ACTHERM INFRARED EAR THERMOMETER ACT 8000R SERIES AND ITS PROBE COVER MED ACCU2010
K093795 · Actherm, Inc. · FLL · Mar 12, 2010 · General Hospital
Device Facts
Record ID
K093795
Device Name
ACTHERM INFRARED EAR THERMOMETER ACT 8000R SERIES AND ITS PROBE COVER MED ACCU2010
Applicant
Actherm, Inc.
Product Code
FLL · General Hospital
Decision Date
Mar 12, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Actherm Infrared Ear Thermometers ACT 8000R Series are used to measure body temperature from auditory canal. This device is intended for household and hospital use on people of all age, and is used with or without probe cover medACCU2010.
Device Story
Device is an infrared ear thermometer using a thermopile sensor to detect thermal radiation from the auditory canal. Input signal from thermopile is processed by a micro-controller to calculate body temperature, which is displayed digitally on an LCD. Device includes a buzzer for audible feedback and optional features like backlight and scan LED. Operated by clinicians or patients in hospital or home settings. Powered by two 1.5V rechargeable alkaline batteries; charging occurs via a dedicated base station. Device cannot be used while recharging. Output provides immediate temperature reading to assist healthcare providers or patients in monitoring body temperature, facilitating clinical decision-making regarding fever management.
Clinical Evidence
Bench testing only. Device demonstrated compliance with ASTM E 1965-98 (Infrared Thermometer Standard), EN 60601-1 (Electrical Safety), EN 60601-1-2 (EMC), EN 12470-5, and ISO 10993 (biocompatibility). Clinical accuracy and precision were validated against requirements of ASTM E 1965-98 and EN 12470-5, showing performance equivalent to the predicate device.
Technological Characteristics
Thermopile-based infrared sensor; micro-controller processing; LCD display; buzzer. Powered by two 1.5V rechargeable alkaline batteries with base station charger. Materials biocompatibility per ISO 10993-1, -5, -10. Complies with ASTM E 1965-98, IEC 60601-1, and IEC 60601-1-2. Software-based temperature calculation.
Indications for Use
Indicated for body temperature measurement via the auditory canal in patients of all ages in household and hospital settings.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Actherm Infrared Ear Thermometer ACT 8000 Series (K090031)
Submission Summary (Full Text)
{1}------------------------------------------------
6F, No.18, Jhanye 2ndRd., Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C. e - mail: actherm@actherm.com.tw
율 886 36 66 95 96 @ 886 36 66 96 97
Image /page/0/Picture/3 description: The image shows a logo with the word "Aachen" written in a stylized font. Below the logo is a graphic element that resembles a stylized wing or swoosh. To the right of the wing, there is a certification mark with the letters "TUV" and "SOD" inside an octagon, and the letters "CE" followed by the number "0123" are present to the right of the certification mark.
1093795
MAR 1 2 2010
# 510(k) Summary
### Submitter's Identification
Actherm Inc. 6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, Taiwan, R.O.C. +886 3 666 9596 Tel : Fax: +886 3 666 9697
Contact Person : Richard Hsieh Date the summary was prepared : December 04, 2009
### Name of the device
Trade Name – Actherm Infrared Ear Thermometer ACT 8000R Series and Its Probe Cover medACCU 2010 Common Name - Infrared Ear Thermometer Classification name - Clinical Electronic Thermometer (21CFR 880.2910, Product FLL)
### Predicate Device Information
Actherm Infrared Ear Thermometer ACT 8000 Series and Its Probe Cover medACCU 2010 510 (k) number : K090031.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ન્દ્ર
{2}------------------------------------------------
6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C. e - mail: actherm@actherm.com.tw
Aetherium
#### Device Description
Actherm Infrared Ear Thermometers (Model ACT 8000R Series) are electronic thermometers using a thermopile as the temperature sensor. The sensor's electric signal is then calculated and displayed by a Micro-Controller. These thermometers display the temperature decimal. To compare with ACT 8000 Series, Model ACT 8000R Series has the same indication for use. ACT 8000R Series can be made with various functions, such as Dual scale, Backlight, Scan LED, W/O cover, following with the model number indicates scale switchable, with Backlight, with Scan completed indicator, without probe cover, will be added following the model number.
