Microlife Digital Infrared Ear Thermometer

K183663 · Microlife Intellectual Property GmbH · FLL · Jul 16, 2019 · General Hospital

Device Facts

Record IDK183663
Device NameMicrolife Digital Infrared Ear Thermometer
ApplicantMicrolife Intellectual Property GmbH
Product CodeFLL · General Hospital
Decision DateJul 16, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Microlife Digital Infrared Ear Thermometer, Model IR1DN1 (IR210) device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

Device Story

The Microlife Digital Infrared Ear Thermometer (Model IR1DN1/IR210) is a handheld electronic device for home use. It utilizes an infrared thermopile sensor to detect natural thermal radiation from the tympanic membrane and surrounding ear canal tissue. The device captures infrared energy, which is processed by an internal application-specific integrated circuit (ASIC) to calculate temperature values. The user operates the device via three buttons (Start/Power, M, Mode) and views results on an LCD screen. It features a position indicator to assist with proper probe placement and provides a high-temperature alarm (10 beeps and red backlight) for readings ≥37.5°C. The device stores up to 30 readings. By providing accurate, non-invasive temperature monitoring, it assists users in tracking health status and identifying potential fever, facilitating timely clinical decision-making.

Clinical Evidence

Clinical study conducted per ASTM E1965-98 and ISO 80601-2-56. Study design: randomized, simple blind, homologous control, pairing design. Sample size: 114 subjects (1/3 infants, 1/3 children, 1/3 adults). Primary endpoints: clinical bias, uncertainty of bias, and clinical repeatability. Results met all acceptance criteria for clinical study protocol.

Technological Characteristics

Infrared thermopile sensor (TPS23B). Materials: PCTG (probe), PMMA (buttons), ABS707 (housing), PP (rigid probe cover). Power: 3.0V DC (CR2032 battery). Dimensions: 159mm x 43mm x 60mm. Standards: ASTM E1965-98, ISO 80601-2-56, ISO 10993, IEC 60601-1, IEC 60601-1-2. Connectivity: None. Sterilization: Not applicable.

