DIGITAL THERMOMETER, MODELS BT-A11CN, BT-A21CN, BT-A41CN
K101387 · Fudakang Industries Co., Ltd. · FLL · Aug 3, 2010 · General Hospital
Device Facts
Record ID
K101387
Device Name
DIGITAL THERMOMETER, MODELS BT-A11CN, BT-A21CN, BT-A41CN
Applicant
Fudakang Industries Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Aug 3, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Fudakang Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home. BT-A11CN, BT-A21CN, BT-A41CN can be used for axillary measurement, oral measurement and rectal measurement.
Device Story
Hand-held, battery-operated digital thermometers (models BT-A11CN, BT-A21CN, BT-A41CN) for clinical or home use. Device utilizes thermistor sensor to detect body temperature; ASIC processes analog signal from thermistor by comparing against reference resistor; digital output displayed on LCD. Includes buzzer for audible feedback. Operated by clinicians or patients for axillary, oral, or rectal temperature monitoring. Provides rapid temperature readings to assist in clinical decision-making regarding patient health status.
Clinical Evidence
Bench testing only. Compliance with ASTM E1112-2006 for accuracy, precision, and repeatability confirmed. Biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10 passed. Electrical safety and electromagnetic compatibility verified per IEC 60601-1 and IEC 60601-1-2.
Indicated for measurement and monitoring of human body temperature in patients of all ages in hospital or home settings via axillary, oral, or rectal routes.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Famidoc Digital Clinical Thermometer FDTH-V0-1, FDTH-V0-2, FDTH-V0-3, FDTH-V0-4 (K072641)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY
AUG D& 2010
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
#### 1.0 Submitter's Information
Establishment Registration Name:
Fudakang Industrial Co., Ltd
No.8 Yinghe Road, Yuanjiangyuan Management Zone, Changping Town, Dongguan, Guangdong, China . .
### Contact Person of applicant:
Mr. Bob Yu
No.8 Yinghe Road, Yuanjiangyuan Management Zone, Changping Town, Dongguan, Guangdong, China
TEL: +86-769-81098181 FAX: +86-769-81098187 Email: bobyu@fudakang.com
#### Contact Person of the Submission:
Mr. Stephen Ling
MEDLAB (Shenzhen) Information Service Co., LTD
Room 2706, Block A, ZhongFang JinYuan Buiding, Xinwen Road Shenzhen, Guangdong, P.R. China, 518034
TEL: +86-755-83089699 FAX: +86-755-83089760
Email: lingbensuo(@gmail.com, please CC sabrinawei(@hotmail.com
Section 5, Page 5-1
{1}------------------------------------------------
## 2.0 Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|-----------------------------------|
| Device Common Name: | Thermometer, electronic |
| Trade Name: | Digital Thermometer |
| Model: | BT-A11CN, BT-A21CN, BT-A41CN |
| Classification name: | Thermometer, electronic, clinical |
| Review Panel: | General Hospital |
| Product Code: | FLL |
| Regulation Class: | II |
| Regulation Number: | 880.2910 |
#### 3.0 Predicate Device Information
| Sponsor: | Famidoc Technology Co., Ltd. |
|----------------|-------------------------------------------------------------------------------------|
| Device: | Famidoc Digital Clinical Thermometer FDTH-V0-1,<br>FDTH-V0-2, FDTH-V0-3, FDTH-V0-4. |
| 510(K) Number: | K072641 |
### 4.0 Device description
Fudakang digital thermometers are hand-held, reusable, battery operated. The models include BT-A11CN, BT-A21CN, and BT-A41CN. All the models have the same indication for use. The difference points are their shapes, colors, voice function, and measurement speed.
The operation principle is based on thermosensor and ASIC technology. A thermistor is used as thermosensor. The ASIC gets the sensor's signal from human body, then processes the signal and calculates the result, after that displays the temperature result by a LCD. The digital thermometer comprises a thermistor for getting temperature signal, a reference resistor for comparing the resistance of the thermistor, a buzzer for sounding effect, an ASIC for processing the target temperature digitally, and a LCD for displaying the temperature result.
#### 5.0 Intended Use
Fudakang Digital The mometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home.
#### 6.0 Performance Summary
The devices meet the ASTM Standard Specification for Electronic Thermometer for intermittent Determination of Patient Temperature ASTM E1112-2006, as well
Section 5, Page 5-2
{2}------------------------------------------------
as IEC 60601-1 and IEC 60601-1-2 requirements. The bench test confirmed that the accuracy, precision and repeatability of device are compliance with ASTM E1112-2006 requirements.
The thermometers can be used for axillary, oral and rectal measurement. For axillary use, it directly contacts body skin. For oral or rectum use, it directly contacts membrane tissue. The patient contact materials in the device are made of stainless steel, ABS plastic, and thermoplastic rubber. They are all compliance with ISO10993-1 biocompatibility requirements, and pass the ISO10993-5 and ISO10993-10 tests.
#### 7.0 Comparison to predicate device and conclusion
Compare with predicate device, they are very similar in design principle, intended use, functions, material and the adopting applicable standards. The differences between applicant device and predicate device do not raise and new questions of safety or effectiveness.
l ). Indication for use
Famidoc Digital Clinical Thermometer can be used for axillary measurement, oral measurement and rectal measurement. Fudakang Digital Thermometers, BT-A11CN, BT-A21CN, BT-A41CN have the same measurement application.
2). Environment conditions
The environment conditions of Fudakang Digital Thermometers are a little different from Famidoc Digital Clinical Thermometer. But they are both compliance with IEC 60601-1 and ASTM E1112 requirements. So the difference of environment conditions will not raise any safety or effectiveness issue.
3). Accuracy
The accuracy of subject device is: 95.0 - 102.0 F : +/- 0.2 F 35.0 - 39.0℃: +/- 0.1℃ The rest: +/- 0.2°C - - -
The accuracy of predicate device is: +/- 0.1℃ (0.2F ) for whole range
Although there is a little difference, but the subject device accuracy is compliance with ASTM E 1112 requirements. So the difference of accuracy will not raise any safety or effectiveness issue.
4). Response time
{3}------------------------------------------------
The Subject Device is 60s. The Predicate Device is 30s.
The Response time of Fudakang Thermometers is a little different from Famidoc Digital Clinical Thermometer. They are all compliance with IEC 60601-1 and ASTM E 1112 requirements. The difference of Response time will not raise any safety or effectiveness issue.
Conclusions: Fudakang Digital Clinical Thermometers, model: BT-A11CN, BT-A21CN, BT-A41CN are substantially equivalent to the predicate devices.
Submission prepared date: March 23, 2010 8.0
{4}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Fudakang Industries Company, Limited C/O Mr. Marc M. Mouser Responsible Third Party Official Underwriters Laboratories, Incorporated 2600 NW Lake Road Camas, Washington 98607-9526
## AUG 03 2010
Re: K101387
Trade/Device Name: Digital Thermometer, Models BT-A11CN, BT-A21CN, BT-A41CN Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: July 9, 2010 Received: July 19, 2010
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2-Mr. Mouser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Buono
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Statement of Indications for Use
510(k) Number (if known):
K101387
Device Name: Digital Thermometer Models: BT-A11CN, BT-A21CN, BT-A41CN
AUG-0 3 2010
Indications For Use:
Fudakang Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home.
BT-A11CN, BT-A21CN, BT-A41CN can be used for axillary measurement, oral measurement and rectal measurement.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
8/3/10
Page 1 of 1
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Powises Division of Anton, Dental Devices
510(k) Number: k101387
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.