← Product Code [FLL](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL) · K223044

# Digital Thermometer (Model QT001) (K223044)

_Shen Zhen Rong Feng Technology Co., Ltd. · FLL · May 10, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K223044

## Device Facts

- **Applicant:** Shen Zhen Rong Feng Technology Co., Ltd.
- **Product Code:** [FLL](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL.md)
- **Decision Date:** May 10, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

The Digital Thermometer QT001 is intended to measure the human body temperature in armpit, orally or rectally, and the device is reusable for clinical or home use on people of all ages.

## Device Story

Handheld digital thermometer; measures human body temperature via armpit, oral, or rectal sites. Input: thermistor sensor (Type 503F) resistance changes based on temperature. Processing: MCU converts resistance to frequency/voltage; calculates temperature; internal firmware manages logic. Output: temperature displayed on LCD; includes backlight color-coded fever alerts (green/yellow/red). Used in clinical or home settings by patients or clinicians. Benefits: provides accurate, reusable temperature monitoring for all ages.

## Clinical Evidence

Bench testing only. Device complies with ASTM E1112-00(2018), ISO 80601-2-56:2017/AMD 1:2018, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5, and ISO 10993-10. No clinical data presented.

## Technological Characteristics

Materials: ABS, TPE, stainless steel. Sensing: Thermistor (Type 503F). Energy: 3.0V CR1225 battery. Form factor: Handheld with flexible tip. Connectivity: Standalone. Software: Firmware-based measurement and display logic. Standards: ISO 80601-2-56, ASTM E1112-00, IEC 62304.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Digital Thermometer ([K200599](/device/K200599.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 10, 2023

Shen Zhen Rong Feng Technology Co., Ltd Jett Lee Regulation Manager 3/F R Building ShaSi Industrial Park, Shajing Town BaoAn District Shenzhen, Guangdong China

Re: K223044

Trade/Device Name: Digital Thermometer (Model QT001) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: April 9, 2023 Received: April 10, 2023

Dear Jett Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)

regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Walloschek

David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223044

Device Name Digital Thermometer (Model: OT001)

Indications for Use (Describe)

The Digital Thermometer OT001 is intended to measure the human body temperature in armpit, or rectally, and the device is reusable for clinical or home use on people of all ages.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="font-family: sans-serif;"> <input disabled="disabled" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)       </span>                  |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <input checked="checked" disabled="disabled" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)       </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K223044

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

### Summary Prepared Date: 25 April 2023

#### 1. Submitter Information

Sponsor Company Name: SHEN ZHEN RONG FENG TECHNOLOGY CO.,LTD

Address: 3/F R building ShaSi Industrial Park,Shajing Town,BaoAn district,ShenZhen city,GuangDong

Province.China

Phone: +86-13715151187

Contact Person (including title): Hong Xiao Lan (Engineer)

E-mail: 844352259@qq.com

# Application Correspondent: SHEN ZHEN RONG FENG TECHNOLOGY CO.,LTD

Address: 3/F R building ShaSi Industrial Park,Shajing Town,BaoAn district,ShenZhen city,GuangDong Province.China

- � Contact Person: Iris Fung
- � Tile: Requlation Manager
- Tel: +86-13211147965 �
- � Email: jianda-lee@foxmail.com; mdc-fs@foxmail.com;

# 2. Subject Device Information

Type of 510(k) submission: Traditional Common Name: Clinical Electronic Thermometer Trade Name: Digital thermometer (Model: QT001) Classification Name: Clinical electronic thermometer Review Panel: General Hospital Product Code: FLL Regulation Number: 21 CFR 880.2910 Requlation Class: II

# 3. Predicate Device Information

Sponsor: JOYTECH HEALTHCARE CO., LTD. Trade Name: Digital Thermometer

{4}------------------------------------------------

510(k) number: K200599 Review Panel: General Hospital Product Code: FLL Regulation Number: 21 CFR 880.2910 Regulation Class: II

# 4. Device Description

The digital thermometer QT001 is hand held device which can measure human body's temperature at the site of armpit, orally or rectally, the device is reusable for clinical or home use on people of all ages. The results can be displayed on LCD. The QT001 has only one operating mode: direct mode.

The digital thermometer measures temperatures of human body by placing the probe tip into the armpit, oral, rectum directly with the measuring time about 3 minutes.

