← Product Code [FEB](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEB) · K201920

# Endoscope Reprocessor OER-Elite (K201920)

_Olympus Medical Systems Corp. · FEB · Sep 1, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/FEB/K201920

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [FEB](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEB.md)
- **Decision Date:** Sep 1, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Intended Use

The OER-Elite is intended for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories. Safe use requires detergent and an FDA-cleared high-level disinfectant/sterilant that Olympus has validated to be efficacious and compatible with the materials of the OER-Elite and Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories. Use of a detergent or high-level disinfectant/sterilant that has not been validated by Olympus may be ineffective and can damage the OER-Elite components and the endoscopes being reprocessed. Endoscopes must be cleaned by the user prior to reprocessing; however, use of the OER-Elite enables the user to perform modified manual cleaning of some endoscopes prior to automated cleaning and high-level disinfection in the OER-Elite.

## Device Story

Automated endoscope reprocessor (AER) for high-level disinfection of heat-sensitive Olympus flexible endoscopes; utilizes immersion method for cleaning/disinfection; performs leak testing, cleaning, disinfection, rinsing, and alcohol/air drying. Features ultrasonic cleaning for debris removal; automated dispensing of detergent and disinfectant (e.g., Acecide-C); 0.2-micron bacteria-retentive filters for water/air. Includes RFID for scope/user identification; touch screen interface; built-in printer for results. Operated by clinical staff in healthcare facilities. Modifications include extending 'extra disinfection process' to ultrasound endoscopes with balloon channels, optimizing error code thresholds, replacing cooling fan, and changing fluid pathway materials. Output includes reprocessed endoscopes and printed/digital reprocessing logs. Benefits include standardized, automated cleaning/disinfection to reduce manual labor and ensure consistent reprocessing outcomes.

## Clinical Evidence

Bench testing only. Testing included process parameter verification per FDA guidance, software verification and validation, electrical safety review (IEC 61010-1:2010/IEC 61010-2-040:2015), and durability testing of new fluid pathway materials exposed to reprocessing chemicals.

## Technological Characteristics

One-basin automated reprocessor; ultrasonic cleaning; fluid flushing/immersion; RFID-enabled; 0.2-micron air/water filtration. Materials: modified fluid pathway components. Energy: electrical. Connectivity: touch screen, printer, portable memory export. Software: embedded control system with GUI.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- ENDOSCOPE REPROCESSOR OER-Elite ([K190969](/device/K190969.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 1, 2020

Olympus Medical Systems Corp. % Lisa Boyle Regulatory Affairs Specialist II Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, Pennsylvania 18034-0610

Re: K201920

Trade/Device Name: Endoscope Reprocessor OER-Elite Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FEB Dated: July 6, 2020 Received: July 10, 2020

Dear Lisa Boyle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

### K201920

Device Name ENDOSCOPE REPROCESSOR OER-Elite

The OER-Elite is intended for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories. Safe use requires detergent and an FDA-cleared high-level disinfectant/sterilant that Olympus has validated to be efficacious and compatible with the materials of the OER-Elite and Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories. Use of a detergent or high-level disinfectant/sterilant that has not been validated by Olympus may be ineffective and can damage the OER-Elite components and the endoscopes being reprocessed. Endoscopes must be cleaned by the user prior to reprocessing; however, use of the OER-Elite enables the user to perform modified manual cleaning of some endoscopes prior to automated cleaning and high-level disinfection in the OER-Elite.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The font is sans-serif and the letters are closely spaced. The word is centered and takes up most of the frame. The background is white.

