← Product Code [NDC](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/NDC) · K071713

# CLARIAN GLUCOSE STABILIZER INSULIN DOSING CALCULATOR (K071713)

_Medical Decisions Network · NDC · Oct 12, 2007 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/NDC/K071713

## Device Facts

- **Applicant:** Medical Decisions Network
- **Product Code:** [NDC](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/NDC.md)
- **Decision Date:** Oct 12, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1890
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Software as a Medical Device, Therapeutic

## Intended Use

The MDN-CGS™ Insulin Dosing Calculator is intended to evaluate the current patient blood glucose values. Based on the measurements, the software will calculate either a dose of glucose or insulin to drive the blood glucose level up or down toward a predetermined target range. Once that target blood glucose range has been reached, the software program calculates dosing of insulin or glucose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of glucose or insulin. The device is intended for use with patients with no known insulin allergies and for patients over the age of 18. The MDN-CGSTM Insulin Dosing Calculator's programmed logic is a tool and not a substitute for, but rather an assist to, clinical reasoning. The measurements and calculations generated by the MDN-CGSTM Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel directed by physician order. No medical decisions are to be made solely on the guidance provided by this software program, but are to include consideration of clinical history, symptoms, other diagnostic measurements, and the professional's clinical judgment.

## Device Story

Web-based software tool for clinical decision support; automates intravenous insulin and glucose dosing calculations. Inputs: current patient blood glucose values. Processing: programmed logic calculates required insulin/glucose doses to reach or maintain target blood glucose ranges. Outputs: dosing recommendations and alerts for blood glucose monitoring. Used in clinical settings by trained medical personnel. Acts as an assist to clinical reasoning; does not replace professional judgment. Benefits: facilitates glucose management in patients requiring intravenous therapy.

## Technological Characteristics

Web-based software application; standalone architecture. Provides automated intravenous insulin/glucose dosing calculations and monitoring alerts. Operates as a clinical decision support tool.

## Regulatory Identification

A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.

## Predicate Devices

- EndoTool™ Drug Dose Calculator ([K053137](/device/K053137.md))
- Glucommander Plus™ ([K061110](/device/K061110.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for MDN, which stands for Medical Device Network. Next to the logo is the text "510(k) Summary". The text is in a simple, sans-serif font and is aligned to the right of the logo. The image is likely a document header or title page for a medical device summary.

K071713
Medical Decisions Network

## 5. 510(k) Summary

| Date:                  | June 21, 2007<br>OCT 12 2007                                                                                                                                                                                                                                                                                                                                                                                                                   |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:             | Medical Decisions Network, Inc.<br>2000 Holiday Drive, Suite 200<br>Charlottesville, VA 22901                                                                                                                                                                                                                                                                                                                                                  |
| Contact:               | Medical Decisions Network, Inc.<br>c/o Mr. Jonathan S. Kahan<br>Hogan & Hartson LLP<br>555 Thirteenth Street, NW<br>Washington, DC 20004                                                                                                                                                                                                                                                                                                       |
|                        | Phone: (202) 637-5600<br>Fax: (202) 637-5910<br>Email: jskahan@hhlaw.com                                                                                                                                                                                                                                                                                                                                                                       |
| Trade Name:            | Clarian Glucose Stabilizer Insulin<br>Dosing Calculator                                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name:           | MDN-CGSTM Insulin Dosing Calculator                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification Name:   | Drug Dose Calculator                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification Number: | Class II (per 21 CFR 868.1890)                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code:          | NDC                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicate Device:      | EndoTool <sup>TM</sup> Drug Dose Calculator<br>(K053137)<br>Glucommander Plus <sup>TM</sup> (K061110)                                                                                                                                                                                                                                                                                                                                          |
| Description of Device: | The MDN-CGS <sup>TM</sup> Insulin Dosing<br>Calculator is a web-based software<br>solution that automates calculations used<br>by healthcare professionals to determine<br>the appropriate intravenous insulin drip<br>rate necessary to manage blood glucose<br>levels across a variety of patient<br>populations. The MDN-CGSTM Insulin<br>Dosing Calculator also provides alerts<br>for subsequent blood glucose testing and<br>monitoring. |
| Intended Use:          | The MDN-CGS <sup>TM</sup> Insulin Dosing<br>Calculator is intended to evaluate the                                                                                                                                                                                                                                                                                                                                                             |

September 20, 2007

-- - ---

-----------

Page of this document of MDN CONFIDENTIAL AND PROPRIETARY
The information in this document is restricted for use by MDN personnel and its Partners and Customers under non-disclosure.

{1}------------------------------------------------

current patient blood glucose values. Based on the measurements, the software will calculate either a dose of glucose or insulin to drive the blood glucose level up or down toward a predetermined target range. Once that target blood glucose range has been reached, the software program calculates dosing of insulin or glucose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of glucose or insulin. The device is intended for use with patients with no known insulin allergies and for patients over the age of 18.

