MICROLIFE DIGITAL ELECTRONIC THERMOMETER, MODEL MT16K1
K091589 · Microlife Intellectual Property GmbH · FLL · Jun 10, 2009 · General Hospital
Device Facts
Record ID
K091589
Device Name
MICROLIFE DIGITAL ELECTRONIC THERMOMETER, MODEL MT16K1
Applicant
Microlife Intellectual Property GmbH
Product Code
FLL · General Hospital
Decision Date
Jun 10, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Microlife Digital Electronic Thermometer MT16K1 is used for the intermittent measurement and monitoring of human body temperature, oral, rectal and under the arm. The device is for the adult and pediatric population.
Device Story
Digital electronic thermometer; measures body temperature via thermistor; converts resistance changes to frequency changes in an R-C oscillator circuit; pulse number corresponds to temperature; displays results on LCD. Used for intermittent monitoring in clinical or home settings; operated by clinicians or patients. Provides numerical temperature output to assist in clinical decision-making regarding patient health status.
Clinical Evidence
Clinical studies conducted per ASTM E1965-03. Evaluated clinical bias, clinical uncertainty, and clinical repeatability. Bench testing performed for compliance with ASTM E1112, IEC60601-1, and IEC60601-1-2.
Technological Characteristics
Thermistor-based temperature sensing; R-C oscillator circuit for frequency conversion; LCD display; microprocessor-based. Complies with ASTM E1112, IEC60601-1, and IEC60601-1-2 standards.
Indications for Use
Indicated for intermittent measurement and monitoring of human body temperature via oral, rectal, or axillary routes in adult and pediatric populations.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Microlife Digital Electronic Thermometer, Model MT3001 (K851146)
Related Devices
K062867 — MICROLIFE INSTANT DIGITAL ELECTRONIC THERMOMETER, MODEL QT1JC1 · Microlife Intellectual Property GmbH · Dec 6, 2006
K031958 — MICROLIFE INSTANT DIGITAL ELECTRONIC THERMOMETER, MODEL QT1JA1 · Microlife Intellectual Property GmbH · Aug 15, 2003
K023500 — DIGITAL CLINICAL THERMOMETER, MODEL BABY TEMP/WL-3305 AND BABY TEMP (+) /WL-3306 · Innovatech, Inc. · Dec 13, 2002
K011585 — FAST THERMOMETER MODELS ST713C AND ST713F · Mesure Technology · Jun 13, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
EXHIBIT #1
# K491589
## 510 (K) SUMMARY
JUN 1 0 2009
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510 (k) number is:
#### 1. Submitter's Identification:
Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland
Date Summary Prepared: May 27, 2009
#### 2. Name of the Device:
Microlife Digital Electronic Thermometer, Model MT16K1
#### 3. Predicate Device Information:
Microlife Digital Electronic Thermometer, Model MT3001, K#851146 .
#### 4. Device Description:
This Digital Electronic Thermometer enables easy and highly accurate readings over the body temperature range and has been designed to provide maximum user-friendliness. The body temperature is measured by the thermistor and displayed as numbers on the LCD (liquid crystal display) through microprocessor of IC.
The basic principle of this thermometer is that change of thermistor resistance, caused by changes of temperature, are converted to changes of frequency of R-C oscillator circuit, Therefore, temperature can be given by measuring the frequency of oscillator.
For a given time period, by applying the R-C oscillator circuit, changes of temperature will correspond to changes of pulse number.
{1}------------------------------------------------
## 5. Intended Use:
# 5091589
The Microlife Digital Electronic Thermometer MT16K1 is used for the intermittent measurement and monitoring of human body temperature, oral, rectal and under the arm. The device is for the audit and pediatric population.
## 6. Comparison to Predicate Devices:
The Microlife Digital Electronic Thermometer, Model MT16K1 is substantially equivalent to Microlife Digital Electronic Thermometer, Model MT3001, K#851146, which has the same intended use and is similar in design to the predicate device.
The Microlife Digital Electronic Thermometer MT16K1 and the predicate device are identical in the temperature measurements algorithm and fundamental scientific technology, differing mostly in probe tip and case material.
## Discussion of Non-Clinical Tests Performed for 7. Determination of Substantial Equivalence are as follows :
Compliance to applicable voluntary standards includes ASTM E1112, as well as IEC60601-1 and IEC60601-1-2 requirements.
Guidance documents included the ''FDA Guidance on the Content of Premarket Notification 510(K) Submissions for Clinical Electronic Thermometers''.
## 8. Discussion of Clinical Tests Performed:
Controlled human clinical studies were conducted in accordance with ASTM E1965-03 using the Microlife Digital Electronic Thermometer, Model MT16K1. Clinical data was presented which evaluated clinical bias, clinical uncertainty and clinical repeatability per the Microlife Clinical Test Protocol outline.
### 9. Conclusions :
The Microlife Digital Electronic Thermometer, Model MT16K1 has the same intended use and similar technological characteristics as the Microlife Digital Electronic Thermometer, Model MT3001. Moreover, bench testing
{2}------------------------------------------------
contained in this submission demonstrates that differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, the Microlife Digital Electronic Thermometer, Model MT16K1 is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. In the center of the seal is an emblem featuring a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and wings at the top.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Microlife Intellectual Property GmbH C/O Ms. Susan D. Goldstein-Falk mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
# JUN 1.0 2009
Re: K091589
Trade/Device Name: Microlife Digital Electronic Thermometer, Model MT16K1 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: May 29, 2009 Received: June 2, 2009
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2- Ms. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/Centers Offices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Anthony, D. m
Susan Runner, D.D.S., M.A.
Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## EXHIBIT B
## Indications for Use
510(k) Number (if known): Komissy
Device Name: Microlife Digital Electronic Thermometer, Model MT16K1
Indications For Use:
The Microlife Digital Electronic Thermometer MT16K1 is used for the intermittent measurement and monitoring of human body temperature, oral, rectal and under the arm. The device is for the adult and pediatric population.
Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page で of -
Nade
:
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: JC 091589
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.