8 886 36 66 95 96
886 36 66 96 97
The infrared ear thermometer comprises: a thermopile for temperature sensing, a buzzer for sounding effect, a Micro Controller and a LCD display for calculating and displaying the target temperature digitally.
Actherm Infrared Ear Thermometers ACT 8000R Series can be combined with accessories, base, wall mount and power adapter. The base is a tray for holding probe cover box, power input, power/charge indicator, portable/ fixed probe-cover dispenser and the area for placing digital thermometer and its probe cover. For the wall mount, it can be fixed on the wall and the base can be put on this wall mount.
The system uses two 1.5V DC alkaline rechargeable batteries for the power supply and the battery power is automatically the Micro-Controller and displayed in LCD if the battery is exhausted. The batteries can be recharged only when the ear thermometer ACT8000R put on the base. ACT 8000R cannot be used when recharging.
#### (Labeling please see Attachment 1)
#### Intended Use
Actherm Infrared Ear Thermometers ACT 8000R Series have the same intended use as the predicate device. They are used to measure body temperature from auditory canal. This device is intended for household and hospital use on people of all age, and is used with or without probe cover.
{3}------------------------------------------------
6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, TAIWAN. R.O.C. e - mail: actherm@actherm.com.tw
Image /page/2/Picture/2 description: The image shows two phone numbers: 886 36 66 95 96 and 886 36 66 96 97. There is a phone icon next to the first number and a document icon next to the second number. The image also contains a logo and a certification mark.
#### Technological Characteristics
Actherm Infrared Ear Thermometers ACT 8000R Series have the same mode of operation, design principle, and biological specifications as the predicate device. ACT 8000R Series can be made with various functions, such as Dual scale, Backlight, Scan LED, W/O cover, following with the model number indicates scale switchable, with Backlight, with Scan completed indicator, without probe cover, will be added following the model number.
#### Comparison to 510(k) Predicate Devices
The Actherm Infrared Ear Thermometers, the ACT 8000R Series are substantially equivalent to the following digital thermometers: Infrared Ear Thermometer Model ACT 8000 Series. Its 510(k) number is K090031.
To compare with ACT 8000 Series, the Actherm Infrared Ear Thermometers ACT 8000R Series are similar in design and intended use to the predicates differing only in supplied batteries.
The ACT8000 Series use two 1.5V DC alkaline batteries and the ACT 8000R Series use two 1.5V alkaline rechargeable batteries that can be recharged only when the infrared ear thermometer ACT 8000R is put on the base. ACT 8000R cannot be used when recharging.
All products use the same temperature sensing element - a thermopile, an LCD display, Micro-Controller and a buzzer.
#### Substantial Equivalence
Infrared Ear Thermometers ACT 8000R Series, and the claimed device are the same in intended use, essential component; sensor, power requirement, PCB circuit, software program, biological specification, performances, design principle, dimension of probe cover, and are also close in electrical safety, and materials, etc. However, in these characters, the slight difference of measuring range, battery type, labeling, won't affect product safety and effectiveness.
{4}------------------------------------------------
r
6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, TAIWAN. R.O.C. e - mail: actherm@actherm.com.tw
会 886 36 66 95 96 图 886 36 66 96 97
The change is in power supply. The ACT 8000 Series use two 1.5V DC alkaline batteries and the ACT 8000R Series use two 1.5V alkaline rechargeable batteries that can be recharged only when the infrared ear thermometer ACT 8000R is put on the base is a tray for holding probe cover box, power input, power/charge indicator, portable/ fixed probe-cover dispenser and the area for placing digital thermometer and its probe cover. ACT 8000R cannot be used when recharging. It has passed electrical safety performed test, therefore, it is safe for human body.
According to the description above, the two devices are substantial equivalent.
