Infrared Thermometer

K172874 · Shenzhen Jiacom Technology CO ., Ltd. · FLL · Jul 12, 2018 · General Hospital

Device Facts

Record IDK172874
Device NameInfrared Thermometer
ApplicantShenzhen Jiacom Technology CO ., Ltd.
Product CodeFLL · General Hospital
Decision DateJul 12, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

Infrared Thermometer is intended to detect body temperature from forehead in the population including infant (above 6 months), child, adolescent, and adult.

Device Story

Hand-held, battery-powered, non-contact infrared thermometer; measures forehead temperature at 3cm-5cm distance. Sensor detects object, environment, and internal sensor temperatures; signals amplified and converted to digital via MCU AD module. MCU calculates body temperature; displays result on LCD. Used by clinicians or patients in home/clinical settings. Provides rapid temperature readings to assist in fever monitoring and clinical decision-making.

Clinical Evidence

Controlled human clinical studies conducted per ASTM E 1965-98(2003). Evaluated clinical bias, uncertainty, and repeatability. Results demonstrate clinical accuracy is not inferior to the predicate device.

Technological Characteristics

Enclosure: Acrylonitrile Butadiene Styrene (ABS) with Titanium dioxide. Sensing: Infrared temperature sensor. Power: 3Vdc. Connectivity: None. Standards: ISO 80601-2-56, ASTM E 1965-98 (R 2016), IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10.

