Glooko Mobile Insulin Dosing System

K171450 · Glooko, Inc. · NDC · Feb 2, 2018 · Anesthesiology

Device Facts

Record IDK171450
Device NameGlooko Mobile Insulin Dosing System
ApplicantGlooko, Inc.
Product CodeNDC · Anesthesiology
Decision DateFeb 2, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1890
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The Glooko Mobile Insulin Dosing System (MIDS) is indicated for the management of type 2 diabetes by calculating appropriate long-acting basal insulin doses for titrating insulin levels based on configuration by a physician or healthcare provider knowledgeable in the care and management of diabetes. The physician or healthcare provider must activate the MIDS dose calculator and configure the patient-specific parameters. The system is not intended to provide treatment decisions or to be used as a substitute for professional healthcare advice.

Device Story

Glooko MIDS is a software module for the Glooko mobile application and web platform; supports titration of long-acting basal insulin for type 2 diabetes patients. Input data is sourced exclusively from blood glucose (BG) meters via cable or Bluetooth; no manual BG entry. HCPs configure patient-specific titration parameters via the web interface; the mobile app provides dosing directions to the patient based on this pre-planned treatment program. System functions as a decision-support tool for insulin management; does not replace professional clinical judgment. Used in home or clinical settings; operated by patients and HCPs. Benefits include structured, physician-guided insulin titration to improve diabetes management.

Clinical Evidence

Bench testing only. No clinical data provided. Validation included human factors testing per FDA guidance (February 3, 2016) and functional software verification testing per FDA guidance (May 11, 2005). All results met pre-determined specifications and acceptance criteria.

Technological Characteristics

Software-only device; operates on iOS/Android mobile platforms and web browsers. Connectivity via Bluetooth or cable to BG meters. Major level of concern software. No hardware components.

Indications for Use

Indicated for management of type 2 diabetes in patients requiring long-acting basal insulin titration. Requires physician/HCP configuration of patient-specific parameters. Not for use as a substitute for professional medical advice.

