BD Vacutainer® Luer-Lok™ Access Device, BD Vacutainer® Blood Transfer Device

K222478 · Becton, Dickinson and Company · JKA · Mar 10, 2023 · Clinical Chemistry

Device Facts

Record IDK222478
Device NameBD Vacutainer® Luer-Lok™ Access Device, BD Vacutainer® Blood Transfer Device
ApplicantBecton, Dickinson and Company
Product CodeJKA · Clinical Chemistry
Decision DateMar 10, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Intended Use

BD Vacutainer® Luer-Lok™ Access Device: The BD Vacutainer® Luer-Lok™ Access Device is a sterile, single-use, noninvasive device intended to be used by healthcare professionals for safe, closed-system venous blood collection from a female Luer of a catheter port or blood collection set, directly into an evacuated blood collection tube(s) for in vitro diagnostic testing. This device may also be used with a blood collection set with a female Luer to collect blood cultures. BD Vacutainer® Blood Transfer Device: The BD Vacutainer® Blood Transfer Device is a sterile, single use, noninvasive medical device intended to be used by healthcare professionals for the safe, closed-system, needleless transfer of venous blood from a BD Luer-Lok™ tip syringe into evacuated blood collection tube(s) or blood culture bottle(s) for in vitro diagnostic testing.

Device Story

Devices facilitate closed-system venous blood collection/transfer for in vitro diagnostic testing. BD Vacutainer® Luer-Lok™ Access Device connects to female Luer catheter ports/blood collection sets; BD Vacutainer® Blood Transfer Device connects to BD Luer-Lok™ syringes. Both devices utilize male/female Luer connectors to transfer blood into evacuated tubes or blood culture bottles. Used by healthcare professionals in clinical settings. Devices are sterile, single-use, non-invasive. Operation involves manual connection to blood source and evacuated container; vacuum pressure draws blood into container. Benefits include safe, needleless, closed-system blood handling, reducing exposure risk for clinicians.

Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical performance testing, including torque to break/unseat, cannula pull tests, air leakage, sleeve function, package integrity, and Luer compatibility per ISO 594/80369-7.

Technological Characteristics

Materials: Polycarbonate hub, stainless steel cannula, isoprene rubber sleeve, polypropylene holder. Connectivity: ISO 594 compliant 6% Luer taper (male/female). Sterilization: Ethylene Oxide (EO) to 10^-6 SAL. Form factor: Single-use, non-invasive, manual assembly. No software or electronic components.

