Digital Thermometer

K183393 · Joytech Healthcare Co. , Ltd. · FLL · Nov 18, 2019 · General Hospital

Device Facts

Record IDK183393
Device NameDigital Thermometer
ApplicantJoytech Healthcare Co. , Ltd.
Product CodeFLL · General Hospital
Decision DateNov 18, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Digital Thermometers DMT series(Except DMT-455) are intended to measure the human body temperature in regular mode orally, rectally or under the arm. And the devices are reusable for clinical or home use on people of all ages, including children under 8 with adult supervision. The device model DMT-455 is intended to measure temperature orally, and the device is reusable for clinical or home use for children less than 4 years old with adult supervision.

Device Story

Digital thermometers measure human body temperature via thermistor sensor; resistance changes with temperature; integrated circuit (IC) converts resistance to frequency; MCU calculates temperature; result displayed on LCD. Reusable devices for clinical/home use; operated by patients or clinicians. Models vary by tip type (flexible/rigid), response time, and features (backlight, predictive mode). Output informs clinical decision-making regarding fever management. Patient-contacting materials include stainless steel, TPE, and ABS.

Clinical Evidence

Clinical testing conducted on models DMT-4726 and DMT-4735 per ISO 80601-2-56. Study included 450 subjects across three age groups: infants (newborn to 1 year), children (>1 to 5 years), and adults (>5 years). Results demonstrated compliance with ISO 80601-2-56 requirements.

Technological Characteristics

Materials: Stainless steel, TPE, ABS. Sensing: Thermistor-based resistance change. Energy: 1.5V button battery (LR41/SR41) or 3.0V CR2032. Connectivity: None. Sterilization: Not specified. Standards: ISO 80601-2-56:2017, ASTM E1112, IEC 60601-1, IEC 60601-1-2, ISO 10993-5/10.