Indications for Use

Indicated for neonatal, pediatric, and adult populations for body temperature measurement via the auditory canal in a home setting.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable. July 16, 2019 Microlife Intellectual Property GmbH % Susan D. Goldstein-Falk Official Correspondent for Microlife Intellectual Property GmbH mdi Consultants, Inc. 55 Northern Boulevard, Suite 200 Great Neck, New York 11021 Re: K183663 Trade/Device Name: Microlife Digital Infrared Ear Thermometer, Model IR1DN1 (IR210) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 14, 2019 Received: June 17, 2019 Dear Susan D. Goldstein-Falk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K183663 Device Name Microlife Digital Infrared Ear Thermometer, Model IR 1DN1(IR210) Indications for Use (Describe) The Microlife Digital Infrared Ear Thermometer, Model IRIDNI (IR210) device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediation used in the home setting. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY # The assigned 510(k) number is K183663 | Manufacturer's Name: | Microlife Intellectual Property GmbH, Switzerland<br>Espenstrasse 139<br>9443 Widnau / Switzerland | |-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | Corresponding Official: | Mr. Gerhard Frick<br>Vice President of Technical and Service<br>Microlife Intellectual Property GmbH,<br>Switzerland | | Telephone Number: | +41 79 216 0070 | | Email: | gerhard.frick@microlife.ch | | Preparation Date: | June 26, 2019 | | Trade Name: | Microlife Digital Infrared Ear Thermometer,<br>Model IR1DN1 (IR210) | | Common or Usual Name: | Clinical Electronic Thermometer | | Regulation Name:<br>Regulation Number:<br>Product Code:<br>Device Class:<br>Primary Predicate Device: | Clinical Electronic Thermometer<br>21 CFR 880.2910<br>FLL<br>Class II<br>K003308 Microlife Digital Infrared Ear<br>Thermometer, Model IR1DB1 | # Device Description: The Microlife Digital Infrared Ear Thermometer, Model IR1DN1 (IR210) is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces. {4}------------------------------------------------ # Mode of operation: The unit measures the infrared energy emitting from the middle ear and the surrounding tissue. This energy is absorbed by lenses and converted into temperature values. The Microlife Digital Infrared Ear thermometer, Model IR1DN1 (IR210), consists of the following parts: - a) Thermopile Sensor - b) Application-Specific Integrated Circuitry - c) EEPROM IC - d) Capacitance-touch IC - e) LCD and Backlight - f) 3 buttons ("START/POWER" button, "M" button, "MODE" button) - g) 1 battery CR2032 3.0V - h) Probe cover - I) Lens The new Model IR1DN1 (IR210) has the same intended use and temperature measurement fundamental algorithm as the predicate device 510(k) K#003308 Microlife Model IR1DB1. #### Indications for Use The Microlife Digital Infrared Ear Thermometer, Model IR1DN1 (IR210) device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting. #### Substantial Equivalence Discussion Both subject device and predicate device use infrared technology to measure and monitor the body temperature by the site of ear canal. Microlife Digital Infrared Ear Thermometer IR1DN1 (IR210) has been compared to the cleared "Microlife Digital Infrared Ear Thermometer IR1DB1" (510(k) K#003308). A table comparing the two devices is provided as follows: {5}------------------------------------------------ | Item | 1. Subject Device<br>Microlife Digital Infrared Ear<br>Thermometer IR1DN1(IR210)<br>K183663 | 2. Predicate Device<br>Microlife Digital Infrared<br>Ear Thermometer IR1DB1<br>K003308 | Comparison | |-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | Indications for<br>Use | The Microlife Digital Infrared Ear<br>Thermometer, Model IR1DN1(IR<br>210) device is an electronic<br>clinical thermometer using an<br>infrared sensor to detect body<br>temperature from the auditory<br>canal in the neonatal, pediatric<br>and adult population used in the<br>home setting. | The device is an electronic<br>clinical thermometer using<br>an infrared sensor to<br>detect body temperature<br>from the auditory canal in<br>the neonatal, pediatric and<br>adult population used in<br>the home setting. | Same | | Thermometer<br>type | Infrared thermometer | Infrared thermometer | Same | | Device<br>Measurement<br>Technology | Infrared | Infrared | Same | | Measurement<br>target | Human Ear | Human Ear | Same | | Measuring<br>location | Ear Canal | Ear Canal | Same | | Physical<br>dimension | 159mm(L)*43mm(W)*60mm(H) | 128mm(L)*48mm(W)*30mm<br>(H) | Different | | Power supply | 3.0V DC with 1 CR2032 battery | 3.0V DC with 1 CR2032<br>battery | Same | | Display<br>resolution | 0.1°C or 0.1°F | 0.1°C or 0.1°F | Same | | Item | 1. Subject Device<br>Microlife Digital Infrared Ear<br>Thermometer IR1DN1(IR210)<br>K183663 | 2. Predicate Device<br>Microlife Digital Infrared<br>Ear Thermometer<br>IR1DB1,K003308 | Comparison<br>1 vs 2 | | Inspect the<br>accuracy of<br>blackbody | The sample must be pressed<br>"Start" button for 8 seconds<br>when the unit turns off, after two<br>short beep sounds, release<br>"Start" button, enters the<br>blackbody measurement mode.