← Product Code [FLL](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL) · K222588

# Temp Pal (Smart Thermometer Patch) Model Number: STP-MB01-1 (K222588)

_Iweecare Co., Ltd. · FLL · Feb 17, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K222588

## Device Facts

- **Applicant:** Iweecare Co., Ltd.
- **Product Code:** [FLL](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL.md)
- **Decision Date:** Feb 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

Temp Pal is a battery-operated electronic device with intended use of measuring and monitoring human armpit temperature continuously via wireless signal transmission of the measuring result. This system is reusable and intended for armpit temperature monitoring for persons of all age. The temperature data of device is not intended to replace the advice, diagnosis, nor treatment recommendations of doctor. Temp Pal can be used at home and healthcare center.

## Device Story

Temp Pal is a wearable, battery-operated electronic thermometer patch worn in the axilla. It uses an NTC resistor to measure body temperature continuously. The device transmits temperature data via Bluetooth (2.4G BLE) to a smartphone running the Temp Pal App. The app displays real-time temperature, provides low battery alerts, and allows data export as CSV files. It is intended for use by patients or caregivers in home or clinical environments. The device provides direct mode temperature readings to assist in monitoring; it does not replace professional medical diagnosis or treatment. The system is reusable and powered by a 3.7V/10mAh rechargeable battery.

## Clinical Evidence

Bench testing only. Testing included shelf-life reliability, biocompatibility (ISO 10993-5, -10), performance testing per ISO 80601-2-56, electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1, -11), and usability engineering (EN 62366).

## Technological Characteristics

Materials: TPE, 304 stainless steel probe, medical adhesive. Sensing: NTC resistor. Energy: 3.7V/10mAh rechargeable battery. Connectivity: Bluetooth BLE (2.4G). Form factor: 28.5 x 26.8 x 3.52 mm, 3.6g. Standards: ISO 80601-2-56, IEC 60601-1, IEC 60601-1-11, IEC 60529 (IP Code).

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Temp Pal ([K202603](/device/K202603.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font below.

February 17, 2023

iWEECARE Co., Ltd. Julia Yang Regulatory Manager 1F., No. 11, Ln. 382, Zhonghe St., Beitou Dist., Taipei City, 112 Taiwan

Re: K222588

Trade/Device Name: Temp Pal (Smart Thermometer Patch) Model Number: STP-MB01-1 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: January 18, 2023 Received: January 20, 2023

Dear Julia Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Walloschek

David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

#### K222588

#### Device Name

Temp Pal (Smart Thermometer Patch) Model Number: STP-MB01-1

Indications for Use (Describe)

Temp Pal is a battery-operated electronic device with intended use of measuring and monitoring human armpit temperature continuously via wireless signal transmission of the measuring result. This system is reusable and intended for armpit temperature monitoring for persons of all age. The temperature data of device is not intended to replace the advice, diagnosis, nor treatment recommendations of doctor. Temp Pal can be used at home and healthcare center.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K222588 - 510(k) SUMMARY

- 5.1 Preparation Date: Feb. 9, 2023

| 5.2 Submitter: | iWEECARE Co., Ltd.                                                          |
|----------------|-----------------------------------------------------------------------------|
| Address:       | 1F., No. 11, Ln. 382, Zhonghe St., Beitou Dist..<br>Taipei City 112, Taiwan |
| Phone:         | +886-2-2891-8636                                                            |
| Contact:       | Glen Tseng<br>glen.tseng@iweecare.com                                       |

#### Identification of the Device: 5.3

| Proprietary/Trade Name: | Temp Pal (Smart Thermometer Patch) |
|-------------------------|------------------------------------|
| Model Number:           | STP-MB01-1                         |
| Regulation Name:        | Clinical Electronic Thermometer    |
| Review Panel:           | General Hospital                   |
| Regulation Number:      | 880.2910                           |
| Product Code:           | FLL                                |
| Device Class:           | II                                 |

# 5.4 Identification of the Predicate Device:

| Predicate Device Name: | Temp Pal           |
|------------------------|--------------------|
| Model Number:          | STP-MB01-1         |
| 510(k) Number:         | K202603            |
| Manufacturer:          | iWEECARE Co., Ltd. |
| Regulation Number:     | 880.2910           |
| Product Code:          | FLL                |
| Device Class:          | Class II           |

## 5.5 Intended Use/ Indications for Use of the Device

Temp Pal is a battery-operated electronic device with intended use of measuring and monitoring human armpit temperature continuously via wireless signal transmission 

{4}------------------------------------------------

of the measuring result. This system is reusable and intended for armpit temperature monitoring for persons of all age. The temperature data of device is not intended to replace the advice, diagnosis, nor treatment recommendations of doctor. Temp Pal can be used at home and healthcare center.

