PROCESS INDICATOR TAPE FOR STEAM STERILIZATION

K140940 · Canadian Technical Tape, Ltd. · JOJ · Oct 27, 2014 · General Hospital

Device Facts

Record IDK140940
Device NamePROCESS INDICATOR TAPE FOR STEAM STERILIZATION
ApplicantCanadian Technical Tape, Ltd.
Product CodeJOJ · General Hospital
Decision DateOct 27, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

A physical/chemical sterilization process indicator is a single use device intended to be used by a health care provider to distinguish between sterilization processed and unprocessed units.

Device Story

Saturated crepe paper tape with printed chemical indicator lines; used by healthcare providers to secure sterilization packs. Tape adheres to packaging; exposed to steam sterilization cycles. Indicator agent (sulfur, lead carbonate hydroxide, magnesium oxide) undergoes color change from white to dark brown/black upon exposure to specific temperature and moisture conditions. Provides visual confirmation to staff that items have undergone a sterilization process. Benefits include simple, reliable identification of processed versus unprocessed medical equipment.

Clinical Evidence

Bench testing only. Testing included performance per ANSI/AAMI/ISO 11140-1:2005(R)2010, resistometer performance, biocompatibility/leach-off, endpoint stability, shelf-life, and tape adhesion (PSTC-101/131). All tests passed.

Technological Characteristics

Saturated crepe paper tape; indicator agent contains sulfur, lead carbonate hydroxide, and magnesium oxide. Dimensions: 0.5", 0.75", 1" widths. Complies with ANSI/AAMI/ISO 11140-1:2005(R)2010. Passive chemical indicator; no energy source or software.

Indications for Use

Indicated for holding sterilization packs together and as a process indicator in gravity sterilizers (121ºC for 30 minutes) or pre-vacuum sterilizers (132ºC for 4 minutes and 135ºC for 3 minutes).

