As a Sterilizer Monitoring Device, the 3M™ Attest™ eBowie-Dick Test Card 10135 and 3M™ Attest™ eBowie-Dick Card Holder 10135CH is used in conjunction with the 3M™ Attest™ eBowie-Dick Auto-reader 1190 to monitor the equipment performance of steam sterilizers each day, as well as after a steam sterilizer repair. The 3M™ Attest™ eBowie-Dick Test Card 10135 will detect steam penetration as a means of measuring air removal efficiency. The 3M™ Attest™ eBowie-Dick Test Card 10135 and 3M™ Attest™ eBowie-Dick Card Holder 10135CH is designed for testing air removal efficiency of 132°C (270°F), 134°C (273°F), and 135°C (275°F) for 3.5 minute test cycles in dynamic-air-removal steam sterilizers with chamber volumes greater than 54L (approximately 2 cubic feet). The 3M™ Attest™ eBowie-Dick Auto-reader is designed to automatically interpret, store, and display the result of the 3M™ Attest™ eBowie-Dick Test Card.
Device Story
System monitors steam sterilizer air removal efficiency; comprises test card, card holder, and auto-reader. Test card detects steam penetration; auto-reader automatically interprets, stores, and displays results. Used in clinical settings to verify sterilizer performance; daily or post-repair. Provides objective digital readout to replace manual interpretation; reduces risk of false results; ensures proper sterilization of instruments.
Clinical Evidence
No clinical data. Device classification supported by non-clinical performance testing, including sensitivity, specificity, accuracy, precision, range, and end-point stability testing, alongside software verification/validation and electrical/EMC safety testing.
Technological Characteristics
Digital physical/chemical sterilization process sensor. System includes test card, holder, and electronic auto-reader. Features automated interpretation of chemical indicator results. Requires electrical safety and EMC compliance. Software-based analysis of sterilization parameters.
Indications for Use
Indicated for monitoring steam sterilizer equipment performance (air removal efficiency) in dynamic-air-removal steam sterilizers (>54L) at 132-135°C. Intended for daily use and post-repair testing. Marketed for Over-The-Counter (OTC) use.
Regulatory Classification
Identification
A digital physical/chemical sterilization process sensor is a device intended to monitor one or more parameter(s) of the sterilization process. The adequacy of the sterilization conditions as measured by the parameter(s) is/are indicated by a digital readout of the result. The 3M™ Attest™ eBowie-Dick Test system (consisting of a Test Card, Card Holder, and Auto-reader) is used to monitor the equipment performance of steam sterilizers by detecting steam penetration as a means of measuring air removal efficiency in dynamic-air-removal steam sterilizers.
Special Controls
In combination with the general controls of the FD&C Act, the digital physical/chemical sterilization process sensor is subject to the following special controls:
- (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must fulfill the following:
- Testing must compare performance with a scientifically justified comparator; (i)
- (ii) Testing must demonstrate that the measured response is dependent on the sterilization cycle parameter(s);
- Testing must identify relevant validation parameters (analytical performance (iii) characteristics) and include scientifically justified samples, conditions, and data analysis to evaluate said parameters with established acceptance criteria for each. parameter:
- (iv) Testing must evaluate sensitivity and specificity to determine conditions for passing and failing results:
- Testing must determine and validate accuracy, precision, and range of the (v) measuring device:
- Testing must evaluate end point stability; and (vi)
- (vii) Testing must demonstrate that the reader maintains a defined range of accuracy over a defined time interval.
- (2) Performance data must support the shelf life of the device by demonstrating continued device functionality over the labeled shelf life under the proposed storage conditions. Transport stability testing must demonstrate device resilience to transport conditions.
- Software verification, validation, and hazard analysis must be performed for any (3) software components.
- (4) Performance data must demonstrate the electromagnetic compatibility (EMC) and electrical safety of the device.
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR 3MTM ATTESTTM EBOWIE-DICK TEST
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Digital physical/chemical sterilization process sensor. A digital physical/chemical sterilization process sensor is a device intended to monitor one or more parameter(s) of the sterilization process. The adequacy of the sterilization conditions as measured by the parameter(s) is/are indicated by a digital readout of the result.