## 5.6 Device Description

The Temp Pal is designed for the following: A comprehensive Android and iOS App are provided to access Temp Pal from a smart device. It is used for measuring and monitoring armpit temperature in real-time continuously and remotely via Bluetooth to smart device.

- · The subject device could measure and monitor temperature in real-time continuously and remotely via Bluetooth to smart phone.
- . The Temp Pal is the combination device of thermometer and Bluetooth communication unit intended to be worn at axilla to monitor the armpit temperature continuously. The subject device is a direct mode clinical thermometer where the output temperature is not adjusted. For the monitoring operation, switch the thermometer on and stick the thermometer in the user's axilla. The thermometer will make a Bluetooth connection between the thermometer and the receiver automatically (User should setup Bluetooth properly on receiver). Then the thermometer starts to measure the body temperature. The wireless thermometer uses a rechargeable battery for operation. When the battery is low, internal circuit will detect the low battery condition automatically and send "low battery" signal through Bluetooth communication unit to receiver.
- . power source Rechargeable Battery 3.7V/10mAh
- . operation mode Direct mode
- . measuring results display method The measuring results are transmitted to smart phone and display by APP.
- data communication method .

{5}------------------------------------------------

Wireless 2.4G Bluetooth BLE.

# 5.7 Substantial Equivalence Determination

The Temp Pal (Smart Thermometer Patch) submitted in this 510(k) file is substantially equivalent in intended use, principles of operation, safety and performance to the cleared Temp Pal (K202603). Differences between the devices are cited.

