← Product Code [FLL](/productcode/FLL) · K172894

# Infrared Thermometer (K172894)

_Shenzhen Jiacom Technology CO ., Ltd. · FLL · May 17, 2018 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172894

## Device Facts

- **Applicant:** Shenzhen Jiacom Technology CO ., Ltd.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** May 17, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

Infrared Thermometer is intended to detect body temperature from the forehead or auditory canal in the population including infant (above 6 months), child, adolescent, and adult.

## Device Story

Hand-held, battery-powered infrared thermometer (IFR 600); detects body temperature via forehead or auditory canal; non-contact infrared measurement principle; user-operated via scan button; LCD display provides temperature readout; used in clinical or home settings; assists healthcare providers or patients in monitoring body temperature; aids in clinical decision-making regarding fever management.

## Clinical Evidence

Clinical study conducted per ASTM E1965-98(2003) evaluated clinical bias, uncertainty, and repeatability. Study included subjects across age ranges (infant to adult), balanced by gender, with 35%–37.5% febrile. Results demonstrated clinical accuracy and repeatability not inferior to the predicate device.

## Technological Characteristics

Hand-held infrared thermometer; ABS enclosure with titanium dioxide/phthalocyanine blue additives; 3Vdc power; LCD display; non-contact infrared sensing. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 80601-2-56, and ASTM E1965-98. Biocompatibility per ISO 10993-5 and ISO 10993-10.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Valeo VT-601 Series ZR Thermometer ([K111463](/device/K111463.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

May 17, 2018

Shenzhen Jiacom Technology Co., Ltd % Diana Hong Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 CHINA

Re: K172894

Trade/Device Name: Infrared Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: April 16, 2018 Received: April 18, 2018

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina Kiang
S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172894

Device Name Infrared Thermometer

Indications for Use (Describe)

Infrared Thermometer is intended to detect body temperature from the forehead or auditory canal in the population including infant (above 6 months), child, adolescent, and adult

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K172894

- 1. Date of Preparation: 5/14/2018
- Sponsor Identification 2.

### Shenzhen Jiacom Technology Co ., Ltd

No. A6 Building, Silicon Valley Power, Qinghu Park, Longhua Street, Bao'an District, Shenzhen, Guangdong, 518109, China

Establishment Registration Number: Not yet registered.

Contact Person: Luo Tianfeng Position: Registration specialist Tel: +86-755-29015600 Fax: +86-755-29575792 Email: zhongjingming@szjiakang.com

#### 3. Designated Submission Correspondent

Ms. Diana Hong (Primary Contact Person) Ms. Betty Xiao (Alternative Contact Person)

### Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

#### 4. Identification of Proposed Device

Trade Name: Infrared Thermometer Common Name: Clinical electronic thermometer Model: IFR 600

Regulatory Information Classification Name: Clinical electronic thermometer Classification: Class II Product Code: FLL Regulation Number: 21 CFR 880.2910 Review Panel: General Hospital

Indication for Use Statement:

Infrared Thermometer is intended to detect body temperature from the forehead or auditory canal in the population including infant (above 6 months), child, adolescent, and adult.

Device Description

The proposed device, Infrared Thermometers IFR 600 is hand-held, reusable, battery powered device, which is intended to detect body temperature from forehead or auditory canal in the pediatric and adult population.

IFR 600 infrared thermometer can measure human body temperature by both ear and forehead measurement.

- న్. Identification of Predicate Device(s)
510(k) Number: K111463 Product Name: Valeo VT-601 Series ZR Thermometer

- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- > ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- > ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

{5}------------------------------------------------

- A AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012, c1:2009/(R)2012 and A2:2010/(R)2012 medical electrical equipment - part 1: general requirements for basic safety and essential performance.
- > IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests.
- > ISO 80601-2-56 First Edition 2009-10-01, Medical Electrical Equipment - Part 2-56: Particular Requirements For Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement.
- A ASTM E1965-98 (R 2016), Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature

The proposed device is tested to evaluate its electrical safety and electromagnetic compatibility. The test results demonstrated that the proposed device complies with the standards.

- 7. Clinical Test Conclusion
Controlled human clinical studies were conducted in accordance with ASTM E 1965-98(2003), clinical bias, clinical uncertainty and clinical repeatability have been evaluated per clinical validation for infrared thermometer .the clinical trial results verify that the clinical accuracy of the proposed device is not inferior to that of predicate device.

40 of each age range of subjects are included in the clinical studies, the amounts of male subjects are the same with that of female subjects, and 35%~37.5% of total subjects are febrile person. Compared statistical result of clinical bias and clinical repeatability of two comparison groups, the result of proposed device is not being inferior to that of predicate device. The result of proposed device was not inferior to that of predicate device, and the proposed device complies with ASTM E1965-98.

