FORA DIAMOND NEXUS MINI INFRARED FOREHEAD THERMOMETER

K132457 · Biocare Asia Corporation , Ltd. · FLL · Feb 21, 2014 · General Hospital

Device Facts

Record IDK132457
Device NameFORA DIAMOND NEXUS MINI INFRARED FOREHEAD THERMOMETER
ApplicantBiocare Asia Corporation , Ltd.
Product CodeFLL · General Hospital
Decision DateFeb 21, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

FORA Diamond Nexus Mini Thermometer is intended for the intermittent measurement and monitoring of human body temperature from forehead measurement at home.

Device Story

FORA Diamond Nexus Mini Infrared Forehead Thermometer measures human body temperature by detecting infrared energy emitted from the skin surface. The device is intended for home use by patients or caregivers. It captures infrared radiation, converts it into a temperature value, and displays the result on an LCD screen. The device features a white backlight and is powered by a 3V CR2032 lithium battery. It provides a non-invasive method for temperature monitoring, allowing users to track body temperature intermittently. The output assists users in identifying fever or monitoring health status, facilitating timely clinical decision-making or self-care.

Clinical Evidence

Clinical study conducted with 120 subjects. Primary endpoints included clinical bias and clinical repeatability. Results showed a pooled clinical bias of -0.08 (Bias+1.96 SD: 0.44; Bias-1.96 SD: -0.60) and pooled clinical repeatability of 0.11. Bench testing confirmed compliance with ASTM E1965-98 and EN12470-5 standards for accuracy, shock, and environmental conditions.

Technological Characteristics

Infrared forehead thermometer; measures infrared energy emitted from skin surface. Powered by 3V CR2032 lithium battery. Features white LCD backlight. Complies with ASTM E1965-98, EN12470-5, IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Measurement range: 89.6°F to 109.4°F (32°C to 43°C).

