Chemical Indicator for enspire CLCSPS (LCC015)

K243475 · STERIS Corporation · JOJ · Dec 4, 2024 · General Hospital

Device Facts

Record IDK243475
Device NameChemical Indicator for enspire CLCSPS (LCC015)
ApplicantSTERIS Corporation
Product CodeJOJ · General Hospital
Decision DateDec 4, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The enspire 3000 Chemical Indicator is a peracetic acid concentration indicator for routine monitoring of the liquid chemical sterilization cycle of the enspire 3000 employing S40 Sterilant. The unprocessed enspire 3000 Chemical Indicator is blue. When exposed in the enspire 3000 processor to a concentration of >1820 ppm (mg/L) peracetic acid found in the S40 use dilution during a controlled 6-minute exposure at 45.5 - 60°C, the indicator color changes from the start to the pass color. See reference colors on the bottle.

Device Story

Chemical indicator strip for monitoring liquid chemical sterilization cycles. Consists of indicator ink (Lissamine green B and savinyl orange) printed on a polypropylene substrate, covered by a sterilant-permeable polyether block amide laminate. Used in the enspire 3000 processor with S40 Sterilant. During a 6-minute cycle at 45.5–60°C, peracetic acid (>1820 ppm) oxidizes the dye, causing a color change from blue to a pass color. Provides visual confirmation of sterilization conditions to the operator. Modification involves updated labeling with additional pass reference colors and gradients to improve user interpretation. No physical device changes.

Clinical Evidence

Bench testing only. Human factors study conducted to validate the updated labeling graphics; results showed users performed all critical tasks successfully.

Technological Characteristics

Polypropylene substrate with screen-printed indicator ink (Lissamine green B and savinyl orange). Covered by a sterilant-permeable polyether block amide laminate. Mechanism: oxidative bleaching of dye. Dimensions/form factor: strip. Disposable. Shelf-life: 26 months; open bottle shelf-life: 6 months.

