← Product Code [FRC](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC) · K070595

# SPORVIEW STEAM SELF-CONTAINED BIOLOGICAL INDICATOR (K070595)

_Spsmedical Supply Corp. · FRC · May 24, 2007 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K070595

## Device Facts

- **Applicant:** Spsmedical Supply Corp.
- **Product Code:** [FRC](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC.md)
- **Decision Date:** May 24, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

SporView" Steam is a self-contained biological indicator inoculated with viable Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of saturated steam sterilization processes operating at 121℃ and 132℃ gravity displacement, 132℃ flash gravity displacement and 121°C - 134°C prevacuum cycles.

## Device Story

SporView® Steam is a self-contained biological indicator (SCBI) used to monitor steam sterilization efficacy. Device contains Geobacillus stearothermophilus spores on a paper carrier and a glass ampoule of Tryptic Soy Broth with Bromocresol Purple pH indicator, all within a plastic vial. After sterilization, the user activates the device by crushing the ampoule to release media onto the spores. The vial is then incubated. A color change indicates spore growth, signaling sterilization failure; no color change indicates successful sterilization. Used by healthcare personnel to verify sterilization cycle performance.

## Clinical Evidence

Bench testing only. Validation included resistance testing, spore population counts, media recovery in extended cycles, holding time effects, incubation period reduction, color change stability, media evaporation, and carrier/packaging material integrity. All results met defined acceptance criteria.

## Technological Characteristics

Self-contained biological indicator. Components: Geobacillus stearothermophilus ATCC #7953 spores, paper filter carrier, Tryptic Soy Broth, Bromocresol Purple pH indicator, plastic vial. Activation: Manual crushing of internal glass ampoule. Incubation: Required for colorimetric readout. Sterilization monitoring range: 121°C to 134°C.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- SGM EZTest® - Steam ([K963841](/device/K963841.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070595

MAY 2 4 2607

## 510(k) Summary of Safety and Effectiveness

### Submitter:

.

- SPSmedical Supply Corp. . 6789 West Henrietta Road Rush, NY 14543 U.S.A. Phone: (585)-359-0130 Fax: (585)-359-0167
- Establishment FDA Registration No.: 1319130 .
- Date Summary was Prepared May 24th, 2007 .
- Gary J. Socola . Printed name of person submitting for 510(k)

Gary J. Socola

Signature of person submitting for 510(k)

- Vice President, Scientific Affairs . Title of person submitting for 510(k)
Device Name and Classification

| Trade Name:            | SporView® Steam Self-Contained Biological Indicator |
|------------------------|-----------------------------------------------------|
| Classification Name:   | Sterilization Process Biological Indicator          |
| Common Name:           | Self-Contained Biological Indicator                 |
| Device Classification: | Class II, Regulation Number 880.2800                |
| Product Code:          | FRC                                                 |
| Predicate Device:      | SGM EZTest® - Steam (K963841)                       |

{1}------------------------------------------------

Device Description:

The SPSmedical SporView® biological indicator consists of a self-contained unit that includes bacterial spores of Geobacillus stearothermophilus ATCC #7953 inoculated onto a paper filter carrier and a small glass ampoule containing Tryptic Soy Broth with Bromnocresol Purple acting as a pH indicator encased in a plastic vial that serves as the culture tube.

#### Intended Use:

SporView" Steam is a self-contained biological indicator inoculated with viable Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of saturated steam sterilization processes operating at 121℃ and 132℃ gravity displacement, 132℃ flash gravity displacement and 121°C - 134°C prevacuum cycles.

Statement of Similarity to the Legally Marketed Predicate Device:

- Both are intended to monitor steam sterilization cycles from 121°C to134°C. .
- . Both utilize the same strain of bacterial spores.
- . Both utilize the same carrier material.
- . Both use virtually the same media.
- . Both are similar in size and shape.
- . Both are activated in the same manner.
- . Both are incubated at the same temperature

#### Non-Clinical Testing:

Testing was performed in order to validate the indicators label claims and performance characteristics. Multiple lots of indicators were tested for;

- . Resistance
- . Spore population
- Media recovery in extended steam sterilization cycles .
- Effects of holding time
- . Reduced incubation period
- . Stability of the color change
- . Media Evaporation
- . Effects of carrier and package materials

All test results met the defined acceptance criteria.

#### Conclusion:

Supportive data has demonstrated that the SPSmedical SporView® Steam Self-Contained Biological Indicator is equivalent to the legally marketed predicate device and therefore, as safe and effective as the legally marketed device.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the caduceus.

#### Public Health Service

ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# MAY 2 4 2007

SPSmedical Supply Corporation C/O Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services 2307 East Aurora Road Twinsburg, Ohio 44087

Re: K070595

Trade/Device Name: SporView® Steam Self Contained BI Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: May 15, 2007 Received: May 16, 2007

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2 - Mr. Lehtonen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Clues

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### INDICATIONS for USE STATEMENT

Applicant: SPSmedical Supply Corp.

f Chemical Supply Corp.

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K070595

SporView® Steam Self Contained BI Device Name:

Indications For Use:

SporView® Steam is a self-contained biological indicator inoculated with viable Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of saturated steam sterilization processes operating at 121°C and 132°C gravity displacement, 132°C flash gravity displacement and 121°C - 134°C prevacuum cycles.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shula H. Murphy MD

ന of Anesthesiology, General Hospital, on Control, Dental Devices

*k) NumberPage 1 of 1 K020595

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K070595](https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FRC/K070595)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