Indications for Use

Indicated for healthcare professionals performing venous blood collection or transfer into evacuated tubes or blood culture bottles for in vitro diagnostic testing. Used with catheter ports, blood collection sets, or Luer-Lok™ syringes.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". March 10, 2023 Becton Dickinson and Company Sravan Rajamani Senior Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417 ## Re: K222478 Trade/Device Name: BD Vacutainer® Luer-Lok™ Access Device, BD Vacutainer® Blood Transfer Device Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA Dated: December 9, 2022 Received: December 12, 2022 Dear Sravan Rajamani: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Daniel Walloschek David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K222478 Device Name BD Vacutainer® Luer-Lok™ Access Device, BD Vacutainer® Blood Transfer Device #### Indications for Use (Describe) BD Vacutainer® Luer-Lok™ Access Device: The BD Vacutainer® Luer-Lok™ Access Device is a sterile, single-use, noninvasive device intended to be used by healthcare professionals for safe, closed-system venous blood collection from a female Luer of a catheter port or blood collection set, directly into an evacuated blood collection tube(s) for in vitro diagnostic testing. This device may also be used with a blood collection set with a female Luer to collect blood cultures. BD Vacutainer® Blood Transfer Device: The BD Vacutainer® Blood Transfer Device is a sterile, single use, noninvasive medical device intended to be used by healthcare professionals for the safe, closed-system, needleless transfer of venous blood from a BD Luer-Lok™ tip syringe into evacuated blood collection tube(s) or blood culture bottle(s) for in vitro diagnostic testing. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K222478 - 510(K) SUMMARY #### Summary Preparation Date 03/10/2023 #### Submitted by: Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417-1885 Phone: (201) 847-6800 #### Contact: Sravan Rajamani Senior Regulatory Affairs Specialist email: sravan.rajamani@bd.com Work: (201) 847 4719 #### Proprietary Names: BD Vacutainer® Luer-Lok™ Access Device BD Vacutainer® Blood Transfer Device #### Common or Usual Names: Blood Collection Tubes, Vials, Systems, Serum Separators #### Regulatory Information Classification Name: Blood specimen collection device Classification Regulation: 21 CFR §862.1675 Review Panel: Clinical Chemistry Class: II Product Code: JKA #### Predicate Device(s) Luer Access Device-holder with Preattached Multiple Sample Adapter (K200027) {4}------------------------------------------------ #### Device Description The BD Vacutainer® Luer-Lok™ Access Device and the BD Vacutainer® Blood Transfer Device are sterile, single use, non-invasive devices for the safe transfer of blood into an evacuated BD Vacutainer® blood collection tube or BD BACTEC™ blood culture bottle for in vitro diagnostic testing. The only difference between the devices is the BD Vacutainer® Luer-Lok™ Access device features a male Luer lock connection with compatible female Luer catheter port (blood collection set) and the BD Vacutainer® Blood Transfer Device features a female Luer lock for connection with BD Luer-Lok™ syringes. | Proposed Device<br>BD Vacutainer® Luer-LokTM<br>Access Device | Proposed Device<br>BD Vacutainer® Blood Transfer<br>Device | K200027<br>Luer Access Device-holder with<br>preattached multiple sample<br>adapter | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | The BD Vacutainer® Luer-LokTM Access Device is a sterile, single- use, noninvasive device intended to be used by healthcare professionals for safe, closed-system venous blood collection from a female Luer of a catheter port or blood collection set, directly into an evacuated blood collection tube(s) for in vitro diagnostic testing. This device may also be used with a blood collection set with a female Luer to collect blood cultures. | The BD Vacutainer® Blood<br>Transfer Device is a sterile, single use, noninvasive medical device intended to be used by healthcare professionals for the safe, closed-system, needleless transfer of venous blood from a BD Luer-LokTM tip syringe into evacuated blood collection tube(s) or blood culture bottle(s) for in vitro diagnostic testing. | The Luer access device- holder with preattached multiple sample adapter is a sterile, non-invasive device used to connect devices with male or female Luer connectors to blood collection tubes for the collection of blood. | | | Characteristic | Subject Device | Predicate Device | Comparison | | | BD Vacutainer® Luer-Lok™<br>Access Device, BD<br>Vacutainer® Blood Transfer | Luer Access Device-holder<br>with preattached multiple<br>sample adapter | | | | Device | K200027 | | | Product Code | JKA | JKA | Same | | Regulation<br>Number | 21 CFR 862.1675 | 21 CFR 862.1675 | Same | | Class | II | II | Same | | Number of Uses | Single Use | Single Use | Same | | Material | Hub (Polycarbonate)<br>Non-patient (NP) Cannula<br>(Stainless Steel)<br>Rubber Sleeve (Isoprene<br>Rubber)<br>Holder (Polypropylene (PP))<br>Bonding agent/adhesive* | Luer Lock Male Hub (MABS)<br>Non-patient Needle/Tube<br>(Stainless Steel)<br>Rubber Sleeve (Isoprene<br>Rubber)<br>Holder (PP) | Difference in Hub<br>Material. The use of a<br>different polymer for<br>the hub component<br>does not raise new<br>questions of safety or<br>effectiveness of the<br>device. | | Label/Labeling | Conform with 21 Part 801 | Conform with 21 Part 801 | Same | | Length | BTD: 61.98 mm (2.44 in),<br>LLAD: 63.42 mm (2.50 in) | 60.8±0.2mm | Difference in Length.