← Product Code [FLL](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL) · K120671

# DIGITAL PACIFIER THERMOMETER (MODEL NO.: TM-03) (K120671)

_Cotronic Technology Limited · FLL · May 17, 2013 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K120671

## Device Facts

- **Applicant:** Cotronic Technology Limited
- **Product Code:** [FLL](/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL.md)
- **Decision Date:** May 17, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

Digital Pacifier Thermometer TM-03 is a non-sterile, reusable & battery-operated clinical electronic thermometer for use by medical professional or at home. The TM-03 is intended to measure the oral temperature of infant to children up to 5 years old.

## Device Story

Digital Pacifier Thermometer TM-03 is a battery-operated, reusable clinical thermometer shaped as a pacifier for oral temperature measurement in infants/children up to 5 years old. Device features a medical-grade silicon rubber nipple containing an NTC thermistor and an ABS hard plastic body. Principle of operation: NTC thermistor resistance changes with external temperature; resistance changes are converted to frequency changes via an R-C oscillator circuit to calculate temperature. Used by clinicians or parents at home. Output is a digital temperature reading. Benefits include ease of use for infants/children compared to traditional electronic thermometers.

## Clinical Evidence

Clinical testing validated performance against ASTM E1112 requirements. Results demonstrated accuracy of +/-0.2°F within the 95°F – 107.6°F range. Additional bench testing confirmed electrical safety (EN60601-1), electromagnetic compatibility (FCC Part 15, IEC60601-1-2), and biocompatibility (ISO 10993).

## Technological Characteristics

Materials: medical silicon rubber (nipple), ABS hard plastic (body). Sensing: NTC thermistor. Energy: 1 x LR41 battery. Form factor: pacifier-shaped. Standards: ASTM E1112, EN60601-1, IEC60601-1-2, ISO 10993, CFR Title 16 Part 1511. Connectivity: None (standalone).

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Microlife Digital Pacifier Thermometer, Model MT1751Q ([K051100](/device/K051100.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Section IV: 510k Summary

### Applicant's Identification

| Applicant             | Cotronic Technology Limited        |
|-----------------------|------------------------------------|
| Phone Number          | +852-2602-2021 / +86-755-8145-3681 |
| Fax Number            | +852-2609-4485 / +86-755-8145-3080 |
| Contact Person        | Ms. Tani Mok                       |
| Date of Application · | 1st March, 2012                    |
| Date of Summary       | 13th March, 2013                   |

#### Device's Identification

Digital Pacifier Thermometer (Model No.: TM-03) Device Proprietary Name Clinical Electronic Thermometer Common Name Thermometer, Electronic, Clinical Classification Name (Class II per 21 CFR 880.2910)

### Marketed Devices to which Equivalence is Claimed

| DEVICE                                  | MANUFACTURER               | 510(k) Number |
|-----------------------------------------|----------------------------|---------------|
| Microlife Digital Pacifier Thermometer, | Microlife Intellectual     | K051100       |
| Model MT1751Q                           | Property GmbH, Switzerland |               |

#### Device Description

Digital Pacifier Thermometer TM-03 is a reusable, battery-operated clinical electronic thermometer, with which a thermistor installed inside the nipple, for use by medical professional or at home. Different from the traditional electronic thermometer, our TM-03 is designed specifically for the measuring the oral temperature of baby or little child. Therefore, our TM-03 is molded as a pacifier. The patient contact portion is the thermometer's nipple which is made of medical silicon rubber. The body of TM-03 is then manufactured by ABS hard plastic.

TM-03 utilizes the phenomenon of thermal conductivity changes of our selected NTC thermistor when there are changes of external environment temperature. Whenever the external temperature changed, the resistance of our thermistor changed uniformly. Then, the changes in resistance are converted to the changes of frequency of R-C oscillator circuit. This means TM-03 keep monitoring the frequency of the oscillator and the temperature can be measured accurately.

# MAY 1 7 2013

{1}------------------------------------------------

### Intended Use (Indication for Use)

Digital Pacifier Thermometer TM-03 is a non-sterile, reusable & battery-operated clinical electronic thermometer for use by medical professional or at home. The TM-03 is intended to measure the oral temperature of infant to children up to 5 years old.

