← Product Code [FEB](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEB) · K152189

# EVOTECH ECR (K152189)

_Advanced Sterilization Products (Asp) · FEB · Jul 20, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/FEB/K152189

## Device Facts

- **Applicant:** Advanced Sterilization Products (Asp)
- **Product Code:** [FEB](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEB.md)
- **Decision Date:** Jul 20, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Intended Use

The EVOTECH® ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high-level disinfectant CIDEX® OPA Concentrate Solution and an enzymatic detergent (CIDEZYME XTRA) to achieve cleaning and high level disinfection of heat sensitive (>60°C) semi-critical endoscopes. Manual cleaning of medical devices (endoscopes) is not required prior to placement in the EVOTECH® ECR Endoscope Cleaner and Reprocessor System when selecting those cycles that contain a wash stage. (Manual cleaning of medical devices (endoscopes) is required when selecting the Disinfect only or Disinfect/Alcohol Flush Cycle.)

## Device Story

Automated two-basin washer/disinfector for flexible endoscopes; utilizes enzymatic detergent and CIDEX® OPA Concentrate high-level disinfectant. System performs automated cleaning, disinfection, rinsing, and air-drying; includes alcohol flush option. Operates via user-selected cycles; monitors temperature (50°C) and disinfectant concentration. Onboard printer provides cycle documentation; Data Transfer Interface (DTI) enables network connectivity (DHCP) for server/tracking system integration. Safety features trigger cycle cancellation and user notification if parameters fail. Used in clinical settings by trained personnel to ensure endoscope reprocessing; eliminates manual cleaning requirement for specific cycles, reducing variability and improving patient safety by ensuring standardized disinfection.

## Clinical Evidence

Bench testing only. Simulated use and in-use studies demonstrated 6-log10 reduction of Mycobacterium terrae. Cleaning efficacy validated via protein and total organic carbon (TOC) quantification, showing non-inferiority to SGNA manual cleaning procedures. Biocompatibility testing confirmed residual OPA levels are non-toxic. Duodenoscope testing (open/closed elevator wire channels) confirmed cleaning and disinfection efficacy.

## Technological Characteristics

Automated two-basin washer/disinfector. Materials: powder-coated aluminum, glass, stainless steel, polyethylene, Neoprene, polypropylene, polyamide, Ultem, polycarbonate. Energy: electrical. Connectivity: DHCP network capability. Software: moderate level of concern. Sterilization: high-level disinfection via CIDEX® OPA Concentrate.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- EVOTECH® ECR Endoscope Cleaner and Reprocessor ([K140977](/device/K140977.md))

## Reference Devices

- CIDEX® OPA Solution ([K991487](/device/K991487.md))
- EVOTECH® ECR Endoscope Cleaner and Reprocessor ([K040883](/device/K040883.md))
- EVOTECH® ECR Endoscope Cleaner and Reprocessor ([K061889](/device/K061889.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three overlapping profiles facing to the right. The profiles are arranged in a way that creates a sense of depth and movement. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 20, 2016

Advanced Sterilization Products (ASP) Nazanin Yacobi Senior Regulatory Affairs Program Lead 33 Technology Drive Irvine, California 92618

Re: K152189

Trade/Device Name: EVOTECH® ECR Endoscope Cleaner and Reprocessor Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: FEB Dated: July 7, 2016 Received: July 11, 2016

Dear Nazanin Yacobi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Michael D. Ryan -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152189

### Device Name

EVOTECH® Endoscope Cleaner and Reprocessor

### Indications for Use (Describe)

The EVOTECH® ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high-level disinfectant CIDEX® OPA Concentrate Solution and an enzymatic detergent (CIDEZYME XTRA) to achieve cleaning and high level disinfection of heat sensitive (>60°C) semi-critical endoscopes. Manual cleaning of medical devices (endoscopes) is not required prior to placement in the EVOTECH® ECR Endoscope Cleaner and Reprocessor System when selecting those cycles that contain a wash stage. (Manual cleaning of medical devices (endoscopes) is required when selecting the Disinfect only or Disinfect/Alcohol Flush Cycle.)

