RELIANCE ADVANCE ENDOSCOPE PROCESSING SYSTEM

K123768 · STERIS Corporation · NZA · Jul 26, 2013 · Gastroenterology, Urology

Device Facts

Record IDK123768
Device NameRELIANCE ADVANCE ENDOSCOPE PROCESSING SYSTEM
ApplicantSTERIS Corporation
Product CodeNZA · Gastroenterology, Urology
Decision DateJul 26, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K123768 · Jul 26, 2013RELIANCE ADVANCE ENDOSCOPE PROCESSING SYSTEMSTERIS CorporationUS hospital clinical in-use studyThe sponsor conducted a clinical in-use study in a hospital setting to evaluate the cleaning efficacy of the device on endoscopes used in routine clinical procedures.Clinical in-use study; Cleaning efficacy; Routine clinical practice

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Clinical In-Use Study; Clinical in-use evaluation; Follow-up/Duration: Not applicableFour types of flexible endoscopes (bronchoscopes, gastroscopes, colonoscopies, and duodenoscopes) and their accessories used in clinical procedures.; Number of Sites: 1Not applicable for this studyVisual cleanliness and soil levels (protein, carbohydrates, and total organic carbon) post-processing.

Indications for Use

The Reliance™ Advance Endoscope Processing System is intended for cleaning and high level disinfection of up to two immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. During the system's standardized Endoscope Processing Cycle, cleaning is achieved within the Cleaning phase, and high level disinfection is achieved within the 50 - 57°C HLD phase (4 minute generation sequence followed by a 6-minute exposure sequence). Manual cleaning is not required prior to processing in Reliance Advance Endoscope Processing System.

Device Story

Automated endoscope reprocessor (AER) for cleaning and high-level disinfection (HLD) of up to two flexible endoscopes/accessories; utilizes proprietary enzymatic detergent and Reliance DG Dry Germicide (peracetic acid-generating oxidative chemistry). System features microprocessor-controlled cycles: cleaning phase (enzymatic detergent injection, flow-meter verified), HLD phase (50-57°C, 6-minute exposure), rinse phase (0.2-micron bacterial-retentive filter with integrity check), and HEPA-filtered air purge. Operated by healthcare staff in clinical settings; provides printed cycle summaries (optional barcode reader for tracking). Eliminates need for manual pre-cleaning. Includes self-decontamination cycles (D-SHORT/D-LONG) to prevent biofilm. Benefits include standardized processing, reduced manual labor, and validated HLD efficacy.

Clinical Evidence

Bench testing included simulated-use studies on 15 heat-sensitive endoscopes/accessories under worst-case conditions (soil levels, drying, detergent concentration). All articles visually clean; protein, carbohydrate, and TOC levels met acceptance criteria. Clinical in-use study at a US hospital evaluated four endoscope types (bronchoscopes, gastroscopes, colonoscopes, duodenoscopes) processed after clinical use; all met visual and chemical cleanliness endpoints. Microbiological efficacy (sporicidal, tuberculocidal, virucidal, bactericidal, fungicidal) confirmed >6 log10 reduction of M. terrae. Biocompatibility testing confirmed non-cytotoxic residuals.

Technological Characteristics

Electromechanical washer/disinfector; microprocessor-controlled. Uses proprietary enzymatic detergent and Reliance DG (peracetic acid). Features: peristaltic pump with flow meter, 0.2-micron bacterial-retentive water filter, HEPA air filter. Connectivity: printer output, optional barcode reader. Electrical safety: ANSI/UL 61010-1, CAN/CSA C22.2 61010-1, IEC 61010-1, IEC 61010-2-40, IEC 61326-1. Software: Major Level of Concern.