Indications for Use

Indicated for forehead body temperature detection in infants (>6 months), children, adolescents, and adults.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. July 12, 2018 Shenzhen Jiacom Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O Box 120-119 Shanghai, 200120 China Re: K172874 Trade/Device Name: Infrared Thermometer, Model FR800 and FR850 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 5, 2018 Received: June 14, 2018 Dear Diana Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely. Digitally signed by Alan M. Alan M. Stevens -S DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=1300 Stevens 189211, cn=Alan M. Stevens -S Date: 2018.07.12 14:55:51 -04'00' for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172874 Device Name Infrared Thermometer Model: FR800 and FR850 Indications for Use (Describe) Infrared Thermometer is intended to detect body temperature from forehead in the population including infant (above 6 months), child, adolescent, and adult. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K172874 - 1. Date of Preparation: 6/19/2018 - 2. Sponsor Identification ### Shenzhen Jiacom Technology Co ., Ltd No. A6 Building, Silicon Valley Power, Qinghu Park, Longhua Street, Bao'an District, Shenzhen, Guangdong, 518109, China Establishment Registration Number: Not yet registered. Contact Person: Luo Tianfeng Position: Registration specialist Tel: +86-755-29015600 Fax: +86-755-29575792 Email: zhongjingming@szjiakang.com - 3. Designated Submission Correspondent Mr. Chengyu Wang (Consulting Engineer) Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China Tel: +86(0)21 2281-5850 Fax: +1(0)360 925-3199 Email: info@mid-link.net - Identification of Proposed Device 4. Trade Name: Infrared Thermometer, Model(s): FR800 and FR850 Common Name: Clinical electronic thermometer Model(s): FR800 and FR850 {4}------------------------------------------------ Regulatory Information Classification Name: Clinical electronic thermometer Classification: Class II Product Code: FLL Regulation Number: 21 CFR 880.2910 Review Panel: General Hospital Indications for Use: Infrared Thermometer is intended to detect body temperature from forehead in the population including infant (above 6 months), child, adolescent, and adult. #### Device Description The proposed device, Infrared Thermometers, which includes model FR800 and FR850 are hand-held, reusable, battery powered device, which are intended to detect body temperature from forehead for infant (above 6 months), child, adolescent, and adult. The proposed device is intended for non-contacting infrared temperature measurement. The distance of the measurement is 3cm~5cm. The proposed device uses a temperature sensor, which can detect the object temperature [human body temperature], environment temperature and temperature of sensor itself; these temperatures are then transfer to electronic signal and amplified; and then it is transferred to digital signal by AD module in MCU of the proposed device. MCU will calculate the body temperature, and then transfer to screen for display. - న్. Identification of Predicate Device(s) 510(k) Number: K161728 Device Name: Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1 - Substantially Equivalent (SE) Comparison 6. | ITEM | Proposed device | Predicate Device | |----------------------------------------------|--------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------| | | K172874 | K161728 | | Product Code | FLL | FLL | | Regulation Number | 21 CFR 880.2910 | 21 CFR 880.2910 | | Indications for Use | Infrared Thermometer is intended to<br>detect body temperature from forehead<br>in the population including infant | The Microlife Digital Infrared<br>Forehead Thermometer, Model<br>FR1MN1-1 is intended for the | | | (above 6 months), child, adolescent, and adult. | intermittent measurement and<br>monitoring of human body<br>temperature. The device is indicated<br>for use by people of all ages in the<br>home. | | Measurement Site | Forehead | Forehead | | Principle of<br>Operation | Non-contacting, Infrared Temperature<br>Measurement | Non-contacting, Infrared Temperature<br>Measurement | | Device features | Display type: LCD | Display type: LCD | | | Activation: Scan button | Activation: Scan button | | | Have Memory function | Have Memory function | | | Have backlight | Have backlight | | | Have beeper indication | Have beeper indication | | Range | 32.0°C-42.9°C | 34.0°C-42.2°C | | Accuracy | ±0.2°C at 35.0°C-42.0°C<br>±0.4°F at 95°F-107.6°F<br>±0.3°C at others<br>±0.5°F at others | 0.4 °F, 96.8 ~ 102.2 °F | | Display type | LCD | LCD | | Activation | Scan button | Scan button | | Power requirements | 3Vdc | 3Vdc. | | Electrical Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | | EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | | Performance | Complied with ISO 80601-2-56 | Complied with ISO 80601-2-56 | | | Complied with ASTM E 1965 -98<br>(2003) | Complied with ASTM E 1965 -98<br>(2003) | | Patient-contact<br>component and<br>material | Enclosure: Acrylonitrile Butadiene<br>Styrene (ABS)<br>Color additive: Titanium dioxide | Unknown | | Biocompatibility | Complied with ISO 10993-5<br>Complied with ISO 10993-10 | Complied with ISO 10993-5<br>Complied with ISO 10993-10 | #### Table 1 Comparison of Technology Characteristics {5}------------------------------------------------ The subject device is used for above 6 months patients. However, the predicate is used for all ages. The user population of the subject device is subset of the predicate. Therefore, this difference does not raise new safety and effectiveness questions. The measurement range and accuracy of proposed device and predicate device is different. {6}------------------------------------------------ Measurement range of proposed device is wider than that of the predicate device. Bench test included in this submission demonstrated that the proposed device can measure this range accurately. The measurement accuracy of proposed device is slightly different from that of the predicate device. However, the specifications of the proposed device comply with the current ISO and ASTM Standards. Therefore, these differences do not raise new questions of safety and effectiveness. Although we do not know the patient-contact component and material of predicate device, the biocompatibility of proposed device meets ISO 10993 and is equivalent to the predicate. - 7. Non-Clinical Tests Non clinical tests were conducted to verify that the proposed device met all design specifications and is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - > ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity - > ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization - > AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012, c1:2009/(R)2012 and A2:2010/(R)2012 medical electrical equipment - part 1: general requirements for basic safety and essential performance. - A IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests. - A ISO 80601-2-56 First Edition 2009-10-01, Medical Electrical Equipment - Part 2-56: Particular Requirements For Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement. - > ASTM E 1965-98 (R 2016). Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature Software verification and validation data was completed as recommended in the FDA guidance document, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. All the tests results of above tests, complied with the requirements of its pre-defined acceptance criteria and intended uses. #### 8. Clinical Test Conclusion Controlled human clinical studies were conducted in accordance with ASTM E 1965-98(2003), clinical bias, clinical uncertainty and clinical repeatability have been evaluated per clinical validation for infrared thermometer. The clinical trial results verify that the clinical accuracy of the proposed device is not inferior to that of predicate device. {7}------------------------------------------------ - Substantially Equivalent (SE) Conclusion 9. Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...