← Product Code [JOJ](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/JOJ) · K241836

# Kem Medical Lead-free Chemical Indicators for Steam Sterilization (K241836)

_Kem Medical Products Corporation · JOJ · Sep 30, 2024 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/JOJ/K241836

## Device Facts

- **Applicant:** Kem Medical Products Corporation
- **Product Code:** [JOJ](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/JOJ.md)
- **Decision Date:** Sep 30, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Kem Medical Lead-free Chemical Indicators for Steam Sterilization are designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed and unprocessed units or loads. Use the Kem Medical Lead-free Chemical Indicators for Steam Sterilization in the validated steam sterilization processes described below: *Gravity: 121 C/250 F - 30 minutes (wrapped/porous) *Gravity: 132 C/270 F - 3 minutes (unwrapped/nonporous) *Gravity: 132 C/270 F - 15 minutes (wrapped/porous) *Gravity: 135 C/275 F - 3 minutes (unwrapped/nonporous) *Gravity: 135 C/275 F - 10 minutes (wrapped/porous or unwrapped/nonporous, mixed load) *Vacuum assisted (prevacuum): 132 C/270 F - 3 minutes (unwrapped/nonporous) *Vacuum assisted (prevacuum): 132 C/270 F - 4 minutes (wrapped/porous) *Vacuum assisted (prevacuum): 134 C/273 F - 4 minutes (wrapped/porous) *Vacuum assisted (prevacuum): 135 C/275 F - 3 minutes (wrapped/porous or unwrapped/nonporous, mixed load)

## Device Story

Device consists of lead-free chemical indicator ink printed on paper substrate, with or without adhesive. Used by healthcare providers to monitor steam sterilization cycles. Input is exposure to steam sterilization conditions; transformation is a chemical reaction causing a visual color change from light (white/off-white) to dark (cocoa/black). Output is a visual indicator on the unit or load. Benefits include clear differentiation between processed and unprocessed items, ensuring sterilization process awareness. Used in clinical settings during standard autoclave cycles.

## Clinical Evidence

Bench testing only. Performance validated per ISO 11140-1:2014 for Type 1 chemical indicators. Testing included BIER vessel runs and standard autoclave cycles (gravity and vacuum-assisted). Results confirmed color change to cocoa/black at pass criteria and no/minimal change at fail criteria. Endpoint color stability verified for 8 months; shelf life validated for 3 years.

## Technological Characteristics

Type 1 chemical process indicator per ISO 11140-1:2014. Consists of lead-free indicator ink on paper substrate (with/without adhesive). Visual color change endpoint (white to cocoa/black). No energy source or software.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- 3M Comply Process Indicators for Steam ([K101528](/device/K101528.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 30, 2024

Kem Medical Products Corporation Lisa Kruger Vice President 400 Oser Avenue Suite 2400 Hauppauge, New York 11788

Re: K241836

Trade/Device Name: Kem Medical Lead-free Chemical Indicators for Steam Sterilization Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: June 25, 2024 Received: June 25, 2024

Dear Lisa Kruger:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Digitally signed by Stephen A. Stephen A. Anisko -S Anisko -S Date: 2024.09.30 16:22:13 -04'00'

for: Christopher Dugard Assistant Director DHT4C: Division of Infection Control OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K241836

Device Name

Kem Medical Lead-free Chemical Indicators for Steam Sterilization

Indications for Use (Describe)

The Kem Medical Lead-free Chemical Indicators for Steam Sterilization are designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed and unprocessed units or loads. Use the Kem Medical Lead-free Chemical Indicators for Steam Sterilization in the validated steam sterilization processes described below:

| *Gravity:                     | 121 C/250 F - 30 minutes (wrapped/porous)                                    |
|-------------------------------|------------------------------------------------------------------------------|
| *Gravity:                     | 132 C/270 F - 3 minutes (unwrapped/nonporous)                                |
| *Gravity:                     | 132 C/270 F - 15 minutes (wrapped/porous)                                    |
| *Gravity:                     | 135 C/275 F - 3 minutes (unwrapped/nonporous)                                |
| *Gravity:                     | 135 C/275 F - 10 minutes (wrapped/porous or unwrapped/nonporous, mixed load) |
| *Vacuum assisted (prevacuum): | 132 C/270 F - 3 minutes (unwrapped/nonporous)                                |
| *Vacuum assisted (prevacuum): | 132 C/270 F - 4 minutes (wrapped/porous)                                     |
| *Vacuum assisted (prevacuum): | 134 C/273 F - 4 minutes (wrapped/porous)                                     |
| *Vacuum assisted (prevacuum): | 135 C/275 F - 3 minutes (wrapped/porous or unwrapped/nonporous, mixed load)  |