Indications for Use

Indicated for cleaning and high-level disinfection of up to two immersible, reusable, heat-sensitive, semi-critical devices, including GI flexible endoscopes, bronchoscopes, and their accessories.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the word "STERIS" in a bold, sans-serif font. There is a registered trademark symbol to the right of the word. Below the word is a graphic of several horizontal lines that are stacked on top of each other. The lines are not straight, but have a slight wave to them. JUL 2 6 2013 # 510(k) Summary For Reliance™ Advance Endoscope Processing System STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198 Contact: Marcia L. Benedict Senior Director, Regulatory Affairs Telephone: (440) 392-7063 Fax No: July 26, 2013 Summary Date: STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {1}------------------------------------------------ #### 1. Device Name | Trade Name: | Reliance Advance Endoscope Processing System | |----------------------|-------------------------------------------------------------| | Common/usual Name: | Automated Endoscope Reprocessor | | Classification Name: | 21 CFR 876.1500<br>Endoscope Cleaning Germicide Accessories | | Device Class: | II | | Product Code: | NZA | #### 2. Predicate Device Reliance Endoscope Processing System (EPS) - K102244 EVOTECH Endoscope Cleaner and Reprocessor (ECR) - K061899 #### 3. Description of Device The Reliance Advance Endoscope Processing System is an economical, easy-to-use cleaning and high level disinfection system that can clean and high level disinfect up to two immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes and related accessories. The system utilizes Reliance™ DG Dry Germicide, a proprietary, safe, dry peracetic acid generating oxidative chemistry. The Reliance Advance Endoscope Processing System was designed to be versatile in meeting the growing demands of the modern flexible endoscope processing department, while offering the highest level of patient and staff safety. The Reliance Advance Endoscope Processing System is a combination of products that are used to clean and high level disinfect flexible endoscopes and their accessories. - The Reliance Advance Endoscope Processor is an electromechanical . washer/high level disinfector with a microprocessor-based controller that provides for automated endoscope processing cycles and processor selfdecontamination cycles. - . The proprietary enzymatic detergent is dispensed automatically during the Cleaning phase of the cycle to effectively clean endoscope(s) and accessories. - . Reliance DG Dry Germicide is a proprietary, two-part, dry, single-use oxidative chemistry, designed to generate the high level disinfection solution upon automatic dilution in water within the Reliance Endoscope Processor. - . Various accessories are available to accommodate the processing needs of specific endoscopes and endoscopic accessories. - VERIFY Process Indicator for Reliance EPS is available to monitor for the . effective dose of the Reliance DG active ingredient, peracetic acid. {2}------------------------------------------------ - CIP 200 Acid-Based Process and Research Cleaner, a general cleaning agent, is . used in one of the two self-decontamination cycles provided by the processor. The Reliance Advance Endoscope Processing Cycle can perform endoscope leak testing and integrity monitoring throughout the following standardized cycle: - > The first part of this cycle is a standardized, non-optional cleaning phase that consists of a dynamic wash using proprietary enzymatic detergent, followed by a rinse. In this phase, a controlled quantity of the concentrated enzymatic detergent is dispensed into the circulating warm water volume in the processor. forming an effective cleaning solution that is directed through endoscope lumens and onto all device surfaces. The Reliance cleaning phase can be used in place of manual cleaning by the user. It does not replace the bedside pre-cleaning performed in the procedure room. - > The second part is a non-optional high level disinfection phase; its parameters cannot be changed by the user. In this phase, the proprietary Reliance DG components, provided in a single use container, are dissolved with water at ~50℃ to create the peracetic acid disinfection solution that is circulated throughout the processor and through device lumens for a 6 minute exposure time. - => Following high level disinfection, the Reliance Advance Endoscope Processor removes the germicide use-dilution through a rinse phase which is nonoptional; the parameters cannot be changed by the user. The processor filters the rinse water (as well as all of the water used throughout the cycle) through a 0.2 micron bacterial-retentive filter. There is an automatic internal integrity check of this filter at the end of each processing cycle. If the integrity check fails, an alarm alerts the user, and