← Product Code [FRC](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC) · K140620

# SPORVIEW PLUS BI TEST PACK (K140620)

_Crosstex International · FRC · Aug 7, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K140620

## Device Facts

- **Applicant:** Crosstex International
- **Product Code:** [FRC](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC.md)
- **Decision Date:** Aug 7, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The SPSmedical SporView® Plus Steam BI Test Pack is designed to monitor sterilization cycles in both gravity displacement and pre-vacuum steam sterilizers. It is to be used for routine monitoring and qualification testing of steam sterilizers.

## Device Story

Single-use biological test pack for steam sterilization monitoring; contains biological indicator (BI) and chemical integrator. Device provides challenge to air removal and steam penetration equivalent to AAMI reference PCD. Used in clinical settings for routine monitoring and qualification of gravity displacement and pre-vacuum steam sterilizers. Operator places pack in sterilizer; after cycle, user retrieves BI for incubation and checks chemical integrator for immediate performance observation. BI uses Geobacillus stearothermophilus spores; 10-hour incubation confirms sterilization efficacy. External process indicator on pack alerts user to exposure. Benefits include assurance of sterilization process performance.

## Clinical Evidence

Bench testing only. Validation performed in steam gravity and pre-vacuum cycles at 121°C/250°F and 132°C/270°F. Testing included pass, fail, and partial cycles at the end of the 18-month shelf life. Results met all predetermined acceptance criteria per FDA guidance for Biological Indicator (BI) Premarket Notification submissions.

## Technological Characteristics

Biological indicator (Geobacillus stearothermophilus ATCC 7953, 10^5 spores) on Whatman paper carrier; STEAMPlus™ chemical integrator; paper layers for steam challenge; external process indicator label. Dimensions: 4-5/8" x 7/8" x 6-1/4". Storage: 15-30°C. Shelf life: 18 months.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- SporView® Plus BI Test Pack ([K051173](/device/K051173.md))

## Reference Devices

- SporView® 10 Steam Self Contained Biological Indicator ([K122024](/device/K122024.md))
- SporView® Steam Self-Contained Biological Indicator ([K070595](/device/K070595.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 7, 2014

Crosstex International Mr. Michael Nolan Research and Development Coordinator 6789 West Henrietta Road Rush, NY 14543

Re: K140620

Trade/Device Name: SporView® Plus BI Test Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Biological Sterilization Process Indicator Regulatory Class: II Product Code: FRC July 9, 2014 Dated: Dated: Received: July 11, 2014

Dear Mr. Nolan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Nolan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Statement of Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K140620

SporView® Plus Steam BI Test Pack Device Trade Name:

# INDICATIONS FOR USE:

The SPSmedical SporView® Plus Steam BI Test Pack with STEAMPlus Integrator is indicated for use in routine monitoring and sterilizer qualification testing of steam gravity displacement cycles at 121°C/250°F for 30 minutes exposure time and for use in pre-vacuum steam sterilization cycles at 132°C/270°F for 4 minutes exposure time.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) and/or

Over the Counter Use X (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation

# Sreekanth Sigland Signaly Signed by Sieckanth Gutala -

Digitally signed by Sreekanth ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=2000 540490, cn=Sreekanth Gutala -S Date: 2014.08.06 11:54:54 -04'00'

{3}------------------------------------------------

# 510(k) Summary

# SUBMITTER INFORMATION:

| SPSmedical Supply Corp.              | Contact: | Michael G. Nolan                     |
|--------------------------------------|----------|--------------------------------------|
| a division of Crosstex International |          | Research and Development Coordinator |
| 6789 West Henrietta Road             | Phone:   | (800) 722-1529                       |
| Rush, NY 14543 U.S.A.                | Fax:     | (585) 359-0167                       |
| Date of Summary:<br>August 5, 2014   |          |                                      |

# DEVICE INFORMATION:

| Device Trade Name:   | SporView® Plus Steam BI Test Pack                               |
|----------------------|-----------------------------------------------------------------|
| Common Name:         | Biological Test Pack                                            |
| Classification Name: | Biological Sterilization Process Indicator 21 CFR § 880.2800(a) |
| Review Panel:        | General Hospital                                                |
| Product Code:        | FRC                                                             |
| Device Class:        | II                                                              |

# PREDICATE DEVICE:

The SPSmedical SporView® Plus BI Test Pack cleared under K051173.

# DEVICE DESCRIPTION:

Description of change to the legally marketed predicate device—The SporView® Plus BI Test Pack cleared under K051173 contained a biological which required 24 hours of incubation time. This submission aims to replace that biological with the Spor View® 10 Steam Self Contained Biological Indicator within the SporView® Plus BI Test Pack.

