K131155 · Medisim, Ltd. · FLL · Aug 28, 2013 · General Hospital
Device Facts
Record ID
K131155
Device Name
UGFT3; FHT7
Applicant
Medisim, Ltd.
Product Code
FLL · General Hospital
Decision Date
Aug 28, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The UGFT3 is a non-sterile, reusable clinical thermometer intended for the determination of human's body temperature using the temple (forehead); and/or underarm (axilla); and/or behind the ear as measurement sites.
Device Story
The UGFT3 is a non-sterile, reusable electronic clinical thermometer. It uses a thermistor sensor to sample heat flow from blood vessels to the skin surface (forehead, underarm, or behind the ear). The device employs Rapid Accurate Temperature Establishment (R.A.T.E.) technology, a heat conduction-based principle, to calculate blood vessel temperature and correct it to core body temperature. Operated by the user via front-panel buttons, the device displays results on an LCD screen in Celsius or Fahrenheit. It includes a Medication Timer Reminder (MTR) feature. The device is intended for home or professional use. Healthcare providers or patients use the output to monitor body temperature for clinical decision-making. The device benefits patients by providing a non-invasive, multi-site method for core temperature assessment.
Clinical Evidence
Clinical study (n=239, ages 0-91, 138 male/101 female) evaluated behind-the-ear temperature measurement accuracy. The study compared UGFT3 (FHT7) against the Vicks Behind Ear Thermometer. Results demonstrated the UGFT3 is as reliable as the predicate. No adverse effects were reported. Bench testing performed per ASTM E1112-00 included display range, lab accuracy, operating/storage environment, and drop tests; the device passed all criteria.
Technological Characteristics
Materials: ABS plastic, silicone rubber, stainless steel. Sensing: Thermistor. Energy: 2x 1.5V AAA batteries. Form factor: Handheld with LCD (20mm x 19.6mm). Connectivity: None. Sterilization: Non-sterile. Software: Proprietary R.A.T.E. algorithm for predictive temperature calculation.
Indications for Use
Indicated for determination of human body temperature in all age groups using forehead, underarm, or behind-the-ear sites. No specific contraindications listed.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Dual Mode Up Grade Forehead/Underarm Thermometer (K051422)
{1}------------------------------------------------
Kr3155
# SECTION 5 - 510(k) SUMMARY UGFT3
Date prepared: 20th of April, 2013
1. Submitter information:
Medisim Ltd.
r
G.G Communications Center
Neve Ilan
Israel 90850
Tel: 972-2-5791921
Fax: 972-2-5791926
Contact person: Moshe Yarden
Submission: 7.4.2013
#### 2. Device information:
Trade Name: Up-Grade Forehead/ Underarm/Behind Ear Thermometer, UGFT3, FHT7**
Common name: Electronic Clinical Thermometer
Product code: FLL; Product class: II
Product classification: Clinical electronic thermometer, § 880.2910
Product Panel: 80 General hospital
#### 3. Substantial equivalent:
a. Medisim's Dual Mode Up Grade Forehead/Underarm Thermometer, cleared under K051422
b. Vicks (KAZ) Behind Ear Thermometer, cleared under K103839
c. NIR Thermometer, cleared under K090386
**AUG 2 8 2013**
{2}------------------------------------------------
#### 4. Description of the new device:
The UGFT3 is a non-sterile, reusable clinical thermometer intended for the determination of human's body temperature using the forehead and/or underarm and/or behind the ear as measurement sites. The UGFT3 is utilizing the Rapid Accurate Temperature Establishment (R.A.T.E.TM) technology, a technology based on the heat conduction principle, for determination of body temperature. Utilizing the R.A.T.E.TM technology, the device rapidly samples the heat flowing from the blood vessels to the skin surface. Then, based on the heat flow, the device calculates the blood vessel temperature, and then corrects this value to the body core temperature.
User operation of the UGFT3 is identical to the operation of the Dual Mode Up Grade Forehead/Underarm Thermometer, cleared under K051422. The device allows user control by use of buttons located on the front panel of the device and the measured body temperature is displayed on an LCD screen, in Co or Fo units. The UGFT3 has an identical hardware as the UGFT2, as it uses the same thermistor sensor and electronic processing unit, however, the UGFT3 has a modified proprietary software algorithms.
