← Product Code [FRC](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC) · K181434

# SporView Rapid Read Biological Indicator SteamPlus Test Pack (K181434)

_Crosstex/Spsmedical, A Division of Cantel Medical · FRC · Nov 14, 2018 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K181434

## Device Facts

- **Applicant:** Crosstex/Spsmedical, A Division of Cantel Medical
- **Product Code:** [FRC](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC.md)
- **Decision Date:** Nov 14, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 270°F (132°C) for 4 minutes exposure time.

## Device Story

Test pack monitors dynamic-air-removal steam sterilization cycles; consists of paper cards, chemical process indicators, and SporView Rapid Read Biological Indicator (BI). BI contains Geobacillus stearothermophilus spores in polypropylene vial with media ampoule. Used in clinical settings to verify sterilizer performance. After sterilization, BI is incubated at 60±2°C for 3 hours. Absence of growth indicates successful sterilization. Provides challenge to sterilization process equivalent to AAMI towel pack. Healthcare providers use results to confirm equipment readiness for patient use.

## Clinical Evidence

Bench testing only. Performance testing confirmed complete kill of G. stearothermophilus spores after 4-minute exposure at 132°C. Resistance of the test pack was verified to be greater than or equal to the AAMI towel pack. Chemical indicators were validated per FDA guidance.

## Technological Characteristics

Test pack composed of paper index cards, chemical process indicators, and a self-contained biological indicator (SCBI). SCBI uses G. stearothermophilus (ATCC 7953, >90% genetic similarity) with 1.0x10^5 spore population. D-value is 10.0 seconds at 132°C. Incubation at 60±2°C for 3 hours. No direct patient contact.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- 3M Attest 41382 Rapid Readout Steam-Plus Process Challenge Device ([K101910](/device/K101910.md))

## Reference Devices

- SporView Plus BI Test Pack ([K140620](/device/K140620.md))
- SporView Rapid Read Biological Indicator ([K172432](/device/K172432.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

November 14, 2018

Crosstex/SPSmedical, A Division of Cantel Medical Megan Skaar Regulatory Affairs Specialist 6789 West Henrietta Road Rush, New York 14543

Re: K181434

Trade/Device Name: SporView Rapid Read Biological Indicator SteamPlus Test Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: October 11, 2018 Received: October 15, 2018

Dear Megan Skaar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known)

K181434

Device Name

SporView Rapid Read Biological Indicator SteamPlus Test Pack

Indications for Use (Describe)

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 270°F (132°C) for 4 minutes exposure time.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

> Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "CANTEL" in blue font next to a blue circular logo. The logo is a circle with several horizontal lines running through it. The word "CANTEL" is in all capital letters and appears to be a sans-serif font. The image is likely a company logo.

## 510(k) Summary - K181434

| Manufacturer:        | Crosstex/SPSmedical, a Cantel Medical Company                                                                                                                                                   |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:             | 6789 W. Henrietta Road<br>Rush, NY 14543<br>(800) 722-1529                                                                                                                                      |
| Official Contact:    | Megan Skaar<br>Regulatory Affairs Specialist, Cantel Medical                                                                                                                                    |
| Date:                | 9 November 2018                                                                                                                                                                                 |
| Trade Name:          | SporView Rapid Read Biological Indicator SteamPlus Test Pack                                                                                                                                    |
| Common Name:         | Sterilization Biological Test Pack                                                                                                                                                              |
| Classification Name: | Sterilization Process Indicator                                                                                                                                                                 |
| Product Code:        | FRC                                                                                                                                                                                             |
| Device Class:        | II                                                                                                                                                                                              |
| Regulation No:       | Subject Device – SporView Rapid Read Biological Indicator Test Pack,<br>880.2800<br>Predicate Device – 3M Attest 3M Attest 41382 Rapid Readout Steam-Plus<br>Process Challenge Device, 880.2800 |

#### 1. Device Description

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 270°F (132°C) for 4 minutes exposure time.

The test pack consists of multiple layers of paper cards with a SporView Rapid Read Biological Indicator, chemical process indicators and load record card. The load record card is used to record the detailed information from the sterilization cycle.

The SporView Rapid Read Biological Indicator, cleared under 510(k) K172432, is composed of a polypropylene vial containing a spore carrier and media ampoule enclosed with a vented cap.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Cantel Medical. The logo consists of a blue circle with several horizontal lines running through it on the left, and the word "CANTEL" in blue on the right. The font is sans-serif and the overall design is simple and modern.

### 2. Indications for Use

The SporView Rapid Read Biological Indicator SteamPlus Test Pack is intended for routine monitoring and sterilizer qualification testing of dynamic-air-removal (pre-vacuum) steam sterilization cycles at 132°C for 4 minutes exposure time.