NEW REGULATION NUMBER: 21 CFR 880.2801
CLASSIFICATION: Class II
PRODUCT CODE: SBE
#### BACKGROUND
DEVICE NAME: 3MIM Attest™ eBowie-Dick Test Card, 3M™ Attest™ eBowie-Dick Card Holder, 3MTM Attest™ eBowie-Dick Auto-reader
SUBMISSION NUMBER: DEN230068
DATE DE NOVO RECEIVED: September 29, 2023
#### SPONSOR INFORMATION:
3M Company 2510 Conway Ave. St. Paul. MN 55144 USA
## INDICATIONS FOR USE
## 3MTM Attest eBowie Dick Test Card and 3MTM Attest eBowie-Dick Card Holder
As a Sterilizer Monitoring Device, the 3M™ Attest™ eBowie-Dick Test Card 10135 and 3M[™ Attest™ eBowie-Dick Card Holder 10135CH is used in conjunction with the 3MTM Attest™ eBowie-Dick Auto-reader 1190 to monitor the equipment performance of steam sterilizers each. day, as well as after a steam sterilizer repair. The 3MTM Attest™ eBowie-Dick Test Card 10135 will detect steam penetration as a means of measuring air removal efficiency. The 3M™ Attest™ eBowie-Dick Test Card 10135 and 3MTM Attest™ eBowie-Dick Card Holder 10135CH is designed for testing air removal efficiency of 132°C (270°F). 134°C (273°F), and 135°C (275°F) for 3.5 minute test cycles in dynamic-air-removal steam sterilizers with chamber volumes greater than 54L (approximately 2 cubic feet).
{1}------------------------------------------------
## 3MTM Attest eBowie-Dick Auto-reader
The 3MTM Attest™ eBowie-Dick Auto-reader is designed to automatically interpret, store, and display the result of the 3MTM Attest™ eBowie-Dick Test Card.
## LIMITATIONS
The 3MTM Attest™ eBowie-Dick Test is intended to be marketed for Over-The-Counter (OTC) use.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS
## DEVICE DESCRIPTION
The 3M™ Attest™ eBowie-Dick Test Card consists of a laminated construction with a lead-free steam-sensitive chemical indicator positioned at the end of a lumen challenge. The 3MTM Attest™ eBowie-Dick Test Card undergoes a change in electrical resistance upon exposure to saturated steam. The test card includes a lumen challenge which is designed to restrict air removal and steam perfusion. At the completion of a Bowie-Dick test cycle, the resistance of the 3MTM Attest™ eBowie-Dick Test Card is read out using the 3MTM Attest™ eBowie-Dick Autoreader. The 3M™ Attest™ eBowie-Dick Test Card will reach a PASS value when adequate air removal and steam profusion is achieved. An air removal failure and or a steam penetration failure is indicated by a FAIL result.
The 3MTM Attest™ eBowie-Dick Card Holder is a reusable sterilizable plastic container for the Test Card which serves to support the laminated construction of the Test Card during the vacuum and pressure cycles of the air removal test, as well as anchor the Test Card in place inside the sterilization chamber through the increased mass of the Card Holder plus Test Card combination.
The 3M™ Attest™ eBowie-Dick Auto-reader is a wall powered electrical resistance measurement device which receives a processed Test Card via the card slot and contains pairs of internal mechanical contacts to probe the metallic contact pads on the Test Card. The Autoreader detects and confirms proper insertion of the Test Card, measures the PANI sensor electrical resistance, and returns a visible PASS, FAIL, or ERROR state indication to the user. The Auto-reader utilizes embedded firmware to detect and read the resistance of a Test Card. The device also includes USB or BLE connectivity for integrating with the optional 3M™ AttestTM Connect Software platform.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the 3M Attest system for Bowie-Dick testing. The system includes the 3M Attest Connect PC application, an eBowie-Dick test card, an eBowie-Dick card holder, and an eBowie-Dick auto-reader. The image also shows a close-up of the eBowie-Dick test card, highlighting the process indicator, steam inlet opening, top paper graphic layer, measurement contact pads, challenge channel, PANI coating, and electrode pads.
{3}------------------------------------------------
#### SUMMARY OF BENCH STUDIES
#### REPROCESSING, STERILITY AND SHELF-LIFE
The 3MTM Attest™ eBowie-Dick Test Card is single use and does not require reprocessing or sterilization. A shelf-life study and a hot and cold excursion stability study are provided to demonstrate that the product performance (e.g., Fail Cycle, Pass Cycle, and Process Indicator Dot Color) is maintained throughout the 6-month shelf life of the device.
3M™ Attest™ eBowie-Dick Auto-reader do not require reprocessing or sterilization. Cleaning instructions for the 3MTM Attest™ eBowie-Dick Auto-reader are included in the labeling.
The 3MTM Attest™ eBowie-Dick Card Holder is a device that does not have a designated shelf life due to its ability to be replaced. A re-use study is provided to support the re-use of the Card Holder for up to 500 uses in steam sterilization. The Instructions for Use has information pertaining to the maintenance of the Card Holder.