Indications for Use

Indicated for human body temperature measurement (oral, rectal, or axillary) in clinical or home settings for all ages. DMT-455 indicated for oral use only in children <4 years. All use requires adult supervision for children.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 18, 2019 Joytech Healthcare Co., LTD Mr. Ren Yunhua General Manager No. 365 Wuzhou Road, Yuhang Economic Development Zone Hangzhou City Zhejian, 311100 China Re: K183393 Trade/Device Name: Digital Thermometer DMT series (DMT-101, DMT-411, DMT-301, DMT-102, DMT-202, DMT-412, DMT-423, DMT-433, DMT-455, DMT-108, DMT-418, DMT-106. DMT-206. DMT-416. DMT-427. DMT-437. DMT-4218. DMT-4318. DMT-1019, DMT-2019, DMT-4119, DMT-4220, DMT-4320, DMT-4340, DMT-4343, DMT-4139, DMT-2021, DMT-4121, DMT-209, DMT-1030, DMT-2030, DMT-4130, DMT-1031, DMT-2031, DMT-4131, DMT-4226, DMT-4326, DMT-4726, DMT-1027, DMT-2027, DMT-4127, DMT-1032, DMT-2032, DMT-3032, DMT-4132, DMT-3033, DMT-4233, DMT-4333, DMT-4235, DMT-4335, DMT-4735, DMT-4236, DMT-4336, DMT-3018, DMT-4238, DMT-4338) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: October 14, 2019 Received: October 17, 2019 Dear Mr. Ren Yunhua: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {1}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K183393 #### Device Name Digital Thermometer DMT series(DMT-101.DMT-411.DMT-301.DMT-202.DMT-412.DMT-423.DMT-433.DMT-455.DMT-108.DMT-418.DMT-106.DMT-416.DMT-427.DMT-437.DMT-4218.DMT-4318.DMT-1019,DMT-2019,DMT-4119,DMT-4220,DMT-4320,DMT-4340,DMT-4343,DMT-4139,DMT-4121,DMT-209,DMT-1030,DMT-2030,DMT-4130,DMT-1031,DMT-4131,DMT-4226,DMT-4326,DMT-4726,DMT-1027,DMT-2027,DMT-4127,DMT-1032,DMT-2032,DMT-4132,DMT-3033,DMT-4233,DMT-4233,DMT-4235,DMT-4335,DMT-4735,DMT-4236,DMT-4336,DMT-3018,DMT-4238,DMT-4338) #### Indications for Use (Describe) The Digital Thermometers DMT series(Except DMT-455) are intended to measure the human body temperature in regular mode orally, rectally or under the arm. And the devices are reusable for clinical or home use on people of all ages, including children under 8 with adult supervision. The device model DMT-455 is intended to measure temperature orally, and the device is reusable for clinical or home use for children less than 4 years old with adult supervision. Type of Use (Select one or both, as applicable) □ Prescription Use (Part 21 CFR 801 Subpart D) 区 Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary The assigned 510(k) number is: K183393 - 1. Date Prepared: 2019.11.18 ## 2. Submitter's Identification: Name: JOYTECH HEALTHCARE CO., LTD. Add.:No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou city, 311100 Zhejiang,China Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: RENYH@SEJOY.COM ## 3. Name of the Device: Trade Name: Digital Thermometer DMT series Model:DMT-101、DMT-411、DMT-301、DMT-102、DMT-202、DMT-412、DMT-423、 DMT-433、DMT-455、DMT-108、DMT-418、DMT-106、DMT-206、DMT-416、 DMT-427、DMT-437、DMT-4218、DMT-4318、DMT-1019、DMT-2019、 DMT-4119、DMT-4220、DMT-4320、DMT-4340、DMT-4343、DMT-4139、 DMT-2021、DMT-4121、DMT-209、DMT-1030、DMT-2030、DMT-4130、 DMT-1031、DMT-2031、DMT-4131、DMT-4226、DMT-4326、DMT-4726、 DMT-1027、DMT-2027、DMT-4127、DMT-1032、DMT-2032、DMT-3032、 DMT-4132、DMT-3033、DMT-4233、DMT-4333、DMT-4235、DMT-4335、DMT-4735、DMT-4236、DMT-4336、DMT-3018, DMT-4238、DMT-4338 Common Name: Digital Thermometer Classification name: Clinical Electronic Thermometer ## 4. Classification Information: Product Code: FLL Device Class: II Panel: 80 Regulation number:21 C.F.R.880.2910 Regulation Description: Clinical electronic thermometer {4}------------------------------------------------ ## 5. Predicate Device Information: The Digital Thermometers are substantially equivalent to the following devices: Digital Thermometers, Models: DMT series; FDA 510(k)number: K 163518; Manufactured