K130231 · Jinxinbao Electronic Co.,Ltd · FLL · Oct 15, 2013 · General Hospital
Device Facts
Record ID
K130231
Device Name
NON-CONTACT INFRARED THERMOMETER
Applicant
Jinxinbao Electronic Co.,Ltd
Product Code
FLL · General Hospital
Decision Date
Oct 15, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
JXB-182 Non-contact Infrared Thermometer is an infrared thermometer intended to measure forehead temperature of infants and adults without contacting human body. It can be used by consumers in household environment and doctor in clinic as reference.
Device Story
Non-contact infrared thermometer; measures forehead temperature via infrared radiation emitted by arterial blood flow. Device includes external probe for ambient temperature analysis to ensure measurement accuracy. Operator holds device 3-5cm from forehead; activates sensor to capture infrared heat. Output displayed on LCD with three-color LED backlight indicating temperature ranges; includes three-language prompt function. Used in home or clinic by consumers or physicians. Provides instant temperature readings to assist in clinical decision-making; non-contact design avoids cross-contamination.
Clinical Evidence
Bench testing only. Device demonstrated compliance with ASTM E 1965-98 (Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature). No clinical trial data provided.
Technological Characteristics
ABS plastic housing. Infrared radiation sensing principle. Powered by 2 AA batteries (DC 3V). LCD display with 0.1°C resolution. Features three-color LED backlight and multi-language audio prompts. Measurement distance 3-5cm. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for measuring forehead temperature in infants and adults. Suitable for use by consumers in household settings and by clinicians in medical facilities.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Remote Infrared Thermometer, Model RC002 (K090361)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo with the letters J, X, and B. The letters are in a bold, sans-serif font. The letter B has a design on the upper right side of the letter. There is a horizontal line under the letters.
Jinxinbao Electronic Co., Ltd
Section 5 510(k) Summary
# (As required by 21 CFR 807.92(a))
#### 5.1 Submitter Information
- Company: Jinxinbao Electronic Co., Ltd
- · Address: No.38, Huanzhenxi Road, Dagang Town, Nansha, Guangzhou, China
- · Phone: 86-20-34938449
- · Fax: 86-20-22865301
- · Email: xhoucc@jxb-htb.net
· Contact: Jiacheng Guo, General Manager
· Date: November 25, 2012
#### 5.2 Device Information
· Trade/Proprietary Name: Non-contact Infrared Thermometer-JXB 182
· Common Name: Infrared Thermometer
· Classification Information: Classification: 2
Review Panel: General Hospital
Name: Thermometer, Electronic, Clinical
Regulation Number: 21 CFR 880.2910
Product Code: FLL
• Predicate Device: Remote Infrared Thermometer, Model RC002, K090361
- · Device Description:
All objects, solid, liquid or gas, emit energy by radiation. The intensity of this energy depends on the temperature of the object. The JXB-182 infrared thermometer is therefore able to measure the temperature of a person by the energy the person emits. This measurement can be taken thanks to an external temperature probe on the device which permanently analyses and registers the ambient temperature. Therefore, as soon as the operator holds the thermometer near the forehead and activates the radiation sensor, the measurement is taken instantly by detection of the infrared heat generated by the arterial blood flow. Body heat can therefore be measured without any interference
OCT 15 2013
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters J, X, and B in a bold, sans-serif font. The letters are black and appear to be slightly distressed or textured. The letter B is partially obscured by a pattern of small dots or speckles, which adds a visual element to the design.
from the heat of the surrounding environment.
The JXB-182 thermometer without contact has been developed using latest infrared technology . This technology allows temporal artery (TA) temperature to be taken at a distance of about 5cm away from the forehead. Then the result will be displayed on LCD. Besides, it owns Led backlight which can show three different colors in different temperature ranges and three-language prompt function. • Intended Use:
JXB-182 Non-contact Infrared Thermometer is an infrared thermometer intended to measure forehead temperature of infants and adults without contacting human body. It can be used by consumers in household environment and doctor in clinic as reference.
| Item | Subject Device | Predicate Device |
|--------------------|----------------------------------|------------------------------|
| Device Name | Non-contact Infrared Thermometer | Remote Infrared Thermometer |
| | Model JXB-182 | Model RC002 |
| Measurement | Infrared Radiation Detection | Infrared Radiation Detection |
| Method | | |
| Range | In body mode: 32-42.9° C | In body mode: 32-42.9° C |
| Accuracy | ± 0.3°C | ± 0.3°C |
| Precision | ± 0.3°C(34-35.9° C) | ± 0.3°C(34-35.9° C) |
| | ±0.2°C(36-39°C) | ±0.2°C(36-39°C) |
| | ±0.3°C(39-42.5°C) | ±0.3°C(39-42.5°C) |
| Measurement | 3cm~5cm | 5cm-8cm |
| Distance | | |
| Power Supply | DC 3V (2AA Batteries) | DC 3V (2AA Batteries) |
| Material | ABS | ABS |
| Operating | Temperature: 10° C to 40° C | Temperature: 10° C to 40° C |
| Condition | Humidity rate: ≤ 95% | Humidity rate: ≤ 85% |
| Display | LCD | LCD |
| Display Resolution | 0.1° C | 0.1° C |
# 5.3 Comparison of Required Technology Characteristics
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters J, X, and B. The letters are in a bold, sans-serif font and are arranged horizontally. The letter B has a pattern of small circles on it. The logo is black and white.