Indications for Use

Indicated for routine monitoring of liquid chemical sterilization cycles in the enspire 3000 processor using S40 Sterilant. Intended for use by healthcare personnel operating the sterilization system.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 3, 2024 STERIS Corporation Gregory Land Lead Regulatory Affairs Specialist 5960 Heisley Rd Mentor, Ohio 44060 Re: K243475 Trade/Device Name: Chemical Indicator for enspire CLCSPS (LCC015) Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization process indicator Regulatory Class: Class II Product Code: JOJ Dated: November 8, 2024 Received: November 8, 2024 Dear Gregory Land: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for: Christopher Dugard Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K243475 Device Name Chemical Indicator for enspire CLCSPS (LCC015) Indications for Use (Describe) The enspire 3000 Chemical Indicator is a peracetic acid concentration indicator for routine monitoring of the liquid chemical sterilization cycle of the enspire 3000 employing S40 Sterilant. The unprocessed enspire 3000 Chemical Indicator is blue. When exposed in the enspire 3000 processor to a concentration of >1820 ppm (mg/L) peracetic acid found in the S40 use dilution during a controlled 6-minute exposure at 45.5 - 60°C, the indicator color changes from the start to the pass color. See reference colors on the bottle. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for STERIS. The word "STERIS" is written in a serif font at the top of the image. Below the word, there is a graphic of several horizontal wavy lines in blue. # 510(k) Summary For Celerity Chemical Indicator for enspire 3000 CLCSPS K243475 STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 | Contact: | Gregory Land | |------------|------------------------------------| | | Lead Regulatory Affairs Specialist | | Telephone: | (440) 392-7424 | | Email: | greg_land@steris.com | Submission Date: December 2, 2024 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {5}------------------------------------------------ ## STERIS Special 510(k) PREMARKET NOTIFICATION - K243475 Celerity Chemical Indicator for enspire 3000 CLCSPS #### 1. Device Name | Trade Name: | Celerity Chemical Indicator for enspire 3000 CLCSPS | |----------------------|-----------------------------------------------------| | Models: | LCC015 | | Common Name: | Chemical Indicator. | | Classification Name: | Physical/chemical sterilization process indicator | | Classification | 21 CFR 880.2800 | | Product Code | JOJ | #### 2. Predicate Device Celerity Chemical Indicator for enspire 3000 CLCSPS (K240032) #### 3. Device Description The enspire 3000 Chemical Indicator is a chemical indicator strip consisting of indicator ink containing the reactive chemicals printed on one end of a polypropylene strip. The product is manufactured by application of the indicator ink by screen printing to a substrate with the indicator ink printed thereon. A clear, sterilant-permeable polyether block amide laminate is adhesively bonded to the polypropylene strip following printing of the ink, completely covering the ink. Modification for Which Clearance is Being Sought: There were no physical changes to the device and are no new indications for use. The labeling was modified to include an additional pass reference color and a gradient between the original and new pass reference colors to improve user interpretation of chemical indicator. #### Intended Use/Indications for Use: 4. The enspire 3000 Chemical Indicator is a peracetic acid concentration indicator for routine monitoring of the liquid chemical sterilization cycle of the enspire 3000 employing S40 Sterilant. The unprocessed enspire 3000 Chemical Indicator is blue. When exposed in the enspire 3000 processor to a concentration of >1820 ppm (mg/L) peracetic acid found in the S40 use dilution during a controlled 6-minute exposure at 45.5 - 60°C, the indicator color changes from the start to the pass color. See reference colors on the bottle. #### న్. Technological Characteristic Comparison The proposed device and its predicate are physically identical and have the same Indications for Use. They differ only in the product labelling. The proposed device has additional graphics to aid the user in identifying Pass conditions. {6}------------------------------------------------ ## STERIS Special 510(k) PREMARKET NOTIFICATION – K243475 Celerity Chemical Indicator for enspire 3000 CLCSPS | Table 1 | | | | |--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Feature | Proposed<br>Celerity Chemical Indicator<br>for enspire CLCSPS | Predicate K240032<br>Celerity Chemical Indicator<br>for enspire CLCSPS | Comparison | | | The enspire 3000 Chemical<br>Indicator is a peracetic acid<br>concentration indicator for<br>routine monitoring of the<br>liquid chemical sterilization<br>cycle of the enspire 3000<br>employing S40 Sterilant. | The enspire 3000 Chemical<br>Indicator is a peracetic acid<br>concentration indicator for<br>routine monitoring of the<br>liquid chemical sterilization<br>cycle of the enspire 3000<br>employing S40 Sterilant. | | | Intended<br>use | The unprocessed enspire 3000<br>Chemical Indicator is blue.<br>When exposed in the enspire<br>3000 processor to a<br>concentration of >1820 ppm<br>(mg/L) peracetic acid found in<br>the S40 use dilution during a<br>controlled 6-minute exposure<br>at 45.5 - 60°C, the indicator<br>color changes from the start to<br>the pass color. See reference<br>colors on the bottle. | The unprocessed enspire<br>3000 Chemical Indicator is<br>blue. When exposed in the<br>enspire 3000 processor to a<br>concentration of >1820 ppm<br>(mg/L) peracetic acid found<br>in the S40 use dilution during<br>a controlled 6-minute<br>exposure at 45.5 - 60°C, the<br>indicator color changes from<br>the start to the pass color. See<br>reference colors on the bottle. | Identical | | Device<br>design -<br>component<br>s | Printed indicator ink printed<br>onto polypropylene overlaid<br>with a clear, permeable<br>laminate | Printed indicator ink printed<br>onto polypropylene overlaid<br>with a clear, permeable<br>laminate | Identical | | Indicator<br>agent | Lissamine green B (active<br>dye) and savinyl orange<br>(background dye) | Lissamine green B (active<br>dye) and savinyl orange<br>(background dye) | Identical | | Sterilizatio<br>n method<br>and cycles | Used in a liquid chemical<br>sterilant processing system<br>employing S40 Sterilant<br>Concentrate to form a use<br>dilution concentration of<br>≥1820 ppm (mg/L) peracetic<br>acid and provide 6 minutes<br>exposure at 45.5 – 60°C. | Used in a liquid chemical<br>sterilant processing system<br>employing S40 Sterilant<br>Concentrate to form a use<br>dilution concentration of<br>≥1820 ppm (mg/L) peracetic<br>acid and provide 6 minutes<br>exposure at 45.5 – 60°C. | Identical | | Mechanis<br>m of<br>action | Bleaching of lissamine green<br>B dye as a result of oxidation,<br>resulting in a color change. | Bleaching of lissamine green<br>B dye as a result of oxidation,<br>resulting in a color change. | Identical | | Peracetic<br>acid | > 1820 mg/L PAA | > 1820 mg/L PAA | Identical | | Feature | Proposed<br>Celerity Chemical Indicator<br>for enspire CLCSPS | Predicate K240032<br>Celerity Chemical Indicator<br>for enspire CLCSPS | Comparison | | concen-<br>tration for<br>the<br>endpoint<br>color<br>change | | | | | Disposable | Yes | Yes | Identical | | Shelf-life | 26 months; | 26 months | Identical | | Open<br>bottle shelf<br>life | 6 months | 6 months | Identical | {7}------------------------------------------------ ### STERIS Special 510(k) PREMARKET NOTIFICATION - K243475 Celerity Chemical Indicator for enspire 3000 CLCSPS #### 6. Performance Testing The following table summarizes the non-clinical testing that has demonstrated that the product is substantial equivalent to the predicate device (K240032). | Testing | Acceptance Criteria | Results | |------------------------|-------------------------------------------------|---------| | Human Factors<br>Study | Users performed all critical tasks successfully | PASS | #### 7. Conclusion Based on the intended uses, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device (K240032), Class II (CFR 880.2800), product code JOJ.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...