<br>The unit display 0.01°C or<br>0.01°F resolution in this mode. | The sample must be<br>pressed "Start" button for 8<br>seconds when the unit turns<br>off, after two short beep<br>sounds, release "Start"<br>button, enters the blackbody<br>measurement mode. The<br>unit display 0.01°C or<br>0.01°F resolution in this<br>mode. | Same | | Measuring<br>range(body<br>mode) | 32.0-43.0 °C<br>(89.6-109.4 °F) | 32.0°C~42.2°C<br>(89.6°F~108.0°F) | Different<br>(1) See below | | Accuracy(black<br>body mode) | ±0.2 °C: 35.0 ~ 42.0 °C<br>±0.3 °C: 32.0 ~ 34.9 °C<br>& 42.1 ~ 43.0 °C<br>±0.4 °F: 95.0 ~ 107.6 °F<br>±0.5 °F: 89.6 ~94.8 °F<br>& 107.8~109.4 °F | ±0.2ºC: 32.0~42.2ºC<br>±0.4°F:89.6~108.0°F | Different<br>(2) See below | | Operating<br>conditions | 10°C~40°C (50°F~104°F)<br>15-95 % relative maximum<br>humidity | 10°C~40°C (50°F~104°F)<br>15-95 % relative maximum<br>humidity | Same | | Storage<br>conditions | -25°C ~ 55 °C/-13°F ~131°F<br>15-95 % relative maximum<br>humidity | -25°C ~ 55 °C/-13°F ~131°F<br>15-95 % relative maximum<br>humidity | Same | | Display type | LCD display | LCD display | Same | | Memory | 30 sets memories | One set memory | Different<br>(3) See below | | Error | Display $Er6$ if<br>system malfunction. | Display $Err$ if<br>system<br>malfunction. | Different<br>(4) See below | | Power auto off<br>in clinical mode<br>and blackbody<br>mode | Approx. 1 minute after last<br>measurement has been taken | Approx. 1 minute after last<br>measurement has been<br>taken | Same | | Item | 1. Subject Device<br>Microlife Digital Infrared Ear<br>Thermometer IR1DN1(IR210)<br>K183663 | 2. Predicate Device<br>Microlife Digital Infrared Ear<br>Thermometer<br>IR1DB1,K003308 | Comparison | | | | | 1 vs 2 | | Sensor type | TPS23B | OTP-668D2 | Different<br>(5) See below | | Lens type | Transparent | Transparent | Same | | High Temperature<br>Alarm | 10 short beeps and a red LCD<br>backlight alert the patient that<br>he/she may have a temperature<br>equal to or higher than 37.5 °C. | 10 short beeps and a red<br>LCD backlight alert the<br>patient that he/she may have<br>a temperature equal to or<br>higher than 37.5 °C. | Same | | Measuring probe<br>material | PCTG(TX2001) | ABS707 | Different<br>(6) See below | | Probe cover | Rigid cover (PP) | Soft cover (PE) | Different<br>(7) See below | | Positioning<br>indication | Yes | No | Different<br>(8) See below | | IC(Integrated<br>Circuitry) | HY11P14 | SN8P1919 | Different<br>(9) See below | | Offset | With | With | Same | | Clinical Study<br>Support | Yes. Clinical test report | Yes. Clinical test report | Same | | "Start "/ "Power"<br>button material | PMMA | ABS707 | Different<br>(10) See below | | Housing and<br>battery cover<br>material | ABS707 | ABS707 | Same | | Patient contact Materials | | | | | Probe Cover | Rigid cover<br>(Polypropylene (PP)) | Soft cover<br>(Polyethylene (PE)) | Different<br>(7) See below | | Measuring probe material | PCTG(TX2001) | ABS707 | Different<br>(6) See below | | "Start"/ "Power" button<br>material | PMMA | ABS707 | Different<br>(10) See below | | Housing and battery cover<br>material | ABS707 | ABS707 | Same | {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ The major differences between the subject modified device IR1DN1 (IR210), and the predicate device IR1DB 1 are the measurement range, probe cover, position indicator, measuring probe and buttons material. The Indications for Use statement of the subject device is identical to the predicate device. Other differences between the devices are: # 1. Measurement range (body mode) The subject device IR1DN1 (IR210) body mode measurement range is 32.0-43.0 ℃ (89.6-109.4 °F), whereas the predicate device IR1DB1 measurement range is 32.0°C ~42.2°C (89.6 ~108.0 °F). The subject device has a wider body mode measurement range to accommodate more physiologically relevant temperatures. This does not introduce any new risk to the device. # 2. Accuracy (blackbody mode) The modified device's accuracy: ±0.2 °C: 35.0 ~ 42.0 °C, ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------±0.4°F:95.0~107.6 °F ±0.5°F:89.6~94.8°F&107.8~109.4 °F The predicate device's accuracy: ±0.2°C: 32.0~42.2°C, ±0.4°F, 89.6~108.0°F. The modified device's measurement range and accuracy are wider than the predicate device. ; these are based on device's performance standards and patient's requirement. These differences do not affect the performance and accuracy according to our Reliability Test Report (ASTM E1965- 98), ISO 80601-2-56 Test Report, and our Clinical Test Report of IR1DN1 (IR210). {9}------------------------------------------------ # 3. Memory The modified device IR1DN1 (IR210) can recall the last 30 readings, whereas the predicate device IR1DB1 can only recall the last reading. Only storage reading quantity is different which does not affect the device's performance or safety and effectiveness. # 4. Error Both modified device IR1DN1 (IR210) and the predicate device IR1DB1 have self-test function. The principle of self-test is the same. Devices can perform a self-test every time when it is switched on to always