K221800 · Zhangzhou Easepal Medical Science and Technology Co., Ltd. · FLL · Dec 14, 2022 · General Hospital
Device Facts
Record ID
K221800
Device Name
Infrared Thermometer (Model: Y20001, Y20002)
Applicant
Zhangzhou Easepal Medical Science and Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Dec 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infrared Thermometer (Model: Y20001, Y20002) is intended for the intermittent measurement and monitoring of human body temperature from forehead. The device is indicated for use for people of two months and above at homecare and in hospital.
Device Story
Hand-held, battery-powered infrared thermometer; measures forehead temperature via non-contact infrared energy detection (1-3cm distance). Device uses thermopile sensor for target reading and thermistor for ambient temperature compensation. Operates by converting infrared radiant energy into electrical signals and then into temperature values. Used by clinicians or patients in home/hospital settings. Features include LCD display, memory for 32 measurements, buzzer, and unit switching (°C/°F). Output provides body temperature readings to assist in clinical decision-making and patient monitoring.
Clinical Evidence
Clinical accuracy testing conducted per ISO 80601-2-56:2017. Study evaluated 150 subjects across three age groups: infants (2 months to 1 year), children (>1 to 5 years), and adults (>5 years), with 50 subjects per group. Fever prevalence exceeded 30% in each group. Results demonstrated compliance with ISO 80601-2-56:2017+AMD2018 requirements.
Technological Characteristics
Hand-held infrared thermometer; thermopile sensor with integrated thermistor; ABS plastic enclosure; battery-powered (2x 1.5V AAA); LCD display. Complies with ISO 80601-2-56, ASTM E1965-98, IEC 60601-1, and IEC 60601-1-2. Biocompatibility per ISO 10993-5, -10, -23.
Indications for Use
Indicated for intermittent measurement and monitoring of human body temperature from the forehead in patients two months of age and older in homecare and hospital settings.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a larger font size than the name. The name is written as "U.S. FOOD & DRUG ADMINISTRATION".
December 14, 2022
Zhangzhou Easepal Medical Science And Technology Co.,Ltd. % Jet Li Regulation Manager Guangdong Jianda Medical Technology Co., Ltd. 906 Room, Longxiang Garden, Tianhe District Guangzhou, Guangdong 510630 China
Re: K221800
Trade/Device Name: Infrared Thermometer (Model: Y20001, Y20002) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: November 14, 2022 Received: November 14, 2022
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
(24. SER. B. — 967) 1. hdl., (24. SER. B.
Page 2
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Danil Walloschek
David Wolloscheck, Ph.D. For Joyce M. Whang, Ph.D. Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221800
Device Name Infrared Thermometer (Model: Y20001, Y20002)
Indications for Use (Describe)
The Infrared Thermometer (Model: Y2001, Y20002) is intended for the intermittent measurement and monitoring of human body temperature from forehead. The device is indicated for use for people of two months and above at homecare and in hospital.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K221800
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 870.1130. Summary Prepared Date:14 December 2022
#### 1. Submitter Information
- Sponsor Company Name: Zhangzhou Easepal Medical Science And Technology Co.,Ltd. �
- � Address: West of the 4th Floor, No. 7 Building, Easepal Factory, No. 228 Jiaosong Road, Jiaomei Town, Zhangzhou Taiwanese Investment Zone, Fujian, China
- � Phone: +86-18905963518
- Contact Person (including title): Flower Huang/ Engineer �
- E-mail: flower@easepal.com.cn �
#### Application Correspondent:
- Guangdong Jianda Medical Technology Co.,Ltd. �
- Address: 906 Room, Longxiang Garden, Tianhe district, Guangzhou, China �
- Contact Person: Mr. Jet Li �
- Tile: Regulation Manager �
- � Tel: +86- 13512755282
- Email: jianda-lee@foxmail.com �
#### 2. Subject Device Information
| Common Name: | Infrared Thermometer |
|----------------------|-----------------------------------|
| Trade Name: | Infrared Thermometer |
| Model names: | Y20001,Y20002 |
| Classification Name: | Thermometer, electronic, clinical |
| Review Panel: | General Hospital |
| Product Code: | FLL |
| Regulation Number: | 21 CFR 880.2910 |
| Regulation Class: | II |
{4}------------------------------------------------
### 3. Predicate Device Information
| Sponsor: | SHENZHEN EVERBEST MACHINERY INDUSTRY CO.,<br>LTD. |
|----------------------|---------------------------------------------------|
| Common Name: | Infrared Thermometer |
| Trade Name: | Infrared Thermometer |
| Classification Name: | Clinical Electronic Thermometer |
| Model names: | DT-8806S, DT-8807S |
| 510(k) number: | K203170 |
| Review Panel: | General Hospital |
| Product Code: | FLL |
| Regulation Number: | 21 CFR 880.2910 |
| Regulation Class: | II |
### 4. Device Description
Infrared thermometer (Model: Y20001 , Y20002) are a hand-held, battery powered, infrared Thermometer that coverts a user's forehead temperature, using the infrared energy emitted in the area around the user's forehead when measure from 1-3cm of the subject's forehead with no contact. The temperature will displayed on the screen, and the reference body site is axilla. Infrared thermometer (Model: Y20001, Y20002) use a thermopile sensor with integrated thermistor for the target reading and a thermistor mounted in the thermometer for ambient temperature readings. The device consists of measuring sensor, PCB, buttons, a LCD and an enclosure.