- Substantially Equivalent (SE) Comparison 8.

| ITEM                | Proposed device                                                                                        | Predicate Device                                                                                                                                                |
|---------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code        | FLL                                                                                                    | K111463                                                                                                                                                         |
| Regulation Number   | 21 CFR 880.2910                                                                                        | 21 CFR 880.2910                                                                                                                                                 |
| Indications for Use | Infrared Thermometer is intended to detect body temperature from the forehead or auditory canal in the | The Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D, VT-601E, VT-601F) is infrared thermometers intended for the intermittent measurement of human body |

Table 1 Comparison of Technology Characteristics

{6}------------------------------------------------

|                              | population including infant<br>(above 6 months), child,<br>adolescent, and adult.                                                | temperature in people of all ages. |                                                                     |                               |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------|---------------------------------------------------------------------|-------------------------------|
| Measurement<br>Site          | Forehead and eardrum                                                                                                             |                                    | Forehead and eardrum                                                |                               |
| Principle<br>of<br>Operation | Non-contacting,<br>Infrared<br>Temperature Measurement                                                                           |                                    | Non-contacting,<br>Infrared<br>Measurement                          |                               |
| Ear Mode                     |                                                                                                                                  |                                    |                                                                     |                               |
| Range                        | 32.0°C-42.9°C                                                                                                                    |                                    | 32.0°C-42.9°C                                                       |                               |
| Accuracy                     | ±0.2°C at 35.5°C-42.0°C                                                                                                          |                                    | ±0.2°C at 36.0°C-39.0°C                                             |                               |
|                              | ±0.3°C at 32.0-35.0°C and<br>42.0°C-42.9°C                                                                                       |                                    | Others ±0.3°C                                                       |                               |
| Forehead Mode                |                                                                                                                                  |                                    |                                                                     |                               |
| Range                        | 34. 0°C-42.9°C                                                                                                                   |                                    | 22.0°C-42.2°C                                                       |                               |
| Accuracy                     | ±0.2°C at 35.5°C-42.0°C<br>±0.3°C at 34.0-35.4°C and<br>42.1°C-42.9°C                                                            |                                    | ±0.3°C at 22.0°C-40.0°C                                             |                               |
| Display type                 | LCD                                                                                                                              |                                    | LCD                                                                 |                               |
| Activation                   | Scan button                                                                                                                      |                                    | Scan button                                                         |                               |
| Power<br>requirements        | 3Vdc                                                                                                                             |                                    | 3Vdc.                                                               |                               |
| Material                     | Enclosure and key: ABS<br>(Acrylonitrile Butadiene<br>Styrene)<br>Color additive: Titanium<br>dioxide and Phthalocyanine<br>blue |                                    | Unknown                                                             |                               |
| Electrical Safety            | Complied with IEC 60601-1                                                                                                        |                                    | Complied with IEC 60601-1                                           |                               |
| EMC                          | Complied with IEC 60601-1-2                                                                                                      |                                    | Complied with IEC 60601-1-2                                         |                               |
| Performance                  | Complied with ISO 80601-2-56<br>Complied with ASTM E1965<br>-98 (2003)                                                           |                                    | Complied with ISO 80601-2-56<br>Complied with ASTM E1965 -98 (2003) |                               |
| Biocompatibility             | Cytotoxicity                                                                                                                     | Complied<br>with ISO<br>10993-5    | Cytotoxicity                                                        | Complied with ISO<br>10993-5  |
|                              |                                                                                                                                  |                                    |                                                                     |                               |
|                              | Irritation                                                                                                                       | Complied<br>with ISO<br>10993-10   | Irritation                                                          | Complied with ISO<br>10993-10 |
|                              |                                                                                                                                  |                                    |                                                                     |                               |

{7}------------------------------------------------

|  | with ISO<br>10993-10 | 10993-10 |
|--|----------------------|----------|
|--|----------------------|----------|

The proposed device and predicate device have the same measurement site, principle of operation and power requirement. Although the measurement range and accuracy are different, they comply with ISO 80601-2-56 and ASTM E 1965-98(2006), thus the difference will not arise the issues related to safety and effectiveness. The materials of predicate are not known, however, the proposed device has been conducted cytotoxicity, irritation and sensitization to demonstrate its biocompatibility. And both the proposed device and predicate device are conducted electrical safety and EMC testing according to IEC 60601-1 and IEC 60601-1-2, the test result shows the proposed device meets the requirements of these standards. Therefore, the proposed device and predicate device can be considered as substantially equivalent.

- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K172894](https://fda.innolitics.com/device/K172894)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