### 3. Technological Characteristics Comparison Table

Table 05.1 below provides a detailed comparison between the subject, predicate and reference devices –

| Technological<br>Characteristics | Subject Device -<br>SporView Rapid<br>Read Biological<br>Indicator Test<br>Pack                        | Predicate Device<br>- 3M Attest 41382<br>Rapid Readout<br>Steam-Plus<br>Process<br>Challenge Device<br>(K101910) | Reference Device<br>- SporView Plus<br>BI Test Pack<br>Process<br>Challenge Device<br>(K140620)                                                               | Comparison<br>Summary |
|----------------------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Intended Use                     | Monitor 270°F<br>(132°C)<br>prevacuum steam<br>sterilization cycles<br>for 4 minutes<br>exposure time. | Monitor 270°F<br>(132°C)<br>prevacuum steam<br>sterilization cycles<br>with exposure time<br>of 4 minutes.       | Monitor 250°F<br>(121°C) gravity<br>displacement<br>cycles and 270°F<br>(132°C)<br>prevacuum steam<br>sterilization cycles<br>for 4 minutes<br>exposure time. | Same                  |
| Product Code                     | FRC                                                                                                    | FRC                                                                                                              | FRC                                                                                                                                                           | Same                  |

#### Table 05.1 - Predicate Device Comparison Table

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for CANTEL. The logo consists of a blue circular graphic on the left and the word "CANTEL" in blue on the right. The circular graphic has several horizontal lines running through it, giving it a stylized appearance.

| FDA<br>Regulation                                  | 21 CFR 880.2800                                                                 | 21 CFR 880.2800                                                                 | 21 CFR 880.2800                                                                 | Same                           |
|----------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------|
| Device Class                                       | II                                                                              | II                                                                              | II                                                                              | Same                           |
| Organism                                           | > 90% genetic<br>similarity to G.<br>stearothermophilus<br>ATCC™ strain<br>7953 | > 90% genetic<br>similarity to G.<br>stearothermophilus<br>ATCC™ strain<br>7953 | > 90% genetic<br>similarity to G.<br>stearothermophilus<br>ATCC™ strain<br>7953 | Same                           |
| Viable Spore<br>Population                         | $1.0 x 10^5$                                                                    | $1.0 x 10^5$                                                                    | $1.0 x 10^5$                                                                    | Same                           |
| D-Value of<br>Biological<br>Indicator              | Steam 132°C:<br>10.0 Seconds                                                    | Steam 132°C:<br>10.0 Seconds                                                    | Steam 132°C:<br>10.0 Seconds                                                    | Same                           |
| Includes 510(k)<br>Cleared Chemical<br>Indicators? | Yes                                                                             | Yes                                                                             | Yes                                                                             | Same                           |
| Paper of Test<br>Pack                              | Paper Index Cards                                                               | Paper Index Cards                                                               | Paper Index Cards                                                               | Same                           |
| Shelf life                                         | 12 Months                                                                       | 24 Months                                                                       | 18 Months                                                                       | Similar                        |
| Storage                                            | Controlled room<br>temperature                                                  | Controlled room<br>temperature                                                  | Controlled room<br>temperature                                                  | Same                           |
| Incubation<br>Temperature                          | 60±2°C                                                                          | 60±2°C                                                                          | 55-60°C                                                                         | Same as<br>Predicate<br>Device |
| BI Incubation<br>Duration                          | 3 hours                                                                         | 3 hours                                                                         | 10 hours                                                                        | Same as<br>Predicate<br>Device |
| Biocompatibility                                   | No direct patient<br>contact                                                    | No direct patient<br>contact                                                    | No direct patient<br>contact                                                    | Same                           |

The subject device – the SporView Rapid Read Biological Indicator SteamPlus Test Pack and its predicate device - 3M Attest 41382 Rapid Readout Steam-Plus Process Challenge Device are similar in that they both have the same intended use, fundamental technology and general performance. Both the subject device and the 3M Attest 41382 Rapid Readout Steam-Plus Process Challenge Device have 510(k) cleared self-contained biological indicators with a spore carrier inoculated with the same organism, G. stearothermophilus. The subject and predicate test packs both provide a significant challenge to the steam sterilization process.

The subject device is very similar to its reference device - the SporView Plus Bl Test Pack cleared under K140620. The 510(k) cleared chemical indicators in the subject device, including the locations, of the subject test pack are identical to the reference device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "CANTEL" in blue font. To the left of the word is a blue circular logo with several horizontal lines running through it. The logo and the word "CANTEL" are the only elements in the image. The background is white.

### 4. Summary of Non-Clinical Performance Data

The following testing has been conducted in accordance with the FDA Guidance on Biological Indicators and the FDA Guidance for Industry and FDA Staff – Premarket Notification [510(k)] Submissions for Chemical Indicators and the FDA Guidance for Industry and FDA Staff -Premarket Notification [510(k)] Submissions for Chemical Indicators. Please refer to the table below for a brief description of the subject device testing

Results

Pass

Pass

Pass

Pass

| Performance Testing       | Details and Acceptance Criteria                                                                                                    |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Test Pack Performance     | All test packs have complete kill in a 270°F (132°C)<br>pre vacuum steam sterilization cycle for an<br>exposure time of 4 minutes. |
| Resistance<br>Performance | The test pack has resistance greater than or equal to<br>the AAMI Towel Pack.                                                      |

contained biological indicator itself.

- . Non-Clinical Performance Testing of the Subject:
Performance testing demonstrates that the subject device has more resistance than the self-contained biological indicator itself and the subject device has resistance greater than or equal to the AAMI Towel Pack. There was no growth of the SCBI contained in the subject device after a full sterilization cycle.

The test pack has resistance greater than the self-

Testing to confirm process indicators contained in the

test pack perform as intended per the FDA Guidance

for Industry and FDA Staff - Premarket Notification [510(k)] Submissions for Chemical Indicators.

#### 5. Conclusion

Chemical Indicator

Performance

Based on the intended uses, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as the legally marketed predicate cleared under 510(k) K101910.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K181434](https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K181434)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