Functionality-The SporView® Plus BI Test Pack is a single use device designed to assess the effective performance of both gravity displacement and pre-vacuum steam sterilization processes.

Scientific Concepts—The SporView® Plus BI Test Pack provides a defined challenge to the sterilization process that is equal to or greater than the challenge posed by the AAMI reference PCD.

Physical Characteristics-The SporView® Plus BI Test Pack consists of layers of paper with a biological indicator and a load record card with chemical integrator placed in the center and all placed within an exterior containment box. A process indicator label placed on the containment box alerts users if a pack has been exposed to the sterilization process. The approximate dimensions of the pack are 4-5/8" x 7/8" x 6-1/4". The load record card is used to record all the detailed information from the sterilization cycle.

Performance Characteristics-The performance of the SporView® Plus BI Test Pack has been demonstrated to be equivalent to that of the AAMI reference PCD.

### INTENDED USE:

NOTE: The intended use of the modified device, as described in the labeling, has not changed as a result of the modification.

The SPSmedical SporView® Plus Steam BI Test Pack is designed to monitor sterilization cycles in both gravity displacement and pre-vacuum steam sterilizers. It is to be used for routine monitoring and qualification testing of steam sterilizers.

### TECHNICAL CHARACTERISTICS:

The outer box and layers of paper within the SporView® Plus BI Test Pack provide a significant challenge to air removal and subsequent steam penetration of prevacuum and gravity steam The Load Record Card with STEAMPlus™ Steam Integrator allow for an immediate sterilizers. observation of cycle performance with 10 hour incubation results of the SporView® 10 Steam Self Contained Biological Indicator offers complete assurance of cycle performance. The external process indicator alerts users when a pack has seen the sterilization process.

{4}------------------------------------------------

# RECOMMENDED STORAGE CONDITIONS:

Store in a cool, dry place (15-30°C).

# NON-CLINICAL TESTING:

Validation of the SporView® Plus BI Test Pack was accomplished with performance testing in steam gravity and prevacuum cycles operating at 250°F/121°C and 270°F/132°C. All results from testing meet the predetermined acceptance criteria. All testing followed the FDA Guidance document for Industry and FDA Staff entitled, "Biological Indicator (BI) Premarket Notification [510(k)] Submissions," issued on October 4, 2007. Testing consisted of running test packs and control packs (predicate) within the same cycles for pass, fail and partial cycles at the end of the product's shelf life of eighteen (18) months.

# COMPONENTS:

The Spor View® Plus BI Test Pack Components are equivalent to its original filing with the exception of the Spor View® Steam Self Contained Biological Indicator.

- Subject of this change SporView® 10 Steam Self Contained Biological Indicator (K122024)
- Load Record Card with STEAMPlus™ Steam Integrator (K051173) o
- External Process Indicator with lot and expiration information (K051173) ●
- o Outer box and paper inserts (K051173)

### SUBSTANTIAL EQUIVALENCE DISCUSSION

SPSmedical has identified the SporView® Plus BI Test Pack (K051173) as the (primary) predicate device. We believe the predicate device to be substantially equivalent to the subject device in terms of their intended use and functional characteristics as they are essentially the same device. See a comparison of the subject device to the predicate device (K051173) in Table 1.

### PREDICATE I.D .:

| Trade Name:       | SporView® Plus BI Test Pack                                                                                                  |
|-------------------|------------------------------------------------------------------------------------------------------------------------------|
| Model No.:        | SBT-025 (25 controls) & SBT-255 (5 controls)                                                                                 |
| Submitter/holder: | SPSmedical Supply Corp.<br>6789 West Henrietta Road<br>Rush, NY 14543 U.S.A.<br>Phone: (585) 359-0130<br>Fax: (585) 359-0167 |
| 510(k) No.:       | K051173                                                                                                                      |

### COMPARISON OF INDICATIONS FOR USE (IFU):

Predicate IFU K051173-The SPSmedical SporView® Plus Steam BI Test Pack with STEAMPlus Integrator is indicated for use in routine and challenge testing of steam gravity displacement cycles at 121°C/250°F for 30 minutes exposure time or longer and for use in pre-vacuum steam sterilization cycles at 132°C/270°F for 4 minutes exposure time or longer.

Subject IFU K140620—The SPSmedical SporView® Plus Steam BI Test Pack with STEAMPlus Integrator is indicated for use in routine monitoring and sterilizer qualification testing of steam gravity displacement cycles at 121°C/250°F for 30 minutes exposure time and for use in pre-vacuum steam sterilization cycles at 132°C/270°F for 4 minutes exposure time.

Compared to the predicate the subject device has changes for clarity but is otherwise the same.