The MDN-CGSTM Insulin Dosing Calculator's programmed logic is a tool and not a substitute for, but rather an assist to, clinical reasoning. The measurements and calculations generated by the MDN-CGSTM Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel directed by physician order. No medical decisions are to be made solely on the guidance provided by this software program, but are to include consideration of clinical history, symptoms, other diagnostic measurements, and the professional's clinical judgment.

The submission device and the predicate devices have the same general intended use and similar indications, technological characteristics, and principles of operation. Each of the devices is intended to be used by trained clinicians. The only technological difference between the MDN-CGSTM Insulin Dosing Calculator and its predicates is that the MDN-CGSTM

Image /page/1/Picture/5 description: The image shows the text "September 20, 2007". The text appears to be a date, indicating the 20th day of September in the year 2007. The text is black and the background is white. The text is slightly blurry.

Image /page/1/Picture/6 description: The image shows a long, narrow object against a white background. The object appears to be a thin rod or stick, possibly made of metal or wood. There is a small, round object attached to the rod near the bottom. The rod is slightly curved and has a rough texture.

Comparison with Predicate Device:

ﺮ ﺍﻟﻤﺮﺍﺟﻊ

Image /page/1/Picture/8 description: The image shows a black and white graphic. The graphic appears to be a partial view of a head or object, with a large, dark, solid area dominating the upper portion. The lower part of the image features a curved line, possibly indicating a chin or the base of the object. The overall impression is that of a silhouette or a close-up of a dark, rounded form against a light background.

MDN CONFIDENTIAL AND PROPRIETARY The information in this document is restricted for use by MDN personnel and its Partners and Customers under non-disclosure.

Page 2 of 3

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a long, narrow object that appears to be a metal rod or pole. The object is dark in color and has a smooth, reflective surface. There are some small, round objects attached to the rod at various points. The background is white, which helps to highlight the shape and details of the object.

Conclusion:

Insulin Dosing Calculator is a standalone software device, while the predicate software programs are accessories to automated insulin pumps.

The MDN-CGS™ Insulin Dosing Calculator and the Endotool and Glucommander Plus have the same intended use and similar indications, technological characteristics and principles of operation. The minor technological differences between the MDN-CGS™ and its predicates do not present any new issues of safety or effectiveness. Thus, the MDN-CGSTM Insulin Dosing Calculator is substantially equivalent to the EndoTool™ Drug Dose Calculator (K053137) and the Glucommander Plus™ (K061110).

![Image of a black and white graphic with a dark, irregular shape at the top and a lighter, partially visible shape at the bottom.](image.png)

September 20, 2007

Page 3 of 3 MDN CONFIDENTIAL AND PROPRIETARY The information in this document is restricted for use by MDN personnel and its Partners and Customers under non-disclosure.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the edge. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

007 1 % 2007

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medical Decisions Network, Incorporated C/O Mr. Jonathan S. Kahan Partner Hogan & Hartson LLP 555 Thirteenth Street, North West Washington, DC 20004

Re: K071713

Trade/Device Name: Clarian Glucose Stabilizer Insulin Dosing Calculator Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: II Product Code: NDC Dated: September 21, 2007 Received: September 21, 2007

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

## Page 2 - Mr. Kahan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Clues

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## 4. Indications for Use

510(k) Number (if known): K071713

Device Name: Clarian Glucose Stabilizer Insulin Dosing Calculator

Indications for Use:

The MDN-CGS™ Insulin Dosing Calculator is intended to evaluate the current patient blood glucose values. Based on the measurements, the software will calculate either a dose of glucose or insulin to drive the blood glucose level up or down toward a predetermined target range. Once that target blood glucose range has been reached, the software program calculates dosing of insulin or glucose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of glucose or insulin. The device is intended for use with patients with no known insulin allergies and for patients over the age of 18.

The MDN-CGS™ Insulin Dosing Calculator programmed logic is a tool and not a substitute for, but rather an assist to, clinical reasoning. The measurements and calculations generated by the MDN-CGS™ Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel directed by physician order. No medical decisions are to be made solely on the guidance provided by this software program, but are to include consideration of clinical history, symptoms, other diagnostic measurements, and the professional's clinical judgment.

| Prescription Use<br>(Part 21 CFR 801 Subpart D) |        |
|-------------------------------------------------|--------|
|                                                 | AND/OR |

Over-The-Counter Use _ (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER

| (Division Sigh-Off)                                    | PAGE OF NEEDED)    |
|--------------------------------------------------------|--------------------|
| Division of Anesthesiology, General Hospital           |                    |
| Infection Control, Dental Devices                      |                    |
| Concurrence of CDRH, Office of Device Evaluation (ODE) |                    |
| 510(k) Number:                                         | k071713            |
| Page 1 of 1                                            | September 20, 2007 |

MDN CONFIDENTIAL AND PROPRIETARY The information in this document is restricted for use by MDN personnel and its Partners and Customers under non-disclosure.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/NDC/K071713](https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/NDC/K071713)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