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top of the seal in a circular fashion. The seal is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 27, 2014 Canadian Technical Tape, LTD C/O Mr. Gary Socola Official Correspondent President Highpower Validation Testing and Lab Services 125 Highpower Road Rochester, NY 14623 Re: K140940 Trade/Device Name: Process Indicator Tape for Steam Sterilization Regulation Number: 21 CFR 880.2800 Regulation Name: Physical/Chemical Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: September 19, 2014 Received: September 23, 2014 Dear Mr. Socola: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {2}------------------------------------------------ comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Susan Runno DDS, MA Erin I. Keith, M.S Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K140940 Device Name Process Indicator Tape for Steam Sterilization #### Indications for Use (Describe) The Process Indicator Tape for Steam Sterilization is indicated for use in holding sterilization packs together and can be used in gravity sterilizers operating at 121ºC for 30 minutes or pre-vacuum sterilizers operating at 132ºC for 4 minutes and 1359C for 3 minutes. The indicator stripes turn dark brown black when sterilization conditions, thus providing an indication of processed items. Type of Use (Select one or both, as applicable) - Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 12140940 #### 510K SUMMARY 3.0 | Submission Date: | April 8, 2014 | |------------------------|-------------------------------------------------------------------------------------------------------------------| | Summary Date: | October 15th, 2014 | | Submitter Information: | | | Company Name: | Canadian Technical Tape, Ltd. | | Company Address: | Canadian Technical Tape, Ltd.<br>455 Côte Vertu<br>Montreal, QC, Canada H4N 1E8 | | Contact Person: | Howard Cohen<br>President<br>Canadian Technical Tape, Ltd.<br>514-334-1510<br>514-745-0764<br>hcohen@cttgroup.com | | Device Information: | | | Trade Name: | Process Indicator Tape for Steam Sterilization | | Common Name: | Process Indicator Tape | | Classification Name: | Physical/Chemical Sterilization Process Indicator | | Product Code: | JOJ | | Device Class: | Class II, 21 CFR 880.2800(b) | | Predicate Devices: | Sterilization Process Indicator Tapes Model CI122 &<br>CI123 (K001649) SteriTec Products, Incorporated | #### Device Description: The process indicator tape distinguishes between items processed and unprocessed in both gravity discharge and pre-vacuum steam sterilization cycles. The tape is made of a saturated crepe paper printed with white indicator lines that turn to dark brown/black when proper levels of moisture and temperature have been achieved. The tape adheres on contact and stays in place through live steam pressure. #### Intended Use: A physical/chemical sterilization process indicator is a single use device intended to be used by a health care provider to distinguish between sterilization processed and unprocessed units. {5}------------------------------------------------ #### Indications for Use: The Process Indicator Tape for Steam Sterilization is indicated for use in holding sterilization packs together and can be used in gravity sterilizers operating at 121°C for 30 minutes or prevacuum sterilizers operating at 132°C for 4 minutes and 135°C for 3 minutes. The indicator stripes turn dark brown/black when exposed to steam sterilization conditions, thus providing an indication of processed items. ### Performance Standard Testing: Testing was performed in accordance with ANSI/AAMI/ISO 11140-1:2005(R) 2010 -Sterilization of health care products - Chemical indicators - Part 1: General requirements and the Guidance for Industry and FDA Staff Premarket Notification [510(k)] Submissions for Chemical Indicators. | Characteristic | Predicate Device: | Proposed Device: | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Process indicator tape for steam<br>sterilization | Process indicator tape for steam<br>sterilization | | Device Design | Crepe paper printed with indicator<br>lines. Provided in natural and blue in<br>widths of approximately 0.5", 0.75"<br>and 1" (12mm, 18mm and 24mm). | Crepe paper printed with indicator<br>lines. Provided in natural and blue in<br>widths of approximately 0.5", 0.75"<br>and 1" (12mm, 18mm and 24mm). | | Indicator Agent | Sulfur, lead carbonate hydroxide<br>and magnesium oxide | Sulfur, lead carbonate hydroxide<br>and magnesium oxide | | Sterilization<br>Method | The tape is intended for use as a steam<br>sterilization cycle process indicator in<br>gravity discharge and pre-vacuum<br>steam sterilizers. | The tape is intended for use as a steam<br>sterilization cycle process indicator in<br>gravity discharge and pre-vacuum<br>steam sterilizers. | | Endpoint<br>Specifications | 121° C for 10 minutes<br>132-135° C for 2 minutes. | 121° C for 10 minutes<br>132-135° C for 2 minutes. | | Shelf-life | 3 years | 3 years | | Indications for use | SteriTec Autoclave and Blue<br>Autoclave Tapes are for use holding<br>sterilization packs together and as<br>process indicators in gravity steam<br>sterilizers operating at 121° C and pre-<br>vacuum steam sterilizers operating at<br>132-134° C. The white indicator<br>stripes on the tape turn black when<br>indication of processed items. | The Process Indicator Tape for Steam<br>Sterilization is indicated for use in<br>holding sterilization packs together<br>and can be used in gravity sterilizers<br>operating at 121°C for 30 minutes or<br>pre-vacuum sterilizers operating at<br>132°C for 4 minutes and 135°C for 4<br>minutes. The indicator stripes turn<br>dark brown/black when exposed to<br>steam sterilization conditions, thus,<br>providing an indication of processed<br>items. | | Performance | ANSI/AAMI ST-60 | ISO 11140-1:2005, ANSI/AAMI/ISO | | Standards | | 11140-1:2005(R)2010 | ## Comparison of the Proposed Device to the Predicate Device: {6}------------------------------------------------ The proposed device is an identical product to the predicate device and both are manufactured by Canadian Technical Tape, Ltd. ## Summary of Nonclinical Testing: | Test | Results | |----------------------------------------------------------------------------------------------------------------------|---------| | ANSI/AAMI/ISO 11140-1 - Performance Testing for a Class 1 Steam<br>Process Indicator | Passed | | FDA Chemical Indicator Guidance Document - Resistometer Performance<br>Testing for a Class 1 Steam Process Indicator | Passed | | In Use Testing - Steam Gravity and Pre-Vacuum 510 K Cleared Sterilizers | Passed | | Biocompatibility/Leach Off Testing | Passed | | Endpoint Stability | Passed | | Shelf Life | Passed | | PSTC-101 and PSTC-131 - Pressure Sensitive Tape Council (PSTC) | Passed | | International Standards Test for Tape Adhesion | | | Post Processing Visual Adhesive Test for Wrapped Packages | Passed | ### Conclusion: The Process Indicator Tape for Steam Sterilization is substantially equivalent to the predicate device. The tape meets the performance claims for a class 1 process indicator according to ANSI/AAMI/ISO 11140-1:2005 (R) 2010 and raises no issues related to safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...