Infrared Forehead Thermometer, Model LFR30B, LFR50, LFR60
K202687 · Shenzhen Lepu Intelligent Medical Equipment Co., Ltd. · FLL · Feb 18, 2021 · General Hospital
Device Facts
Record ID
K202687
Device Name
Infrared Forehead Thermometer, Model LFR30B, LFR50, LFR60
Applicant
Shenzhen Lepu Intelligent Medical Equipment Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Feb 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infrared Forehead Thermometer is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead for people of all ages. The device is reusable for home use and clinical use.
Device Story
Non-contact infrared thermometer (models LFR30B, LFR50, LFR60) for intermittent body temperature measurement. Uses thermopile sensor to detect infrared energy from forehead skin; converts energy to temperature (Celsius or Fahrenheit). Battery-powered; handheld device. Used in home or clinical settings by lay users or professionals. Measurement distance 0-5 cm; measurement time 1s. Results displayed on LCD. Data storage up to 99 sets. Provides rapid, non-invasive temperature monitoring to assist in clinical decision-making regarding patient health status.
Clinical Evidence
Clinical comparison and repeatability study conducted per ISO 80601-2-56 across four age groups (0-3 months, 3 months-1 year, 1-5 years, >5 years). Compared subject device against mercury thermometer. Results demonstrated high correlation; clinical bias and repeatability were within acceptable limits defined by ISO 80601-2-56.
Technological Characteristics
Materials: ABS, PC, PMMA. Sensing: Thermopile infrared sensor. Energy: Battery-powered. Connectivity: None. Sterilization: Not applicable. Software: Embedded, developed per IEC 62304. Standards: ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ASTM E1965.
Indications for Use
Indicated for intermittent human body temperature measurement via forehead for all ages. Suitable for home and clinical use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 18, 2021
Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. % Arthur Goddard President FDA Regulatory and Ouality Systems Consultant 31853 Cedar Road Mayfield Heights, Ohio 44124-4445
Re: K202687
Trade/Device Name: Infrared Forehead Thermometer, Model LFR30B, LFR50, LFR60 Regulation Number: 21 CFR 21 CFR 880.2910 Regulation Name: Clinical Electrical Thermometer Regulatory Class: Class II Product Code: FLL Dated: September 3, 2020 Received: September 15, 2020
Dear Arthur Goddard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal. ObGyn. General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202687
#### Device Name
Infrared Forehead Thermometer, Model LFR30B, LFR50, LFR60
Indications for Use (Describe)
The Infrared Forehead Thermometer is a non-contact infermometer intended for the intermittent measurement of human body temperature from forehead for people of all ages. The device is reusable for home use and clinical use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(K) Summary
This summary of 510(K) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(K) Number: K202687
#### ನ. 510(K) Summary
#### 5.1. Date of Preparation: February 18, 2021
#### 5.2. Sponsor
Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. North side of floor 3, BLD 9 BaiWangxin High-Tech Industrial Park Songbai Road, Xili Street, Nanshan District 518055 Shenzhen, Guangdong, CHINA Tel: +86 0755-86952278 Fax: +86 0755-86952278
Contact Person: Aaron Lin Position: General Manager Email: aaron.lin(@lepu-medical.com
#### 5.3. Subject Device Identification
Subject Device Name: Infrared Forehead Thermometer, Model LFR30B, LFR50, LFR60 Common name: Infrared Forehead Thermometer Regulation Name: Clinical Electronic Thermometer Product Code: FLL Regulation Number: 21 CFR 880.2910 Review Panel: General Hospital Classification: II
#### 5.4. Predicate Device
510(k) Number: K191251 Device Name: Infrared Thermometer (DT-8836T, DT-8836P) Manufacturer: Shenzhen Calibeur Industries Co., Ltd.
#### 5.5. Indications for use:
The Infrared Forehead Thermometer is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead for people of all ages. The device is reusable for home use and clinical use.