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary for K152189

#### I. SUBMITTER

Advanced Sterilization Products Division of Ethicon, Inc., a Johnson and Johnson Company 33 Technology Drive Irvine, CA 92618

Contact Person: Nazanin Yacobi Senior Regulatory Affairs Program Lead Tel: (949) 453-8672

Alternate Contact: Laurie Cartwright Associate Director Regulatory Affairs Tel: (949) 789-8545

### Summary Date

July 12, 2016

#### CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME II.

Trade Name: EVOTECH® ECR Endoscope Cleaner and Reprocessor Common/Usual Name: Endoscope Cleaner and Reprocessor Classification Name: Endoscope and Accessories (21 CFR 876.1500) Product Code FEB Product Classification: Class II 510(k) Number: K152189

#### III. PREDICATE DEVICE

EVOTECH® ECR Endoscope Cleaner and Reprocessor (K140977)

#### IV. DESCRIPTION OF DEVICE

EVOTECH® ECR Endoscope Cleaner and Reprocessor is a two-basin The washer/disinfector utilizing an enzymatic detergent and a concentrated high level disinfectant, CIDEX® OPA Concentrate Solution. Both the detergent and high level disinfectant are diluted in the system to in-use concentrations. The EVOTECH® ECR

{4}------------------------------------------------

Endoscope Cleaner and Reprocessor is capable of cleaning endoscopes that have not been manually cleaned prior to placing in the system.

Endoscopes that have been qualified for processing in the EVOTECH® ECR Endoscope Cleaner and Reprocessor must go through manual bedside pre-cleaning according to SGNA and facility guidelines prior to being placed in the EVOTECH® ECR Endoscope Cleaner and Reprocessor. If automatic cleaning is not available, or desired for the device by the user, the endoscope must also be manually cleaned by the user according to the endoscope manufacturer's instructions prior to being placed in the EVOTECH® ECR Endoscope Cleaner and Reprocessor.

After the endoscopes are connected to the EVOTECH® ECR Endoscope Cleaner and Reprocessor, the detergent is sprayed onto the surface of the device, and pumped through the lumens of the device, at sufficient pressure levels to remove soil present on or in the device.

The disinfectant is then mixed with incoming water, which is monitored to ensure that the temperature of the solution that is sprayed onto the surface of the device, and pumped through the lumens of the device, is at sufficient time and temperature to achieve high level disinfection. The temperature is monitored to ensure operating parameters of 50°C are met. A sample of the diluted CIDEX® OPA Concentrate Solution is extracted to ensure minimum effective concentration levels have been achieved. Following disinfection. the endoscopes are rinsed by the machine and air pumped through the lumens to facilitate drying of the endoscope. If selected by the user, an alcohol flush cycle may also be run to facilitate drying of the endoscope. The endoscope may then be removed, dried and used. The onboard printer generates a cycle printout summarizing the cycle results for the user's records. EVOTECH® ECR has Data Transfer Interface (DTI) capability which provide dynamic host configuration protocol (DHCP) for connecting the EVOTECH® ECR system to hospital networks.

The machine has built in safety features which cancel the cycle and notify the user if parameters are not met. These notifications, their causes, and next steps for the user, are defined in the EVOTECH® ECR Endoscope Cleaner and Reprocessor User Guide.

The network connectivity software revision part of this 510(k) premarket notification allows the Hospital IT Department to connect the EVOTECH® ECR Endoscope Cleaner and Reprocessor to a Hospital Local Area Network (LAN) for transfer of cycle parameters to a server and if desired, to an Instrument Tracking System.

{5}------------------------------------------------

#### v. INDENTED USE

The EVOTECH® ECR Endoscope Cleaner and Reprocessor is designed to automatically clean and high level disinfect flexible, submersible video or fiber-optic endoscopes. High level disinfection requires that the system be used with CIDEX® OPA Concentrate Solution.

The new device has the same intended use as the predicate device.

#### VI. INDICATION FOR USE

The EVOTECH® ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high level disinfectant CIDEX® OPA Concentrate and an enzymatic detergent (CIDEZYME GI) to achieve cleaning and high level disinfection of heat sensitive (>60 ℃) semi-critical endoscopes. Manual cleaning of medical devices (endoscopes) is not required prior to placement in the EVOTECH® ECR System when selecting those cycles that contain a wash stage. (Manual cleaning of medical devices (endoscopes) is required when selecting the Disinfect only or Disinfect/Alcohol Flush Cycle.)

The new device has the same indication for use as the predicate device.

#### VII. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS

The EVOTECH® ECR Endoscope Cleaner and Reprocessor that is this subject of this submission is the same model as the predicate device cleared in K140977. The fundamental technological characteristics of the predicate devices are same as the device that is the subject of this 510(k). Design, materials, and functions of the Endoscope Cleaner and Reprocessor have not changed.