{6}------------------------------------------------

| Item                                  | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comparison                                                                                                                                                                                                                                                                                                          |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No.                            | K222588                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | K202603                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                     |
| Proprietary Name                      | Temp Pal (Smart<br>Thermometer Patch)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Temp Pal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | N/A                                                                                                                                                                                                                                                                                                                 |
| Model                                 | STP-MB01-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | STP-MB01-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                     |
| Manufacturer                          | iWEECARE Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | iWEECARE Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                     |
| Product Code                          | FLL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | FLL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                |
| Classification                        | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                |
| Indications for use                   | Temp Pal is a battery-<br>operated electronic<br>device with intended<br>use of measuring and<br>monitoring human<br>armpit temperature<br>continuously via<br>wireless signal<br>transmission of the<br>measuring result. This<br>system is reusable and<br>intended for armpit<br>temperature monitoring<br>for persons of all age.<br>The temperature data of<br>device is not intended to<br>replace the advice,<br>diagnosis, nor treatment<br>recommendations of<br>doctor. Temp Pal can be<br>used at home and<br>healthcare center.                                                                          | The Temp Pal is a<br>battery-operated<br>electronic device with<br>intended use of<br>measuring and<br>monitoring human<br>armpit temperature<br>continuously via<br>wireless signal<br>transmission of the<br>measuring result.<br>This system is reusable<br>and intended for armpit<br>temperature monitoring<br>for<br>persons over two years<br>old.                                                                                                                                                                                                                                                                     | Similar<br>Both devices are designed<br>to measure and monitor<br>human armpit temperature<br>continuously via wireless<br>signal transmission of the<br>result.<br>The subject device expands<br>the patient population to<br>include child under 2 years<br>old (comparing with<br>predicate device<br>(K202603). |
| Thermometer                           | Clinical Electronic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Clinical Electronic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                |
| Item                                  | Subject device<br>Temp Pal (Smart<br>Thermometer Patch)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate device<br>Temp Pal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison                                                                                                                                                                                                                                                                                                          |
| Type                                  | Thermometer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Thermometer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                     |
| Principle of<br>Operation             | For the monitoring<br>operation, switch the<br>thermometer on and<br>stick the thermometer in<br>the user's axilla.<br>The thermometer will<br>make a Bluetooth<br>connection between<br>the<br>thermometer and the<br>receiver automatically<br>(User should setup<br>Bluetooth properly on<br>receiver).<br>Then the thermometer<br>starts to measure the<br>body temperature by<br>means of testing the<br>NTC resistor's<br>resistance value and<br>calculates the body<br>temperature every 10<br>seconds continuously<br>and sends the<br>temperature data to the<br>receiver through<br>Bluetooth connection. | For the monitoring<br>operation, switch the<br>thermometer on and<br>stick the thermometer<br>in the user's axilla.<br>The thermometer will<br>make a Bluetooth<br>connection between<br>the<br>thermometer and the<br>receiver automatically<br>(User should setup<br>Bluetooth properly on<br>receiver).<br>Then the thermometer<br>starts to measure the<br>body temperature by<br>means of testing the<br>NTC resistor's<br>resistance value and<br>calculates the body<br>temperature every<br>10/30/60 seconds<br>continuously and<br>sends the temperature<br>data to the receiver<br>through Bluetooth<br>connection. | Similar<br>Both devices measure the<br>body temperature<br>continuously by Bluetooth<br>connection between the<br>patch and smart device.<br>Subject device keeps only<br>10 seconds measure<br>frequency.                                                                                                          |
|                                       | Anatomical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Patients' armpit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Patients' armpit                                                                                                                                                                                                                                                                                                    |
| Item                                  | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comparison                                                                                                                                                                                                                                                                                                          |
| Application                           | Temp Pal (Smart Thermometer Patch)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Temp Pal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                     |
| Sensor                                | NTC Resistor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | NTC Resistor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                |
| Signal                                | Wireless 2.4G                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Wireless 2.4G                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                |
| Transmission                          | Bluetooth BLE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Bluetooth BLE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                |
| Display                               | iOS device display,<br>Android device display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | iOS device display,<br>Android device display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                |
| Working Voltage                       | DC3.7V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | DC3.7V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                |
| Power Requirements                    | Rechargeable Battery<br>3.7V/10mAh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Rechargeable Battery<br>3.7V/10mAh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                |
| Materials                             | TPE, 304 Stainless Steel Probe and double-sided medical adhesive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | TPE, 304 Stainless Steel Probe and double-sided medical adhesive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                |
| Maximum continuous measuring time     | 24 hours                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 24 hours                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                |
| Response time                         | Require up to 5 minutes to reach stable reading                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Require up to 5 minutes to reach stable reading                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                |
| Device size                           | Size: 28.5 x 26.8 x 3.52 mm, Weight: 3.6 g                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Size: 28.5 x 26.8 x 3.52 mm, Weight: 3.3 g                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similar                                                                                                                                                                                                                                                                                                             |
| Reusability                           | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                |
| Temperature range (Measurement Range) | 77-113°F (25-45°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 77-113°F (25-45°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                |
| Operation Environment                 | 41-113°F (5-45°C), 15-95%RH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 41-113°F (5-45°C), 15-95%RH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                |
| Storage condition                     | -4-131°F (-20-55°C), 15-95%RH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | -4-131°F (-20-55°C), 15-95%RH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                |
| Accuracy                              | 0.09 °F (± 0.05 °C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 0.09 °F (± 0.05 °C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                |
| Display                               | 0.01 °F (± 0.01 °C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 0.01 °F (± 0.01 °C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                |
| Item                                  | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comparison                                                                                                                                                                                                                                                                                                          |
| resolution                            | Temp Pal (Smart Thermometer Patch)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Temp Pal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                     |
| Valid transmission distance           | Up to 5 meters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Up to 5 meters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                |
| Shelf Life                            | 24 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 13 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Different<br>Reliability up to 24 months for subject device is tested.                                                                                                                                                                                                                                              |
| APP name                              | Temp Pal APP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Temp Pal APP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                                                                                                                                 |
| APP description                       | The Temp Pal APP provides an interface to receive and display the temperature data transmitted from the Temp Pal device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | The Temp Pal APP provides an interface to receive and display the temperature data transmitted from the Temp Pal device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                |
| APP Operation Platform                | iOS and Android                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | iOS and Android                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                |
| APP Temperature display               | Display in either °C or°F                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Display in either °C or°F                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                |
| APP Data Export                       | Enables temperature data to be exported via e-mail as csv file.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Enables temperature data to be exported via e-mail as csv file.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                |
| APP Battery Capacity                  | Display with 5 scale                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Display with 5 scale                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                |
| Operation mode                        | Direct mode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Direct mode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                |
| Biocompatibility                      | Biocompatibility, including In Vitro Cytotoxicity, Skin Irritation, and Skin Sensitization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Biocompatibility, including In Vitro Cytotoxicity, Skin Irritation, and Skin Sensitization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                |
| Item                                  | Subject device…

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K222588](https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K222588)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