by JOYTECH HEALTHCARE CO., LTD. #### Intended use / Indication for Use: 6. The Digital Thermometers DMT series(Except DMT-455) are intended to measure the human body temperature in regular mode orally, rectally or under the arm. And the devices are reusable for clinical or home use on people of all ages, including children under 8 with adult supervision. The device model DMT-455 is intended to measure temperature orally, and the device is reusable for clinical or home use for children less than 4 years old with adult supervision. #### 7. Device Description: The Digital Thermometers DMT series(Except DMT-455) are used to measure the human body temperature in regular mode orally, rectally, or under the arm.The device model DMT-455 is intended to measure temperature orally. The Digital Thermometers DMT series consists of a temperature sensor, low power consumption integrated circuit (IC), LCD display, and buzzer. The resistance of sensor changes with temperature and the IC converts the resistance to frequency and calculates the temperature according to the relation of resistance and frequency. The calculated temperature is displayed on the LCD.Only model DMT-4726 and DMT-4735 are predictive optional thermometers.The patient contacting materials are Stainless Steel,TPE and Acrylonitrile Butadiene Styrene (ABS). {5}------------------------------------------------ # 8.Model List | Features | A | B | C | D | E | F | G | H | I | J | K | |----------|---|---|---|---|---|---|---|---|-------------|------------------|----------------| | Models | | | | | | | | | | | | | DMT-101 | ○ | ● | ○ | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.3 x 1.8x 0.9 | Approx. 10 g | | DMT-411 | ○ | ● | ○ | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 12.3 x 1.8x 0.9 | Approx. 10 g | | DMT-301 | ○ | ● | ○ | ○ | ○ | ● | ○ | ○ | ≤60 Seconds | 12.3 x 1.8x 0.9 | Approx. 10 g | | DMT-102 | ○ | ● | ○ | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.4 x 1.8x 1.1 | Approx. 10 g | | DMT-202 | ○ | ● | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.4 x 1.8x 1.1 | Approx. 10 g | | DMT-412 | ○ | ● | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 12.4 x 1.8x 1.1 | Approx. 10 g | | DMT-423 | ● | ○ | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.4 x 2.4x 1.2 | Approx. 12 g | | DMT-433 | ● | ○ | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 12.4 x 2.4x 1.2 | Approx. 12 g | | DMT-455 | ○ | ● | ○ | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 5.8x 4.1x 4.3 | Approx. 13 g | | DMT-108 | ○ | ● | ○ | ○ | △ | ○ | ○ | ○ | ≤60 Seconds | 13.9 x 2.3x 1.2 | Approx. 12 g | | DMT-418 | ○ | ● | ○ | ● | △ | ● | ○ | ○ | ≤45 Seconds | 13.9 x 2.3x 1.2 | Approx. 12 g | | DMT-106 | ○ | ● | ○ | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.2 x 1.9x 1.1 | Approx. 11 g | | DMT-206 | ○ | ● | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.2 x 1.9x 1.1 | Approx. 11 g | | DMT-416 | ○ | ● | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 12.2 x 1.9x 1.1 | Approx. 11 g | | DMT-427 | ● | ○ | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.4 x 1.9x 1.1 | Approx. 11 g | | DMT-437 | ● | ○ | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 12.4 x 1.9x 1.1 | Approx. 11 g | | DMT-4218 | ● | ○ | ● | ○ | △ | ○ | ○ | ○ | ≤60 Seconds | 13.9 x 2.2x 1.3 | Approx. 11 g | | DMT-4318 | ● | ○ | ● | ● | △ | ● | ○ | ○ | ≤45 Seconds | 13.9 x 2.2x 1.3 | Approx. 11 g | | DMT-3018 | ● | ○ | ● | ○ | △ | ● | ○ | ○ | ≤60 Seconds | 13.9 x 2.2x 1.3 | Approx. 11 g | | DMT-1019 | ○ | ● | ○ | ○ | △ | ○ | ○ | ○ | ≤60 Seconds | 13.9 x 2.3x 1.3 | Approx. 11 g | | DMT-2019 | ○ | ● | ● | ○ | △ | ○ | ○ | ○ | ≤60 Seconds | 13.9 x 2.3x 1.3 | Approx. 12 g | | DMT-4119 | ○ | ● | ● | ● | △ | ● | ○ | ○ | ≤45 Seconds | 13.9 x 2.3x 1.3 | Approx. 