TEMPTELLER - HIGH SPEED DIGITAL THERMOMETERS, MODELS DT-302, DT-312, DT-412, DT-502
K042202 · Oriental System Technology, Inc. · FLL · Sep 10, 2004 · General Hospital
Device Facts
Record ID
K042202
Device Name
TEMPTELLER - HIGH SPEED DIGITAL THERMOMETERS, MODELS DT-302, DT-312, DT-412, DT-502
Applicant
Oriental System Technology, Inc.
Product Code
FLL · General Hospital
Decision Date
Sep 10, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is an electronic clinical thermometer using a thermistor to detect body temperature from the oral , armpit , and rectal in neonatal , pediatric and adult population used in the clinical and home testing .
Device Story
Electronic clinical thermometer using thermistor sensor to detect body temperature; ASIC (Application Specific IC) processes thermistor electrical signals; compares against reference resistor; calculates target temperature; displays on LCD. Features include buzzer for audible alerts, battery level monitoring, and water-resistant housing. Models vary by tip type (flexible vs. rigid) and button placement. Used by clinicians or patients in clinical or home environments. Provides rapid temperature readings (within 10 seconds in 37°C water bath). Output assists healthcare providers in assessing patient febrile status.
Clinical Evidence
Controlled human clinical studies compared the subject thermometers to predicate devices. Readings were validated against conventional/accepted oral or rectal temperature measurement methods. Bench testing confirmed compliance with ASTM E 1112-2000, IEC 601-1-1, and IEC 601-1-2 standards.
Technological Characteristics
Thermistor-based temperature sensing; ASIC-driven signal processing; 1.55V battery power; LCD display; water-resistant housing; flexible or rigid tip configurations. Complies with ASTM E 1112-2000, IEC 601-1-1, and IEC 601-1-2.
Indications for Use
Indicated for body temperature measurement via oral, axillary, or rectal routes in neonatal, pediatric, and adult populations in clinical or home settings.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
TempTeller- Digital Thermometers, Model DT-102 (K992601)
VICKS - Speed Read Digital Thermometer, Model V911, V965
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Oriental System Technology Inc. The logo consists of a stylized eye-like symbol on the left, followed by the company name in a simple, sans-serif font. The text is arranged in two lines, with "ORIENTAL SYSTEM" on the top line and "TECHNOLOGY INC." on the bottom line.
K042202
#### SEP 1 0 2004 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA 1990 and CFR§807.92.
The assigned 510(k) number is:
#### 1. Submitter's Identifications:
Oriental System Technology Inc. 2F No.23, Industry E, Road 9th, Science Based Industrial Park Hsinchu, Taiwan, R.O.C
# Contact:
Mr. Herman Lee General Manager
Date of Summary Preparation: August 12, 2004
#### 2. Name of the Device:
TempTeller-High Speed Digital Thermometer , Model DT302 , DT312 , DT412,DT502
# 3. Information of the 510(k) Cleared Device (Predicate Device):
- 1. TempTeller- Digital Thermometers, Model DT-102
510(k) Number: K#992601
2.VICKS - Speed Read Digital Thermometer , Model V911 , V965
# 4. Device description:
The OSTI Temp-Teller High Speed Digital Thermometer(Model DT-302 DT-312 , DT-412 , DT-502) is an electronic thermometer using a thermistor as the temperature sensor . The sensor's electric signal is calculated and displayed by an ASIC (Application Specific IC)
The High Speed digital thermometer comprises: a thermistor for temperature sensing , a reference resistor for comparing the resistance of the thermistor , a buzzer for sounding effect , an ASIC and a LCD display for calculating and displaying the target temperature digitally . The system uses a 1.55 V battery
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a black and white logo. The logo consists of a circle with a spiral design inside. The spiral design has a small circle in the center. The logo is simple and modern.
ORIENTAL SYSTEM
TECHNOLOGY INC
for the power supply and the battery power is automatically check by the ASIC and displayed in LCD if the battery is exhausted .
The OSTI Temp-Teller High Speed Digital Thermometer (Model DT-302 , DT-312 . DT-412 . DT-502) can makes temperature taking faster , its temperature stabilize reading within 10 sec in a 37℃ water bath at ambient temperature 23+1-2°C .