### DESIGN DIFFERENCES PREDICATE VS. SUBJECT DEVICE:

The Predicate Device contains the SporView® Steam Self-Contained Biological Indicator which was cleared under K070595. The Subject Device contains the SporView® 10 Steam Self Contained Biological Indicator which was cleared under K122024.

{5}------------------------------------------------

# FUNCTIONAL CHARACTERISTICS:

The SporView® Plus Steam BI Test Pack is designed to create a significant challenge to air removal and steam penetration. Like the AAMI test pack, the SPSmedical SporView® Plus Steam BI Test Pack provides a significant challenge to the steam sterilization process and is used for routine monitoring and qualification testing of gravity displacement and pre-vacuum steam sterilizers. Both have internal steam monitoring components located within the center of the pack and chemical indicator located on the pack.

# DISCUSSION:

SPSmedical is claiming substantial equivalence for its SporView® Plus Steam BI Test Pack to the original SporView® Plus Steam BI Test Pack (K051173) based on test data obtained during validation studies. We have demonstrated with testing that the Spor View® Plus Steam BI Test Pack performs consistently when run in steam gravity and prevacuum cycles operating at 250°F/121°C and 270°F/132°C for pass, fail and partial cycles.

# SUBSTANTIAL EQUIVALENCE CONCLUSIONS:

The SporView® Plus Steam BI Test Pack has the same intended use and characteristics as the predicate Spor View® Plus Steam BJ Test Pack. They both provide a significant challenge to the steam sterilization process. Testing data demonstrates the subject device is substantially equivalent to the predicate device.

| Element                                                        | Subject Device<br>(K140620)                                                                                                                                                                                                                                                                                                                                                | Predicate<br>(K051173)                                                                                                                                                                                                                                                                                                                                      |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use: method<br>of sterilization<br>process parameters | Steam<br>250°F/121°C 30 minutes<br>270°F/132°C 4 minutes                                                                                                                                                                                                                                                                                                                   | Steam<br>250°F/121°C 30 minutes<br>270°F/132°C 4 minutes                                                                                                                                                                                                                                                                                                    |
| Indications for Use<br>(IFU)                                   | The SPSmedical SporView® Plus Steam BI Test<br>Pack with STEAMPlus Integrator is indicated<br>for use in routine monitoring and sterilizer<br>qualification<br>testing<br>of steam<br>gravity<br>displacement cycles at 121°C/250°F for 30<br>minutes exposure time and for use in pre-vacuum<br>steam sterilization cycles at 132°C/270°F for 4<br>minutes exposure time. | The SPSmedical SporView® Plus Steam BI Test<br>Pack with STEAMPlus Integrator is indicated<br>for use in routine and challenge testing of steam<br>gravity displacement cycles at 121°C/250°F for<br>30 minutes exposure time or longer and for use<br>in pre-vacuum steam sterilization cycles at<br>132°C/270°F for 4 minutes exposure time or<br>longer. |
| Labeling                                                       | Equivalent, except for 10 hour<br>incubation time                                                                                                                                                                                                                                                                                                                          | Equivalent, except for 24 hour<br>incubation time                                                                                                                                                                                                                                                                                                           |
| Organism: spore<br>species<br>& strain                         | Geobacillus Stearothermophilus<br>ATCC 7953                                                                                                                                                                                                                                                                                                                                | Geobacillus Stearothermophilus<br>ATCC 7953                                                                                                                                                                                                                                                                                                                 |
| Viable spore<br>population                                     | $10^5$ or greater                                                                                                                                                                                                                                                                                                                                                          | $10^5$ or greater                                                                                                                                                                                                                                                                                                                                           |
| Resistance<br>characteristics                                  | Equivalent, calculated per USP                                                                                                                                                                                                                                                                                                                                             | Equivalent, calculated per USP                                                                                                                                                                                                                                                                                                                              |
| Culture Conditions                                             | 55-60°C                                                                                                                                                                                                                                                                                                                                                                    | 55-60°C                                                                                                                                                                                                                                                                                                                                                     |
| Carrier Materials                                              | Whatman Paper                                                                                                                                                                                                                                                                                                                                                              | Whatman Paper                                                                                                                                                                                                                                                                                                                                               |
| Packaging: primary &<br>secondary                              | Equivalent                                                                                                                                                                                                                                                                                                                                                                 | Equivalent                                                                                                                                                                                                                                                                                                                                                  |
| Storage                                                        | Equivalent                                                                                                                                                                                                                                                                                                                                                                 | Equivalent                                                                                                                                                                                                                                                                                                                                                  |
| Shelf Life                                                     | Eighteen (18) months                                                                                                                                                                                                                                                                                                                                                       | Eighteen (18) months                                                                                                                                                                                                                                                                                                                                        |

# TABLE 1-COMPARISON OF THE SUBJECT DEVICE TO THE PREDICATE

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K140620](https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K140620)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