Date Prepared: September 1, 2020

# K201920 510(k) Summary

# 1. GENERAL INFORMATION

| ■ 510(k) Submitter: | OLYMPUS MEDICAL SYSTEMS CORP.                          |  |  |
|---------------------|--------------------------------------------------------|--|--|
|                     | 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 |  |  |

- Contact Person: Lisa Boyle Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: lisa.boyle@olympus.com
- Manufacturing site: Aizu Olympus Co., Ltd., 500 Muranishi, Niidera, Monden-machi, Aizuwakamatsu-shi, Fukushima 965-8520, Japan

# 2. DEVICE IDENTIFICATION

- ENDOSCOPE REPROCESSOR OER-Elite ■ Device Name:
- Model Name: OER-Elite
- Common Name: Endoscope washer/disinfector
- Regulation Number: 876.1500
- Regulation Name: Endoscope and accessories
- Regulatory Class: II
- Product Code: FEB - Accessories, Cleaning, For Endoscope
- Gastroenterology/Urology ■ Classification Panel:

Olympus Corporation of the Americ 3500 CORPORATE PARKWAY, P.O. BOX 610, CENTER VALLEY, PA 18034-0610 TELEPHONE (484) 896-5000

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. A thin, horizontal, yellow line is located beneath the word. 

# 3. PREDICATE DEVICE

### ■ Predicate device

| Device name                        | 510(k) Submitter                 | 510(k) No. |
|------------------------------------|----------------------------------|------------|
| ENDOSCOPE REPROCESSOR<br>OER-Elite | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K190969    |

### 4. DEVICE DESCRIPTION

### ■ General Description of the subject device

The OER-Elite Endoscope Reprocessor is an automated endoscope reprocessor intended for high-level disinfection of Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories, utilizing both a detergent and FDA cleared high-level disinfectant validated by Olympus to be efficacious and compatible with the materials of the OER-Pro and Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories.

The OER-Elite originally featured an "extra disinfection process" which can provide an extra three-minute process for delivering fluid to both the forceps elevator area and elevator wire channel when reprocessing endoscopes with a forceps elevator.

The software of the OER-Elite was changed to apply the "extra disinfection process" to the ultrasound endoscope with balloon channel in addition to the endoscopes with a forceps elevator.

The proposed OER-Elite differs from the predicate OER-Elite in the following minor modifications:

Add "Extra disinfection process" to ultrasound endoscope with balloon a) channel.

- Optimize the threshold values of existing error codes. b)
- Replace the electronic cooling fan due to discontinuing of the product. c)
- Change the material used for fluid pathways. d)

### ■ Principle of Operation

The principle of operation has not been changed from that of the predicate OER-Elite.

The OER-Elite is a one-basin automatic endoscope reprocessor that performs leak testing, cleaning, disinfection, rinsing, and an alcohol flush to render a high-level disinfected endoscope, their accessories, and endoscope reprocessor accessories.

Olympus Corporation of the Americas

{5}------------------------------------------------

# OLYMPUS

The OER-Elite utilizes an immersion method for cleaning, disinfecting, and rinsing of endoscope and accessory external surfaces, and connectors for endoscope channel cleaning, disinfecting, and rinsing. Two endoscopes, with several exceptions, can be reprocessed simultaneously in the basin during one reprocessing cycle. The OER-Elite's cleaning cycle includes ultrasonic cleaning. which helps remove debris and dirt from endoscope surfaces.

The OER-Elite is capable of fully automated detergent/disinfectant solution dispensing and alcohol/air drying of endoscope channels. The 0.2-micron air/water filters are bacteria retentive and produce suitable rinse water and air for reprocessing. Built-in sensors detect fluid levels, fluid temperature, air/fluid pressure, fluid flow, and the operating states of the components within the OER-Elite.

The OER-Elite is also equipped with a RFID (Radio-Frequency Identification) function. With a built-in antenna, the OER-Elite is capable of reading user and scope ID data from the proprietary ID tag/chip. The scope/user ID information and each reprocessing result can be printed out with a built-in printer, displayed on a touch screen, or exported to a portable memory.

The OER-Elite is capable of using a concentrated disinfectant (e.g., Acecide-C) sealed in dedicated cassette bottles. The concentrated disinfectant is automatically diluted by filtered water until it reaches a specified quantity in the device.

### 5. INDICATIONS FOR USE

The OER-Elite is intended for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories. Safe use requires detergent and an FDA-cleared high-level disinfectant/sterilant that Olympus has validated to be efficacious and compatible with the materials of the OER-Elite and Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories. Use of a detergent or high-level disinfectant/sterilant that has not been validated by Olympus may be ineffective and can damage the OER-Elite components and the endoscopes being reprocessed. Endoscopes must be cleaned by the user prior to reprocessing; however, use of the OER-Elite enables the user to perform modified manual cleaning of some endoscopes prior to automated cleaning and high-level disinfection in the OER-Elite.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The letters are evenly spaced and appear to be a sans-serif font. There is a thin yellow line underneath the word, adding a subtle accent to the logo.