Indications for Use

Indicated for intermittent measurement and monitoring of human body temperature via forehead in a home setting for all ages.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ FEB 2 1 2014 Image /page/0/Picture/2 description: The image shows the logo for Biocare Asia Co., LTD. The logo includes the letters 'bca' in a stylized font, with a heart shape inside the letter 'c'. To the right of the letters is the company name in Chinese characters. Below the logo is the full company name, 'BIOCARE ASIA CO., LTD.', in a bold, sans-serif font. # 510(k) Summary Jan.8 2014 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. The Assigned 510(k) number is: K132457 #### 1. Submitter's Name: BioCare Asia Co., LTD Address: No.260, Mayun Road, New District Suzhou, Jiangsu, 215129, P.R.C Contact person: Shuchi Chang / Administrator Phone: +886-3-2616678#6533 +886-3-2616679 Fax: E-mail:shuchichang@biocare.com.tw #### 2. Device Name and Classification Trade name: FORA Diamond Nexus Mini Infrared Forehead Thermometer Common name: Clinical electronic thermometer Regulation section: 21 CFR 880.2910 Classification Code: FLL, Clinical electronic thermometer Classification Panel: General Hospital (80) Device Classification: Class II #### 3. Predicate Device The predicate device is URIGHT TD-1240 Thermometer (K113159) marketed by TaiDoc Technology Corporation. #### 4. Device Description FORA Diamond Nexus Mini Infrared Forehead Thermometer is characterized by measuring human body temperature from the surface of human skin. It utilizes infrared {2}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for BIOCARE ASIA CO., LTD. The logo features the letters 'bca' in a stylized font, with a heart shape inside the 'c'. To the right of the letters is Chinese text, and below the letters is the company name in English. technology to measure infrared energy emitted from the skin surface when making a temperature measurement. ### 5. Intended Use FORA Diamond Nexus Mini Thermometer is intended for the intermittent measurement and monitoring of human body temperature from forehead measurement at home. # 6. Comparison with predicate: FORA Diamond Nexus Mini Infrared Forehead Thermometer has the following similarities to the predicate device: - V Same Indications for use. - A Same operating principle and fundamental scientific technology. - 公 Incorporates the same materials. - A Same shelf life. - Packaged using the same materials. A - Has the same memory storage capacity. V The FORA Diamond Nexus Mini Infrared Forehead Thermometer has the following dissimilarity to the predicate device: | Item | FORA Diamond Nexus Mini<br>Infrared Forehead<br>Thermometer | U-RIGHT TD-1240<br>Thermometer (k113159) | |---------------------------|-------------------------------------------------------------|------------------------------------------| | Body Measuring Range | 89.6°F to 109.4°F<br>(32°C to 43°C). | 73.4°F to 111.2°F | | Operating temperature | 60.8°F to 104°F<br>(16°C to 40°C). | 50°F to 104°F | | Storage temperature range | -13°F to 131°F<br>(-25°C to 55°C) | -4°F to 140°F | | power down time | 15 seconds | 3 minutes | {3}------------------------------------------------ | battery type | One 3V CR2032 lithium<br>battery | 1.5V AA | |---------------|----------------------------------|---------| | LCD Backlight | white | blue | The differences between the two devices list as: display, parameters and appearance. There are no significant differences that affect the safety and effectiveness. Therefore, BioCare Asia FORA Diamond Nexus Mini Infrared Forehead Thermometer is substantially equivalent to the legally marketed device U-RIGHT TD-1240 Thermometer manufactured by TaiDoc Technology Corporation, K113159. ## 7. Performance Characteristics: (Non-clinical testing data) The FORA Diamond Nexus Mini Infrared Forehead Thermometer was validated by the tests According to ASTM E1965-98 standard and met the requirements of EN12470-5:2000 standard. A brief description for each test was given in this section. Following table lists items of tests, related standard complied and acceptance criteria. | Item | Standard complied | Acceptance criteria | Attachment | |-----------------------------------------------------|------------------------------|----------------------------|------------| | Clinical accuracy | ASTM E 1965-98<br>EN-12470-5 | $\pm$ 0.3°C ( $\pm$ 0.5°F) | 23 | | Laboratory accuracy | ASTM E 1965-98<br>EN-12470-5 | $\pm$ 0.3°C ( $\pm$ 0.5°F) | 23 | | Shock test | ASTM E 1965-98 | $\pm$ 0.3°C ( $\pm$ 0.5°F) | 25 | | Storage environment test | ASTM E 1965-98 | $\pm$ 0.3°C ( $\pm$ 0.5°F) | 26 | | Displayed temperature range | ASTM E 1965-98 | $\pm$ 0.3°C ( $\pm$ 0.5°F) | 27 | | Operating condition range<br>36-39°C (96.8-102.2°F) | ASTM E 1965-98<br>EN-12470-5 | $\pm$ 0.2°C ( $\pm$ 0.4°F) | 28 | {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for BIOCARE ASIA CO., LTD. The logo consists of the letters "bca" in a stylized font, with a heart shape inside the letter "c". To the right of the letters is the company name in Chinese characters. Below the logo is the company name in English, "BIOCARE ASIA CO., LTD." | Operating condition range<br><36°C(96.8°F) or >39°C (102.2°F) | ASTM E 1965-98<br>EN-12470-5 | $\pm$ 0.3°C ( $\pm$ 0.5°F) | 28 | |---------------------------------------------------------------|------------------------------|----------------------------|----| | | | | | | Safety | IEC 60601-1 | Evaluated by SGS | 19 | | Electromagnetic<br>compatibility (EMC) | IEC 60601-1-2 | Evaluated by SGS | 22 | | Home Healthcare Equipment | IEC 60601-1-11 | Evaluated by SGS | 20 | ### 8. Performance Characteristics: (Clinical testing data) Two kind of errors were represented in clinical accuracy: clinical bias and clinical repeatability. Following table is the pooled clinical bias and its standard deviation of 1267. | Device | Sample size | The pooled clinical bias | Bias+1.96 SD | Bias-1.96 SD | Uncertainty | |--------|-------------|--------------------------|--------------|--------------|-------------| | 1267 | 120 | -0.08 | 0.44 | -0.60 | 0.02 | The following show the pooled clinical repeatability of 1267. | Device | Sample size | The pooled clinical bias | |--------|-------------|--------------------------| | 1267 | 120 | 0.11 | #### 7. Test Principle: The thermometer measures temperature by reading infrared radiation emitting from the skin and converts it into a temperature value. ### 8. Conclusions Based on the information provided in this submission, the FORA Diamond Nexus Mini Infrared Forehead Thermometer is substantially equivalent to the predicate U-RIGHT 1240 Thermometer, model TD-1240. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 21, 2014 BioCare Asia Corporation, Limited Ms. Shuchi Chang Administrator, R&D Department No.260, Mayun Road New District Suzhou, Jiangsu 215129 P.R. China Re: K132457 Trade/Device Name: FORA Diamond Nexus Mini Infrared Forehead Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: January 15, 2014 Received: January 17, 2014 Dear Ms. Chang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Ms. Chang Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Image /page/5/Picture/7 description: The image shows the name "Kwame Ulmer's" in a stylized font. The letters are bold and have a textured appearance, giving them a unique and eye-catching look. The name is arranged horizontally, with "Kwame" stacked above "Ulmer's". The overall design is simple yet effective, making it easily recognizable. for Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K132457 #### Device Name FORA Diamond Nexus Mini Infrared Forehead Thermometer #### Indications for Use (Describe) FORA Diamond Nexus Mini Infrared Forehead Thermittent messurement and monitoring of human body temperature from forehead measurement at home. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over - The - Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. THE CHILIFORNIA CONTRACTOR FOR FOR FOR FOR FOR FOR FOR FOR FOR THE FOR FOR THE FOR A FOR A FOR Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/6/Picture/12 description: The image shows the logo for the Food and Drug Administration (FDA). The logo consists of the letters "FDA" in a stylized font. The letters are outlined with a thick black line, and the interior of the letters is filled with a pattern of lines. The logo is simple and recognizable. Digitally signed by Richard C. Chapman Date: 2014.02.19 14:49:37 -05'00' Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. {8}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or spansor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." .
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...