Jinxinbao Electronic Co., Ltd
| Dimension | 175×50×52mm(L×W×H) | 196×150×50mm(L×W×H) |
|--------------------|----------------------------|---------------------|
| Weight | 148g | 220g |
| With Led Backlight | Yes, with three colors | Yes |
| Automatic Stop | 5 seconds | 5 seconds |
| Language<br>Prompt | Yes, three languages | No |
| Function | (French, English, Spanish) | |
#### Brief Summary:
The subject device - JXB 182 Non-contact Infrared Thermometer enjoys almost the same design and technological characteristics with the predicate device, such as the same measurement method, accuracy, precision, power supply, display, material and measurement range, and the similar operating condition and measurement distance.
Though they differ in dimension and weight, such trivial difference in appearance will not affect the core function of the device, let alone affecting the comparison of substantial equivalence. A lso they are different in Led backlight and language prompt function, that is, the subject device has two more functions - three-color backlight and three-language prompt function than the predicate device. But such addition will not affect their core usage of the device. Besides, the tests mentioned below have demonstrated the subject device is as safe and effective as the predicate device.
#### 5.4 Discussion of Tests Performed
#### Clinical Tests:
Subject device JXB 182 is compliant to the ASTM E 1965-98(Reapproved 2009), Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.
#### Non-clinical Tests:
The subject device was tested to evaluate its safety and effectiveness according to the following applicable standards:
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters J, X, and B. The letters are in a bold, sans-serif font and are arranged horizontally. The letter B has some decorative elements on the right side. The logo is black and white.
# · Electrical Safety:
IEC 60601-1 Medical electrical equipment Part 1: General requirement for basic safety and essential performance
### · Electromagnetic Compatibility:
IEC 60601-1-2: 2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
### · Performance Effectiveness:
ASTM E 1965-98(Reapproved 2009), Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature.
#### 5.5 Conclusion
The subject device, JXB-182 Non-contact Infrared Thermometer has the same intended use and similar characteristics with the predicate device. Besides, the safety of the subject has been demonstrated by the following standards: IEC 60601-1(Electrical Safety), IEC 60601-1-2(Electromagnetic Compatibility). And since the subject is composed of the same material as the predicate device, they enjoy the same biocompatibility. Moreover, the subject device meets the specific requirements of ASTM E 1965, which further embodies that the subject device is both safe and effective.
So from the above information, it is reasonable to conclude that the subject device, JXB-182 Non-contact Infrared Thermometer, is substantially equivalent to the predicate device, RC002 Remote Infrared Thermometer.
4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 0903 New Hampshire Avenue cument Control Center - WO66-G60 Silver Spring, MD 20993-0002
# October 15, 2013
Jinxinbao Electronic Company, Limited C/O Ms. Helen Nan, General Manager Wenzhous Cytech Information Service Company, Limited Room 404, Building 7, Jinhuichang Homeland, Liuhongqiao Road WENZHOUS CITY, 325000, Zhejiang Province CHINA
Re: K130231
Trade/Device Name: JXB182 Non-contact Infrared Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: II Product Code: FLL Dated: November 25, 2012 Received: August 7, 2013
Dear Ms. Nan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Nan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Mary S. Brunner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters J, X, B, and D. The letters are in a bold, sans-serif font. The letters are arranged horizontally, with the J on the left and the D on the right. The letters are all the same size. The logo is black and white.
# Section 4 Indications for Use Statement
510(k) Number (if known): K 13023 !
Device Name: JXB182 Non-contact Infrared Thermometer
Indications for Use:
JXB182 Non-contact Infrared Thermometer is an infrared thermometer intended to measure forehead temperature of infants and adults without contacting human body. It can be used by consumers in household environment and doctor in clinic as reference.
Prescription Use
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 80) Subpart C) ·
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
l
AND/OR
Image /page/6/Picture/13 description: The image shows the logo of the Food and Drug Administration (FDA). The logo is in black and white and appears to be a stylized version of the letters "FDA". The letters are blocky and have a slightly pixelated appearance. The logo is relatively small and centered in the image.
Richard C. Chapman 2013.10.15 10:16:38 -04'00'
Page _1_ of _1_
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.