{4}------------------------------------------------
# 5.6. Device Description
The subject device Infrared Forehead Thermometer has three models: LFR30B, LFR50 and LFR60. The three modes have same operation principle. A thermopile sensor is employed to detect the infrared energy emitted from the surface of the skin of the forehead which is converted into temperature measurement with the unit of ℃ or ºF. The results can be displayed on LCD. The measurement is non-contact with the patient. The measuring distance is 0~5 cm to the forehead and the measuring time is 1s. The Infrared Forehead Thermometer is battery powered.
| Item | Subject Device<br>Infrared Forehead<br>Thermometer | Predicate Device<br>Infrared Thermometer<br>(DT-8836T, DT-8836P)<br>K191251 | Remark |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Device<br>Common/Usual<br>Name | Infrared Forehead<br>Thermometer | Infrared Thermometer | / |
| Device Class | Class II | Class II | SE |
| Product Code/<br>Regulation<br>Number | FLL<br>21 CFR 880.2910 | FLL<br>21 CFR 880.2910 | SE |
| Classification<br>Name(s) | Clinical Electronic<br>Thermometer | Clinical Electronic<br>Thermometer | SE |
| Indications for<br>Use | The Infrared Forehead<br>Thermometer is a non-contact<br>infrared thermometer intended<br>for the intermittent<br>measurement of human body<br>temperature from forehead for<br>people of all ages. The device<br>is reusable for home use and<br>clinical use. | The Infrared thermometer is a<br>non-contact infrared<br>thermometer intended for the<br>intermittent measurement of<br>human body temperature<br>from forehead for people of<br>all ages. The device is<br>reusable for home use and<br>clinical use. | SE |
| Intended Users | Lay user and professional | Lay user and professional | SE |
| Measurement<br>Method | Infrared radiation detection | Infrared radiation detection | SE |
| Measuring Site | Forehead | Forehead | SE |
#### 5.7. Predicate Devices and Subject Device Comparison
{5}------------------------------------------------
| Item | Subject Device<br>Infrared Forehead<br>Thermometer | Predicate Device<br>Infrared Thermometer<br>(DT-8836T, DT-8836P)<br>K191251 | Remark |
|-----------------|----------------------------------------------------|-----------------------------------------------------------------------------|--------------|
| Scale Selection | °C/°F | °C/°F | SE |
| Measuring Range | 32-43°C/89.6-109.4°F | 32.0°C ~42.5°C<br>(89.6 to 108.5 °F) | Discussion 1 |
#### Discussion 1 :
The measuring range of subject device is different with predicate device. The performance testing shows that the subject device complies with performance standard IEC 80601-2-56. The difference with predicate device K191251 doesn't raise any new issues of safety or efficacy.
| Display<br>Resolution | 0.1°C/0.1°F | 0.1°C/0.1°F | SE |
|-----------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------|
| Measuring<br>Accuracy | ±0.3°C (±0.5°F) within<br>34.0~43°C (93.2~109.4°F),<br>±0.4°C (±0.7°F) within<br>32.0~33.9°C (89.6~93.0°F) | ±0.2°C (0.4°F) within<br>35.0°C ~ 42.0°C<br>(95.0°F ~ 107.6°F),<br>±0.3°C(0.5°F) other range | Discussion 2 |
# Discussion 2:
The measuring accuracy of subject device is different with predicate device. The performance testing shows that the subject device complies with performance standard IEC 80601-2-56 and ASTM E 1965. The difference with predicate device K191251 doesn't raise any new issues of safety or efficacy.
| Measure Time | 1s | 1s | SE |
|------------------|-------|------|--------------|
| Measure Distance | 0~5cm | ≤3cm | Discussion 3 |
#### Discussion 3:
The measure distance of subject device is different with predicate device. The performance testing shows that the subject device complies with performance standard IEC 80601-2-56 and ASTM E 1965. The difference with predicate device K191251 doesn't raise any new issues of safety or efficacy.
| Sensor | Thermopile | Thermopile | SE |
|----------------|---------------|------------|--------------|
| Audio Reminder | Buzzer | Buzzer | SE |
| Data Storage | Up to 99 sets | 60 sets | Discussion 4 |
# Discussion 4:
The data storage of subject device is different with predicate device. However, the software was validation according to FDA's software guidance. The performance testing shows that the subject device complies with performance standard IEC 80601-2-56 and ASTM E 1965. The difference with predicate device K191251 doesn't raise any new issues of safety or efficacy.
| Power Supply | Batterv | Batterv | CITY |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------|
| Display Screen | C | T AT | CTT |
| | ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ | UND | DI |
{6}------------------------------------------------
| Item | Subject Device<br>Infrared Forehead<br>Thermometer | Predicate Device<br>Infrared Thermometer<br>(DT-8836T, DT-8836P)<br>K191251 | Remark |
|--------------------------------|----------------------------------------------------|-----------------------------------------------------------------------------|--------------|
| Operating<br>Condition | 5°C~40°C, Relative<br>Humidity 24~90% | 10~40°C (50°F ~104 °F)<br>RH 15~95% | |
| Storage/transport<br>Condition | -20°C~+55°C, Relative<br>Humidity < 95% | -25 ~+55°C (-13~+131°F)<br>RH:15~95% | Discussion 5 |
# Discussion 5:
The operating and storage/transport condition of subject device are different with predicate device. The performance testing shows that the subject device complies with performance standard IEC 80601-2-56 and ASTM E 1965. The difference with predicate device K191251 doesn't raise any new issues of safety or efficacy.