Device is made of ABS plastic and silicone rubber
Image /page/1/Picture/4 description: The image shows a digital thermometer. The thermometer has a screen that displays the number 98.6. There are three buttons on the thermometer, one with a power symbol, one with the word "SET", and one with a clock symbol.
#### 5. Intended use of the device
The device indication for use is:
The UGFT3 is a non-sterile, reusable clinical thermometer intended for the determination of human's body temperature using the temple (forehead) and/or underarm (axilla) and/or behind the ear as measurement sites.
The UGFT3 is substantially equivalent regarding the intended use to the Dual Mode Up Grade Forehead/Underarm Thermometer cleared under K051422 and the Vicks Behind Ear
{3}------------------------------------------------
Thermometer cleared under K103839. The UGFT3 and it's predicate devices are intended for use for determination of core -body temperature through measurements taken on the skin above a blood vessel.
The Dual Mode Up Grade Forehead/Underarm Thermometer is intended to measure using the forehead and underarm sites only; the Vicks Behind Ear Thermometer is intended to measure using the behind the ear site only; the new UGFT3 is intended to measure either in the underarm, and/or behind the ear and/or the forehead.
The UGFT3 is substantially equivalent to NIR Thermometer regarding the presence and function of MTR Feature (Medication Timer Reminder).
| Intended Use and | UGFT3 | Dual Mode Up Grade | Vicks Behind Ear | NIR Thermometer |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | | Forehead/Underarm<br>Thermometer cleared<br>under K051422 | Gentle Touch<br>Thermometer<br>cleared under<br>K103839 | K090386 |
| Product Code /<br>Class | FLL / II | FLL / II | FLL / II | FLL / II |
| Indications | A non-sterile, reusable<br>clinical thermometer<br>intended for the<br>determination of<br>human's body<br>temperature using the<br>forehead and/or axilla<br>and/or behind the ear<br>sites as measurement<br>sites. | A non-sterile, reusable<br>clinical thermometer<br>intended for the<br>determination of<br>human's body<br>temperature using the<br>forehead and/or axilla as<br>measurement sites. | A non-sterile,<br>reusable clinical<br>thermometer<br>intended for the<br>determination of<br>human's body<br>temperature using<br>behind the ear as<br>measurement site. | The NIR is non<br>sterile non-invasive<br>infrared<br>thermometer<br>intended for the<br>intermittent<br>calculation of<br>human body<br>temperature of<br>people of all ages<br>for home and<br>professional<br>use. |
| Population | All age groups | All age groups | All age groups | All age groups |
| Operational Modes<br>(Measurement<br>Sites) | Forehead<br>Axilla<br>Behind the ear- above<br>the PAA (posterior<br>auricular artery). | Forehead<br>Axilla | Behind the ear | Forehead |
| Thermometer Type | Clinical Electronic | Clinical Electronic | Clinical Electronic | Clinical Electronic |
| Intended Use and<br>Technological<br>Characteristics | UGFT3 | Dual Mode Up Grade<br>Forehead/Underarm<br>Thermometer cleared<br>under K051422 | Vicks Behind Ear<br>Gentle Touch<br>Thermometer<br>cleared under<br>K103839 | NIR Thermometer<br>K090386 |
| Principle of<br>Operation | Thermometer<br>Predictive<br>Thermometer | Thermometer<br>Predictive Thermometer | Thermometer<br>Predictive<br>Thermometer | Thermometer<br>Infra Red<br>Thermometer |
| Labeling | Instructions For Use | Instructions For Use | Instructions For<br>Use | Instructions For Use |
| Major Components | Main (Control &<br>Processing) Unit<br>including display, and<br>Probe Tip | Main (Control &<br>Processing) Unit,<br>including display, and<br>Probe Tip | Main (Control &<br>Processing) Unit,<br>including display,<br>and Probe Tip | Main (Control &<br>Processing) Unit<br>including display,<br>and Probe housing |
| Sensor | Thermistor | Thermistor | Thermistor | Thermopile |
