FORA IR20B EAR THERMOMETER, MODEL TD-1261F, FORA IR21B EAR THERMOMETER, MODEL TD-1261E
K090395 · Taidoc Technology Corporation · FLL · Mar 10, 2009 · General Hospital
Device Facts
Record ID
K090395
Device Name
FORA IR20B EAR THERMOMETER, MODEL TD-1261F, FORA IR21B EAR THERMOMETER, MODEL TD-1261E
Applicant
Taidoc Technology Corporation
Product Code
FLL · General Hospital
Decision Date
Mar 10, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The FORA IR20b/FORA IR21b Ear Thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal on people of all ages and for use in the home. It is also available to detect object's surface temperature including human skin.
Device Story
FORA IR20b/IR21b are infrared ear thermometers for home use by patients of all ages. Device uses infrared sensor to detect thermal radiation from ear canal; also measures surface temperature of objects/skin. Operation involves positioning probe in ear canal; device captures infrared signal, processes data, and displays temperature reading on integrated screen. Provides non-invasive, rapid temperature assessment to assist users in monitoring health status. No remote processing or cloud connectivity described.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison to legally marketed predicate devices.
Technological Characteristics
Infrared sensor-based electronic thermometer; handheld form factor; battery-powered; intended for ear canal and surface temperature measurement.
Indications for Use
Indicated for human body temperature measurement via ear canal for all ages in home settings; also indicated for surface temperature measurement of objects, including human skin.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Related Devices
K090294 — FORA IR22 EAR THERMOMETER, MODEL TD-1262 · Taidoc Technology Corporation · Feb 26, 2009
K083299 — TAIDOC TD-11 SERIES AND CLEVER TD-1112 EAR/SKIN/ SURFACE IR THERMOMETERS & FORA IR16/17 EAR THERMOMETER · Taidoc Technology Corporation · Dec 10, 2008
K081445 — FORA COMFORTSCAN EAR THERMOMETER, MODELS TD-1261A; TD-1261B · Taidoc Technology Corporation · Aug 11, 2008
K042631 — EAR THERMOMETER, MODEL WL-320X · Innovatech, Inc. · Dec 22, 2004
K030324 — RADIANT INNOVATION INC. INFRARED EAR THERMOMETER, MODELS TH1 SERIES · Radiant Innovation, Inc. · Feb 27, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPT. OF HEALTH & HUMAN SERVICES - USA" is arranged around the caduceus in a circular fashion. The logo is black and white.
MAR 1 0 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
1
Mr. Yuhua Chen Assistant Manager TaiDoc Technology Corporation 6F, Number 127, Wugong 2nd Road Wugu Township Taipei County 248 CHINA (TAIWAN)
K090395 Re:
> Trade/Device Name: FORA IR20b//FORA IR21b Ear Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: February 13, 2009 Received: February 17, 2009
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Chen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Snitte H. Hickam Onis
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K090395
## Attachment 2
## Indications for Use
510(k) Number:
Device Name: FORA IR20b/FORA IR21b Ear Thermometer
Indications for Use:
The FORA IR20b/FORA IR21b Ear Thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal on people of all ages and for use in the home. It is also available to detect object's surface temperature including human skin.
Prescription Use ______________ AND/OR (21 CFR Part 801 Subpart D)
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
KORDS 85 510(k) Number: __
A2
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.