older<br>version, the<br>factory mode was<br>triggered without<br>any user input.<br>This was changed<br>such that a user<br>has to<br>intentionally put<br>in a password to<br>enable the factory<br>mode. This was<br>done to reduce<br>the risk related to<br>any inaccurate<br>measurements<br>during normal<br>clinical use. This | 1. After starting<br>the unit, press and<br>hold the ON<br>switch (power<br>key) for more<br>than 3 seconds.<br>The signal "- - -"<br>will appear. Then<br>key in a right one<br>digit password, by<br>pressing the same<br>ON switch, which<br>will put the unit<br>into the factory<br>mode.<br>2. Verify accuracy<br>of the product in a<br>water bath. | 1. Verify that the<br>unit goes into the<br>factory mode.<br>2.The product<br>should have the<br>following<br>accuracy in water-<br>bath mode:<br>±0.2 °F 98.0°F-<br>102.0°F<br><br>±0.3°F 96.4°F-<br>97.9°F,<br>102.1°F-106.0°F<br><br>±0.5 °F <96.4°F<br>or >106°F | PASS. Traceable to<br>V901US Design<br>Verification Tests<br>1 thru 28. |
| profile of the<br>thermometer by<br>eliminating any<br>inaccurate<br>temperature<br>measurements. | | | | |
{4}------------------------------------------------
Digital Clinical Thermometer Model V901US
б
{5}------------------------------------------------
Digital Clinical Thermometer Model V901US
"
ﻪ
{6}------------------------------------------------
Digital Clinical Thermometer Model V901US
.
| Changes to device sensors. | | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| NONE | | | | |
| Manufacturing Process Changes | | | | |
| NONE | | | | |
| Vendor Changes | | | | |
| NONE | | | | |
| Material Changes in Outer Casing of Device | | | | |
| There was no change to any patient contacting material. The probe cover is the only part that comes in contact with the patient or end user. Biocompatibility Testing for the probe cover was addressed in 510(K) K082266 of predicate device. | | | | |
| Labeling Changes | | | | |
Changes done may have been to clarify or further elaborate existing language only. Legal entity and format only.
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Amperor Electronics (SZ) Company, Limited % Mr. Raj Kasbekar Regulatory Affairs Kaz Incorporated 250 Turnpike Road Southborough, Massachusetts 01772
OCT - 4 2011
Re: K112116
Trade/Device Name: Digital Clinical Thermometer Model V901US Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: September 13, 2011 Received: September 14, 2011
Dear Mr. Kasbekar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Kasbekar
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan Quon
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Digital Clinical Thermometer Model V901US
### Intended Use/Indications for Use Statement
510(k) Number (if known): K | 1 2 \ 1 2 \ 1 2 \ (b
Digital Clinical Thermometer Model V901US Device Name:
Intended Use/Indications For Use:
The Digital Thermometer Model V901US is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm. This thermometer is intended for the adult and pediatric population.
The Digital Thermometer Model V901US is a hand- held, non-sterile, reusable clinical thermometer intended for the determination of human temperature in either predictive or Instant mode ( 10-seconds in oral, rectal mode and under the arm mode-predicative temperature ), or standard mode (actual determination of temperature).
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ AND/OR Prescription Use _ _ _ _ _ (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ald C. Ch 9/30/4
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112116
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Date: 2010-10-14 Revision: 1.1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.