| Display | LCD (20 mm x 19.6<br>mm) | LCD (20 mm x 19.6<br>mm) | LCD (20.4mm x<br>20.4 mm) | LCD (20 mm x 19.6<br>mm) |
| Power<br>Requirements | 2 x 1.5V AAA<br>Batteries | 2 x 1.5V AAA Batteries | 1 x CR2032 (3V<br>lithium battery) | 3x1.5V AAA<br>Batteries |
| Patient Contact<br>Materials | Case - ABS Plastic<br>Probe Tip - Stainless<br>Steel , ABS plastic<br>and silicon rubber | Case - ABS Plastic<br>Probe Tip - Stainless<br>Steel , ABS plastic and<br>silicon rubber | Case - ABS<br>Plastic, rubber | ABS plastics |
| Clinical<br>Measurement<br>Display | Temperature | Temperature | Temperature | Temperature |
| Units of<br>Measurement | °C, °F | °C, °F | °C, °F | °C, °F |
| Clinical<br>Measurement<br>Range | 35.0°C to 42.0°C<br>95.5°F to 107.6°F | 35.0°C to 42.0°C<br>95.5°F to 107.6°F | 35.5°C to 42.0°C<br>95.9°F to 107.6°F | 22.0°C to 40.0<br>°C (71.6°F to<br>104.0°F) |
| Ambient<br>Temperature for<br>measurement | 16°C to 40°C<br>60.8°F to 104.0°F | 16°C to 40°C<br>60.8°F to 104.0°F | 10°C to 40°C<br>50°F to 104.0°F | 16.0°C to 40.0°C<br>(60.8 °F to 104 °F) |
| Environment | 15% to 95% RH non- | 15% to 95% RH non- | 15% to 95% RH | 15% to 95% RH |
| Intended Use and<br>Technological<br>Characteristics | UGFT3 | Dual Mode Up Grade<br>Forehead/Underarm<br>Thermometer cleared<br>under K051422 | Vicks Behind Ear<br>Gentle Touch<br>Thermometer<br>cleared under<br>K103839 | NIR Thermometer<br>K090386 |
| | condensing | condensing | non-condensing | non-condensing |
| Medication Timer<br>Reminder with<br>alert (MTR) | Timer to alert the<br>patient when to take<br>next dosage | None | None | Timer to alert the<br>patient when to take<br>next dosage |
| Technical<br>Accuracy | ASTM E1112-00 | ASTM E1112-00 | ASTM E1112-00<br>ASTM E1965-98 | ASTM E1112-00<br>ASTM E1965-98 |
| Response Time | 6-8 seconds | 6-8 seconds | 1 second | 2 seconds |
| Activation | Automatic body<br>detection | Automatic body<br>detection | Manually | Automatic body<br>detection |
| Dimensions | Length: 10.0cm (3.9")<br>Width: 5.2cm (2.05")<br>Depth: 3.2cm (1.25") | Length: 10.0cm (3.9")<br>Width: 5.2cm (2.05")<br>Depth: 3.2cm (1.25") | Length: 14.4cm<br>(6")<br>Width: 4.2cm<br>(1.65")<br>Depth: 2.4cm<br>(0.95") | Length: 10.5cm<br>(4.1")<br>Width: 4.5 cm<br>(1.77")<br>Depth: 3.5 cm<br>(1.37") |
| Weight (including<br>batteries) | 44gr | 44gr | 76gr | 70gr |
## 6. Comparison to the predicate devices:
{4}------------------------------------------------
{5}------------------------------------------------
The differences and similarity between UGTF3 Device and its predicates:
# UGTF3 Vs Dual Mode Up Grade Forehead/Underarm:
## The devices are similar in:
- Similar principal of operation and user interface .
- Measuring sites: Forehead/Under arm .
# Devices are different:
- Additional measurement location for UGFT3: Behind-ear; .
{6}------------------------------------------------
# UGTF3 Vs Vicks Gentle Touch device:
## Devices are similar in:
- . Measuring location: Behind Ear
## Devices are different in:
- . Technological principal of operation
- . Additional measuring site of the UGFT3: Forehead /under arm
# UGTF3 Vs NIR Thermometer:
## Devices are similar in:
- . MTR- Medication Reminder Timer feature
- . Measuring location Forehead
## Devices are different in:
- Technological principal of operation .
- Additional measuring site of the UGFT3: Forehead /under arm .
Essentially, the intended use between the two predicate thermometers, Upgrade dual mode and Vicks gentle touch. are due to the fact that the new UGFT3 device is the only device that is capable of measuring temperature in all three locations.
Comment: The predicate NIR thermometer is presented in this application to demonstrate similarity to the new UGFT3 device MTR (Medication Reminder Timer) feature only. The other characteristics of the NIR thermometer are is irrelevant to the comparison to the UGTF3 and it.