# 6. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE

The OER-Elite has the same technological characteristics and design as the predicate device except for the subject of the "extra disinfection process" to the ultrasound endoscope with balloon channel. All other technological characteristics of both the subject and predicate device are identical.

| Features                          | Subject Device: OER-Elite                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device: OER-Elite<br>(K190969)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comment on<br>difference                                                                                              |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Intended Use                      | The OER-Elite is intended for use in<br>cleaning and high-level disinfection of<br>heat sensitive Olympus flexible<br>endoscopes, their accessories, and<br>endoscope reprocessor accessories.<br>Safe use requires detergent and an<br>FDA-cleared high-level<br>disinfectant/sterilant that Olympus has<br>validated to be efficacious and<br>compatible with the materials of the<br>OER-Elite and Olympus flexible<br>endoscopes, their accessories, and<br>endoscope reprocessor accessories.                                          | The OER-Elite is intended for use in<br>cleaning and high-level disinfection of<br>heat sensitive Olympus flexible<br>endoscopes, their accessories, and<br>endoscope reprocessor accessories.<br>Safe use requires detergent and an<br>FDA-cleared high-level<br>disinfectant/sterilant that Olympus has<br>validated to be efficacious and<br>compatible with the materials of the<br>OER-Elite and Olympus flexible<br>endoscopes, their accessories, and<br>endoscope reprocessor accessories.                                          | Same                                                                                                                  |
|                                   | Use of a detergent or high-level<br>disinfectant/sterilant that has not been<br>validated by Olympus may be<br>ineffective and can damage the OER-<br>Elite components and the endoscopes<br>being reprocessed. Endoscopes must<br>be cleaned by the user prior to<br>reprocessing; however, use of the<br>OER-Elite enables the user to perform<br>modified manual cleaning of some<br>endoscopes prior to automated<br>cleaning and high-level disinfection in<br>the OER-Elite.                                                          | Use of a detergent or high-level<br>disinfectant/sterilant that has not been<br>validated by Olympus may be<br>ineffective and can damage the OER-<br>Elite components and the endoscopes<br>being reprocessed. Endoscopes must<br>be cleaned by the user prior to<br>reprocessing; however, use of the<br>OER-Elite enables the user to perform<br>modified manual cleaning of some<br>endoscopes prior to automated<br>cleaning and high-level disinfection in<br>the OER-Elite.                                                          |                                                                                                                       |
| Disinfectant                      | Olympus validated, FDA cleared<br>High-Level Disinfectant (Acecide-C);                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Olympus validated, FDA cleared<br>High-Level Disinfectant (Acecide-C);                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                  |
| Detergent                         | Olympus validated Detergent<br>(EndoQuick)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Olympus validated Detergent<br>(EndoQuick)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                  |
| Features                          | Subject Device: OER-Elite                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device: OER-Elite<br>(K190969)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comment on<br>difference                                                                                              |
| Wash/HLD<br>Methods               | Cleaning method:<br>- Exterior surfaces Ultrasonic<br>cleaning, turbulent bath Channel<br>interiors Fluid flushing<br>- Valves Ultrasonic cleaning, fluid flushing<br><br>Disinfection method:<br>- Exterior surfaces Disinfectant<br>solution immersion Channel<br>interiors Disinfectant solution<br>flushing and filling<br>- Valves Disinfectant solution<br>immersion                                                                                                                                                                  | Cleaning method:<br>- Exterior surfaces Ultrasonic<br>cleaning, turbulent bath Channel<br>interiors Fluid flushing<br>- Valves Ultrasonic cleaning, fluid flushing<br><br>Disinfection method:<br>- Exterior surfaces Disinfectant<br>solution immersion Channel<br>interiors Disinfectant solution<br>flushing and filling<br>- Valves Disinfectant solution<br>immersion                                                                                                                                                                  | Same                                                                                                                  |
| Independent<br>sub functions      | Functions:<br>- Heat LCG<br>- Mix LCG<br>- Rinse<br>- Air Purge<br>- Water Line Disinfection<br>- Self-Disinfection & Water<br>Sampling<br>- Detergent Line Disinfection<br>- Alcohol Line Disinfection<br>- Manual Leak Test<br>- Auto Leak Test<br>- ALT Self-Check<br>- Alcohol Flush<br>- Leaking scope<br>decontamination<br>- Heat LCG Timer<br><br>Replacement of Consumable Items:<br>- Drain LCG<br>- Load LCG<br>- Replace Detergent<br>- Replace Water Filter<br>- Replace Air Filter<br>- Replace Gas Filter on the<br>lid/tank | Functions:<br>- Heat LCG<br>- Mix LCG<br>- Rinse<br>- Air Purge<br>- Water Line Disinfection<br>- Self-Disinfection & Water<br>Sampling<br>- Detergent Line Disinfection<br>- Alcohol Line Disinfection<br>- Manual Leak Test<br>- Auto Leak Test<br>- ALT Self-Check<br>- Alcohol Flush<br>- Leaking scope<br>decontamination<br>- Heat LCG Timer<br><br>Replacement of Consumable Items:<br>- Drain LCG<br>- Load LCG<br>- Replace Detergent<br>- Replace Water Filter<br>- Replace Air Filter<br>- Replace Gas Filter on the<br>lid/tank | Same                                                                                                                  |
| Leak test<br>method               | Manual leak testing or Auto leak<br>testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Manual leak testing or Auto leak<br>testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                  |
| Channel<br>monitoring<br>function | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                  |
| User Interface                    | Graphical User Interface (GUI)<br>and manual control buttons                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Graphical User Interface (GUI)<br>and manual control buttons                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                  |
| Features                          | Subject Device: OER-Elite                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device: OER-Elite<br>(K190969)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comment on<br>difference                                                                                              |
| Extra<br>Disinfection<br>Process  | Applied to endoscope models with<br>forceps<br>elevator and/or balloon channel                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Applied to endoscope models<br>with forceps elevator only.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Endoscope<br>models with<br>balloon channel<br>was added to<br>the subject of<br>the extra<br>disinfection<br>process |

