K130211 · Spsmedical Supply Corp. · JOJ · Jul 5, 2013 · General Hospital
Device Facts
Record ID
K130211
Device Name
AIRVIEW BOWIE DICK; TEST PACK, INDICATOR SHEETS
Applicant
Spsmedical Supply Corp.
Product Code
JOJ · General Hospital
Decision Date
Jul 5, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are designed to detect the presence of residual air in pre-vacuum steam sterilizers. When tested in a prevacuum sterilizer operating at 134°C the indicator will demonstrate a uniform color change from cream or blue to dark brown/black when proper sterilization conditions have been met and no air is detected. It is designed to be used for daily Bowie Dick testing of steam sterilizers as described in ANSVAAMI ST79.
Device Story
Disposable test pack and indicator sheets used to monitor air removal in pre-vacuum steam sterilizers; operates by trapping air within paper layers to challenge sterilizer performance; indicator ink changes from cream/blue to uniform dark brown/black upon successful steam penetration and air removal; non-uniform color change indicates failure (residual air present); used by sterile processing staff in clinical environments; provides visual confirmation of sterilization conditions; ensures patient safety by verifying equipment efficacy before use on surgical instruments.
Clinical Evidence
Bench testing only. Validation studies confirmed the indicator's sensitivity to detect residual air sufficient to create a 2°C or greater temperature difference in the test pack compared to the sterilizer chamber during 134°C pre-vacuum cycles. Biocompatibility testing confirmed non-toxicity.
Technological Characteristics
Consists of paper sheets, foam, containment box/wrap, and chemical indicator ink. Sensing principle: chemical color change triggered by steam penetration and air removal. Dimensions/form factor: test pack. Sterilization method: steam pre-vacuum. Shelf life: 3 years. No software or electronic components.
Indications for Use
Indicated for daily monitoring of air removal in pre-vacuum steam sterilizers operating at 134°C to detect residual air. Intended for use in clinical settings following ANSI/AAMI ST79 guidelines.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets (K041017)
Related Devices
K041017 — SPSMEDICAL AIRVIEW BOWIE DICK TEST PACK · Spsmedical Supply Corp. · Jun 25, 2004
K990246 — STERI-PAK LF BOWIE-DICK TEST PACK · Steritec Products Mfg Co, Inc. · Jun 10, 1999
K093199 — 3M COMPLY BOWIE- DICK TYPE LEAD FREE TEST PACK · 3M Company · Dec 4, 2009
K133204 — AIRVIEW MINI OWIE DICK TEST PACK · Crosstex International · May 5, 2014
K120592 — DANA REUSABLE BOWIE TEST PACK FOR DANA BOWIE DICK INDICATORS · Dana Products, Inc. · Feb 27, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
# KISOZIT
# 510(k) Summary
## SUBMITTER INFORMATION:
JUL 5 2013
SPSmedical Supply Corp. a division of Crosstex International 6789 West Henrietta Road Rush, NY 14543 U.S.A. Phone: (800) 722-1529 Fax: (585) 359-0167
| Contact Person: | Michael Nolan<br>Research and Development Coordinator |
|------------------------|---------------------------------------------------------------------------|
| Date of Summary: | July 5, 2013 |
| DEVICE INFORMATION: | |
| Device Trade Name: | AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator<br>Sheets |
| Common Name: | Bowie Dick Test Pack and Bowie Dick Indicator Sheets |
| Device Classification: | Indicator, Physical/Chemical Sterilization Process<br>21 CFR § 880.2800 |
| Device Class: | 2 |
| Product Code: | JOJ |
## PREDICATE DEVICE:
AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets K041017
### DEVICE DESCRIPTION:
The SPSmedical AirView™ Bowie Dick Test Pack is a disposable device which is intended for use as a replacement for the ANSI/AAMI ST79 Bowie Dick Test Pack. The SPSmedical AirView™ Bowie Dick Indicator Sheets are designed to be used within the ANSI/AAMI ST79 Bowie Dick test pack
#### INTENDED USE:
The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are designed to detect the presence of residual air in pre-vacuum steam sterilizers operating at 134°C. The Bowie Dick test pack and indicator sheets are reliable tools used for the monitoring of air removal in steam pre-vacuum processes and provides a visual indication if residual air was left in the chamber during sterilization. Its internal indicator sheet changes to a uniform dark brown/black signal color under proper sterilization and air removal conditions. A failure would result in a nonuniform color change on the indicator sheet.
#### TECHNICAL CHARACTERISTICS:
The layers of paper within the SPSmedical AirView™ Bowie Dick Test Pack provide resistance to steam penetration and also trap air between them which is difficult for marginal performing prevacuum sterilizers to remove.
### RECOMMENDED STORAGE CONDITIONS:
Store in a cool, dry place (15-30°C)
{1}------------------------------------------------
# 510(k) Summary (Continued)
#### INTERFERING SUBSTANCES OR CONDITIONS:
Testing verified that the indicators in their unprocessed form are not sensitive to an acidic or basic environment. Testing verified that the indicators in their processed form are not sensitive to an acidic or basic environment.
#### SHELF LIFE:
The Shelf Life of the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets shall be three (3) years from the date of manufacture, when stored in a cool, dry place (15-30°C)
#### BIOCOMPATIBILITY:
The purpose of a biocompatibility test is to demonstrate that the active ingredients of the chemical indicator do not release any substance known to be toxic onto surgical instruments. Although the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are intended to be run in an empty chamber it was necessary to address this concern. Biocompatibility testing on the indicators concluded that the SPSmedical AirView™M Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets were non-toxic.