## 7. Device Testing (Software Validation, Bench and Clinical test)
The new device UGFT3 was tested in order to fully validate its performance under the following tests:
- A. Software Validation Tests.
- B. Performance testing:
- 1. Bench testing according to ASTM E-1112-00
2. Clinical testing behind the ear temperature measurement accuracy validation through a clinical study according to ISO80601-2-56.
## A. Software Validation:
The software was validated to assure the full functionality of the device UGFT3.
The software validation protocol and relevant ECO report can be found in section 14 of this submission. The software validation information includes the following:
- · Hardware architecture diagram of up-grade forehead thermometer
- · General software flowchart of up-grade forehead thermometer
{7}------------------------------------------------
- · Summary of system requirement specification (SRS)
- · Summary of system test plan (STP)
- · Device Hazard Analysis
The device software was validated successfully and changes in software are implemented successfully as device is fully functional.
B. 1. Behind the ear temperature measurement accuracy validation through Performance of a clinical study:
The primary objective of the study was to demonstrate the effectiveness of the new device UGFT3 (also referred to in the clinical study report as FHT7) for measurements behind the ear. Clinical accuracy of the UGFT3 was evaluated by comparing to temperature measurements taken using the Vicks behind ear Thermometer to validate the compliance of the behind the ear site. The secondary objective was to test,the similarity between temperature measured on the forehead site (above the STA) compared to the temperature measured on the behind the ear site (above the PAA) using the UGFT3.
Total subjects were 239, Subjects were 138 males and 101 females, within ages range of 0-91 years (The full clinical study report is detailed in section 17).
Subjects with and without fever were included in the clinical study, Subjects from all groups were covered.
The UGFT-3 was used safely without causing any discomfort to the patients. There were no observations or reports of adverse effects. The UGFT-3 was found to be as reliable as Vicks Behind the Ear thermometer, cleared under K103839.
B. 2. Bench testing Bench testing was performed in accordance to ASTM E-1112-00 The following tests were performed on the UGTF3 device:
o Display Temperature Range Test
o Lab Accuracy Test
- o Operating Environment Test
- o Storage Environment Test
.
o Drop Test
The device passed all bench tests successfully as it withstands acceptance criteria for each and every bench test performed.
{8}------------------------------------------------
#### 8. Conclusions
The clinical and non-clinical testing, as well as the clinical research and previous testing done on the original device prove that the UGFT3 device is safe and effective as the cleared predicate devices.
The UGFT3 thermometer (K131155) is substantially equivalent to the UGFT2 thermometer , cleared under K051422, in the principal of operations and the measuring sites of forehead and underarm: the UGFT3 was found substantially equivalent to the Vicks thermometer, cleared under K103839 ,in regards of the measuring site - behind ear; and The UGFT3 thermometer is substantially equivalent to the NIR thermometer, cleared under K090386, in regards to the Medication Timer Reminder feature.
Based on the safety and performance testing and compliance with acceptable voluntary standards, we believe that the UGFT3 Thermometer is substantially equivalent to its predicate devices cited above without raising new safety and/or effectiveness issues.
**Notes:
1. Throughout the submission the device is referred to as UGFT3 or FHT7 which is another trade name for the device, used in the device's labeling.
2. The dual mode upgrade forehead thermometer is referred to as UGFT2/FHT2/FHT5.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the right.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
August 28, 2013
Medisim, Limited Mr. Moshe Yarden Quality Assurance Manager G.G Communications Center Neve Ilan JERUSALEM IL 90850
Re: K131155
Trade/Device Name: Up-grade forehead/underarm/behind car thermometer (UGFT3) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: July 11, 2013 Received: July 18, 2013
Dear Mr. Yarden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
#### Page 2 - Ms. Yarden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/9/Picture/7 description: The image shows the name "Mary S. Runner - S" in a bold, sans-serif font. The text is black against a white background. The letters are clear and legible, although there is some noise around the letters.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
510(k) Number: K131155
Device Name : Up-grade forehead/underarm/behind ear thermometer (UGFT3)
Indications For Use:
The UGFT3 is a non-sterile, reusable clinical thermometer intended for the determination of human's body temperature using the temple (forehead); and/or underarm (axilla); and/or behind the ear as measurement sites.
AND/OR Over-The-Counter Use_ V Prescription Use ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
: :
. '
·
Page 1 of 1
Richard C. Chapman 2013.08.28 11:55:47 · -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.