VERIFY 270FP CHALLENGE PACK

K103053 · STERIS Corporation · JOJ · Mar 10, 2011 · General Hospital

Device Facts

Record IDK103053
Device NameVERIFY 270FP CHALLENGE PACK
ApplicantSTERIS Corporation
Product CodeJOJ · General Hospital
Decision DateMar 10, 2011
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The Verify® 270FP Challenge Pack is a test pack consisting of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier, intended for use in SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization. The Verify® 270FP Challenge Pack indicators change color from yellow to blue/purple when exposed to the following conditions at 270°F (132°C): - 4 minute indicator SFPP and pre-vacuum steam sterilization for 4 minutes. - 10 minute indicator Pre-vacuum steam sterilization for 10 minutes. - 20 minute indicator Pre-vacuum steam sterilization for 20 minutes.

Device Story

Verify® 270FP Challenge Pack is a single-use chemical indicator test pack for steam sterilization monitoring. Device consists of three emulating indicator inks on a test sheet enclosed within a steam penetration barrier. Inks change color from yellow to blue/purple upon exposure to saturated steam at 270°F (132°C) for specific durations (4, 10, or 20 minutes). An external process indicator changes from pink to dark purple when exposed to steam between 250°F and 275°F. Used in clinical settings to verify sterilization cycle efficacy. Healthcare personnel observe color changes to confirm sterilization parameters were met. Benefits include assurance of steam penetration and cycle completion, reducing risk of using improperly sterilized instruments.

Clinical Evidence

Bench testing only. Performance verified using a resistometer per ANSI/AAMI/ISO 18472 to meet Class 6 indicator requirements. Efficacy of barrier material demonstrated via 16-towel test pack comparison using biological indicators.

Technological Characteristics

Single-use chemical indicator test pack. Emulating indicator inks (Class 6 per ANSI/AAMI ISO 11140-1:2005). Steam penetration barrier material. Passive colorimetric sensing principle. No energy source or software.

Indications for Use

Indicated for monitoring SFPP and pre-vacuum steam sterilization cycles at 270°F (132°C) using three specific emulating indicator inks (4, 10, and 20-minute durations).

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ STERIS® MAR | Q 2011 # 510(k) Summary For Verify® 270FP Challenge Pack STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459 Contact: Robert Sullivan Senior Director, Regulatory Affairs Telephone: (440) 392-7695 Fax No: (440) 357-9198 Submission Date: March 10, 2011 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {1}------------------------------------------------ K103053 p² of 3 #### 1. Device Name | Indicator Pack Model: | Verify® 270FP Challenge Pack | |-----------------------|-----------------------------------------------------------------------------------------------| | Common Name: | Chemical Indicator | | Classification Name: | Physical/chemical sterilization process indicator (21<br>CFR 880.2800 (b), Product Code JOJ). | #### Predicate Device 2. - STERIS Verify® Challenge Pack (Model 270F 4) K073683 . - Getinge ChemSix Test Pack K080136 . #### 3. Device Description The proposed Verify® 270FP Challenge Pack consists of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier. The indicator inks on the proposed Verify® 270FP Challenge Pack test sheet change from yellow to blue/purple color when exposed to saturated steam at 270°F (132°C) for the following times. - 4 minute indicator The 4 minute indicator ink on the Verify® 270FP . Challenge Pack can be used to monitor a 4 minute SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization cycle. - 10 minute indicator The 10 minute indicator ink on the Verify 270FP . Challenge Pack can be used to monitor a 10 minute pre-vacuum steam sterilization cycle. - 20 minute indicator The 20 minute indicator ink on the Verify® 270FP . Challenge Pack can be used to monitor a 20 minute pre-vacuum steam sterilization cycle. The process indicator outside of the packs undergoes a visual color change from pink to dark purple when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C). #### 4. Intended Use The Verify® 270FP Challenge Pack is a test pack consisting of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier, {2}------------------------------------------------ K103053 ### K103053/S001 STERIS Response to 3/10/11 Request for Clarification Verify 270FP Challenge Pack intended for use in SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization. The Verify® 270FP Challenge Pack indicators change color from yellow to blue/purple when exposed to the following conditions at 270°F (132°C): - 4 minute indicator SFPP and pre-vacuum steam sterilization for 4 minutes. . - 10 minute indicator Pre-vacuum steam sterilization for 10 minutes. . - 20 minute indicator Pre-vacuum steam sterilization for 20 minutes. . #### રાં Description of Safety and Substantial Equivalence The proposed and predicate devices are all single use indicator test packs for use in steam sterilization cycles. The differences between the proposed Verify® 270FP Challenge Pack and predicate devices are limited to differences in design, materials and parameters of the sterilization cycles these indicator test packs are designed to monitor. These differences do not raise any new issues of safety or efficacy. #### Performance Testing 6. Performance testing was conducted to verify that the proposed emulating indicator inks within the test pack meet the requirements for Class 6 indicators as defined in ANSI/AAMI ISO 11140-1:2005 using a resistometer to ANSI/AAMI/ISO 18472. Further testing was conducted to demonstrate the efficacy of the barrier material using a 16-towel test pack with biological indicators as a reference. The predicate test packs were included in this testing. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Robert Sullivan Senior Director, Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060 MAR 1 Q 2011 Re: K103053 Trade/Device Name: Verify® 270FP Challenge Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: February 4, 2011 Received: February 7, 2011 Dear Mr. Sullivan; We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class.III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Sullivan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vou Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: Verify® 270FP Challenge Pack Indications For Use: The Verify® 270FP Challenge Pack is a test pack consisting of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier, intended for use in SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization. The Verify® 270FP Challenge Pack indicators change color from yellow to blue/purple when exposed to the following conditions at 270°F (132°C): - 4 minute indicator SFPP and pre-vacuum steam sterilization for 4 minutes. . - 10 minute indicator Pre-vacuum steam sterilization for 10 minutes. . - 20 minute indicator Pre-vacuum steam sterilization for 20 minutes. ● Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device:Evaluation (ODE) Page 1 of 1 Eldith S. (Clavis. William (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control and Dental Devices, 510(k) Number: K103053
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...