The digital thermometer consists of a temperature sensor, low power consumption integrated circuit(IC),LCD display and buzzer. The resistance of sensor changes with temperature and the IC converts the resistance to frequency and calculates the temperature according to the relation of resistance and frequency. The calculated temperature is displayed on the LCD.

The thermistor sensor(Type:503F) reacts to temperature. In different temperature, the thermistor has a different resistance value. The essence of temperature measurement is actually measuring the resistance of the sensor, usually is to change the resistance into voltage or current and other analog signals, and digital signals, and then the processor converts the signals to the corresponding temperatures.

QT001 has several functions, such as beep alarm, unit switchable, low battery detection, memories, backlight, auto power off functions.

# 5. Indication for use

The Digital Thermometer QT001 is intended to measure the human body temperature in armpit, orally or rectally, and the device is reusable for clinical or home use on people of all ages.

# 6. Comparison to Predicate Device

A comparison of key technological characteristics between the subject devices and predicate device was listed as below:

| Elements of<br>Comparison                            | Subject Device                                                                                                                                                                                                                                                                                                   | Predicate Device                                                                                                                                                                                                                                                                         | Verdict                                                                                                                                                                                                                                              |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                         | SHEN ZHEN RONG FENG<br>TECHNOLOGY CO., LTD                                                                                                                                                                                                                                                                       | Joytech Healthcare Co., Ltd                                                                                                                                                                                                                                                              | --                                                                                                                                                                                                                                                   |
| Elements of<br>Comparison                            | Subject Device                                                                                                                                                                                                                                                                                                   | Predicate Device                                                                                                                                                                                                                                                                         | Verdict                                                                                                                                                                                                                                              |
| 510 (k) Number                                       | K223044                                                                                                                                                                                                                                                                                                          | K200599                                                                                                                                                                                                                                                                                  | --                                                                                                                                                                                                                                                   |
| Product Name                                         | Digital Thermometer                                                                                                                                                                                                                                                                                              | Digital Thermometer                                                                                                                                                                                                                                                                      | --                                                                                                                                                                                                                                                   |
| Models                                               | QT001                                                                                                                                                                                                                                                                                                            | DMT-4756                                                                                                                                                                                                                                                                                 | --                                                                                                                                                                                                                                                   |
| Intended Use                                         | The Digital Thermometer QT001<br>is intended to measure the<br>human body temperature in<br>armpit, orally or rectally, and the<br>device is reusable for clinical or<br>home use on people of all ages.                                                                                                         | The Digital Thermometer DMT-<br>4756 is intended to measure the<br>human body temperature in regular<br>mode orally, rectally or under the<br>arm, and the device is reusable<br>for clinical or home use on people<br>of all age, including children under 8<br>with adult supervision. | Similar<br>There is minor<br>difference, but the<br>user manual have<br>been indicated in<br>the instructions that<br>children should not<br>be allowed to<br>perform the test<br>unsupervised<br>It will not affect<br>safety and<br>effectiveness. |
| Fundamental<br>technology&<br>Operating<br>principle | A change of thermistor<br>resistance,<br>caused by changes of<br>temperature. The resistance is<br>measured by MCU, so changes<br>of temperature will correspond to<br>changes of resistance                                                                                                                     | A change of thermistor resistance,<br>caused by changes of temperature.<br>The resistance is measured by<br>MCU, so changes of temperature<br>will correspond to changes of<br>resistance                                                                                                | Identical                                                                                                                                                                                                                                            |
| Sensor                                               | Thermistor                                                                                                                                                                                                                                                                                                       | Thermistor                                                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                                                            |
| Signal<br>processing<br>and display                  | Internal firmware and local LCD<br>display                                                                                                                                                                                                                                                                       | Internal firmware and local LCD<br>display                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                                                            |
| Measurement<br>Site                                  | armpit, orally or rectally                                                                                                                                                                                                                                                                                       | orally, rectally or under the arm                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                            |
| Measuring<br>Range                                   | $32.0~42.9 °C (89.6°F ~ 109.2°F)$                                                                                                                                                                                                                                                                                | $32.0°C~43.9°C (89.6°F~111.0°F)$                                                                                                                                                                                                                                                         | Different(Note 1)                                                                                                                                                                                                                                    |
| Accuracy                                             | $±0.1°C,34.0°C-42.0°C (±0.2°F,93.2°F -107.6°F )± 0.2 °C under34.0 °C or over42.0°C( ± 0.4°F under 93.2°F or over107.6°F )$                                                                                                                                                                                       | $±0.1°C between 35.5°C to42.0°C(±0.2°F,95.9°F-107.6°F),±0.2°C under 35.5°C or over42.0°C(±0.4°F under 95.9°F or over107.6°F )$                                                                                                                                                           | Different(Note 2)                                                                                                                                                                                                                                    |
| Display<br>resolution                                | $0.1 °C/0.1 °F$                                                                                                                                                                                                                                                                                                  | $0.1 °C/0.1 °F$                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                                                            |
| Components                                           | Sensor, buzz film, housing,<br>stainless steel cap, LCD display,<br>measurement control module.                                                                                                                                                                                                                  | Sensor, buzz film, housing,<br>stainless steel cap, LCD display,<br>measurement control module.                                                                                                                                                                                          | Identical                                                                                                                                                                                                                                            |
| Sensor type                                          | Thermopile                                                                                                                                                                                                                                                                                                       | Thermopile                                                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                                                            |
| Material                                             | ABS, TPE, Stainless steel                                                                                                                                                                                                                                                                                        | ABS, TPE, Stainless steel                                                                                                                                                                                                                                                                | Identical                                                                                                                                                                                                                                            |