The comparison to the Predicate Device was summarized in the Table shown below.

Page 6 of 6

{7}------------------------------------------------

# OLY MPUS

3500 CORPORATE PARKWAY, P.O. BOX 610, CENTER VALLEY, PA 18034-0610

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered horizontally and takes up most of the image space.

# 7. NON-CLINICAL PERFORMANCE TESTING

Performance testing of the OER-Elite was performed to evaluate the modified device and the results are summarized as follows.

### 1) Process Parameter Test

Process parameter testing was conducted on the OER-Elite in accordance with recommendations in the FDA Guidance: "FDA guidance "Guidance on Premarket [510(k)] Submissions for Notification Washers, Washer/Disinfectors, and Disinfectors Intended for Use in Health Care Facilities" to demonstrate that the machine achieves and maintains the specified physical process parameters, including detection of the defined fault conditions and execution of automatic response/processing following fault detection.

### 2) Software verification and validation testing

Software verification and validation testing for the OER-Elite was conducted and documentations were provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

### 3) Electrical Safety

A Design/Specification review of the OER-Elite cooling fan was conducted to demonstrate the new cooling fan does not deviate from the technical specifications of the previous cooling fan which was previously evaluated in the electrical safety testing in accordance with IEC 61010-1:2010 and IEC 61010-2-040:2015.

### 4) Durability

The components with new fluid pathway materials of the OER-Elite were exposed to the reprocessing chemicals. No functional degradation was observed.

### 8. CONCLUSIONS

Based on the indications for use, technological characteristics, non-clinical performance testing and technological comparison to the predicate device, the OER-Elite is as safe, as effective, and performs as well or better than the legally marketed predicate device (K190969).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/FEB/K201920](https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/FEB/K201920)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