Type of Use (Select one or both, as applicable)

Type of Use (**Select one or both**, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

> Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

Prepared on: 2024-09-25

| Contact Details                    | 21 CFR 807.92(a)(1)                                               |                     |  |
|------------------------------------|-------------------------------------------------------------------|---------------------|--|
| Applicant Name                     | Kem Medical Products Corporation                                  |                     |  |
| Applicant Address                  | 400 Oser Avenue Suite 2400 Hauppauge NY 11788 United States       |                     |  |
| Applicant Contact Telephone        | 631-454-6565                                                      |                     |  |
| Applicant Contact                  | Lisa Kruger                                                       |                     |  |
| Applicant Contact Email            | lisa@kemmed.com                                                   |                     |  |
| Device Name                        | 21 CFR 807.92(a)(2)                                               |                     |  |
| Device Trade Name                  | Kem Medical Lead-free Chemical Indicators for Steam Sterilization |                     |  |
| Common Name                        | Sterilization process indicator                                   |                     |  |
| Classification Name                | Indicator, Physical/Chemical Sterilization Process                |                     |  |
| Regulation Number                  | 880.2800                                                          |                     |  |
| Product Code(s)                    | JOJ                                                               |                     |  |
| Legally Marketed Predicate Devices |                                                                   | 21 CFR 807.92(a)(3) |  |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first)          | Product Code        |  |
| K101528                            | 3M Comply Process Indicators for Steam                            | JOJ                 |  |

#### Device Description Summary

21 CFR 807.92(a)(4)

The Kem Medical Lead-free Chemical Indicators for Steam Sterilization are designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed and unprocessed units or loads.

They are Type 1 Process Indicators per ISO 11140-1:2014 Sterilization of healthcare products-Chemical Indicators - Part 1 General Requirements, consisting of chemical indicator ink and paper substrate with a visual color change from light color (white/off-white) to dark color (cocoa/black) when exposed to a steam sterilization process.

The Kem Medical Lead-free Chemical Indicators for Steam Sterilization consist of indicator ink printed on a paper substrate with or without adhesive. The indicator ink is lead free. An example of this indicator is the ink printed onto adhesive-coated paper for autoclave labels, or "dots", which are removed from a poly or paper backing for use.

Kem Medical manufactures in various shapes and sizes printed on paper substrates with and without adhesive on the underside, and with or without labeling on each individual indicator.

{5}------------------------------------------------

The Kem Medical Lead-free Chemical Indicators for Steam Sterilization are designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed and unprocessed units or loads. Use the Kem Medical Lead-free Chemical Indicators for Steam Sterilization in the validated steam sterilization processes described below:

*Gravity: 121 C/250 F - 30 minutes (wrapped/porous)

*Gravity: 132 C/270 F - 3 minutes (unwrapped/nonporous)

*Gravity: 132 C/270 F - 15 minutes (wrapped/porous)

*Gravity: 135 C/275 F - 3 minutes (unwrapped/nonporous)

* Gravity: 135 C/275 F - 10 minutes (wrapped/porous or unwrapped/nonporous, mixed load)

* Vacuum assisted (prevacuum): 132 C/270 F - 3 minutes (unwrapped/nonporous)

* Vacuum assisted (prevacuum): 132 C/270 F - 4 minutes (wrapped/porous)

* Vacuum assisted (prevacuum): 134 C/273 F - 4 minutes (wrapped/porous)

*Vacuum assisted (prevacuum): 135 C/275 F - 3 minutes (wrapped/porous or unwrapped/nonporous, mixed load)

#### Indications for Use Comparison

Indications for Use for the Kem Medical Lead-free Chemical Indicators for Steam Sterilization are the same as the Predicate Device, the 3M Comply Lead-free Indicator Tape for Steam Sterilization. They are both designed for use by a healthcare provider to accompany individual units or loads (wrapped or unwrapped packs) to demonstrate that the unit has been exposed to the sterilization process and to distinguish between processed and unprocessed units or loads. For use in validated 121 - 135 degrees Celsius (250 - 275 degrees Fahrenheit) steam sterilization processes.