Regulatory Classification

Identification

A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 2, 2018 Glooko, Inc. Tejasvi Pasi Senior Regulatory Manager 899 West Evelyn Avenue Mountain View, California 94041 Re: K171450 Trade/Device Name: Glooko Mobile Insulin Dosing System (MIDS) Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: Class II Product Code: NDC Dated: December 29, 2017 Received: January 2, 2018 Dear Tejasvi Pasi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Tina Kiang -S Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171450 Device Name Glooko Mobile Insulin Dosing System (MIDS) ### Indications for Use (Describe) The Glooko Mobile Insulin Dosing System (MIDS) is indicated for the management of type 2 diabetes by calculating appropriate long-acting basal insulin doses for titrating insulin levels based on configuration by a physician or healthcare provider knowledgeable in the care and management of diabetes. The physician or healthcare provider must activate the MIDS dose calculator and configure the patient-specific parameters. The system is not intended to provide treatment decisions or to be used as a substitute for professional healthcare advice. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the word "glooko" in a stylized, cursive font. The word is rendered in a bright blue color, which stands out against the white background. The font has rounded edges and a playful feel, giving the word a friendly and approachable appearance. # 510(K) Summary for K171450 - 5.1 Submitter Information Name: Address: | Phone: | | |------------------|--| | Fax: | | | Contact Name: | | | Date of Summary: | | 5.2 Subject Device > Trade Name: Common Name: Classification Name: Regulation Number: Regulation Name: Requlatory Class: Product code: Classification Panel: #### 5.3 Predicate Device - Trade Name: (K161433) Common Name: Classification Name: Regulation Number: Regulation Name: Requlatory Class: Product code: Classification Panel: #### 5.4 Reference Device Trade Name: (K132272) Common Name: Glooko Inc. 303 Bryant Street, Mountain View, CA 94041, USA (650) 720-5310 (650)-720-5310 Tejasvi Pasi 02/01/2018 Glooko Mobile Insulin Dosing System (MIDS) Diabetes Management Software Calculator, Druq Dose 21 CFR 868.1890 Predictive pulmonary - function value calculator Class II NDC General Hospital Insulia Diabetes Management Companion Diabetes Management Software Calculator, Drug Dose 21 CFR 868.1890 Predictive pulmonary - function value calculator Class II NDC General Hospital Glooko Device System for Glooko Application Blood Glucose Meter and Data Management System {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "glooko" in a stylized, cursive font. The word is rendered in a bright blue color. The letters are connected, giving the word a flowing appearance. Glucose Test System, Calculator Data Classification Name: Processing Module for Clinical Use Requlation Number: 21 CFR 862.1345 Regulatory Class: Product code: Classification Panel: 21 CFR 862.2100 Class II, Class I NBW, JQP Clinical Chemistry #### 5.5 Device Description The Glooko MIDS system will be an optional new module to support the titration of long acting basal insulin doses. Health care providers (HCPs) will be able to opt-in to this new MIDS module and use it with a subset of their patients. Although the Glooko App may interact with Blood Glucose (BG) meters, insulin pumps and Continuous Glucose Meters (CGM), the MIDS interface will get data ONLY from the BG meters Glooko MIDS consists of the following two interfaces: - . MIDS Provider Interface on the Glooko Web Application for use by HCP's to prescribe long acting insulin doses for their patients - MIDS Patient interface on the Glooko mobile application for use by . patients on compatible iOS and Android phones Glooko MIDS provides directions to the patient based on a pre-planned treatment program as suggested by their HCP for titrating long acting insulin doses. Glooko MIDS is for titrating long acting insulin doses only. #### 5.6 Indications for Use The Glooko Mobile Insulin Dosing System (MIDS) is indicated for the management of type 2 diabetes by calculating appropriate long-acting basal insulin doses for titrating insulin levels based on configuration by a physician or healthcare provider knowledgeable in the care and management of diabetes. The physician or healthcare provider must activate the MIDS dose calculator and configure the patient-specific parameters. The system is not {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the word "glooko" in a stylized, cursive font. The word is rendered in a bright blue color, which stands out against the white background. The font has rounded edges and a playful feel, making it appear friendly and approachable. The overall design is simple and clean, with the focus entirely on the brand name. intended to provide treatment decisions or to be used as a substitute for professional healthcare advice. #### 5.7 Comparison to Predicate Device | Feature | Glooko Mobile Insulin Dosing<br>System<br>(Subject device) | Insulia Diabetes Management Companion<br>(Predicate device) | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Code | NDC | NDC | | Class | II | II | | Regulation<br>Name/Regulation | Predictive pulmonary-function value<br>calculator/21 CFR 868.1890 | Predictive pulmonary-function value<br>calculator/21 CFR 868.1890 | | K Number | K171450 | K161433 | | Indications for Use* | The Glooko Mobile Insulin Dosing<br>System (MIDS) is indicated for the<br>management of type 2 diabetes by<br>calculating appropriate long-acting<br>basal insulin doses for titrating insulin<br>levels based on configuration by a<br>physician or healthcare provider<br>knowledgeable in the care and<br>management of diabetes. The physician<br>or healthcare provider must activate the<br>MIDS dose calculator and configure the<br>patient-specific parameters. The system<br>is not intended to provide treatment<br>decisions or to be used as a substitute<br>for professional healthcare advice. | Insulia Diabetes Management Companion is<br>indicated