12 g | | DMT-4220 | ● | ○ | ● | ○ | △ | ○ | ○ | ○ | ≤60 Seconds | 13.9 x 2.3x 1.3 | Approx. 13 g | | DMT-4320 | ● | ○ | ● | ● | △ | ● | ○ | ○ | ≤45 Seconds | 13.9 x 2.3x 1.3 | Approx. 13 g | | DMT-2021 | ○ | ● | ● | ○ | △ | ○ | ○ | ○ | ≤60 Seconds | 12.2 x 1.9x 1.0 | Approx. 11 g | | DMT-4121 | ○ | ● | ● | ● | △ | ● | ○ | ○ | ≤45 Seconds | 12.2 x 1.9x 1.0 | Approx. 11 g | | DMT-209 | ○ | ● | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.8 x 2.0x 1.2 | Approx. 11 g | | DMT-1030 | ○ | ● | ○ | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 13.2 x 2.3x 1.2 | Approx. 11 g | | DMT-2030 | ○ | ● | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 13.2 x 2.3x 1.2 | Approx. 11 g | | DMT-4130 | ○ | ● | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 13.2 x 2.3x 1.2 | Approx. 11 g | | DMT-1031 | ○ | ● | ○ | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 13.0 x 1.9x 1.0 | Approx. 12 g | | DMT-2031 | ○ | ● | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 13.0 x 1.9x 1.0 | Approx. 12 g | | DMT-4131 | ○ | ● | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 13.0 x 1.9x 1.0 | Approx. 12 g | | DMT-4226 | ● | ○ | ● | ○ | △ | ○ | ○ | ○ | ≤60 Seconds | 13.5 x 3.4x 1.7 | Approx. 22 g | | DMT-4326 | ● | ○ | ● | ● | △ | ● | ○ | ○ | ≤45 Seconds | 13.5 x 3.4x 1.7 | Approx. 22 g | | DMT-4726 | ● | ○ | ● | ● | Δ | ● | Δ | Δ | ≤45 Seconds | 13.5 x 3.4x 1.7 | Approx. 22 g | | DMT-1027 | ○ | ● | ○ | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.2 x1.8 x 1.1 | Approx. 10 g | | DMT-2027 | ○ | ● | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 12.2 x1.8 x 1.1 | Approx. 10 g | | DMT-4127 | ○ | ● | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 12.2 x 1.8 x 1.1 | Approx. 10 g | | DMT-1032 | ○ | ● | ○ | ○ | Δ | ○ | ○ | ○ | ≤60 Seconds | 13.8 x 2.2 x 1.2 | Approx. 12 g | | DMT-2032 | ○ | ● | ● | ○ | Δ | ○ | ○ | ○ | ≤60 Seconds | 13.8 x 2.2 x 1.2 | Approx. 12 g | | DMT-3032 | ○ | ● | ● | ○ | ○ | ● | ○ | ○ | ≤60 Seconds | 13.8 x 2.2 x 1.2 | Approx. 12 g | | DMT-4132 | ○ | ● | ● | ● | Δ | ● | ○ | ○ | ≤45 Seconds | 13.8 x 2.2 x 1.2 | Approx. 12 g | | DMT-3033 | ● | ○ | ● | ○ | ○ | ● | ○ | ○ | ≤60 Seconds | 13.9 x 2.2x 1.2 | Approx. 12 g | | DMT-4233 | ● | ○ | ● | ○ | Δ | ○ | ○ | ○ | ≤60 Seconds | 13.9 x 2.2x 1.2 | Approx. 12 g | | DMT-4333 | ● | ○ | ● | ● | Δ | ● | ○ | ○ | ≤45 Seconds | 13.9 x 2.2 x 1.2 | Approx. 12 g | | DMT-4235 | ● | ○ | ● | ○ | Δ | ○ | ○ | ○ | ≤60 Seconds | 13.4 x 3.1 x 1.6 | Approx. 23 g | | DMT-4335 | ● | ○ | ● | ● | Δ | ● | ○ | ○ | ≤45 Seconds | 13.4 x 3.1 x 1.6 | Approx. 23 g | | DMT-4735 | ● | ○ | ● | ● | Δ | ● | Δ | Δ | ≤45 Seconds | 13.4 x 3.1x 1.6 | Approx. 23 g | | DMT-4236 | ● | ○ | ● | ○ | ○ | ○ | ○ | ○ | ≤60 Seconds | 13.1 x 2.0x 1.2 | Approx. 10 g | | DMT-4336 | ● | ○ | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 13.1 x 2.0x 1.2 | Approx. 10 g | | DMT-4343 | ● | ○ | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 14.2 x 2.3x 1.3 | Approx. 11 g | | DMT-4238 | ● | ○ | ● | ○ | ○ | ● | ○ | ○ | ≤60 Seconds | 12.5 x 2.1x 1.2 | Approx. 12.8 g | | DMT-4338 | ● | ○ | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 12.5 x 2.1x 1.2 | Approx. 12.8 g | | DMT-4139 | ○ | ● | ● | ● | ○ | ● | ○ | ○ | ≤60 Seconds | 13.2 x 2.3x 1.3 | Approx. 14.5 g | | DMT-4340 | ● | ○ | ● | ● | ○ | ● | ○ | ○ | ≤45 Seconds | 13.2 x 2.3x 1.3 | Approx. 14.5 g | {6}------------------------------------------------ A=Flexible tip B=Rigid tip C=Waterproof D=Instant E=TEMPERATURE ARROW INDICATING function F=°C/°F switchable G=Backlight function H=Predictive I=Water Bath Response time J=Physical dimensions(L x H x W in cm) K=Weight( including battery in grams ) • = Yes o = No △ =Optional {7}------------------------------------------------ # 9. Comparing to Predicate Device: | SE Comparisons | Subject device application k183393 | Predicate device k163518<br>(Model:DMT series) | Comparison<br>Result | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------| | Intended Use<br>/Indication for use | DMT series(except DMT-455) are<br>intended to measure the human body<br>temperature in regular mode orally,<br>rectally or under the arm. And the<br>devices are reusable for clinical or<br>home use on people of all ages.