the processor does not complete the cvcle. - > The last step in the processing cycle is an air purge phase using HEPAfiltered air. The air purge helps to remove excess rinse water from the · processed devices. The final air purge is preset to run for 4 minutes; additional air purge time may be selected by the operator. - > At the end of each cycle the processor prints a detailed evcle summary with information such as processor number, date, start and stop times, and phase parameters. With an optional bar code reader, the printouts can also include identification numbers for the operator, patient, device, doctor and procedure. The processor features two decontamination cycles that are to be used without endoscopes in the processor: - > D-SHORT consists of hot water circulating through the processor for 10 minutes, followed by a 10-minute hot air purge. This cycle is to be run every 54 hours. D-SHORT is intended to prevent biofilm from forming. - > The second, called D-LONG, consists of a cvcle in which CIP 200 Acid-Based Process and Research Cleaner is added to hot water and circulated through the processor for 20 minutes; this is followed by three rinses to {3}------------------------------------------------ remove the solution, and a 10-minute hot air purge. D-LONG is used when the D-SHORT cycle has not been run within the past 54 hours. #### 4. Intended Use The Reliance Advance Endoscope Processing System is intended for cleaning and high level disinfection of up to two immersible, reusable, heat-sensitive, semicritical devices such as GI flexible endoscopes, bronchoscopes and their accessories. During the system's Endoscope Processing Cycle, cleaning is achieved within the Cleaning Phase, and high level disinfection is achieved within the 50 - 57°C HLD Phase (4 minute generation sequence followed by a 6 minute exposure sequence). Manual cleaning is not required prior to processing in Reliance Advance Endoscope Processing System. #### Technological Characteristics న. The Reliance Advance Endoscope Processing System is not significantly different from the predicate Reliance EPS in technology. Technological modifications of the Reliance EPS design for the Reliance Advance Endoscope Processing System are summarized below: | Property<br>▼ | Proposed Reliance Advance<br>Endoscope Processor | K102244 STERIS<br>Reliance Endoscope Processor | |-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Endoscope<br>Processing<br>Cycle | The standardized Endoscope Processing<br>Cycle (with one or two endoscopes)<br>includes a Cleaning phase and its rinse,<br>which is followed by the HLD phase,<br>final rinses, and air purge. | In the Endoscope Processing Cycle (with<br>one or two endoscopes) the user has the<br>option to select zero, one or two Wash<br>phases of 5 - 10 minutes each including<br>a rinse, which is followed by the HLD<br>phase, final rinses, and air purge. | | Enzymatic<br>Detergent<br>and Dose<br>Control | Peristaltic pump injects a metered<br>volume of the proprietary new enzymatic<br>detergent during the Cleaning phase.<br>A flow meter provides feedback to the<br>microprocessor to verify that the<br>required volume of enzymatic detergent<br>was injected by the peristaltic pump. | Peristaltic pump injects a metered<br>volume of Klenzyme during Wash<br>phases. | {4}------------------------------------------------ The Reliance Advance Endoscope Processing System is technologically similar to the EvoTech Endoscope Cleaner and Reprocessor (ECR) predicate, as shown in the comparison table below. | Property ▼ | Proposed Reliance Advance<br>Endoscope Processor | K061899<br>EvoTech Endoscope Processor | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Control | Microprocessor-controlled<br>programmable processor monitors and<br>controls critical cycle parameters and<br>provides a printed record upon cycle<br>completion. | Microprocessor-controlled<br>programmable processor monitors and<br>controls critical cycle parameters and<br>provides a printed record upon cycle<br>completion. | | Capacity | One or two endoscopes and accessories | One or two endoscopes and accessories | | Mechanism<br>for channel<br>access | Uses a pressurized control handle boot<br>such that operation is largely connector-<br>less. Flow Units are used for specialized<br>device processing needs. | Relies upon multiple tubings and<br>connectors to provide flow to inner<br>channels of endoscopes. | | Enzymatic<br>Detergent | Requires use of a proprietary enzymatic<br>detergent for cleaning. | Requires use of a proprietary enzymatic<br>detergent for cleaning. | | Germicide | Requires Reliance DG Dry Germicide. | Requires Cidex OPA. | | Monitoring | Requires use of a VERIFY Process<br>Indicator for Reliance EPS strip with<br>each Endoscope Processing Cycle. | Has an on-board concentration monitor. | | Rinsing | Devices are rinsed with filtered water. | Devices are rinsed with filtered water. | | Leak