| Elements of<br>Comparison                            | Subject Device                                                                                                                                                                                                                                                                                                   | Predicate Device                                                                                                                                                                                                                                                                         | Verdict                                                                                                                                                                                                                                              |
| Memories                                             | For storing the last measured<br>value                                                                                                                                                                                                                                                                           | 10 memories                                                                                                                                                                                                                                                                              | Different(Note 3)                                                                                                                                                                                                                                    |
| Power Supply                                         | One 3.0 V CR1225 battery                                                                                                                                                                                                                                                                                         | One 3.0V CR2032 battery                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                            |
| Operating range                                      | Temperature: 5°C-40°C (41°F -<br>104°F )<br>Relative humidity:15%RH-<br>95%RH<br>Atmospheric pressure: 70kPa-<br>106kPa                                                                                                                                                                                          | Temperature:<br>41°F ~104°F (5°C ~40°C )<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure :<br>700hPa ~ 1060hPa                                                                                                                                                                   | Different(Note4)                                                                                                                                                                                                                                     |
| Construction                                         | Flexible tip                                                                                                                                                                                                                                                                                                     | The probe is foldable into the body<br>of the thermometer                                                                                                                                                                                                                                | Different(Note5)                                                                                                                                                                                                                                     |
| Measurement<br>method & time                         | Approximate 3 minutes                                                                                                                                                                                                                                                                                            | 10s Predictive read<br>30s Final read.                                                                                                                                                                                                                                                   | Different(Note6)                                                                                                                                                                                                                                     |
| Thermometer<br>Size                                  | 138mm x 25mm x 12mm (L x W x<br>H)                                                                                                                                                                                                                                                                               | 10.6cm*3.2cm*2.5cm(L x W x H)                                                                                                                                                                                                                                                            | Different(Note7)                                                                                                                                                                                                                                     |
| Weight                                               | Approx. 10 grams including<br>battery                                                                                                                                                                                                                                                                            | Approx. 40 grams including battery                                                                                                                                                                                                                                                       | Different(Note7)                                                                                                                                                                                                                                     |
| LCD Size                                             | 2.0cm*0.7cm                                                                                                                                                                                                                                                                                                      | 2.5cm*1.4cm                                                                                                                                                                                                                                                                              | Different(Note7)                                                                                                                                                                                                                                     |
| Color &Coding                                        | White(R9003), Green(P294C)                                                                                                                                                                                                                                                                                       | White(R9003),<br>Green(P564C)                                                                                                                                                                                                                                                            | Different(Note7)                                                                                                                                                                                                                                     |
| Predictive mode                                      | NO                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                      | Different(Note6)                                                                                                                                                                                                                                     |
| Storage<br>and<br>transportation<br>condition        | Temperature:-20℃~55℃(-<br>4°F ~ 131°F )<br>Relative humidity:15%RH-<br>95%RH<br>Atmospheric pressure: 50kPa-<br>106kPa                                                                                                                                                                                           | Temperature:<br>-4°F ~131°F (-20°C~55°C)<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure:<br>700hPa ~ 1060hPa                                                                                                                                                                    | Different(Note4)                                                                                                                                                                                                                                     |
| Safety                                               | Complied with IEC 60601-1                                                                                                                                                                                                                                                                                        | Complied with IEC 60601-1                                                                                                                                                                                                                                                                | Identical                                                                                                                                                                                                                                            |
| EMC                                                  | Complied with IEC 60601-1-2                                                                                                                                                                                                                                                                                      | Complied with IEC 60601-1-2                                                                                                                                                                                                                                                              | Identical                                                                                                                                                                                                                                            |
| Biocompatibility                                     | Complied with ISO 10993-5 and<br>ISO 10993-10                                                                                                                                                                                                                                                                    | Complied with ISO 10993-5 and<br>ISO 10993-10                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                            |
| Performance                                          | Complied with ISO 80601-2-<br>56:2017/AMD 1:2018 and ASTM<br>E1112:00(2018).                                                                                                                                                                                                                                     | Complied with ISO 80601-2-<br>56:2017                                                                                                                                                                                                                                                    | Different(Note8)                                                                                                                                                                                                                                     |
| Backlight<br>warning function                        | 1.When measured temperature is<br>35.9℃,the LCD without<br>backlight<br>2.When measured temperature<br>≥36℃ and ≤37.1℃, the LCD<br>shows green backlight<br>3.When measured temperature<br>≥37.2℃ and ≤37.4℃, the LCD<br>shows yellow backlight.<br>4.When measured temperature<br>>=37.5°C and ≤42.9°C, the LCD | /                                                                                                                                                                                                                                                                                        | Different(Note9)                                                                                                                                                                                                                                     |
| Elements of<br>Comparison                            | Subject Device                                                                                                                                                                                                                                                                                                   | Predicate Device                                                                                                                                                                                                                                                                         | Verdict                                                                                                                                                                                                                                              |
|                                                      | shows red backlight.<br>The laboratory accuracy error no<br>greater than 0.3 °C                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                      |