<br>including children under 8 with adult<br>supervision.<br>DMT-455 is intended to measure<br>temperature orally, and the device is<br>reusable for clinical or home use for<br>children less than 4 years old with<br>adult supervision. | DMT series(except DMT-455) are<br>intended to measure the human body<br>temperature in regular mode orally,<br>rectally or under the arm. And the<br>devices are reusable for clinical or<br>home use on people of all ages.<br>including children under 8 with adult<br>supervision.<br>DMT-455 is intended to measure<br>temperature orally, and the device is<br>reusable for clinical or home use for<br>children less than 4 years old with<br>adult supervision. | Identical | | Measurement Site | DMT series (except DMT-455):<br>orally, rectally or under the arm<br><br>DMT-455:orally | DMT series (except DMT-455):<br>orally, rectally or under the arm<br><br>DMT-455:orally | Identical | | Range | DMT series<br>(except DMT-30XX, DMT-47XX):<br>$32.0°C~43.9°C (90.0°F~111.9°F)$<br>DMT-301,DMT-3032,DMT-3033,<br>DMT-3018:<br>$32.00°C~43.99°C(90.00°F~111.99°F)$<br>DMT-4735,DMT-4726: $32.0°C~43.9°C (89.6°F~111.0°F)$ | DMT series<br>(except DMT-30XX,DMT-47XX):<br>$32.0°C~42.9°C (90°F~109.9°F)$<br>DMT-301,DMT-3032,DMT-3033:<br>$32.00°C~42.99°C (90.00°F~109.99°F)$<br>DMT-4735,DMT-4726: $32.0°C~42.9°C (89.6°F~109.2°F)$ | Similar | | Accuracy | $±0.1°C$ between $35.5°C$ to<br>$42.0°C (±0.2°F,95.9°F-107.6°F)$ ,<br>$±0.2°C$ under $35.5°C$ or over<br>$42.0°C (±0.4°F$ under $95.9°F$ or over<br>$107.6°F$ )<br>DMT-301, DMT-3032, DMT-3033, DMT-<br>3018: $±0.10 °C$ between $35.5°C$ to $42°C$ | $±0.1°C$ between $35.5°C$ to<br>$42.0°C (±0.2°F,95.9°F-107.6°F)$ ,<br>$±0.2°C$ under $35.5°C$ or over<br>$42.0°C (±0.4°F$ under $95.9°F$ or over<br>$107.6°F$ ) | Identical | {8}------------------------------------------------ | | | DMT series(except DMT-301,<br>DMT-3032,DMT-3033,DMT-3018):<br>0.1°C/0.1 °F<br>DMT-301,DMT-3032,DMT-3033,<br>DMT-3018:0.01℃/0.01°F | DMT series(except DMT-301,<br>DMT-3032,DMT-3033):<br>0.1°C/0.1 °F<br>DMT-301,DMT-3032,DMT-3033:<br>0.01 ℃/0.01°F | Identical | | |--------------------------------------------|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-----------| | Display resolution | | | | | | | components | | Sensor, buzz film, housing, stainless<br>steel cap, LCD display, measurement<br>control module. | Sensor, buzz film, housing, stainless<br>steel cap, LCD display,<br>measurement control module. | Identical | | | Principle of operation | | A change of thermistor resistance,<br>caused by changes of temperature.<br>The resistance is measured by<br>microcontroller unit (MCU), so<br>changes of temperature will<br>correspond to changes of resistance. | A change of thermistor resistance,<br>caused by changes of temperature.<br>The resistance is measured by<br>microcontroller unit (MCU), so<br>changes of temperature will<br>correspond to changes of resistance | Identical | | | Operating range | | Temperature: 41°F~104°F (5°C<br>~40°C) Relative humidity:<br>15%~95%RH<br>Atmospheric Pressure : 700hPa ~<br>1060hPa | Temperature: 41°F~104°F (5°C<br>~40°C) Relative humidity:<br>15%~95%RH<br>Atmospheric Pressure : 700hPa ~<br>1060hPa | Identical | | | Patient contact<br>material | | ABS,TPE,Stainless Steel | ABS,TPE,Stainless Steel | Identical | | | Colour coding | ABS | 9003 | ABS | 9003 | Identical | | | TPE | P351C,P288C,P374C,P2<br>84C,P2715C,P109C,P42<br>8C,P427C | TPE | P351C,P288C,P374C,P<br>284C,P2715C,P109C,P<br>428C,P427C | | | Storage and<br>transportation<br>condition | Temperature: | -4°F~131°F (-20℃~55℃)<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure : 700hPa ~ | Temperature: | -4°F~131°F (-20℃~55℃)<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure : 700hPa ~ | Identical | {9}------------------------------------------------ | | 1060hPa | 1060hPa | | |-------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----------| | Battery type | One 1.5 V DC. button battery (size<br>LR41or SR41, UCC 392)<br>DMT-4726,DMT-4735:<br>One 3.0V CR2032 battery | One 1.5 V DC. button battery (size<br>LR41or SR41, UCC 392)<br>DMT-4726,DMT-4735:<br>One 3.0V CR2032 battery | Identical | | Biocompatibility | Comply with ISO 10993-5 and ISO<br>10993-10 | Comply with ISO 10993-5 and ISO<br>10993-10 | Identical | | Electrical Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | Identical | | EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | Identical | The differences between two devices are: 1. Changed thermometer chip to extend temperature measuring range from 32.0~42.9℃ to 32.0~43.9 ℃ ,as new version of standard ISO 80601-2-56:2017 required, no other functions changed. 2.New model DMT-3018: it is a new model that have same appearance and same electrical scheme with predicate device model DMT-4318 but have only different temperature range and PCB with the predictive device. ## 10. Performance data The following performance data were provided in support of the substantial equivalence determination: Performance testing was conducted to validate and verify that Digital Thermometers, DMT series met all requirements of related international standards, including electrical safety, EMC, biocompatibility, software validation and product specifications. Results of these tests demonstrate compliance to the requirements of the below consensus standards. Electrical Safety and performance requirements: - . AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 Medical Electrical Equipment - ISO80601-2-56:2017 Medical electrical equipment Part 2-56 Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement - · ASTM E1112:00(Reapproved 2011) Standard specification for Electronic Thermometer for Intermittent Determination of Patient Temperature {10}------------------------------------------------ Home-used medical equipment requirements and environmental test: - IEC 60601-1-11:2015 General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment Electromagnetic compatibility requirements: - IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests Biocompatibility Evaluation for patient contacting components: - · ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity - · ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization #### Guidance Document: · Guidance on the content of Premarket Notifications [510(k)] Submissions for clinical electronic thermometers The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005. ## 11. Discussion of Clinical Tests Performed: Clinical tests were conducted on the DMT-4726 and DMT-4735. The clinical tests evaluated 450 of subjects. and the thermometer was evaluated in three groups 1) infants-newborn to one year; 2) children-greater than one to five years; and 3) adults-greater than five years old.The clinical performance test protocol and data analysis were conducted in accordance with the requirement of ISO 80601-2-56. The test report showed the clinical performance of the subject devices complied with the requirement of ISO 80601-2-56. ## 12. Conclusions: Based on the information provided in this submission,the subject digital thermometers DMT series are substantially equivalent to the predicate thermometers, model: DMT series.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...