Test | Automated leak test is selectable. | Automated test assures absence of major<br>leaks in the endoscope body. | | Alcohol flush | Not provided | Selectable in this system | | Self-<br>Disinfection | Has capacity to perform self-<br>disinfection. | Has capacity to perform self-<br>disinfection. | #### 6. Description of Safety and Substantial Equivalence The Reliance Advance Endoscope Processing System has the same intended use for cleaning and high level disinfection of flexible endoscopes as the predicate EvoTech ECR. It is substantially equivalent to the EvoTech ECR with respect to achieving cleaning and high level disinfection of a wide range of flexible endoscopes and accessories. {5}------------------------------------------------ #### 7. Performance Testing Summary | Reliance Advance Endoscope Processing System Performance Testing Summary | | |--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Cleaning<br>Phase<br>Efficacy | • Cleaning Phase Effectiveness in Simulated Use Studies<br>Reliance Advance EPS has been tested and shown to achieve effective cleaning<br>of soiled flexible endoscopes and their accessories. Testing was performed on a<br>diverse array of OEM endoscopes incorporating the most challenging design<br>features for cleaning, and under worst case simulated use conditions with respect<br>to soil levels, drying, enzymatic detergent concentration, water condition, and<br>cleaning phase exposure time. | | | Fifteen (15) heat-sensitive flexible endoscopes, dilators, and their accessories<br>were soiled with select test soils at levels appropriate to the intended use of the<br>instrument. After the worst case Reliance Advance EPS cleaning phase in<br>simulated use, 1) all test articles were visually clean, and 2) all samples recovered<br>from previously soiled areas for all test articles met the established acceptance<br>levels defined as "clean" for protein, carbohydrates, and total organic carbon in<br>validated assays for those components. | | | • Cleaning Phase Effectiveness in Clinical In-Use Studies<br>Efficacy of the Reliance Advance EPS Cleaning Phase was evaluated in use in a<br>US hospital. Four types of flexible endoscopes (bronchoscopes, gastroscopes,<br>colonoscopies, and duodenoscopes) and their accessories as listed in Reliance<br>Advance EPS labeling were used in clinical procedures, pre-cleaned at the<br>bedside, allowed to dry for 47 – 100 minutes, and then processed through the<br>Reliance Advance Endoscope Processing Cycle's Cleaning phase only, according<br>to the instructions for use. Cleaning endpoints were evaluated. | | | In triplicate evaluations, each of the tested clinically soiled endoscopes 1) was<br>visually clean, and 2) met the soil pre-determined "clean" acceptance levels for<br>protein, carbohydrates, and total organic carbon in validated assays. | | Enzymatic<br>Detergent<br>Stability | • Enzymatic Detergent Stability:<br>The proprietary enzymatic detergent has been shown to be stable in its Bag-in-Box packaging through 12 months under representative storage conditions. This protocol of testing continues; the shelf life may be extended based on additional data meeting established acceptance criteria. | | Software<br>Validation | The Reliance Advance EPS software was developed and validated in accordance with the FDA requirements for a Major Level of Concern determination based on the recommendation in the FDA Guidance on Premarket Notification [510(k)] Submissions for Automated Endoscope Washers, Washer/Disinfectors, and Disinfectors Intended for Use in Health Care Facilities dated August 1993.<br>All acceptance criteria were met. | . {6}------------------------------------------------ T | Reliance Advance Endoscope Processing System Performance Testing Summary | | | | | |--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | Material<br>Compatibility | Enzymatic Detergent Material Compatibility | | | | | | The proprietary enzymatic detergent was evaluated for its effect on common<br>materials of endoscope and accessory construction. After 300 processing cycles,<br>no deleterious effects were observed other than minor cosmetic changes. No<br>functional changes in flexible endoscopes were observed. | | | | | | Reliance DG Material Compatibility: | | | | | | | The system was previously evaluated for its effect on intact medical devices,<br>including flexible endoscopes and/or common materials of device construction.