#### BIOCOMPATIBILITY
The 3MTM AttestTM eBowie-Dick Test does not have any direct tissue contacting components.
## SOFTWARE
The 3MTM Attest™ eBowie-Dick Auto-reader utilizes embedded firmware to detect and read the resistance of a 3MTM Attest™ eBowie-Dick Test Card. The embedded firmware stores measurements, which determine individual card results by comparing the measured resistance against a predefined threshold. Adequate software development, test plans, and documentation were demonstrated according to FDA guidance document, "Content of Premarket Submissions for Device Software Functions", Content of Premarket Submissions for Device Software Functions (fda.gov).
The hardware interfaces of the device provide users with access to the most recent 1000 records. They also allow remote firmware update. Additionally, they offer the ability to enable or disable an audible alarm and configure the date and time settings. These interfaces include:
- 1) CDC over USB connection.
- (b)(4) Bluetooth (b)(4) Module. 2)
The 3MTM Attest™ eBowie-Dick Auto-reader can be connected via USB and BLE connection to a Windows PC. This allows communications between the auto-reader and PC. where such communications are directed to a PC application that collects results and configuration. The name of the PC application is 3MTM Attest™ Connect. Attest Connect
{4}------------------------------------------------
can also facilitate firmware updates over the USB or BLE connection and enable or disable the audible alarm.
A backend server with a database, web services, and a registration portal developed and hosted in 3M's Azure Cloud infrastructure will be utilized as part of the Attest™ Connect app. This cloud server will be used to collect some customer information and data on cards read. It will also keep all the Auto-reader configurations from manufacturing and will be checked real time when card results are uploaded. With this check, 3M can monitor if any Attest™ Auto-reader configurations have changed. Additionally, the backend cloud server will facilitate application and firmware updates.
Adequate cybersecurity and software functionality were demonstrated according to FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".
3M controls deployment of software updates as per change control and procedures within the Software Development Plan. All potential firmware and software updates are reviewed against the most recent versions of the SRS.
## ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY
EMC Immunity testing was conducted on the Auto-reader using consensus standard 60601-1-2 and includes testing to IEC 61000-4-8 with levels directed by IEC 60601-1- 2. Electrical Safety testing was conducted per IEC 61010-1.
# HUMAN FACTORS
The 3MTM Attest™ eBowie-Dick Test System was evaluated via simulated-use testing with users representative of the use specification in accordance with the guidance "Applying Human Factors and Usability Engineering to Medical Devices." This study included identification of known risks for commercially available similar devices, analysis of similarities and differences between potential use errors, risks, and harms, and evaluation of the risk mitigations and residual risks. No use errors, close calls or difficulties were observed during the use of the 3MTM Attest™ Electronic Bowie-Dick (eBD) System for Critical Task related Use Scenarios. In conclusion, the results demonstrate that the 3MTM Attest™ eBowie-Dick Test System, with the associated risk control measures, are safe and effective, and no design or labeling modifications are necessary.
# PERFORMANCE TESTING - BENCH
The sponsor conducted the following performance tests to support that the device can achieve its intended use:
- . Comparison of air pocket size and position in a reference test pack to the sensor size
{5}------------------------------------------------
and location in eBD Test card for an inadequate cycle to demonstrate comparable pass/fail testing performance to the test pack.
- . Comparison of air pocket size and position in a reference test pack to the sensor size and location in eBD Test card to determine the impact of the presence of noncondensable gasses.
- . Characterization of air removal from eBD Test card lumen construction and demonstration of equivalence to reference test pack.
- Testing of the sensor in the standard test pack in adequate and inadequate cycles to . show expected pass/fail results.
- . Quantifying the relationship between the resistance and the color of the sensor to show that they are related to the oxidation state of the polymer.
- . Sensitivity and specificity testing as a function of changes in the vacuum level and pressure reached during the conditioning phase of a dynamic air removal cycle to support low occurrences of false positive and/or false negative results.
- . Determination of equivalent performance to the Standard Test Pack as defined in ISO 11140:2007-5.
- Determination of accuracy, precision, and dynamic range of the 3MTM AttestTM . eBowie-Dick Auto-reader 1190 for measuring resistance.
- Testing showing endpoint stability for Adequate and Inadequate Cycles as a function . of Dwell Time in the sterilizer at the completion of a test cycle as well as a function of cooling time (referred to as Measurement Time in the report) once the device is removed from the sterilizer and before reading the test result.
