K152690 · Pinghu Sama Medical Packing Co., Ltd. · FLL · Jan 15, 2016 · General Hospital
Device Facts
Record ID
K152690
Device Name
DISPOSABLE THERMOMETER PROBE SHEATHS
Applicant
Pinghu Sama Medical Packing Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Jan 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
The Disposable Thermometer Probe Sheaths are intended for use as barriers between digital thermometers and users' oral cavities to avoid the possible contamination and infection during temperature measuring. The sheaths are non-sterile and intended for single use only.
Device Story
Disposable, non-sterile, single-use shell-like barrier; constructed from PE/EVA film sandwiched between two layers of medical kraft paper. Paper layers removed prior to use. Device covers digital thermometer probe during oral temperature measurement. Used in clinical or home settings by patients or clinicians. Acts as physical barrier to prevent cross-contamination between patient and thermometer. Does not alter thermometer function; performance verified via temperature accuracy testing per ASTM E1112.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ASTM E1104 (probe cover specs), and ASTM E1112 (electronic thermometer accuracy). Testing confirmed device integrity and that it does not interfere with thermometer accuracy requirements.
Technological Characteristics
Materials: PE/EVA film, medical kraft paper. Dimensions: 115mm x 30mm. Density: 0.87-0.88g/cm3. Thickness: 23.5um. Non-sterile. Single-use. Mechanical barrier principle. Complies with ASTM E1104 and ASTM E1112.
Indications for Use
Indicated for use as a barrier between digital thermometers and oral cavities to prevent contamination/infection during temperature measurement in patients requiring oral temperature assessment.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Disposable Thermometer Covers and Sheaths (K112289)
K211931 — Medline Digital Thermometer Non-Lubricated Probe Sheath · Medline Industries, Inc. · Sep 20, 2021
K180714 — Probe Covers for Thermometers · K-Jump Health Co., Ltd. · Sep 24, 2018
K112289 — DISPOSABLE THERMOMETER COVERS AND SHEATHS · Kang ZE Industrial Co. Limited · Dec 16, 2011
K173113 — Medline Disposable Electronic Thermometer Probe Cover · Medline Industries, Inc. · Jan 17, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2016
Pinghu Sama Medical Packing Co., Ltd. c/o Mr. Ray Wang Beijing Believe Tech. Service Co., Ltd. 1-202, Build 3, Beijing New World No.5 Chaogyang Rd., Chaogyang District Beijing, 100024 CHINA
Re: K152690
Trade/Device Name: Disposable Thermometer Probe Sheath Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: December 15, 2015 Received: December 18, 2015
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Ray Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
lina Kiana -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152690
Device Name Disposable Thermometer Probe Sheath
#### Indications for Use (Describe)
The Disposable Thermometer Probe Sheaths are intended for use as barriers between digital thermometers and users' oral cavities to avoid the possible contamination and infection during temperature measuring. The sheaths are non-sterile and intended for single use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K152690
Date of Preparation: 9/17/2015
Sponsor Identification PINGHU SAMAMEDICAL PACKING CO., LTD. (Establishment Reg: 3010521050) 1 Qunfeng Road, South of Lindai, PingHu, Zhejiang, 314202, China
Contact Person: Ma JianZhong Position: General Manager Tel: +86-573-85923333 Fax:+86-573-85924444 Email: mjz1966@163.com
#### Submission Correspondent: Mr. Ray Wang
Beijing Believe Tech. Service Co., Ltd 1-202, Build 3, Beijing New World No.5 Chaogyang Rd., Chaogyang District Beijing, 100024 CN Tel: +86-21-50313932, Fax: +86-21-68093116 Email: Ray.Wang@believe-med.com
#### Identification of Proposed Device
Trade name: Disposable Thermometer Probe Sheath Common Name: Thermometer Probe Covers and Sheaths Model: Oral
#### Regulatory Information
Classification Name: Clinical Electronic Thermometer Classification: 2 Product Code: FLL Regulation Number: 880.2910 Review Panel: General Hospital
#### Proposed Predicate Device
K112289 Disposable Thermometer Covers and Sheaths KANG ZE INDUSTRIAL Co., Ltd.