Table below contains a summary of the technological characteristics of the new device compared to the predicate device.

| Feature       | Predicate Device: K140977<br>EVOTECH® ECR Endoscope<br>Cleaner and Reprocessor                                                                                                                                                                                                  | New Device: K152189<br>EVOTECH® ECR Endoscope<br>Cleaner and Reprocessor |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| System Design | Automated, stand-alone system<br>with two operating reprocessing<br>basin comprising hardware,<br>software, and consumables,<br>designed to perform cleaning<br>and high level disinfection of<br>flexible endoscopes. The<br>system can reprocess two<br>endoscopes at a time. | Same as predicate device                                                 |

{6}------------------------------------------------

| Disinfectant Used                                                                                                              | CIDEX® OPA Concentrate<br>Solution (Single Use)                                                                                                                                                                                                                                                     | Same as predicate device                                             |
|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| System<br>Performance                                                                                                          | The system can perform block<br>test, leak and endoscope<br>connection test, washing<br>(cleaning) and disinfection of<br>flexible endoscopes. The<br>system has a self-disinfect<br>cycle. The cycle steps for each<br>basin are monitored and<br>separately controlled by the<br>system software. | Same as predicate device                                             |
| Endoscope and the<br>Endoscope Cleaner<br>and Reprocessor<br>Materials and<br>composition in<br>contact with<br>fluid pathways | powder coated aluminum glass,<br>stainless steel, polyethylene,<br>Neoprene, polypropylene,<br>polyamide, Ultem,<br>polycarbonate, polypropylene                                                                                                                                                    | Same as predicate device                                             |
| System<br>compatibility with<br>associated chemical<br>consumable                                                              | System material is compatible<br>with CIDEX® OPA Concentrate<br>high level disinfectant and an<br>enzymatic detergent<br>(CIDEZYME GI) for cleaning<br>medical instruments                                                                                                                          | Same as predicate device                                             |
| Network<br>Connectivity                                                                                                        | Static IP network connectivity.                                                                                                                                                                                                                                                                     | Dynamic host configuration<br>protocol (DHCP) network<br>capability. |

#### VIII. PERFORMANCE DATA

The modified device has the same intended use, indications for use, materials, design features and fundamental technological characteristics as the predicate device. The following performance data were provided in support of the substantial equivalence determination:

### Software Verification and Validation Testing

Software verification and validation testing were conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern since prior to mitigation of hazards a

{7}------------------------------------------------

failure of the software could result in Minor Injury, either to a patient or to a user of the device.

# Electrical Safety and Electromagnetic Compatibility (EMC)

The modified device has been tested for Radiated and Conducted Emissions according to the Standards EN 60601-1-2:2007 Class A and IEC 60601-1-2:2014 Class A, method CISPR 11:2009 (Amended by A1: 2010) Class A. All test results met the requirements of the Standards.

The modified device has been evaluated for safety in accordance with CAN/CSA-C22.2 No.: 61010-1 2004, UL 61010-1/R: 2008-10 and EN 61010-1:2001. The system meets the standards' requirements.

## Other Testing

The EVOTECH® ECR Endoscope Cleaner and Reprocessor is a washer/disinfector for use with an enzymatic detergent and the high level disinfectant CIDEX® OPA Concentrate Solution at 50° with an exposure time of 5 min to achieve cleaning and high level disinfection of flexible endoscopes.

The EVOTECH® ECR Endoscope Cleaner and Reprocessor was tested in K040883 using the standard array of tests defined in the FDA Guidance on Premarket Notification [510(k)] Submission for Automated Endoscope Washer / Disinfectors, and Disinfectors Intended for Use in Health Care Facilities, August, 1993.

For simulated use testing, the high level disinfectant indicated for use with the EVOTECH® ECR Endoscope Cleaner and Reprocessor was tested for efficacy in a simulated use environment. A 6 Logio reduction of Mycobacterium terrae was achieved when flexible endoscopes, contaminated with Mycobacterium terrae and artificial soil and were exposed to CIDEX® OPA Concentrate in use solution without cleaning. The diluted CIDEX® OPA Concentrate at an MEC of 0.055% OPA concentration at 50°C is effective against Mycobacterium terrae in artificial soil.

For in use testing, endoscopes used in a clinical environment were reprocessed in the EVOTECH® ECR Endoscope Cleaner and Reprocessor without manual cleaning. Sterility testing demonstrated no growth.

For biocompatibility, an endoscope reprocessed in the EVOTECH® ECR Endoscope Cleaner and Reprocessor was evaluated for residue levels. The analysis indicates that the 

{8}------------------------------------------------

level of OPA residual remaining on an endoscope following the rinse phase is not likely to cause toxic effects in humans.

