← Product Code [FRC](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC) · K971432

# PROSPORE (K971432)

_Raven Biological Laboratories, Inc. · FRC · May 20, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K971432

## Device Facts

- **Applicant:** Raven Biological Laboratories, Inc.
- **Product Code:** [FRC](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC.md)
- **Decision Date:** May 20, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

ProSpore® is a biological steritization process indicator for steam sterilization at 121°C is a device intended for use by a health care provider to accompany products being sterilized through a steam sterilization procedure and to monitor adequacy of sterilization.

## Device Story

ProSpore is a self-contained biological indicator for monitoring 121°C steam sterilization efficacy. Device consists of a hermetically sealed glass ampoule containing Bacillus stearothermophilus spores and Tryptic Soy Broth with Bromocresol Purple pH indicator. Ampoule is placed in sterilization chamber with load; subjected to steam cycle; then removed and cooled. Processed ampoule and unprocessed control ampoule are incubated at 55-60°C for 48 hours. If spores survive, they germinate and consume nutrients, releasing nitrogen waste; this lowers media pH, causing color change from purple to yellow. Healthcare providers interpret results by comparing processed ampoule color/turbidity against control. Color change indicates sterilization failure; absence of change indicates successful sterilization.

## Clinical Evidence

Bench testing only. Testing validated labeled claims and performance characteristics using three separate lots. Parameters evaluated included stability of resistance characteristics, spore population, recovery of low numbers of injured spores, effect of pH indicator on recovery, and 48-hour incubation period. Effectiveness in monitoring routine steam sterilization cycles was demonstrated.

## Technological Characteristics

Self-contained biological indicator. Components: Bacillus stearothermophilus ATCC#7953 spores, Tryptic Soy Broth, Bromocresol Purple pH indicator. Form factor: Hermetically sealed glass ampoule. Principle: Biological growth detection via pH-mediated colorimetric change. Sterilization monitoring: 121°C steam cycle.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- Kilit™ (BBL) (M452998)
- SporView™ (SPSmedical) ([K905425](/device/K905425.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### Summary of "ProSpore Self-contained Biological Indicator" for steam sterilization at 121°C 510(k) # K971432

| Submitter:                            | Raven Biological Laboratories, Inc.                                |
|---------------------------------------|--------------------------------------------------------------------|
|                                       | 5017 Leavenworth Street<br>Omaha, Nebraska 68106                   |
|                                       | (402) 556-6690<br>(402) 556-4722 FAX                               |
| Contact:                              | Russ Nyberg<br>Production Microbiologist                           |
| Prepared on:                          | 2 April 1997                                                       |
| Device name:                          | ProSpore Self-contained Biological Indicator                       |
| Classification:                       | Class II medical device, General hospital                          |
| Predicate Devices (legally marketed): | Kilit™ (BBL)<br>SporView™ (SPSmedical)                             |
| Predicate Devices 510(k) numbers:     | Kilit™ (BBL) - Grandfather In*<br>SporView™ (SPSmedical) - K905425 |

### DESCRIPTION:

ProSpore is a self-contained biological indicator used for determining the efficacy of a 121°C sterilization cycle. The device is a hermetically sealed glass ampoule containing Bacillus stearothermophilus ATCC#7953 bacterial spores in a recovery medium of Tryptic Soy Broth with Bromocresol Purple pH indicator.

### OPERATIONAL PRINCIPLES

The ProSpore ampoule is placed with a load in the sterilization chamber and subjected to a normal steam sterilization cycle. The ampoule is then removed and cooled to room temperature. The processed ampoule and an unprocessed (control) ampoule are incubated at 55-60°C for 48 hours.

During incubation, the available food supply (Tryptic Soy Broth) and temperature promote growth of any viable spores. As viable spores germinate and consume the nutrients provided nitrogen waste products are released, increasing the acidity of the media which lowers the pH and causes the color to change from purple to yellow.

{1}------------------------------------------------

Evidence of growth by color change and/or turbidity within 48 hours may be interpreted Evidence of growth by Color change and on one of the sterilization, provided signs of growth are present in the control ampoule.

# STATEMENT OF SIMILARITY TO LEGALLY MARKETED PREDICATE DEVICE

ProSpore is similar in composition and function to the "Legally Marketed Predicate Devices" Kilit™ and SporView™.

- Intended for use in monitoring steam sterilization cycles at 121°C
- Intended for use in montoning other of B. stearothermophilus bacterial spore as its organism of choice for steam resistance characteristics
- Combine B. stearcthermophilus spores with a recovery media containing a pH indicator in a flame-sealed glass ampoule

# DESCRIPTION OF TESTING

Testing was performed to validate the labeled claims and performance characteristics for Prospore. These included using three separate lots to show stability of resistance characteristics and spore population, recovery of low numbers of injured spores, the effect of the pH indicator on recovery of spores and a reduced incubation period of 48 hours. For all lots tested, the above parameters and overall effectiveness in monitoring routine steam sterilization cycles has been demonstrated.

## STATEMENT OF SAFETY AND EFFECTIVENESS

Based on similar claims and design and the results from the above mentioned testing, the ProSpore Biological Indicator has been determined to be substantially equivalent to and therefore, as safe and effective as, the legally marketed devices Kilit™ and SporView™

*Kilit™ (BBL) does not have a 510(k) number. It was marketed under the grandfather clause, which exempts devices in commercial distribution prior to May 28, 1976. We have obtained the results of the Tradesmarkscan search of the U.S. Patent Office trademark filing showing that the product was introduced into commerce on June 8, 1955 (Attachment 10). The FDA number is M452998.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 0 1997

Mr. Kerrie Groff Quality Assurance Manager Raven Biological laboratories, Incorporated P.O. Box 6408 5017 Leavenworth Street Omaha, Nebraska 68106

K971432 Re : Prospore Trade Name: Requlatory Class: II Product Code: FRC Dated: April 15, 1997 April 17, 1997 Received:

Dear Mr. Groff:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with --the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic ... GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of

{3}------------------------------------------------

Page 2 - Mr. Groff

the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed i predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, uMisbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

L.J. Celtrist

hy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

| 510(k) Number (if known): |            |
|---------------------------|------------|
| Device Name:              | ProSpore @ |

Indications For Use:

ProSpore® is a biological steritization process indicator for steam sterilization at 121°C is a device intended for use by a health care provider to accompany products being sterilized through a steam sterilization procedure and to monitor adequacy of sterilization.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off)

Division of Dental, Infection Control, and General Hospital Devices 510(k) Number J 5

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use X

(Optional Formal 1-2-96)

3

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K971432](https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K971432)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