guarantee the specified accuracy of any measurement, when it has error, it will display signal. IR1DN1 (IR210) device will display icon "Er6", IR1DB1 will display icon "Err". Although the icons are different, the purpose is the same. This does not affect the device's performance or safety and effectiveness. # 5. Sensor Type The sensor type of the modified device is TPS23B, whereas the sensor type of the predicate device is OTP-668D2. The difference does not affect the performance and accuracy according to Reliability Test Report (ASTM E1965-98), ISO 80601-2-56 Test Report and Clinical Test Report of IR1DN1 (IR210). ### 6. Measuring probe material The measuring probe material of the subject device is PCTG, whereas the measuring probe material of the predicate device is ABS707. The probe is not the patient contacting material. Biocompatibility testing was performed on the PCTG material in accordance with ISO 10993 standard and testing results proved the subject probe material to be biocompatible. # 7. Probe cover The subject device measures with rigid cover (PP), whereas the predicate device measures with a soft cover (PE). Biocompatibility testing was performed on the finished subject probe cover according to the ISO10993 standard. Our Biocompatibility Test Report of the PP material (Rigid Probe Cover) proves the subject probe cover material to be biocompatible. Our Clinical Test Report of IR1DN1 (IR210), showed that the subject probe cover did not affect the subject modified device's accuracy and performance. # 8. Position indication The subject device includes a position indication feature which was not included with the predicate device. This feature has been validated and verified through performance testing and clinical testing. {10}------------------------------------------------ # 9. IC (Integrated Circuitry) The integrated circuit model and its features have changes have been verified through performance testing. # 10. "Start"/ "Power" button The button material of the subject device is PMMA, whereas the button material of the predicate device is ABS707. The PMMA material's biocompatibility testing was performed according to ISO10993 standard. Biocompatibility Test Report of PMMA testing proved that the subject button material to be biocompatible. Based on the aforementioned modifications to the subject device, the subject device does not raise different types of safety and effectiveness questions when compared to the predicate device. # Performance Testing Testing information demonstrating performance of the Microlife Digital Infrared Ear Thermometer, Model IR1DN1 (IR210) in the intended environment of use is supported by testing that was conducted in accordance with Guidance on the Content of Premarket Notification [510(K)] Submissions for Clinical Electronic Thermometers. > The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices: | 1 | AAMI/ANSI ES60601-<br>1:2005/(R2012) and A1:2012,<br>C1:2009/(R)2012 and<br>A2:2010/(R2012) | Passed all testing requirements | |----|---------------------------------------------------------------------------------------------|---------------------------------| | 2 | AAMI/ANSI/IEC 60601-1-2:2014 | Passed all testing requirements | | 3 | ASTM E1965-98 (2016) | Passed all testing requirements | | 4 | AAMI/ANS/ISO 14971:2007/(R)2010 | Passed all testing requirements | | 5 | AAMI/ANSI/ISO 10993-1: 2009/(R)2013 | Passed all testing requirements | | 6 | AAMI/ANSI/ISO 10993-5:2009/(R)2014 | Passed all testing requirements | | 7 | AAMI/ANSI/ISO 10993-10: 2010/(R)2014 | Passed all testing requirements | | 8 | AAMI/ANSI/ISO 10993-12: 2012 | Passed all testing requirements | | 9 | ISO 80601-2-56: 2017 | Passed all testing requirements | | 10 | AAMI/ANSI HA60601-1-11:2015 | Passed all testing requirements | {11}------------------------------------------------ ### Performance Testing Summary A risk assessment was conducted in accordance is ISO 14971:2007/EN ISO 14971:2012 to assess risk management activities due to modifications to our predicate device. Verification activities that were performed included modifications, associated risk, validation method, and results revealed that all acceptance criteria passed. ### Clinical Tests Conducted Clinical testing was conducted according to ASTM E1965-98 and ISO 80601-2-56. This clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation, consists of 114 subjects, of which 1/3 are infants, 1/3 are children and the rest 1/3 are adults (NOTE: Infants---newborn to one year: Children--- greater than one to five years; Adults---greater than five years old.). The test reports demonstrated that the clinical data, represented by clinical bias, with uncertainty of bias, and clinical repeatability, met the acceptance criteria of the clinical study protocol. ### Conclusions Based on the performance testing, comparison, analysis, and risk assessment, it demonstrated that the Microlife Digital Infrared Ear Thermometer IR1DN1 (IR210) is substantially equivalent to the Microlife Digital Infrared Ear Thermometer IR1DB1, cleared under K003308.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...