Infrared Forehead Thermometer, Model JZK-601, JZK-602, JZK-603

K203707 · Shenzhen Zhengkang Technology Co., Ltd. · FLL · Mar 12, 2021 · General Hospital

Device Facts

Record IDK203707
Device NameInfrared Forehead Thermometer, Model JZK-601, JZK-602, JZK-603
ApplicantShenzhen Zhengkang Technology Co., Ltd.
Product CodeFLL · General Hospital
Decision DateMar 12, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Infrared forehead thermometer (Models JZK-601, JZK-602, JZK-603) is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead at a distance of 1-5cm for people of all ages. The device is reusable for home use and clinical use.

Device Story

Non-contact infrared thermometer; measures infrared energy radiation from forehead; distance 1-5cm. Features: body/object modes; LCD display; 32-set memory; buzzer; backlight; low battery indicator; auto power-off. Used in home or clinical settings by patients or clinicians. Input: infrared radiation; Output: temperature reading on LCD. Benefits: non-invasive, rapid temperature monitoring for all ages.

Clinical Evidence

Clinical study conducted per ASTM E1965-98. Randomized, simple blind, homologous control, pairing design. Evaluated 200 subjects across age groups: 50 newborns (0-3 months), 50 infants (>3 months-1 year), 50 children (>1-5 years), and 50 people (>5 years). Results confirmed compliance with ASTM E1965-98 performance requirements.

Technological Characteristics

Materials: ABS plastic. Sensing: Thermopile infrared sensor. Energy: DC 3V (2x AA or AAA alkaline batteries). Form factor: Handheld device with LCD. Connectivity: None. Sterilization: Not applicable. Software: Embedded firmware for signal processing and memory management. Standards: ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ASTM E1965-98, ISO 10993-5/10.