<br>After 300 processing cycles, no deleterious effects were observed other than minor<br>cosmetic changes. No functional changes in flexible endoscopes were observed. | | | | | specifications. | The critical process parameters for the processor (water temperature and volume,<br>fresh DG container detection, boot pressure, water filter integrity testing, and<br>delivery of cleaning and high level disinfection solutions) have each been<br>evaluated in replicate under worst case conditions, and found to be within required | | | | | Each processor phase or cycle has been separately evaluated and shown to be<br>effective under worst case conditions: | | | | | Processor<br>Performance | The standard Cleaning phase using STERIS's proprietary enzymatic<br>î<br>detergent had no effect on generation of the active from Reliance DG in<br>Reliance Advance EPS during the HLD phase of the Endoscope Processing<br>Cycle. | | | | | | The HLD and Rinse phases of the Endoscope Processing Cycle have been<br>ी<br>shown to be effective. The air purge phase effectively removes rinse water<br>from processed devices. The filter integrity test system has been shown to<br>reliably detect filter failure. | | | | | | The two self-decontamination cycles were shown to be effective as follows in<br>ੀ<br>prior studies: | | | | | | D-LONG cycle - can disinfect the Reliance Endoscope Processor after a<br>. .<br>high level challenge with Pseudomonas aeruginosa followed by a 5 day<br>inactive period; | | | | | | ● | D-SHORT cycle - can kill bacteria that have potential to form biofilm. | | | | Certification<br>to electrical<br>safety<br>standards | ANSI/UL -61010-1, 200 | Electrical equipment for measurement, control and<br>laboratory use Part 1, 200 Edition | | | | | CAN/CSA C22.2 61010-1,<br>2" Ed. | Electrical equipment for measurement, control and<br>laboratory use Part 1, 2nd Edition | | | | | IEC 61010-1:2001 | Safety requirements for electrical equipment for<br>measurement, control and laboratory use, Part 1:<br>General requirements | | | | | IEC 61010-2-40, 1st Ed. | Safety requirements for electrical equipment for<br>measurement, control and laboratory use -Part 2-<br>040: Particular requirements for sterilizers and<br>washer-disinfectors used to treat medical materials | | | | | IEC 61326-1:2005, 136 Ed. | Electrical equipment for measurement, control and<br>laboratory use, EMC requirements Part 1: General<br>requirements | | | . {7}------------------------------------------------ | Reliance Advance Endoscope Processing System Performance Testing Summary | | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Reliance DG<br>Dry<br>Germicide<br>Efficacy | Reliance DG Dry Germicide Use Solution Microbial Efficacy: | | | | Potency: Reliance DG has been tested and shown to generate an effective high<br>level disinfection solution using the standard array of microbiological tests for<br>germicidal efficacy. Prior to the current submission, the following testing was<br>performed at conditions of use that were worst case with respect to germicide<br>concentration, contact time, circulation, water hardness, temperature and artificial<br>soiling. | | | | ⇒ Sporicidal: Reliance DG was proven to be sporicidal as defined by AOAC<br>Sporicidal Activity Test with exposure time of 6 minutes. Confirmatory testing<br>was completed successfully and supplemental confirmatory testing was<br>completed in the Reliance Endoscope Processor. Potency was subsequently<br>confirmed in the processor and in vitro using Reliance DG containers that<br>were aged beyond the end of its shelf life. | | | | ⇒ Tuberculocidal: Reliance DG was proven to be tuberculocidal as defined by<br>the AOAC Tuberculocidal Activity Test with an exposure time of 6 minutes.<br>Potency was subsequently confirmed using Reliance DG aged beyond the<br>end of its shelf life. | | | | ⇒ Virucidal: The Reliance Process was proven to reduce the viable population<br>of poliovirus Type 1, adenovirus Type 5, and herpes simplex virus Type 1<br>by > $4 log10$ . | | | | ⇒ Bactericidal: Reliance DG was proven to be bactericidal as defined by the<br>AOAC Bactericidal Activity Test with an exposure time of 6 minutes at worst<br>case conditions, whether performed in situ or in vitro. | | | | ⇒ Fungicidal: Reliance DG was proven to be tuberculocidal as defined by the<br>AOAC Tuberculocidal Activity Test with an exposure time of 6 minutes,<br>whether performed in situ or in vitro. | | | | Simulated-Use: Reliance DG, at its minimum recommended concentration,<br>reproducibly achieved greater than a 6 $log^{10}$ reduction of Mycobacterium terrae in<br>triplicate trials within the Reliance Endoscope Processor for each selected<br>clinically relevant flexible endoscope and its accessory. The test articles<br>represented the range of most challenging devices, accessories, and processing<br>situations. | | | | In-Use: Three flexible endoscopes representing the range of types indicated in the<br>product labeling were used in clinical procedures and processed according to<br>instructions for use. In triplicate evaluations of each endoscope, no organisms<br>were recovered after processing. Bioburden levels on the clinically used<br>endoscopes after manual cleaning and before high level disinfection were<br>determined to be as high as 105 CFU/device. | | | Reliance DG Stability: | | | | Reliance DG<br>Stability | Reliance DG has been shown to be stable for 