- . Independent verification of Auto-reader accuracy is completed during manufacturing final functional testing using external reference cards with verified resistance values.
# LABELING
The labeling consists of a user manual, instructions for use, and packaging labels.
# RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of a digital. physical/chemical sterilization process indicator and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Infection resulting from exposure to<br>unsterile instruments due to false<br>passing results or failure to properly<br>monitor sterilizer performance | Non-clinical performance testing<br>Software verification, validation, and hazard analysis |
| Delayed or cancelled procedure due<br>to false positive results | Non-clinical performance testing<br>Software verification, validation, and hazard analysis |
| Electrical shock | Electrical safety testing<br>Non-clinical performance testing |
## Identified Risks to Health and Mitigation Measures
{6}------------------------------------------------
| Risks to Health | Mitigation Measures |
|---------------------------------|--------------------------------------------------------------------|
| Interference with other devices | Electromagnetic compatibility testing<br>Electrical safety testing |
## SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the digital physical/chemical sterilization process sensor is subject to the following special controls:
- (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must fulfill the following:
- Testing must compare performance with a scientifically justified comparator; (i)
- (ii) Testing must demonstrate that the measured response is dependent on the sterilization cycle parameter(s);
- Testing must identify relevant validation parameters (analytical performance (iii) characteristics) and include scientifically justified samples, conditions, and data analysis to evaluate said parameters with established acceptance criteria for each. parameter:
- (iv) Testing must evaluate sensitivity and specificity to determine conditions for passing and failing results:
- Testing must determine and validate accuracy, precision, and range of the (v) measuring device:
- Testing must evaluate end point stability; and (vi)
- (vii) Testing must demonstrate that the reader maintains a defined range of accuracy over a defined time interval.
- (2) Performance data must support the shelf life of the device by demonstrating continued device functionality over the labeled shelf life under the proposed storage conditions. Transport stability testing must demonstrate device resilience to transport conditions.
- Software verification, validation, and hazard analysis must be performed for any (3) software components.
- (4) Performance data must demonstrate the electromagnetic compatibility (EMC) and electrical safety of the device.
## BENEFIT-RISK DETERMINATION
# Risks:
The risks of the device are based on nonclinical laboratory testing described above.
The main probable risk of the proposed device is false results which may lead to ineffective sterilization of medical devices with a false negative result or may lead to re-running the Bowie-Dick cycle with a false positive result that causes a delayed procedure or procedures. There is an additional risk of electrical shock and interference with other devices. Both are mitigated with the performance testing, warnings, precautions, and special controls.
{7}------------------------------------------------
# Benefits:
The probable benefits of the device are based on non-clinical laboratory testing described above.
The probable benefits of the subject device, include:
1 ) increases testing sensitivity with a digital measurement
2) reduces false results and repeat testing with an automatic reader
3) general patient benefit with improved sterilization process monitoring that reduces the risk of infection from improperly sterilized devices.
## Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
## Benefit/Risk Conclusion
Based on the performance testing provided (in particular; sensitivity, end point stability, shelf life, software validation, labeling, and human factors study), the sponsor has adequately supported that the subject device is a digital physical/chemical sterilization process sensor.
In conclusion, given the testing provided above, for the following indication statement:
As a Sterilizer Monitoring Device, the 3MTM Attest™ eBowie-Dick Test Card and 3M™ Attest eBowie-Dick Card Holder are used in conjunction with the 3MTM Attest™ eBowie-Dick Auto-reader to monitor the equipment performance of steam sterilizers each day, as well as after a steam sterilizer repair. The 3M™ Attest™ eBowie-Dick Test Card will detect steam penetration as a means of measuring air removal efficiency. The 3MTM Attest™ eBowie-Dick Test Card and 3M™ Attest™ eBowie-Dick Card Holder is designed for testing air removal efficiency of 132-135°C (270-275°F) dynamic-airremoval steam sterilizers with chamber volumes greater than 54L (approximately 2 cubic feet).
The 3M™ Attest™ eBowie-Dick Auto-reader is designed to automatically interpret, store, and display the result of the 3MTM Attest™ eBowie-Dick Test Card.
The probable benefits outweigh the probable risks for the 3M™ Attest™ eBowie-Dick Test. The device provides benefits, and the risks can be mitigated using general controls and the identified special controls.
## CONCLUSION
The De Novo request for the 3M™ Attest™ eBowie-Dick Test is granted and the device is classified as follows:
Product Code: SBE
De Novo Summary (DEN230068)
{8}------------------------------------------------
Device Type: Digital physical/chemical sterilization process sensor Regulation Number: 21 CFR 880.2801 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.