TEMPTELLER INFARED EAR THERMOMETERS, MODEL CT-31/31DX/32/32DX
K030299 · Oriental System Technology, Inc. · FLL · Feb 25, 2003 · General Hospital
Device Facts
Record ID
K030299
Device Name
TEMPTELLER INFARED EAR THERMOMETERS, MODEL CT-31/31DX/32/32DX
Applicant
Oriental System Technology, Inc.
Product Code
FLL · General Hospital
Decision Date
Feb 25, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is an electronic clinical thermometer using an infrared sensor to detect the body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.
Device Story
TempTeller® Infrared Ear Thermometer (models CT-31/31DX/32/32DX) measures core body temperature via infrared radiation from the tympanic membrane and adjacent tissue. Device components include an IR sensor, ambient temperature sensor, waveguide, zinc alloy heat sink, LCD display, and internal circuitry. Operation involves inserting the probe into the auditory canal; pressing an activation button initiates measurement. The waveguide confines radiation fluxes between the tympanum, IR sensor, and waveguide surface. The device captures electrical signals from the IR and ambient sensors, which are amplified, digitized, and calculated by internal firmware to determine temperature. Results are displayed on the LCD within one second. Used in home settings by patients or caregivers to monitor body temperature for clinical decision-making.
Clinical Evidence
Bench testing only. Compliance with ASTM E 1965-98, ASTM E 1104, IEC 601-1-1, and IEC 601-1-2 standards was demonstrated to verify performance and safety.
Technological Characteristics
Infrared sensor-based thermometer; zinc alloy heat sink; waveguide with reflective inner surface; ambient temperature sensor; LCD display. Connectivity: None. Power: Electronic circuit. Standards: ASTM E 1965-98, ASTM E 1104, IEC 601-1-1, IEC 601-1-2.
Indications for Use
Indicated for neonatal, pediatric, and adult populations for measuring body temperature via the auditory canal in a home setting.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
TempTeller® Infrared Ear Thermometers, Model CT-30/30DX (K010235)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Oriental System Technology Inc. The logo consists of a circular graphic on the left and the company name on the right. The graphic is a stylized image of a swirl inside of a circle.
# 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA 1990 and CFR§807.92.
The assigned 510(k) number is: צע 030299 FEB 2 5 2003
### 1. Submitter's Identifications:
Oriental System Technology Inc. 2F No.23, Industry E, Road 9th, Science Based Industrial Park Hsinchu, Taiwan, R.O.C
### Contact:
Mr. Herman Lee General Manager
Date of Summary Preparation: 30 December 2002
### 2. Name of the Device:
TempTeller® Infrared Ear Thermometers, Model CT-31/31DX/32/32DX
### 3. Information of the 510(k) Cleared Device (Predicate Device):
TempTeller® Infrared Ear Thermometers, Model CT-30/30DX 510(k) Number: K#010235
### 4. Device description:
The OSTI TempTeller® Infrared Ear Thermometers, model CT-31/31DX/32/32DX, is an electronic thermometer using an infrared sensor to detect human body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane (eardrum) and the adjacent tissue.
OSTI TempTeller® Infrared Ear Thermometers, model CT-31/31DX/32/32DX, consists mainly of five parts: an IR sensor packed together with an ambient temperature sensor, a waveguide, a heat sink made of zinc alloy, a LCD display and the associated circuit.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with a circular design. Inside the circle, there is a spiral shape that resembles an eye or a swirling vortex. To the right of the logo, there is some text, but it is not clear enough to read.
The tympanic membrane (eardrum) is thin and flooded with blood at the core temperature. The wavequide, usually a cylindrical pipe with a highly reflective inner surface for confining the radiation, is adaptive to the outer without contacting the eardrum. When inserting the probe into the ear canal, the radiation fluxes transfer among the tympanum membrane (eardrum), the IR sensor, and the inner surface of the wavequide. The ambient sensor is packed with the IR sensor to monitor the ambient temperature of the IR sensor.
To measure core temperature, an ear thermometer is inserted into a patient's outer ear canal. An activation button is pressed to start the measurement through the radiation exchanges. The electrical signal readouts form the IR sensor and the ambient temperature sensor are fed to the circuit for amplification, digitization and calculation. The measured temperature then appears on the LCD. The total operation takes one second.
The CT-31/31DX/32/32DX is identical in functionality and performance with the essential change being the external shape of the devices. The modifications to our original 510(k) cleared device, model CT-30/30DX, include dimensional and firmware. The fundamental scientific technology of the modified device remains the same as that of the 510(k) cleared device.
# 5. Intended Use:
The device is an electronic clinical thermometer using an infrared sensor to detect the body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.
# 6. Comparison to Predicate Devices:
The OSTI TempTeller® Infrared Ear Thermometers, model CT-31/31DX/32/32DX are substantially equivalent to OSTI TempTeller® Infrared Ear Thermometers, CT-30/30DX. The new models CT-31/31DX/32/32DX have the same intended use as and are similar in design to the 510(k) cleared device.
# 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards include ASTM E 1965-98 and ASTM E 1104, as well as IEC 601-1-1 and IEC 601-1-2 requirements.
FDA Guidance documents include the" Deciding When to Submit a 510(k) for a Change to An Exiting Device" and "The New 510(k) Paradigm: Alternate
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for Oriental System Technology Inc. The logo consists of a circular graphic on the left and the company name on the right. The graphic is a stylized circle with a swirl inside. The text is stacked in two lines, with "ORIENTAL SYSTEM" on the top line and "TECHNOLOGY INC." on the bottom line.
Approaches to Demonstrating Substantial Equivalence in Premarket Notifications".
# 8. Conclusions
The OSTI TempTeller® infrared Ear thermometers, model CT-31/31DX/32/32DX have the same intended use and technological characteristics as the 510(k) cleared device. Moreover, verification and validation tests contained in this submission demonstrate that the modified portions maintained its original safety and effectiveness. Those engineering changes do not: (1) affect the intended use or (2) alter the fundamental scientific of the device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 5 2003
Oriental System Technology, Incorporated C/O Ms. Susan D. Goldstein-Falk Official Correspondent Mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K030299
Trade/Device Name: Oriental System Technology Incorporated Temp Teller® Infrared Ear Thermometer, Model CT-31/31 DX/32/32DX Regulation Number: 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: January 27, 2003 Received: January 29, 2003
Dear Ms. Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Kinner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a circular symbol that resembles a stylized eye or a hurricane. The symbol consists of a dark circle with a smaller, lighter circle in the center. Around the central circle, there are curved, dark shapes that create a swirling or spiral effect, giving the impression of movement or rotation. The overall design is simple and graphic, with a focus on contrast between the dark and light areas.
510(K) Number (If known): K030299
Device Name: Oriental System Technology Inc. TempTeller® Infrared Ear Thermometer, Model CT-31/31DX/32/32DX
Indications for Use
The device is an electronic clinical thermometer using an infrared sensor to detect the human body temperature from the auditory canal in the neonatal, pediatric and adult Populations used in the home setting.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | We livinile in BRC |
|---------------|-----------------------------------------------|
| | (Division Sign-Off) |
| | Division of Anesthesiology, General Hospital, |
| | Infection Control, Dental Devices |
| 510(k) Number | 030299 |
| Prescription Use | Over-The- Counter Use |
|------------------|-----------------------|
|------------------|-----------------------|
(Per 21 CFR 801.109)
ﺎ i ne- Counter U (Optional Format 1-2-96)
OR
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.