DT-312 ,DT-412 is high speed digital thermometers, with flexible front tip and with power /measuring button set aside with LCD display and with water resistant function .
DT502 is high speed digital thermometers, with rigid front tip and with power /measuring button set aside with LCD display and with water resistant function .
DT302 is high speed digital thermometers, with rigid front tip and with power /measuring button on top end and with water resistant function .
The DT-302 , DT-312 , DT-412 , DT-502 is identical in functionality and performance .
The essential change include 1)Metal cap / Thermistor mechanical structure change to improve thermal transmission 2)Firm ware 3)Material 4)Shape layout and dimension .
The fundamental scientific technology of the modified device remains the same as that of the 510(k) cleared device.
# 5. Intended Use:
The device is an electronic clinical thermometer using a thermistor to detect body temperature from the oral , armpit , and rectal in neonatal , pediatric and adult population used in the clinical and home testing .
# 6. Comparison to Predicate Devices:
The OSTI Temp Teller High Speed Digital Thermometer , Model DT-301,DT-312 , DT-412 .DT-502are substantially equivalent to the following digital thermometers.
1. TempTeller- Digital Thermometers, Model DT-102
510(k) Number: K#992601
2.VICKS -- Speed Read Digital Thermometer , Model V911 , V965
The OSTI Temp Teller High Speed Digital Thermometer , Model
DT-301,DT-312 ,DT-412 ,DT-502 have the same intended use as and are similar in design to the 510(k) cleared device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Oriental System Technology Inc. The logo consists of a circular symbol on the left and the company name on the right. The circular symbol is a stylized design with a swirling pattern inside a circle. The company name is written in a simple, sans-serif font.
K042202
3 of 3
# 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards include ASTM E 1112-2000 , as well as IEC 601-1-1 and IEC 601-1-2 requirements.
FDA Guidance documents include the" Deciding When to Submit a 510(k) for a Change to An Existing Device" and "The New 510(k) Paradigm: Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications".
# 8.Discussion of Clinical Tests Performed :
Controlled human clinical studies were conducted using the OSTI Temp Teller Thermometer and predicate devices . Clinical data is presented with readings representing a conventional /currently accepted reading by oral or rectal temperature .
#### 9. Conclusions
The OSTI TempTeller -High Speed Digital thermometers, model DT302 , DT312 , DT412,DT502 have the same intended use and technological characteristics as the 510(k) cleared device. Moreover, verification and validation tests contained in this submission demonstrate that the modified portions maintained its original safety and effectiveness. Those engineering changes do not: (1) affect the intended use or (2) alter the fundamental scientific of the device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 0 2004
Oriental System Technology, Incorporated C/O Ms. Susan D. Goldstein-Falk Official Correspondent MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck. New York 11021
Re: K042202
Trade/Device Name: Temp Teller-High Speed Digital Thermometers, Models DT-302, DT-312, DT-412, DT-502 Regulation Number: 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: September 3, 2004 Received: September 7, 2004
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 wr ), it hay of cash be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not I lease be act rood that 12 a determination that your device complies with other requirements mount that I Drivinederal statutes and regulations administered by other Federal agencies. or the receir any a with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice alle listing (2) es result in the quality systems (QS) regulation (21 CFR Part 820); and if requirences as better in product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mill letter notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), n you don't the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the maj obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Cluts
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K042202
Exhibit B
×
1 of of 1 Page
510(k) Number (if known): _
Device Name: Oriental System Technology Inc. Temp Teller High Speed Digital Thermometers, Models DT-32, DT-312, DT-412, DT502.
Indications For Use:
The device is an electronic clinical thermometer using a thermistor to The device is an olootronify on the oral, armpit and rectal in the neonatal, pediatric and adult population used in clinical and home testing.
Prescription Use (Per 21 CFR 801 Subpart D)
OR
Irene Naveau for ADW 9/10/04
Over-The Counter Use
(21 CFT 807 Subpart C)
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital, Division Control, Dental Devices
510(k) Number: K042202
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.