<br>This difference does<br>not affect the<br>intended use or<br>performance of the<br>device or raise new<br>questions of safety or<br>effectiveness. | | ISO 594 | Compliant | Compliant | Same | | Biocompatibility | ISO 10993 Compliant including:<br>In vitro toxicitySkin sensitizationIntracutaneous<br>ReactivityAcute System ToxicityHemocompatibility | ISO 10993 Compliant including:<br>In vitro toxicitySkin sensitizationIntracutaneous<br>ReactivityAcute System ToxicityHemocompatibility | Same | | Sterilization<br>Method | EO | EO | Same | | Sterilization<br>Assurance Level<br>(SAL) | 10-6 | 10-6 | Same | | Connector Type | Luer-Lok™ Access Device -<br>Male Luer<br>Blood Transfer Device - Female<br>Luer | Male & Female Luers | Same | | Characteristic | Subject Device<br>BD Vacutainer® Luer-Lok™<br>Access Device and BD<br>Vacutainer® Blood Transfer<br>Device | Predicate Device<br>Luer Access Device-holder<br>with preattached multiple<br>sample adapter<br>K200027 | Comparison | | Sample Collected | Blood | Blood | Same | #### Table 1: Indications for Use The proposed and predicate device have the same basic indication for use, the collection of a blood sample using a male or female Luer connector on a multiple sample adapter with a preattached holder. All devices use an ISO 594 connector of various gender (male/female) connected to a compatible 6% Luer to transfer blood into an evacuated container (blood collection tube, blood culture bottle) for in vitro diagnostic testing. The differences in the indications statement are in language and do not pose new questions of safety or effectiveness. #### Substantial Equivalence The subject and predicate device are substantially equivalent as described in Table 3. {5}------------------------------------------------ Substantial Equivalence Comparison Table 2: {6}------------------------------------------------ *The clearance for the predicate K200027 did not list the bonding agent used. Difference in Hub Material: The materials of hub from proposed device are different with predicate device. However, the biocompatibility test for proposed device has been tested and the results comply with the requirements of ISO 10993. Therefore, this difference is not determined to affect substantial equivalence or safety/effectiveness. Difference in Length: The length of between the proposed device and predicate device is different. This difference in length is not substantial. This difference does not affect intended use or the performance of the device. Therefore, this difference does not affect substantial equivalence or safety/effectiveness. Difference in precautions: The proposed Luer-Lok™ Access Device has an added precaution for use in urinalysis to address reported adverse events associated with reasonably foreseeable misuse. This difference does not affect substantial equivalence or safety/effectiveness. This difference does not affect the Blood Transfer Device. Conclusion: The differences between proposed device and predicated device include materials of manufacture, product length and device labeling. These differences do not raise any question regarding its safety and effectiveness. The differences in technological characteristics may be evaluated through performance testing. ## Non-clinical Performance Testing: #### 1. Performance Testing ISO 594-1-1986 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements ISO 594-2-1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -Part 2: Lock fittings EN ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods EN ISO 15223-1:2016 Medical Devices – Symbols to be used with medical device labels, labelling and information to be supplied EN ISO 14971:2012 Medical Devices – Application of risk management to medical devices ## 2. Sterilization, Package Integrity, Shipping and Shelf Life {7}------------------------------------------------ EN ISO 11135:2014 Medical Devices – Validation and Routine Control of Ethylene Oxide Sterilization Residuals EN ISO 11737-1:2019 Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products ISO 11607-1:2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems ISO 11607-2:2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) ASTM F88 Standard Test Method for Seal Strength of Flexible Barrier Materials ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems #### 3. Biocompatibility BD Vacutainer® Luer-Lok™ Access Device and BD Vacutainer® Blood Transfer Device are classified as a surface medical devices that comes in contact with intact skin for a limited (≤24 hours) duration. ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process EN ISO 10993-2:2006 Biological evaluation of medical devices - Part 2: Animal welfare requirements ISO 10993-4:2017 Biological evaluation of medical devices – Part 4: Selection of tests for interactions with blood ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity ISO 10993-7:2008/AC:2009 Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity {8}------------------------------------------------ ISO 10993-12:2012 Biological evaluation of medical devices – Part 12: Sample preparation and reference materials ISO 10993-23:2021, Biological Evaluation of Medical Devices - Part 23: Tests for irritation. #### Non-clinical Performance Summary The non-clinical performance tests below were conducted to verify that the proposed devices met all design specifications and performance standards and are each Substantially Equivalent (SE) to the predicate device. - Torque to Break - . Torque to Unseat - NP Cannula Pull Test • - Air Leakage Test - Torque to Break . - Torque to Unseat - NP Sleeve Function - Package Integrity test ● - Peel Strength Test . - Luer Compatibility as per ISO 594/80369-7. ## Clinical Data Not applicable ## Conclusion Based on the comparison and analysis above and the performance testing conducted, each of the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...