The difference between two models are size, weight and button settings. Y20001 has four buttons(Test button, °C/°F button, "B/S"button, "M" key) and Y20002 has two buttons(Function button, Measurement button). The functions of the two models are same. Infrared thermometer (Model: Y20001,Y20002) have the following features:
1) The device is intended to be reusable for home use and clinical use.
2) Switching of temperature unit between °C and °F.
3) The latest 32 sets of memory for measuring human body and object; the user can view or delete the previous measurement results.
4) Buzzer on or off to set the prompt tone on or off.
{5}------------------------------------------------
5) Prompt tone function.
- 6) Low battery indication, and auto power-off
# 5. Indication for use
The Infrared Thermometer (Model: Y20001, Y20002) is intended for the intermittent measurement and monitoring of human body temperature from forehead. The device is indicated for use for people of two months and above at homecare and in hospital.
# 6. Comparison to Predicate Device
| Elements of<br>Comparison | Subject Device | Predicate Device 1# | Verdict |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Manufacturer | Zhangzhou Easepal Medical<br>Science And Technology Co.,Ltd. | SHENZHEN EVERBEST<br>MACHINERY INDUSTRY CO.,<br>LTD. | -- |
| 510 (k) Number | K221800 | K203170 | -- |
| Product Name | Infrared Thermometer | Infrared Thermometer | -- |
| Models | Y20001, Y20002 | DT-8806S, DT-8807S | -- |
| Intended Use | The Infrared Thermometer<br>(Model: Y20001, Y20002) is<br>intended for the intermittent<br>measurement and monitoring of<br>human body temperature from<br>forehead. The device is indicated<br>for use for people of two months<br>and above at homecare and in<br>hospital. | The Infrared thermometer is intended<br>for the intermittent measurement of<br>human body temperature from<br>forehead for people of all ages. The<br>device is reusable for home use and<br>clinical use. | Different,<br>See Note 1 |
| Measurement<br>Method | Infrared radiation detection,<br>adjusted mode | Infrared radiation detection, adjusted<br>mode | Same |
| Principle of<br>operation | It detects the radiant energy of<br>infrared radiation emitted by<br>objects or human body and<br>converts it into electrical signals,<br>and eventually into a temperature<br>value | It detects the radiant energy of<br>infrared radiation emitted by objects<br>and converts it into electrical signals,<br>and eventually into a temperature<br>value | Same |
| Measurement<br>Place | Forehead<br>Surface | Forehead<br>Surface | Same |
| Sensor type | Thermopile | Thermopile | Same |
| Measurement<br>Distance | 1-3cm | 8806S: 1 cm - 10 cm<br>8807S: 1 cm-4 cm | Different,<br>See Note 2 |
| Measuring<br>Range | 34.0°C to 42.9°C (93.2 to 109.2 °<br>F) | 32.0°C -42.5°C (89.6 to 108.5 ° F) | Different,<br>See Note 3 |
| Accuracy | $\pm$ 0.3°C/0.5°F | ±0.2°C(0.4°F) within 36.0°C -<br>39.0°C, (96.8°F - 102.2°F) | Different,<br>See Note 4 |
| Elements of<br>Comparison | Subject Device | Predicate Device 1# | Verdict |
| | | ±0.3°C(0.6°F) within 32°C -35.9°C<br>(89.6°F -96.6°F) and 39.1 °C-42.5°C<br>(102.3°F -108.5°F) | |
| Display<br>Resolution | 0.1°C/0.1°F | 0.1°C/0.1°F | Same |
| Display Screen | LCD | LCD | Same |
| Buzzer | Yes | Yes | Same |
| °C/°F<br>switchable | Yes | Yes | Same |
| Memory | 32 sets | 32 sets | Same |
| Auto power-off<br>Same while no<br>operation | Yes | Yes | Same |
| Power Supply | 2 * 1.5V AAA | 2 * 1.5V AAA | Same |
| Contact<br>materials | ABS | ABS | Same |
| Operating<br>Conditions | Temperature:10°C-40°C;<br>Relative humidity: ≤95% | Temperature: 10.0°C~40°C (50.0°F<br>~104.0°F);<br>Relative humidity: ≤85% | Different,<br>See Note 5 |
| Storage<br>Conditions | Temperature:-20°C-55°C;<br>Relative humidity: ≤95%RH | 0 to 50°C (32 to 122°F) ;RH≤85% | |
| Dimension | Y20001:145 mm * 93 mm * 38<br>mm<br>Y20002:155 mm *40 mm *40<br>mm | 128 * 74 * 36 mm | Different,<br>See Note 6 |
| Weight | Y20001: 110 g(excluding<br>batteries )<br>Y20002: 59g (excluding batteries) | 104.5 g | |
| Performance | Compliance with ASTME 1965;<br>ISO80601-2-56 | Compliance with ASTME 1965;<br>ISO80601-2-56 | Same |
| Clinical<br>accuracy | Meet the requirements of ISO<br>80601-2-56:2017, ASTM 1965-<br>98(2016) | Meet the requirements of ISO 80601-<br>2-56:2017, ASTM 1965-98(2016) | Same |
| Biocompatibility | All the patient contacting<br>materials are compliance with<br>ISO 10993-1,ISO 10993-5,<br>ISO 10993-10,ISO 10993-23.<br>Under the condition of this study | All the patient contacting materials<br>are compliance with ISO 10993.<br>Under the condition of this study the | Same |
| | the device is non-cytotoxic , non-<br>sensitizing and non-irritating. | device is non-cytotoxic, non-<br>sensitizing and non-irritating. | |