For material compatibility, the EVOTECH® ECR Endoscope Cleaner and Reprocessor using CIDEX® OPA Concentrate In-Use solution was evaluated for its effect on materials commonly used in medical devices. Multiple disinfection cycles over extended periods of time resulted in minimal effect on the test articles. The effects seen were similar to those seen with the predicate device for the high level disinfectant CIDEX® OPA Solution K991487.

The EVOTECH® ECR Endoscope Cleaner and Reprocessor was further tested in K061889 to determine its ability to high level disinfect and clean endoscopes. Both simulated use and in-use studies were completed. In both instances, endoscopes were contaminated and then processed in the "wash only" cycle in the EVOTECH® ECR Endoscope Cleaner and Reprocessor. Residual soil was quantified for both protein and total organic carbon (TOC). In all instances the residuals were below the predefined limit of 8.5ug/cm2.

Additionally, because the wash cycle is intended to eliminate the need for manual cleaning, studies were completed comparing the residual soil after manual cleaning according to the procedures of the Society for Gastroenterology Nurses and Associates (SGNA) and cleaning in the EVOTECH® ECR Endoscope Cleaner and Reprocessor. In a non-inferiority trial, the washing of endoscopes in the EVOTECH® ECR Endoscope Cleaner and Reprocessor was determined to be non-inferior to washing of endoscopes following the SGNA procedure.

Endoscopes used in clinical procedures were processed through the "wash only" cycle of the EVOTECH® ECR Endoscope Cleaner and Reprocessor and then extracted for residual organic material. Processing in the EVOTECH® ECR Endoscope Cleaner and Reprocessor reduced the residual protein and TOC in all channels and surfaces to less than the predefined acceptance criteria, indicating that the EVOTECH® ECR Endoscope Cleaner and Reprocessor can clean clinically used endoscopes.

High level disinfection of cleaned endoscopes was achieved. Endoscopes were contaminated with Mycobacterium terrae imbedded in soil. Soil was inoculated in channels and on the surfaces of endoscopes so that each endoscopes contained ≥8.5ug/cm² protein with >10' cfu/mL M. terrae. Endoscopes were disinfected in the EVOTECH® ECR Endoscope Cleaner and Reprocessor without manual or automated washing. After processing the contaminated endoscopes through the disinfect cycle only

{9}------------------------------------------------

in the EVOTECH® ECR Endoscope Cleaner and Reprocessor, there was a >10° reduction in M. terrae.

Further high level disinfection testing and quantitative tuberculocidal testing were completed for K140977 demonstrating effectiveness of the OPA Concentrate solution at 0.042% concentration in the presence of 5% fetal bovine serum organic soil load within 5 minutes at 50-52°C using CIDEX® OPA Concentrate solution.

Additional cleaning and disinfection studies utilizing duodenoscopes with open and closed elevator wire channels were conducted.

Cleaning: Olympus, Pentax and Fujinon duodenoscopes (preconditioned by repeated articulation of the wire mechanisms with the distal tip immersed in artificial test soil and inoculated at the distal tip with a biopsy forceps in place), were cleaned using the EVOTECH® ECR System wash-only cycle. Residual protein and hemoglobin were extracted from the distal tip of all tested endoscopes and results showed that all test samples met the acceptance criteria of < 6.4 µg/cm- residual protein and "no residue" for Hemoglobin.

Due to the complexity of the manual cleaning instructions for duodenoscopes, it is recommended that endoscopes with open/closed elevator wire channels be manually cleaned as per manufacturer's instructions in addition to using the cleaning cycle of the EVOTECH® System.

High level disinfection: Worst case simulated use tests demonstrated that EVOTECH® ECR Endoscope Cleaner and Reprocessor achieved cleaning and high level disinfection of duodenoscopes with closed and open elevator wire channels. Olympus, Pentax and Fujinon duodenoscopes with closed and open elevator wire channels were contaminated with Mycobacterium terrae and reprocessed using the "disinfect only" cycle in the EVOTECH® ECR without cleaning. A 6 log10 reduction of M. terrae was achieved for all tested duodenoscopes.

#### IX. CONCLUSIONS

There is no change to the proposed device intended use/indication for use or fundamental scientific technology compared to the predicate device. The software verification and validation performed demonstrated that the software changes in the proposed device could not have a significant effect on the safety or effectiveness of the system. Therefore data presented in this submission indicate that the EVOTECH® ECR Endoscope Cleaner and Reprocessor is substantially equivalent to the Predicate Device, EVOTECH® ECR Endoscope Cleaner and Reprocessor (K140977).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/FEB/K152189](https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94diagnostic-devices/FEB/K152189)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