for use by healthcare professionals<br>(HCPs) and their type 2 adult diabetes<br>patients treated with long-acting insulin<br>analog. Insulia Diabetes Management<br>Companion is intended to provide secure<br>capture, storage and transmission of<br>diabetes-related healthcare information, to<br>enhance data management, to display<br>reports and graphs, and to aid the HCP and<br>the patient in the review, analysis, and<br>evaluation of patient data in order to support<br>effective diabetes management. Insulia<br>Diabetes Management Companion includes a<br>basal calculator intended to provide direction<br>to the patient in response to blood glucose<br>and health events, within the scope of a pre-<br>planned treatment program from a healthcare<br>professional for insulin adjustments, similar to<br>the directions provided to patients as a part of<br>routine clinical practice. Insulia Diabetes<br>Management Companion includes software<br>intended for use on commercially available | | Feature | Glooko Mobile Insulin Dosing System<br>(Subject device) | Insulia Diabetes Management Companion<br>(Predicate device) | | | | mobile platforms, personal computers, in the<br>home or in professional healthcare settings,<br>and uses generally available networks and<br>communication protocols. Insulia Diabetes<br>Management Companion is not intended to<br>replace the care provided by a licensed<br>healthcare professional, including<br>prescriptions, diagnosis, or treatment. | | Components | Patient mobile based application,<br>patient and HCP web based application | Software only, patient mobile based<br>application, patient and HCP web based<br>application | | Treatment<br>Guidance | Adjustments to insulin doses within the<br>scope of a pre-planned, physician<br>specified treatment program similar to<br>routine clinical practice | Adjustments to insulin doses within the<br>scope of a pre-planned, physician specified<br>treatment program similar to routine clinical<br>practice | | Type of Calculated<br>Insulin | Basal Insulin | Basal Insulin | | Manual BG Entry | No | Yes | | Logbook | Yes | Yes | | Reports & Statistics | Yes | Yes | | Coaching Messages | No | Yes | | Secure Database | Yes | Yes | | Data Transfer | Internet | Internet | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "glooko" in a stylized, sans-serif font. The word is rendered in a bright blue color. The letters are connected in a cursive-like manner, with rounded edges and a playful appearance. *Glooko MIDS has the same intended use as the predicate device. Both software applications are intended for use by healthcare professionals and their patients with type 2 diabetes for calculating appropriate long-acting insulin doses. Glooko MIDS and the predicate device provide HCP-configured directions within the scope of a pre-planned treatment program for titrating long-acting insulin as a part of routine clinical practice. The proposed indications for use for the subject device are similar to and consistent with that of the predicate device, and do not raise different questions of safety or effectiveness. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "glooko" in a stylized, sans-serif font. The word is rendered in a bright blue color. The letters are connected in a cursive-like manner, giving the word a flowing appearance. #### 5.8 Comparison to Reference Device The reference device supports the BG entry method for the subject device. Both the subject and the reference device do not support manual entry of BG data. In both the Glooko MIDS and the reference device BG data is transferred from supported BG meters to the Glooko Application via cable(s) or Bluetooth enabled BG meters. | Feature | Glooko Mobile Insulin Dosing System<br>(Subject device) | Glooko Device System for Glooko<br>Application<br>(Reference device) | |----------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | K Number | K171450 | K132272 | | BG Entry | BG data is transferred from the BG meter to<br>the Glooko Application via cable(s) or<br>Bluetooth enabled BG meters | BG data is transferred from the BG meter to<br>the Glooko Application via cable(s) or<br>Bluetooth enabled BG meters | #### 5.9 Performance Data Demonstrating Substantial Equivalence The Glooko MIDS underwent Human Factors validation testing and software performance testing to support a determination of substantial equivalence. The Glooko MIDS software was validated pursuant to the Major Level of Concern requirements. Design validation testing and human factors study results confirmed that the Glooko MIDS software performs according to the stated intended use. The Human Factors validation was documented according to FDA Guidance - Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016). Software evaluation consisted of functional testing performed pursuant to Glooko's design verification protocol. All of the software tests were documented according to FDA's guidance document - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005). All test results fell within the pre-determined specification parameters and acceptance criteria. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "glooko" in a stylized, sans-serif font. The word is rendered in a bright blue color. The letters are connected in a cursive-like manner, giving the word a flowing appearance. #### 5.10 Statement of Equivalence The Glooko MIDS is substantially equivalent to the predicate device with regards to its intended use and function. #### 5.11 Conclusion Glooko MIDS has similar indications for use and similar technological characteristics as those of the predicate device Insulia Diabetes Management Companion. Performance testing of the Glooko MIDS software has demonstrated that Glooko MIDS performs as intended and is substantially equivalent to the predicate device. Based on the product technical information, intended use, human factors and validation data provided in this pre-market notification, the Glooko MIDS software has been shown to be substantially equivalent to the currently marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...