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

### Comparison discussion:

### Note 1

The subject device and predicate device have different measurement range, but the measurement range of subject device meets the requirements of ASTM E1112:00 and ISO 80601-2-56. The difference does not raise any new or different safety and effectiveness questions. .

# Note 2

The subject device and predicate device have different accuracy, but the accuracy of the subject device meets the requirements of ASTM E1112:00 and ISO 80601-2-56. The difference does not raise any new or different safety and effectiveness questions.

#### Note 3

The memory function difference: The software verification were conducted to demonstrate the performance of the function as intended. The memory function does not affect accuracy of measurement; so such difference does not impact the performance of subject device.

#### Note 4

Operating range, Storage and transportation condition difference: The subject device has been tested and the test results met the requirements of IEC60601-1 and ISO 80601-2-56. The difference does not raise any new or different safety and effectiveness questions.

# Note 5

Construction difference: Performance testing shows the subject device complies with standard ISO 80601-2-56. The difference does not raise any new questions of safety and effectiveness.

# Note 6

Measurement time, predictive mode difference: Performance test was conducted in accordance with ISO 80601-2-56. The test results demonstrated the subject device complies with performance standard. The differences do not raise any new questions of safety and effectiveness.

# Note 7

Weight, size, color difference: Biocompatibility test and performance bench test met the requirements of ISO 10993-5 and ISO 10993-10. The differences do not raise any new questions of safety and effectiveness.

# Note 8

Performance testing shows the subject device complies with standard ISO 80601-2-56:2017/AMD 1:2018 and ASTM E1112-00 (2018). The difference does not raise any new questions of safety and effectiveness.

{8}------------------------------------------------

# Note 9

Backlight warning function does not affect accuracy of measurement; the software verification and validation were conducted to demonstrate the performance of the function as intended. So such difference does not raise any new safety and effectiveness questions.

# 7. Test Summary

# Non-clinical test:

Digital Thermometer conforms to applicable standards that include:

- � ASTM E1112-00(2018) Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
- � IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- � ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity,
- � ISO 10993-10:2010. Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- � IEC 62304:2006+AMD1:2015 Medical Device Software - Software Life Cycle Processes
- � IEC60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- � ISO 80601-2-56:2017/AMD 1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement

# 8. Conclusion

Based on comparison of the intended use, technological characteristics, applicable safety standards, verification and validation testing, the differences between subject device and the predicate device do not raise new issues of safety and effectiveness. Thus, the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K223044](https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K223044)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