#### SUBSTANTIAL EQUIVALENCE ANALYSIS
SPSmedical has identified the AirView™ Bowje Dick Test Pack & AirView™ Bowie Dick Indicator Sheets (K041017) as the (primary) predicate device. We believe the predicate device to be substantially equivalent to the AirView™ Bowie Dick Test Pack & AirView™ Bowie Dick Indicator Sheets which is the subject of this submission in terms of their intended use and functional characteristics in determining the effectiveness of air removal in pre-vacuum steam sterilizers.
#### FUNCTIONAL CHARACTERISTICS:
The SPSmedical AirView™ Bowie Dick Test Pack and Air View™ Bowie Dick Indicator Sheets are designed to detect the presence of residual air in pre-vacuum steam sterilizers operating at 134°C. The layers of paper within the pack provide resistance to steam penetration and trap residual air which is difficult for marginal performing pre-vacuum sterilizers to remove. The pack is a reliable tool used for the monitoring of air removal in pre-vacuum steam sterilizers and provides a visual indication if residual air was left in the chamber during sterilization. The indicator sheet changes from its initial color to a uniform dark brown/black signal color under proper sterilization and air removal conditions. A failure would result in a non-uniform color change on the indicator sheet.
#### DISCUSSION:
SPSmedical is claiming substantial equivalence for its SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets to the original AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets (K041017) based on test data obtained during validation studies. We have demonstrated with testing that the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets perform consistently with results which indicate that the indicator is sensitive enough to detect when enough air is left within the sterilizer chamber to create a 2°C or greater temperature difference in the test pack as compared to the sterilizer chamber when running pre-vacuum sterilization at 134°C. Under these conditions, the indicator sheet would demonstrate a non-uniform color change. The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are also comparable to other commercially available Bowie Dick test packs cleared by the FDA. See examples of similarly cleared devices in Tab 8.
{2}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE CONCLUSIONS:
The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets have the same intended use and characteristics as the primary predicate device (K041017). They both provide a visual indication that proper air removal and steam penetration conditions have been met within the sterilizer's chamber. Both products are comprised of paper sheets, foam, a containment box or wrap, external chemical indicator label and printed indicator sheet.
SPSmedical believes that the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are substantially equivalent to the predicate device because it has the same intended use, technical characteristics and performance. Because the ability to perform its intended function as been shown through validated testing, the SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets raise no issues related to safety or effectiveness and therefore, the device should be allowed for market in the United states. See Table 1 below for a substantial equivalence comparison.
| Element | Modified Device | Predicate<br>(K041017) |
|------------------------------------------------|-------------------------------|-------------------------------|
| Intended Use | Air Removal Indicator | Air Removal Indicator |
| Device Design | Bowie Dick Test Pack or Sheet | Bowie Dick Test Pack or Sheet |
| Endpoint Color | Dark Brown/Black | Dark Brown/Black |
| Indicator Agent | Indicator Ink | Indicator Ink |
| Sterilization Method | Steam Pre-vacuum | Steam Pre-vacuum |
| Device Materials | Equivalent | Equivalent |
| Air Porosity | Equivalent | Equivalent |
| Performance under<br>standard fault condition | Equivalent | Equivalent |
| Performance under<br>standard Bowie Dick cycle | Equivalent | Equivalent |
| Shelf Life | Three (3) years | Three (3) years |
#### TABLE 1-COMPARISON OF THE MODIFIED DEVICE TO THE PREDICATE
#### CONCLUSION:
Supportive data has demonstrated that the SPSmedical AirView™ Bowie Dick Test Pack and AirView™M Bowie Dick Indicator Sheets are equivalent to the predicate device. The SPSmedical AirView™ Bowie Dick Test pack and AirView™ Bowie Dick Indicator Sheet raise no issues related to safety or effectiveness and therefore should be allowed for market in the United States.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three wavy lines extending from a central point. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle, following its curvature.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## July 5, 2013
SPS Medical Supply Corporation C/O Mr. Michael Nolan Research and Development Coordinator 6789 West Henrietta Road RUSH, NY 14543
Re: K130211
Trade/Device Name: AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: May 30, 2013 Received: June 7, 2013
#### Dear Mr. Nolan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Nolan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Image /page/4/Picture/8 description: The image shows a signature and some text. The text includes the name "Tefashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director". The image also contains the acronym "DAGRID".
Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health'
Enclosure
{5}------------------------------------------------
# Statement of Indications for Use
5 10(k) Number (if known): K130211
Device Trade Name:
AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets
#### INDICATIONS FOR USE:
The SPSmedical AirView™ Bowie Dick Test Pack and AirView™ Bowie Dick Indicator Sheets are designed to detect the presence of residual air in pre-vacuum steam sterilizers. When tested in a prevacuum sterilizer operating at 134°C the indicator will demonstrate a uniform color change from cream or blue to dark brown/black when proper sterilization conditions have been met and no air is detected. It is designed to be used for daily Bowie Dick testing of steam sterilizers as described in ANSVAAMI ST79.
Prescription Use __ (Part 21 CFR 801 Subpart D) and/or
Over the Counter Use X (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
Elizabeth F. Clavette
2013.06.28 21:36:53 04'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number;
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.