#### Performance Data
Actherm Infrared Ear Thermometers ACT 8000R Series meets the ASTM Standard Specification for Infrared Thermometer for Intermittent Determination of Patient Temperature (ASTM E 1965-98: 2003), as well as EN 60601-1 (IEC 60601-1), EN 60601-1-2 (IEC 60601-1-2), EN 12470-5 and ISO 10993-1, ISO 10993-5, ISO 10993-10 requirements. For all body contacting materials, the analysis has been made and the identical materials have been used in other legally marketed devices under the same use conditions.
### Safety and Effectiveness
Actherm Infrared Ear Thermometer ACT 8000 and ACT 8000R and Probe Cover medACCU 2010 fulfill the testing requirements of ASTM E 1965-98, EN 60601-1), EN 60601-1-2 (IEC 60601-1-2) and EN 12470-5. ACT 8000R just add the function of recharge, when recharging ACT 8000R, it must be put on the base and cannot take temperature. In addition, ACT 8000R Series also meets electrical Safety performed test such as EN 60601-1 (IEC 60601-1-2 (IEC 60601-1-2), so it is safe for human body. Moreover, Infrared Ear Thermometers ACT 8000R Series, and the claimed device are the same in intended use, essential component, sensor, power requirement, PCB circuit, Software program, biological specification, performances, design principle, dimension of probe and probe cover, and also close in electrical safety, and materials, etc. The clinical test results of Actherm Infrared Ear Thermometer ACT 8000 and Probe Cover medACCU 2010 conform to the requirements of ASTM E 1965-98 and EN 12470-5 that are the same performance for ACT 8000R Series. The product quality is standard-guaranteed. The compliance tests demonstrate precise measuring function and the clinical accuracy of the thermometer. Therefore, these thermometers are safety and effectiveness.
4
{5}------------------------------------------------
# ACTIFICI
6F, No.18, Jhanye 2ndRd., Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C. e - mail: actherm@actherm.com.tw
Image /page/4/Picture/3 description: The image shows a logo with the word "Acaborn" written in a stylized font. Below the word is a horizontal line. To the right of the line is a certification mark with the letters "TUV" inside of an octagon. Next to the certification mark is the text "CE0123".
### Conclusion
ﺮ ﮐ
Actherm Infrared Ear Thermometers ACT 8000R Series have intended use, principles of operation, and similar technological characteristics as predicate devices. Moreover, bench testing contained in this submission and clinical testing supplied demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Actherm Infrared Ear Thermometers ACT 8000R Series are substantially equivalent to the predicate devices.
5
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Richard Hsieh General Manager Actherm, Incorporated 6F, No. 18 Jhanye 2nd Road Hsinchu Science Park Hsinchu China Taiwan 30078
MAR ] 2 2010
Re: K093795
Trade/Device Name: Actherm Infrared Ear Thermometer ACT 8000R Series and Its Probe Cover medACCU2010 Regulation Number: 21CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: February 4, 2010 Received: February 16, 2010
Dear Mr. Hsieh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
#### Page 2- Mr. Hsieh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm 1.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
ACTHERM Inc.
6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C. e - mail: actherm@actherm.com.tw
☎ 886 36 66 95 96 B 886 36 66 96 97
Image /page/7/Picture/3 description: The image shows a logo with the word "Aetherium" in a stylized font. Below the word is a graphic element that appears to be a stylized representation of a bird or wing. To the right of the graphic element are two certification marks: a TÜV SÜD octagon logo and a CE mark followed by the number 0123.
## Indication for Use
510(k) Number (if known) : K093795
Device Name : Actherm Infrared Ear Thermometer ACT 8000R Series and Its Probe Cover medACCU2010
Indication For Use :
Actherm Infrared Ear Thermometers ACT 8000R Series are used to measure body temperature from auditory canal. This device is intended for household and hospital use on pcople of all age, and is used with or without probe cover medACCU2010.
Prescription Use Over-The-Counrter Use ____x AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Th
(Division Sign-Q (Offision of Anesthesiology. General Hospital Infection Control, Dental Devices
Page 1 of 1
16 093 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.