| Elements of<br>Comparison | Subject Device | Predicate Device 1# | Verdict |
| Electrical Safety | Compliance with IEC 60601-1 | Compliance with IEC 60601-1 | Same |
| EMC | Compliance with IEC 60601-1-2 | Compliance with IEC 60601-1-2 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
#### Note 1
The population of the predicate device is wider than the subject device. The user population of the subject device is subset of the predicate. Therefore, this difference does not raise new safety and effectiveness questions.
#### Note 2
The measurement distance is different. But we had conducted clinical accuracy testing on subject device according to ISO 80601-2-56:2017, and the clinical accuracy test result can demonstrate that the subject device can meet the requirement on this measurement distance. So, the different does not raise different questions of safety and effectiveness.
#### Note 3
The measurement range of subject device is different from the predicate device. The performance testing shows that the subject device complies with the performance standard ISO 80601-2-56 and ASTM E1965-98. Therefore, this difference does not affect the performance and accuracy.
### Note 4
The subject device's accuracy are wider than the predicate device within 36.0℃ - 39.0℃, (96.8°F -102.2°F) ,these are based on device's performance standards. These differences do not affect the performance and accuracy according to our Test Report (ASTM E1965- 98), ISO 80601-2-56 Test Report, and our Clinical Test Report.
#### Note 5
The operation and storage environment of subject device is different from the predicate device, but the measurement accuracy of subject device has been demonstrated to comply with the requirements of standards IEC 60601-1 and ISO 80601-2-56 in operation and storage environment. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness.
#### Note 6
The "Dimensions", "Weight" of subject device is similar with predicate devices, both of them meet the requirement of safety and essential performance standard IEC 60601-1-2 and IEC 60601-1-2 and IEC 60601-1
{8}------------------------------------------------
11. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
#### 7. Non-Clinical Test Summary
#### 7.1. Electromagnetic Compatibility and Electrical Safety Test
The subject device has been tested in compliance with the following standards:
1) AAMI/IEC 60601-1:2005+AMD 1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
2) IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests 3) IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirementsfor basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
#### 7.2. Biocompatibility Test
The subject device is considered for surface contacting, limited exposure (Less than 24 hours), it has been tested in compliance with the following standards:
1) ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
2) ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
3) ISO 10993-23:2021 Biological evaluation of medical devices - Part 23:Tests for irritation
#### 7.3. Performance Test-Bench
The subject device has been tested in compliance with the following standards:
1) ISO 80601-2-56:2017+AMD2018 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement 2) ASTM E1965-98 (Reapproved 2016) Standard Specification for InfraredThermometers for Intermittent Determination of Patient Temperature
#### 7.4. Software Verification
The softw are documentation of the subject device was provided in accordance with FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005.
#### 8. Clinical Accuracy Test Summary
{9}------------------------------------------------
Both models(Y20001, Y20002) have undergone clinical accuracy testing, both clinical studies are based on the requirement of ISO 80601-2-56:2017. The test reports showed the clinical performance of the subject device complied with the requirement of ISO 80601-2-56:2017+AMD2018 .
Each clinical study evaluated 150 subjects, and the infrared thermometer was evaluated for the patient populations for two months and above. All subjects were divided into three age groups with 50 subjects in each group, including (1)Infants –two months to one year, (2)Children - greater than one to five years; (3)Adults - greater than five years old. The proportion of fever patient is more than 30% in each age groups.
# 9. Conclusion
The subject device Infrared Thermometer (Model: Y20002) are substantially equivalent to the predicate device (K203170). This conclusion is based upon comparison of the intended use, technological characteristics and applicable safety standards. Any differences in the technological characteristics do not raise any new issues or different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.