| Human-<br>contacting<br>Material | ABS, PC, PMMA | ABS | Discussion 6 |
|----------------------------------|---------------|-----|--------------|
|----------------------------------|---------------|-----|--------------|
# Discussion 6:
The human-contacting material of subject device is different with predicate device. PC and PMMA are widely used in medical device. The biocompatibility evaluation of the subject device was conducted according to FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Cytotoxicity, sensitization tests are performed. The test reports addressed during the biocompatibility studies (Section 15: Biocompatibility) did not raise any new issues of biocompatible safety. The difference with predicate device K191251 doesn't raise any new issues of safety or efficacy.
| Biocompatibility<br>Information | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | SE |
|---------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|----|
| Reference Body<br>Site | Axilla | Axilla | SE |
| Measuring Mode | Adjusted (Body mode) | Adjusted (Forehead mode) | SE |
| Conformance<br>Standard | ISO 80601-2-56<br>IEC 60601-1<br>IEC 60601-1-2<br>ASTM E1965 | ISO 80601-2-56<br>IEC 60601-1<br>IEC 60601-1-2<br>ASTM E1965 | SE |
{7}------------------------------------------------
#### 5.8. Performance Test Summary
Bench test were conducted to verify that the subject device met all design specifications, as was Substantially Equivalent (SE) to the predicate device.
### 5.8.1. Biocompatibility Testing
The Infrared Forehead Thermometer was assessed against the International Standard ISO 10993-1, "Biological evaluation of medical devices. Part 1. Guidance on selection of tests." The subject device would be classified as a Surface Medical Device in contact with the intact skin for a Limited Duration (<24 hours). The following test were performed for any user contacting material:
| Test | Standard | Results |
|-------------------------------------------------------------|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| In Vitro Cytotoxicity<br>Study (MTT Method) | ISO 10993-5 | Under the conditions of this study, the MEM<br>test extracts would be considered no<br>cytotoxicity potential. The negative controls,<br>blank controls, and the positive controls<br>performed as anticipated. |
| Skin Sensitization Study<br>Guinea Pig Maximization<br>Test | ISO 10993-10 | Under the conditions of this study, the test<br>article extracts showed no evidence of<br>causing delayed dermal contact sensitization<br>in the guinea pig. The test article was not<br>considered a sensitizer in the guinea pig<br>maximization test. |
| Skin Irritation Study | ISO 10993-10 | Under the conditions of this study, the<br>irritation response category of the test article<br>is classified as Negligible for polar extract and<br>Negligible for non-polar extract. |
#### 5.8.2. Non-clinical Tests
The Infrared Forehead Thermometer is tested per the following standard, to evaluate its performance. The test results demonstrated that the proposed device comply with the standard requirements.
IEC 60601-1:2005+AMD1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
IEC 60601-1-11:2015 medical electrical equipment - part 1-11: general requirements
{8}------------------------------------------------
for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
ISO 80601-2-56:2017+A1:2018 Medical electrical equipment - Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement.
ASTM E1965-98 (2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
#### 5.8.3. Clinical Data
A comparison study and clinical repeatability testing was performed on the following four age groups: 0 up to 3 months, 3 months up to 1 year, older than 1 year and younger than 5 years and older than 5 years in accordance with ISO 80601-2-56 to compare the Infrared Forehead Thermometer with Mercury thermometer. This clinical comparison study demonstrated that the temperatures obtained with the Infrared Forehead Thermometer were highly related when compared to the Mercury thermometer, the clinical bias with stated uncertainty and clinical repeatability as defined in the ISO 80601-2-56 standard were within clinical acceptability.
#### 5.8.4. Software
The software embedded in Infrared Forehead Thermometer has been developed, documented and validated in accordance with industry standards (IEC 62304 - Medical device software - Software life cycle processes) and FDA guidance (GUIDANCE FOR CONTENT OF PRE-MARKET SUBMISSIONS FOR SOFTWARE THE CONTAINED IN DEVICES).
#### 5.9. Substantially Equivalent Conclusion
The subject device, Infrared Forehead Thermometer, is determined to be Substantially Equivalent (SE) to the predicate device, K191251, in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.