18 months in the unopened<br>moisture-resistant package at the stated conditions for storage. Once opened, the<br>five Reliance DG containers within each package are to be used within two (2)<br>weeks, or by the expiration date on the container, whichever comes first. | | | Reliance Advance Endoscope Processing System Performance Testing Summary | | | | Bio-<br>compatibility | Biocompatibility:<br>The Reliance chemical formulations, as supplied in packaging as well as in use<br>dilutions, can be safely handled and used by customers. Residues that may<br>remain on medical endoscopes and accessories are below established residue<br>limits and do not pose a risk to patients. Safety statements in product labeling are<br>appropriate to the potential risk. | | | | ⇒ Enzymatic detergent components, reaction products, and residuals remaining<br>on medical devices were evaluated for biocompatibility and possible risks to<br>users including literature reviews of raw material toxicity data. Testing<br>included <i>in vitro</i> cytotoxicity evaluations on extracts of processed devices and<br>rinse water, and analysis for chemical residuals. | | | | ⇒ Reliance DG, its components, reaction products, and residuals remaining on<br>medical devices have previously been evaluated for biocompatibility and<br>possible risks to users. Testing included acute oral and ocular toxicity tests,<br>dermal irritation studies, <i>in vitro</i> bacterial mutation genotoxicity studies,<br>sensitization tests, and <i>in vitro</i> cytotoxicity evaluations; literature reviews of<br>raw material toxicity data were also performed. Certain components in the<br>single-use DG container, which under normal use conditions never contact<br>the user, have the potential for irritation or skin sensitization; therefore<br>appropriate warnings and instructions are displayed on labeling for the<br>unusual event of a spill or container breakage. | | | | ⇒ Use dilution reaches non-cytotoxic levels with minimal dilution. | | | | ⇒ Biocompatibility testing of extracts from processed medical devices<br>demonstrated that no toxic residuals remain on devices under worst case<br>circumstances. The test data indicate that the worst case residue levels for<br>the components of maximum potential risk are far below the allowable limits,<br>and the processor final rinse water was found to be non-cytotoxic. | | | | Performance<br>of VERIFY CI<br>for Reliance<br>EPS | | | NOT colorless (incomplete change) at end of Reliance Advance Endoscope<br>Processing Cycle using the minimum effective PAA dose of 9000 mg/L PAA<br>min, or in a processing cycle without PAA | | | {8}------------------------------------------------ ### Conclusion: The completed nonclinical performance testing has demonstrated that the device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate devices. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with three intertwined strands and a single wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 26, 2013 STERIS Corporation Ms. Marcia L. Benedict Senior Director, Regulatory Affairs 5960 Heislev Road Mentor OH 44060 Re: K123768 Trade/Device Name: Reliance™ Advance Endoscope Reprocessing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NZA Dated: June 27, 2013 Received: June 28, 2013 Dear Ms. Benedict: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {10}------------------------------------------------ ### Page 2 - Ms. Benedict Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Image /page/10/Picture/8 description: The image shows the name "Mary S. Runner - S" in a bold, sans-serif font. The letters are black, and the background is white. The letters "S.R" are partially obscured by a faded, patterned design. The name appears to be a signature or a title. Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {11}------------------------------------------------ ## Indications for Use 510(k) Number (if known): | 23 7 レタ Device Name: Reliance™ Advance Endoscope Processing System Indications For Use: The Reliance™ Advance Endoscope Processing System is intended for cleaning and high level disinfection of up to two immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. During the system's standardized Endoscope Processing Cycle, cleaning is achieved within the Cleaning phase, and high level disinfection is achieved within the 50 - 57°C HLD phase (4 minute generation sequence followed by a 6-minute exposure sequence). Manual cleaning is not required prior to processing in Reliance Advance Endoscope Processing System. Prescription Use (Part 2) CFR 801 Subpart D) AND/OR Over-The-Counter Use x (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) : r Concurrence of CDRH, Office of Device Evaluation (ODE) Sreekanth Gutala -S 2013.07.25 14:42:31 -04'00' (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number. K183768 Page 1 of 1 December 6, 2012
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...