MICROLIFE INSTANT DIGITAL ELECTRONIC THERMOMETER, MODEL QT1JA1

K031958 · Microlife Intellectual Property GmbH · FLL · Aug 15, 2003 · General Hospital

Device Facts

Record IDK031958
Device NameMICROLIFE INSTANT DIGITAL ELECTRONIC THERMOMETER, MODEL QT1JA1
ApplicantMicrolife Intellectual Property GmbH
Product CodeFLL · General Hospital
Decision DateAug 15, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Microlife Instant Digital Thermometer, Model, QT1JA1 is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm. The Microlife Instant Digital Thermometer, Model, QT1JA1 is a hand-held, non-sterile, reusable clinical thermometer intended for the determination of human temperature in either predictive (Instant) mode (5- seconds orally, 7- seconds rectally and under the arm-predictive temperature), or standard mode (actual determination of temperature). The device is for the adult and pediatric population.

Device Story

Hand-held, reusable digital thermometer; measures human body temperature via oral, rectal, or axillary routes. Operates by detecting thermistor resistance changes; converts resistance to frequency via R-C oscillator circuit; frequency changes correspond to pulse number changes to calculate temperature. Offers predictive (instant) mode (5-7 seconds) and standard mode. Used by clinicians or patients for intermittent monitoring. Output displayed for provider/user to assess patient temperature status; aids clinical decision-making regarding fever management or health monitoring.

Clinical Evidence

Controlled human clinical studies conducted. Evaluated clinical bias, clinical uncertainty, and clinical repeatability per Microlife Clinical Test Protocol.

Technological Characteristics

Thermistor-based sensing; R-C oscillator circuit; hand-held form factor; reusable; non-sterile. Compliance with ASTM E1112, IEC60601-1, and IEC60601-1-2.

Indications for Use

Indicated for intermittent measurement and monitoring of human body temperature in adult and pediatric populations via oral, rectal, or axillary routes.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## AUG 1 5 2003 ### Exhibit #1 ### 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: jn 031.958 #### 1. Submitter's Identification: Microlife Intellectual Property GmbH, Switzerland Max Schmidheiny-Strasse 201 9435 Heerbrugg / Switzerland Date Summary Prepared: June 24, 2003 #### 2. Name of the Device: Microlife Instant Digital Electronic Thermometer, Model QT1JA1 #### 3. Predicate Device Information: AVITA Instant Thermometer, Model IT101, K#012136, AVITA International Corporation #### 4. Device Description: This predictive (instant) digital electronic thermometer enables very fast and reliable measurements. With its predicative technology, this thermometer offers very high clinical accuracy and, has been designed to provide maximum userfriendliness. The basic principle of this thermometer is that a change of thermistor resistance, caused by changes of temperature, are converted to changes of frequency of R-C oscillator circuit. Therefore, temperature can be given by measuring the frequency of the oscillator. For a given time period by applying to R-C oscillator circuit, changes of temperature will correspond to changes of pulse number. {2}------------------------------------------------ #### 5. Intended Use: The Microlife Instant Digital Thermometer, Model, QT1JA1 is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm. The Microlife Instant Digital Thermometer, Model, QT1JA1 is a hand-held, nonsterile, reusable clinical thermometer intended for the determination of human temperature in either predictive (Instant) mode (5- seconds orally, 7- seconds rectally and under the arm-predictive temperature), or standard mode (actual determination of temperature). The device is for the adult and pediatric population. #### 6. Comparison to Predicate Devices: The Microlife Instant Digital Electronic Thermometer, Model QT1JA1 is similar in design and intended use to the AVITA Instant Thermometer, Model IT101, K#012136, AVITA International Corporation, differing mostly in response time, physical dimensions, a wider measurement range, display resolution and wider operating conditions. ### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: Compliance to applicable voluntary standards includes ASTM E1112, as well as IEC60601-1 and IEC60601-1-2 requirements. Guidance documents included the "FDA Guidance on the Content of Premarket Notification 510(K) Submissions for Clinical Electronic Thermometers. #### 8. Discussion of Clinical Tests Performed: Controlled human clinical studies were conducted using the Microlife Instant Digital Electronic Thermometer, Model QT1JA1. Clinical data was presented which evaluated clinical bias, clinical uncertainty and clinical repeatability per the Microlife Clinical Test Protocol outline. #### 9. Conclusions: The Microlife Instant Digital Electronic Thermometer, Model QT1JA1 has the same intended use and similar technological characteristics as the AVITA Instant Thermometer, Model IT101. Moreover, bench testing contained in this submission supplied demonstrates that any differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, the Microlife Instant {3}------------------------------------------------ Digital Electronic Thermometer, Model QT1JA1 is substantially equivalent to the predicate device. : : {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with healthcare, with three lines forming the wings of the symbol. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 15 2003 Microlife Intellectual Property GmbH C/O Ms. Susan D. Goldstein-Falk Official Correspondent mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021 Re: K031958 Trade/Device Name: Microlife Instant Digital Thermometer, Model, QT1JA1 Regulation Number: 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: June 24, 2003 Received: June 25, 2003 Dear Ms. Goldstein-Falk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Ms. Goldstein- Falk Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Exhibit B Page # 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Microlife Instant Digital Thermometer, Model, QT1JA1 Indications For Use: The Microlife Instant Digital Thermometer, Model, QT1JA1 is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm. The Microlife Instant Digital Thermometer, Model, QT1JA1 is a hand-held, non-sterile, reusable clinical thermometer intended for the determination of human temperature in either predictive (Instant) mode (5- seconds orally, 7seconds rectally and under the arm-predictive temperature), or standard mode (actual determination of temperature). The device is for the adult and pediatric population. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The Counter Use \$\frac{i}{}\$ (Optional Format 1-2-96) Patricia Crescent n of Anesthesiology, General Hospital, ion Control, Dental Devices 510(k) Number: *K031958*
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...