#### Technological Comparison

#### 21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

| Element                   | Subject Device (K241836)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device (K101528)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Comparison                                  |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Intended Use              | Process indicator for steam<br>sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Process Indicator for steam<br>sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                        |
| Product Code              | JOJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | JOJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                        |
| Regulation                | 21 CFR§ 880.2800                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 21 CFR§ 880.2800                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                        |
| Indications for Use       | The Kem Medical Lead-free<br>Chemical Indicator for<br>Steam Sterilization are<br>designed for use by a health<br>care provider to demonstrate<br>that the unit or load has been<br>exposed to a steam<br>sterilization process, and to<br>distinguish between<br>processed and unprocessed<br>units or loads. Use the Kem<br>Medical Lead-free Chemical<br>Indicators for Steam<br>Sterilization in the validated<br>steam sterilization processes<br>described below:                                                                                                                                                                                                                                                                   | The 3M Comply™ Lead Free<br>Process Indicators for Steam are<br>designed to demonstrate that the<br>unit or load has been exposed to a<br>steam sterilization process and to<br>distinguish between processed and<br>unprocessed units or loads.<br>3M Comply Lead Free Process<br>Indicators for Steam include: *<br>Comply 68200 Lead Free Record<br>Card * Comply 1 322 Lead Free<br>Indicator Tape * Lead Free Test<br>Pack Labels to be used on test<br>packs including the following -<br>Comply 4 13 80 Steam-Plus Test<br>Pack - Comply 4 1360 Steam Test<br>Pack -3M Attest™ 1276 Steam | Same                                        |
|                           | *Gravity: 121 C/250 F -<br>30 minutes<br>(wrapped/porous)<br>*Gravity: 132 C/270 F - 3<br>minutes<br>(unwrapped/nonporous)<br>*Gravity: 132 C/270 F - 15<br>minutes (wrapped/porous)<br>*Gravity: 135 C/275 F - 3<br>minutes<br>(unwrapped/nonporous)<br>*Gravity: 135 C/275 F - 10<br>minutes (wrapped/porous or<br>unwrapped/nonporous,<br>mixed load)<br>*Vacuum assisted<br>(prevacuum): 132 C/270 F -<br>3 minutes<br>(unwrapped/nonporous)<br>*Vacuum assisted<br>(prevacuum): 132 C/270 F -<br>4 minutes (wrapped/porous)<br>*Vacuum assisted<br>(prevacuum): 134 C/273 F -<br>4 minutes (wrapped/porous)<br>*Vacuum assisted<br>(prevacuum): 135 C/275 F -<br>3 minutes (wrapped/porous<br>or unwrapped/nonporous,<br>mixed load) | Test Pack - 3M AttestTM 1296<br>Steam Test Pack - 3M AttestTM 4<br>1382 Steam-Plus Test Pack -<br>Comply 1233LF, 001 32LF, 0013<br>5LF Bowie Dick Test Packs                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                             |
| Intended Use              | Process indicator for steam<br>sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Process indicator for steam<br>sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                        |
| Indicator Agent           | Lead-free Steam Indicator<br>Ink                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Lead Free Steam Indicator Ink                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                        |
| Endpoint<br>Specification | Cocoa/Black color change                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Dark brown/black color change                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Different ink chemical<br>composition       |
| Endpoint Stability        | 8 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 6 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different                                   |
| Shelf Life                | 3 years from date of<br>manufacture, validated real<br>time at 0, 1, 2, 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 18 months from date of<br>manufacture                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Different<br>validation<br>testing schedule |

Technological Characteristics Comparison Table

{6}------------------------------------------------

#### Non-Clinical and/or Clinical Tests Summary & Conclusions

#### 21 CFR 807.92(b)

#### Non-Clinical Testing

Qualification of the Kem Medical Products Type 1 Chemical Indicators for Steam Sterilization was performed by True Indicating LLC for 0-, 1-, 2- and 3-years real time. Testing was done to demonstrate that the CI's (Chemical Indicator's) performance meets the requirements defined in ISO 11140-1: 2014 Sterilization of Healthcare Products- Chemical Indicators- General Requirements for a Type 1 Chemical Indicator, and to determine substantial equivalence to the 3M Comply Lead Free Indicator Tape for Steam Sterilization. The Cls are for use as an external or internal chemical indicator in monitoring exposure to conditions of steam sterilization for 121°C - 135°C gravity and pre-vacuum sterilizer cycles.