Indications for Use

Indicated for intermittent measurement of human body temperature via forehead in patients of all ages (newborns to adults). Suitable for home and clinical use.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. March 12, 2021 ShenZhen ZhengKang Technology Co., Ltd. % Iris Wang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town Shenzhen, Guangdong 518000 China Re: K203707 Trade/Device Name: Infrared Forehead Thermometer, Model JZK-601, JZK-602, JZK-603 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: December 8, 2020 Received: December 18, 2020 Dear Iris Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K203707 Device Name Infrared Forehead Thermometer, Model JZK601, JZK602, JZK603 Indications for Use (Describe) The Infrared forehead thermometer (Models JZK-602, JZK-603) is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead at a distance of 1-5cm for people of all ages. The device is reusable for home use and clinical use. Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92. ## 1. Administrative Information | Date Prepared | March 10, 2021 | |--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | ShenZhen ZhengKang Technology Co., Ltd.<br>Address:<br>2&3/F, Building A, No.3 FuXing Yi Lane, HeHua Community,<br>PingHu Street, LongGang District, ShenZhen, GuangDong,<br>China.<br>Contact person: Huayong Yang<br>TEL: +86 -0755-83260864<br>E-Mail: 893488645@qq.com | | Submission<br>Correspondent | Shenzhen Joyantech Consulting Co., Ltd.<br>1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian<br>Avenue, Xili Town, Nanshan District, Shenzhen, Guangdong<br>Province, China.<br>Contact person: Ms. Iris Wang; Mr. Field Fu<br>E-Mail: iris@cefda.com; field@cefda.com<br>Image: [logo] | | Establishment<br>registration number | 3015697152 | # 2. Device Information | Device Name: | Infrared Forehead Thermometer | |----------------------|-----------------------------------| | Model: | JZK-601, JZK-602, JZK-603 | | Classification Name: | Thermometer, Electronic, Clinical | | Review Panel: | General Hospital | | Device Class: | 2 | | Regulation Number: | 880.2910 | | Product Code: | FLL | {5}------------------------------------------------ ## 3. Predicate Device | Manufacturer: | Shenzhen Calibeur Industries Co., Ltd. | |--------------------|----------------------------------------| | Device Name: | Infrared Thermometer | | Model: | DT-8836T, DT-8836P | | 510(k) Number: | K191251 | | Regulation Number: | 880.2910 | | Product Code: | FLL | ## 4. Device Description Infrared forehead thermometers (includes model JZK-602, JZK-602, JZK-603) are intended to measure the body temperature through receiving infrared energy radiation via the forehead for people of all ages. These thermometers have the capability to measure temperature via body mode or object mode, and the temperature is directly shown on the LCD display. These thermometers have the following features: - The device is intended to be reusable for home use and clinical use. 1) - The device is mainly composed of infrared sensor, signal receiving processor, kevs, buzzer, 2) LCD display. - 3) Switching of temperature unit between °C and °F. - The latest 32 sets of memory for measuring human body and object; the user can view or delete 4) the previous measurement results. - 5) Buzzer on or off to set the prompt tone on or off. - 6) Prompt tone function and backlights function. - 7) The prompt limit setting function. - Low battery indication, and auto power-off. 8) ## 5. Intended Use/ Indications for Use The Infrared forehead thermometer is a non-contact infrared thermometer (Models JZK-601, JZK-602, JZK-603) intended for the intermittent measurement of human body temperature from forehead at a distance of 1-5cm for people of all ages. The device is reusable for home use and clinical use. ## 6. Comparison with Predicate Device The subject device infrared forehead thermometer, Model: JZK-602, JZK-603 is substantially equivalent to the predicate device (K191251). This conclusion is based upon comparison on intended use, technological characteristics, materials and applicable safety standards. The difference between the subject device and predicate device do not raise any issues on the device safety and effectiveness. {6}------------------------------------------------ | Items | Subject Device (K203707)<br>(Model: JZK-601, JZK-602,<br>JZK-603) | Predicate Device (K191251) | Comparison | |--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------| | Regulation number | 880.2910 | 880.2910 | Same | | Product code | FLL | FLL | Same | | Intended Use/<br>Indications for use | The Infrared forehead<br>thermometer (Models JZK-<br>601, JZK-602, JZK-603) is a<br>non-contact infrared<br>thermometer intended for the<br>intermittent measurement of<br>human body temperature from<br>forehead at a distance of 1-<br>5cm for people of all ages.<br>The device is reusable for<br>home use and clinical use. | The Infrared thermometer is a<br>non-contact infrared<br>thermometer intended for the<br>intermittent measurement of<br>human body temperature from<br>forehead for people of all<br>ages. The device is reusable<br>for home use and clinical use. | Comparable<br>(1) See below | | Measurement Method | Infrared radiation detection | Infrared radiation detection | Same | | Measurement Range | 32 - 42.9°C (89.6 - 109.2°F) | 32.0 - 42.5°C (89.6 - 108.5°F) | Comparable<br>(2) See below | | Accuracy | ±0.3°C(0.5°F) within 32 -<br>34.9°C (89.6~94.8°F)<br>±0.2°C (0.4°F) within 35 -<br>42°C (95.0~107.6°F)<br>±0.3°C(0.5°F) within 42.1 -<br>42.9°C (107.8~109.2°F) | ±0.2°C (0.4°F) within 35.0°C -<br>42.0°C (95.0°F ~ 107.6°F),<br>±0.3°C(0.5°F) other range | Same | | Display | 0.1°C /0.1°F | 0.1°C(0.1°F) | Same | | Measurement<br>distance | 1-5cm | ≤3cm | Comparable<br>(3) See below | | Measurement place | Forehead | Forehead | Same | | Response time | ≤3 seconds | 1s | Comparable<br>(4) See below | | Sensor type | Thermopile | Thermopile | Same | | Scale Selection | °C /°F | °C /°F | Same | | Memory | 32 sets | 60 sets | Comparable<br>(5) See below | | Buzzer | Yes | Yes | Same | | Shelf Life | 5 years | Not Available | Acceptable (6)<br>See below | | IP classification | IP22 | Not Available | Acceptable (6)<br>See below | | Auto power-off while<br>no operation | Yes | Yes | Same | | Power supply | JZK-601 and JZK-603: DC 3V<br>(2× AA 1.5V Alkaline<br>batteries)<br>JZK-602: DC 3V (2× AAA) | 2 * 1.5V AAA | Comparable<br>(7) See below | | | 1.5V Alkaline batteries) | | | | Display screen | LCD | LCD | Same | | Contact materials | ABS | ABS | Same | | Backlight | JZK-601 and JZK-603:<br>Three color backlight (green,<br>yellow, and red)<br>JZK-602: two color backlight<br>(green and red) | Not Available | Acceptable (6)<br>See below | | Operation<br>Environment | 10 - 40°C (50 - 104°F)<br>Humidity:≤95% | 10 - 40°C (50 - 104 °F)<br>RH 15 - 95% | Comparable<br>(8) See below | | Storage Environment | -20 - 60°C (-4 - 140°F)<br>Humidity: ≤95% | -25 - 55°C (-13 - 131°F)<br>RH:15 - 95% | Comparable<br>(9) See below | | Dimension | JZK-601 and JZK-603:<br>154 x 96 x 42mm<br>JZK-602: 143 x 81 x 36mm | 153.8 x 62.4 x 62.4 mm | Comparable<br>(10) See below | | Weight | JZK-601 and JZK-603: 93g<br>JZK-602: 86g | 96g | Comparable<br>(10) See below | | Conformance<br>standard | ISO80601-2-56 (performance)<br>IEC60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ASTM E1965-98 | ISO80601-2-56 (performance)<br>IEC60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ASTM E1965-98 | Same | | Patient contact<br>materials | ABS | ABS | Same | | Biocompatibility | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Same | {7}------------------------------------------------ Discussion of Comparable Items: (1) Indications for Use The subject device indications for use includes the models and measurement distance. Therefore, this difference does not affect the performance and accuracy #### (2) Measurement Range The measurement range of subject device is different from the predicate device. The performance testing shows that the subject device complies with the performance standard ISO 80601-2-56 and ASTM E1965-98. Therefore, this difference does not affect the performance and accuracy. #### (3) Measurement distance The measurement distance of subject device is a little different from the predicate device. The performance testing shows that the subject device complies with the performance standard ISO 80601-2-56 and ASTM E1965-98. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness of the subject device. Page 4 of 7 {8}------------------------------------------------ #### (4) Response time The response time of subject device is different from the predicate device. However, the software has been validated according to FDA's software guidance. The performance testing shows that the subject device complies with performance standard during performance testing ISO 80601-2-56 and ASTM E1965-98 and clinical testing. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness of the subject device. #### (5) Memory The memory of subject device is different from predicate device. The software function of memory of subject device has been verified during software verification according to FDA's software guidance, and the performance testing shows that the subject device complies with performance standard ISO 80601-2-56 and ASTM E1965-98performance and clinical testing. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness of the subject device. #### (6) Shelf Life, IP Classification, Backlight The shelf life, ingress protection, and backlight of the predicate were not available. However, the device is comparable to other devices on the market. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness of the subject device. #### (7) Power supply The power supply of subject device for model: JZK-602 is the same with the predicate device, but the power supply of subject device for model: JZK-601/JZK-603 is different from the predicate device. However, the both of the subject device has been demonstrated to comply with the requirements of electrical safety IEC 60601-1 and Electromagnetic Compatibility IEC 60601-1-2. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness. #### (8) Operation Environment The operation environment of subject device is different from the predicate device, but the measurement accuracy of subject device has been demonstrated to comply with the requirements of standards IEC 60601-1 and ISO 80601-2-56 in operation environment. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness. #### (9) Storage Environment The operation and storage environment of subject device is different from the predicate device, but the subject device has been demonstrated to comply with the requirements of standards IEC Page 5 of 7 {9}------------------------------------------------ 60601-1 and ISO 80601-2-56. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness. #### (10) Dimension, Weight The dimension and weight of subject device is different from the subject device. The difference is caused by their different appearance and construction, but the electrical safety, electromagnetic compatibility, performance of subject device has been evaluated to meet the requirements of the standards IECS60601-1, IEC 60601-1-2, ASTM E1965-98 and ISO 80601-2-56. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness. ## 7. Non-Clinical Test Summary ### 7.1. Electromaqnetic Compatibility and Electrical Safety Test The subject device has been tested in compliance with the following standards: 1) AAMI/IEC 60601-1:2005+AMD 1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance 2) IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests 3) IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment ## 7.2. Biocompatibility Test The subject device has been tested in compliance with the following standards: 1) ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity - 2) ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization ## 7.3. Performance Test-Bench The subject device has been tested in compliance with the following standards: 1) ISO 80601-2-56:2017+AMD2018 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement {10}------------------------------------------------ 2) ASTM E1965-98 (Reapproved 2016) Standard Specification for InfraredThermometers for Intermittent Determination of Patient Temperature ## 7.4. Software Verification The software documentation of the subject device was provided in accordance with FDA guidance document "Guidance for the Content of Premarket Submissions for Software in Medical Devices" issued May 11, 2005. # 8. Clinical Study The clinical study is conducted as the requirement of ASTM E1965-98 (Reapproved 2016). The test report showed the clinical performance of the subject device complied with the requirement of ASTM E1965-98(Reapproved 2016). The clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation. The clinical study evaluated 200 subjects, and the infrared forehead thermometer was evaluated in all age groups. All subjects were divided into three age groups, including (1)Infants newborn to one year, (2)Children - greater than one to five years; (3)Adults - greater than five years old. 50 newborns 0 up to 3 months, 50 infants older than 3 months up to 1 year, 50 children older than 1 year and younger than 5 years, and 50 people older than 5 years are selected. The clinical test report showed the clinical performance of the subject devices complied with the requirement of ASTM E1965-98(Reapproved 2016). # 9. Conclusion The subject device Infrared Forehead Thermometer (Model: JZK-601, JZK-602, JZK-603) is substantially equivalent to the predicate device (K191251). This conclusion is based upon comparison on intended use, technological characteristics and applicable safety standards. Any difference in the technological characteristics does not raise any new issues or concerns of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...