#### Intended Use Statement
The Disposable Thermometer Probe Sheaths are intended for use as barriers between digital thermometers and users' oral cavities to avoid the possible contamination and infection during temperature measuring. The sheaths are non-sterile and intended for single use only.
{4}------------------------------------------------
#### K152690 -510K Summary
## Device Description
The Disposable Thermometer Probe Sheaths are used for oral measurements for digital thermometers. The products and packaging are non-sterile, not made with natural rubber latex, and intended for single patient use only.
The proposed device is a shell-like device, made of PE/EVA, use to cover the oral thermometer as barrier for avoiding possible contamination and infection during measurement of temperature. The device is covered by the paper (Medical Kraft Paper) at the top and bottom which to avoid unnecessary touch with the device before usage and will be removed from the device before use. The outer most packing material is carton box. Each outer packing contains 100 pieces of the device.
| ITEM | Proposed Device-K112289 | Predicate Device-K152690 |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Disposable Thermometer Probe Sheaths<br>are intended for use as barriers between digital<br>thermometers and users' oral cavities to avoid<br>the possible contamination and infection<br>during temperature measuring. The sheaths are<br>non-sterile and intended for single use only. | The Disposable Thermometer Covers and<br>Sheaths are intended for use as barriers<br>between digital or mercury thermometers and<br>users' rectum or oral cavities to avoid the<br>possible contamination and infection during<br>temperature measuring. These covers and<br>sheaths are non-sterile and intended for single<br>use only. |
| OTC use | Yes | Yes |
| Basic Design | 3 layer design (Bottom Side Paper +<br>PE/EVA film + Top Side Paper);<br>Removing paper before use; | 3 layer design (Bottom Side Paper +<br>PE/EVA film + Top Side Paper);<br>Removing paper before use; |
| Device Materials<br>Composition | PE/EVA | PE/EVA |
| Size | One size | Varied size |
| Thickness | 23.5 um | NA |
| Outsider<br>Dimension | Length<br>115 mm<br>Width<br>30 mm | Length<br>120 mm<br>Width<br>33 mm |
| Available<br>Dimension | Length<br>90 mm<br>Width<br>30 mm | Length<br>88 mm<br>Width<br>29 mm |
| Density | 0.87-0.88g/cm3 | N/A |
| Strength | Meet the requirements of ASTM 1104<br>Air Pressure of 8.4 kPa (1.2 psi) for 5<br>seconds | Meet the requirements of ASTM 1104 |
### Subject and Predicate Device Comparison:
{5}------------------------------------------------
## K152690 -510K Summary
| Accuracy | Meet the requirements of ASTM E1104/E1112 | Meet the requirements of ASTM E1104/E1112 |
|----------|-------------------------------------------|-------------------------------------------|
| | ±0.3℃, Less than 38 ℃; | |
| | ±0.2℃,35.8℃ to less than 37 ℃; | |
| | ±0.1℃, 37.0 ℃ to 39.0 ℃; | |
| | ±0.2℃, greater than 39.0 ℃ to 41.0 ℃; | |
| Color | None | None |
| | | |
| Sterile | No | No |
## Discussion
The basic intended use is the same for both the predicate and subject device in that the device functions as an outer barrier to the thermometer. The predicate device is indicated for both oral and rectal thermometer whereas the subject device is only indicated for oral thermometers. Although small differences exist with predicate device in terms of dimension specification. whether the dimension differences affects temperature measuring accuracy of covered thermometers was evaluated. Performance testing for temperature accuracy was investigated per ASTM E1104/ASTM E1112, the test results showed compliance with the temperature accuracy requirements, thus the subject device has demonstrated it is substantially equivalent with the predicate device.
### Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications to perform similarly and in a substantially equivalent manner to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro ■ Cytotoxicity
- ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
-ASTM E1104-98: Standard Specification for Clinical Thermometer Probe Covers and Sheaths
-ASTM E1112-00: Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
-Bench Testing for the performance of Dimensions, Integrality
Conclusion: The Disposable Thermometer Probe Sheaths have demonstrated the product is similar in intended use and materials of construction as that of the predicate device. Performance testing showed it performs in a manner that is substantially equivalent to the legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.