Cycles for validation testing included BIER Vessel runs, with "pass/fail" criteria, of Saturated Steam 121°C for 2 minutes "fail" and 10 minutes "pass", Saturated Steam 132°C, 134°C and 135°C for 0.3 minutes "fail" and for 2 minutes "pass". Autoclave cycles were also used to test for reaction to Dry Heat 140°C for 30 minutes, and

{7}------------------------------------------------

performance in the following standard autoclave cycles:

Autoclave Pre- vac cycle 132°C for 3 minutes, Pre-vac cycle 132°C for 4 minutes, Pre-vac cycle 134°C for 3 minutes, Pre-vac cycle 135°C for 3 minutes; and Autoclave Gravity cycle 121°C for 30 minutes, Autoclave Gravity cycle 132°C for 15 minutes, Autoclave Gravity cycle 135°C for 10 minutes. Failure cycles were run for Pre-vac 132°C for 0 minutes and Pre-vac 135°C for 0 minutes.

Chemical indicator results for all testing performed under the acceptance criterialisted and as required in ISO 11140-1:2014 for Type 1 chemical indicators to achieve the target 3-year shelf life. All indicators exposed to either BIER sterilization parameters at "pass" test criteria (121°C for 10 minutes and at 132°C, 134°C and 135°C for 2 minutes) showed a dark color change. All indicators exposed to either BIER sterlization parameters at "fail" test criteria (121°C for 2 minutes and at 132°C, 134°C for .3 minutes) and dry heat (1400 C for 30 minutes) showed either no color change or a slight color change. All samples also met the endpoint/color change requirements per the ISO standard.

| Test                                                | Purpose                                                                                                                | Acceptance Criteria                                                                                                                                                                                                                                  | Result                                                                                |      |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------|
| Saturated<br>Steam Testing;<br>ISO 11140-<br>1:2014 | Meets the<br>requirements for<br>Type I process<br>indicator for steam.                                                | 2 min at 121 °C                                                                                                                                                                                                                                      | No change or change<br>that is markedly<br>different from the<br>visible color change | Pass |
|                                                     |                                                                                                                        | 10 min at 121 °C                                                                                                                                                                                                                                     | Visible color change                                                                  |      |
|                                                     |                                                                                                                        | 0.3 min at 132 °C /<br>134 °C /135 °C                                                                                                                                                                                                                | No change or change<br>that is markedly<br>different from the<br>visible color change |      |
|                                                     |                                                                                                                        | 2 min at 132 °C /<br>134 °C /135 °C                                                                                                                                                                                                                  | Visible color change                                                                  |      |
| Standard<br>Autoclave<br>cycles                     | Meets performance<br>criteria in standard<br>autoclave cycles                                                          | Endpoint color reached when exposed to<br>standard autoclave cycles.<br>Endpoints color not reached when exposed to<br>failing conditions in autoclave cycles.                                                                                       |                                                                                       | Pass |
| Dry heat                                            | Verify device<br>requires the<br>presence of<br>saturated steam to<br>turn to reach<br>endpoint                        | 30 min at 140 °C                                                                                                                                                                                                                                     | No change                                                                             | Pass |
| Transference<br>ISO 11140-<br>1:2014                | Demonstrate the<br>chemical indicators<br>do not bleed or<br>offset to substrate<br>which it's applied<br>according to | The chemical<br>indicators shall not<br>offset or bleed,<br>penetrate the<br>substrate to which it<br>is applied, or<br>materials in which it<br>is in contact before,<br>during or after the<br>sterilization cycles<br>for which it is<br>designed |                                                                                       | Pass |
| Endpoint color<br>stability                         | Endpoint color<br>maintained                                                                                           | Endpoint color<br>remained<br>unchanged for 8<br>months                                                                                                                                                                                              |                                                                                       | Pass |

#### Summary of Non-Clinical Testing

{8}------------------------------------------------

### Clinical Testing

Not Applicable

# Conclusion

The conclusions drawn from the nonclinical testing demonstrate that the subject device, Kem Medical Lead-Free Chemical Indicators for Steam Sterilization, are as safe, as effective and performs as well as or better than the legally marketed predicate, 3M Comply Process Indicators for Steam, K101528.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/JOJ/K241836](https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/JOJ/K241836)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