## Comparison of Technological Characteristics between New Device and Predicate Devices

Our Digital Pacifier Thermometer TM-03 is non-sterile, reusable & battery-operated clinical electronic thermometer and utilizes the phenomenon of thermal conductivity changes of NTC thermistor to measure the oral temperature reading of infant/children. The key components of device are a NTC thermistor, silicon nipple & a ABS hard plastic The predicate device adopts exactly same methodology and key components for bodv. measuring the infant's oral temperature.

| Item                                  | Predicate Device<br>(Microlife MT1751Q<br>Digital Pacifier<br>Thermometer)                                                            | TM-03            |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Product Design                        | Dummy-like Shape                                                                                                                      | Equivalent       |
| Temperature Measurement<br>Technology | NTC Thermistor<br>Resistance Technique                                                                                                | Equivalent       |
| Key Temperature Sensor                | NTC Thermistor                                                                                                                        | Equivalent       |
| Material                              | Resistances, capacitances,<br>transistors, buzzer, MCU, a<br>thermistor, button key,<br>PCB, silicon, ABS plastic<br>cabinet, battery | Equivalent       |
| Energy Source                         | 1 x SR 41battery                                                                                                                      | 1 x LR41 battery |
| Measurement Time                      | Around 5 minutes                                                                                                                      | 3 — 5mins        |

A comparison table between New Device and Predicate Device is listed as below:

### Clinical & Non-clinical Tests

A systematic & independent clinical test was conducted to validate the performance of the Digital Pacifier Thermometer (Model No. TM-03). The results demonstrated that our device (Model No. TM-03) meets the requirement of +/-0.2°F for 95°F – 107.6°F. Also, we have asked an independent laboratory to test the accuracy of our device (Model No. TM-03) based on the ASTME1112 and the results further verified our device meets the requirement of +/-0.2°F for 95°F - 107.6°F.

{2}------------------------------------------------

For the electrical safety of our device, comprehensive testings were carried out according various international standards. Firstly, our devices are tested and compiled with the EN60601-1 (Standard Title: Medical Equipment Part 1: General requirements for safety), which includes a collective of safety tests. Then, the electrical safety of our devices is further verified via conducting the electromagnetic compatibility test according to FCC Part 15 and IEC60601-1-2. The results classified and demonstrated our devices as Class B device and are compiled with the acceptance limit (40dB (uV/m)).

Furthermore, we have conducted various tests on the safety of material to verify our device (Model No. TM-03) was made by safe material. Firstly, we applied the ISO10993 sets to test the biocompatibility of our device. The results showed no evidence of causing delayed dermal contact sensitization. And our device is classified as a non-irritant and not a cytotoxic potential. Secondly, we tested the silicon nipple of our device with US Compliance Policy Guide Part 7117.11 for nitrosamine content in rubber baby bottle nipples. We notice that the testing scope of Policy Guide Part 7117.11 was not applicable to our product category. However, we still would like to have a reference on the safety towards nitrosamine content in our device (Model No. TM-03). The results demonstrated our device is safe and within the corresponding acceptance limit (10ppb). Thirdly, our device/device components not made with BPA. BPA is not used in the manufacture of the accessible parts of the device (TM-03).

- Comprehensive safety and EMC tests were performed and compiled to demonstrate (Model No.TM-03) is safe for use. Tests include
1. Biological evaluation of medical devices --- Part10: Tests for irritation and skin sensitization

2. Biological evaluation of medical devices --- Part12: Sample preparation and reference materials

3. Biological evaluation of medical devices --- Part 5: Tests for In Vitro Cytotoxicity

4. Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature

5. CFR Title 16 Part 1511 - Requirement for pacifiers

6. CPG Sec. 500.450 Volatile N-Nitrosamines in Rubber Baby Bottle Nipples

7. Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Requirement

8. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance

### Conclusion

{3}------------------------------------------------

#### Additional information of K120671

### K120671

Our Digital Pacifier Thermometer TM-03 has the same intended use and similar technological characteristics as predicate device (K051100). Moreover, bench testing contained in this submission and clinical testing supplied demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, the Digital Pacifier Thermometer TM-03 is substantially equivalent to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

#### May 17, 2013

Ms. Tani Mok Cotronic Technology, Limited Floor 4-6, Block 7, West of Zhoushi Road Xixiang Street Baoan Zone, Shenzhen, Guang Dong Province. China 518126

Re: K120671

Trade/Device Name: Digital Pacifier Thermometer (Model No .: TM-03) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: April 24, 2013 Received: May 16, 2013

Dear Ms. Mok:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Mok

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

· http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.

Sincerely yours,

DAGRID

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director

FOR Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and

Dental Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

Enclosure

{6}------------------------------------------------

# Section V: Statement of Indications for Use

1120071 510(k) Number (if known):

Device Name:

Digital Pacifier Thermometer (Model No.: TM-03)

Indications For Use:

Digital Pacifier Thermometer TM-03 is a non-sterile, reusable & battery-operated clinical electronic thermometer for use by medical professional or at home. The TM-03 is intended to measure the oral thermometer of infant to children up to 5.years old.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Richard C. Chapman 2013.05.17 09:50:15 -04'00'

(Division Sign-Off) Tivision of Anesthesiology, General Hospital ection Control, Dental Devices

K12667) 510(k) Number: ___

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K120671](https://fda.innolitics.com/submissions/HO/subpart-c%E2%80%94general-hospital-and-personal-use-monitoring-devices/FLL/K120671)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
