LUMISCOPE MODEL 2018 THERMOMETER

K954792 · Lumiscope Co., Inc. · FLL · Jun 21, 1996 · General Hospital

Device Facts

Record IDK954792
Device NameLUMISCOPE MODEL 2018 THERMOMETER
ApplicantLumiscope Co., Inc.
Product CodeFLL · General Hospital
Decision DateJun 21, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Lumiscope Model 2018 Thermometer has the same intended use as the predicate device. It is indicated for use in the measurement of oral and axillary temperature. It is not intended for rectal (rectum) temperature measurement. The device also provides a clock function when not in the temperature mode. The Lumiscope Model 2018 Thermometer is intended for use by adults and may be used by the infant or pediatric patient population for measurements of axillary body temperature or for measurements of oral body temperature if the child is able to hold the thermometer beneath the tongue while his mouth is closed.

Device Story

Digital electronic thermometer; consists of flexible probe, connecting cord, and display unit. Input: thermal signal via transducer. Operation: standard electrical transducer design; converts thermal input to digital temperature reading. Output: LCD temperature display and audible alert. Secondary function: quartz clock. Used in home or clinical settings; operated by adults or caregivers. Provides temperature data to assist in fever monitoring and clinical decision-making.

Clinical Evidence

Bench testing only. Device meets ASTM E 1112-86 Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature. Accuracy, precision, and repeatability confirmed. Materials identical to those used in other legally marketed devices.

Technological Characteristics

Electrical transducer design; flexible probe; 1.5V DC button battery; LCD display; slide switches for mode selection; temperature alert beeper. Complies with ASTM E 1112-86. Materials identical to predicate.

Indications for Use

Indicated for oral and axillary temperature measurement in adults, infants, and pediatric patients. Contraindicated for rectal use. Pediatric oral use requires child's ability to hold thermometer under tongue with mouth closed.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} LUMISCOPE JUN 21 1996 K954792 # 510(k) SUMMARY - Lumiscope Model 2018 Thermometer This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. **Proprietary Name:** Lumiscope Model 2018 Thermometer **Common/Usual Name:** Digital Fever Thermometer with Quartz Clock Oral and Axillary Thermometer **Classification Name:** Clinical Electronic Thermometer (per 21 CFR 880.2910) **Legally Marketed Device:** Hanson Scale Co Clinical Electronic Thermometer - Model 7616-10 **Device Description:** The Lumiscope Model 2018 Thermometer is a thermometer of standard electrical transducer design, consisting of a flexible probe, connecting cord and display unit. The probe consists of a transducer and supporting plastic tube which houses the connection between the tube and cord. The display unit houses a 1.5V DC button battery, Liquid Crystal Display (LCD), slide switches for controlling the thermometer/clock mode and °F/°C mode, hour and minute set buttons, and temperature alert beeper. **Intended Use:** The Lumiscope Model 2018 Thermometer has the same intended use as the predicate device. It is indicated for use in the measurement of oral and axillary temperature. It is not intended for rectal (rectum) temperature measurement. The device also provides a clock function when not in the temperature mode. The Lumiscope Model 2018 Thermometer is intended for use by adults and may be used by the infant or pediatric patient population for measurements of axillary body temperature or for measurements of oral body temperature if the child is able to hold the thermometer beneath the tongue while his mouth is closed. **Technological Characteristics:** The Lumiscope Model 2018 Thermometer has the same technological features as the predicate device: the mode of operation, physical specifications, operational, and biological specifications are identical. **Performance Data:** The Lumiscope Model 2018 Thermometer meets the ASTM Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature (ASTM E 1112-86). Bench testing confirmed accuracy, precision and repeatability measurements specified in the labeling. For all body or fluid contacting materials, certification is made that the identical materials have been used in other legally marketed devices under the same use conditions. **Substantial Equivalence:** The information provided demonstrates that the Lumiscope Model 2018 Thermometer is substantially equivalent to the predicate device since the devices have the same intended use, principles of operation, and technological characteristics. **Contact Person:** ![img-0.jpeg](img-0.jpeg) Thomas C. Parnell ![img-1.jpeg](img-1.jpeg) E-2a HOME OFFICE: THE LUMISCOPE COMPANY, INC., 400 RARITAN CENTER PARKWAY, EDISON, NJ 08837 PHONE 908/225-5533 FAX 908/225-7047
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...