RIWO SYSTEM TRAYS

K012309 · Richard Wolf Medical Instruments Corp. · FEB · Oct 19, 2001 · Gastroenterology, Urology

Device Facts

Record IDK012309
Device NameRIWO SYSTEM TRAYS
ApplicantRichard Wolf Medical Instruments Corp.
Product CodeFEB · Gastroenterology, Urology
Decision DateOct 19, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The RIWO System Trays 38201.xxx, 38202.xxx, 38301.xxx have been designed to hold devices during sterile storage, transport, and while undergoing steam and ethylene oxide sterilization.

Device Story

RIWO System Trays are sterilization, storage, and transport containers for medical instruments and endoscopes. Trays feature perforated bases and lids, latched covers, and silicone peg mat inlays to secure instruments and prevent damage. Used in clinical settings; operated by healthcare staff. Trays must be wrapped in sterile packaging material (per DIN 58953) prior to sterilization. Compatible with steam and ethylene oxide (EtO) sterilization processes. Benefits include organized instrument handling, protection during transport, and maintenance of device integrity during sterilization cycles.

Clinical Evidence

No clinical data. Bench testing only. Steam and ethylene oxide sterilization tests performed by the manufacturer demonstrated that sterilization processes have no influence on the functional performance of the devices.

Technological Characteristics

Materials: PSU granulate Udel P-1700, PPSU granulate Radel R 5000, silicone, medical grade stainless steel, polyester foil. Design: Perforated container with latched cover and silicone peg mat inlay. Dimensions: Inner length 300-705mm, width 120-200mm; outer length 366-766mm, width 150-266mm, height 47-103mm. Sterilization: Steam and EtO compatible. Requires external sterile wrapping (DIN 58953).

Indications for Use

Indicated for use as a container to hold medical instruments and endoscopes during steam or ethylene oxide sterilization, sterile storage, and transport.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # OCT 1 9 2001 K612309 Pg 1 of v 353 Corporate Woods Parkway Vernon Hills, IL 60061 Phone: 847-913-1113 Customer Service: 800-323-WOLF www.richard-wolf.com Medical Instruments Corporation ## 510(k) Summary of Safety and Effectiveness | Submitter: | | Date of Preparation: | July 18, 2001 | |---------------------------------------------------------------------|------------------------------------------------------------------------------|--------------------------------------|----------------| | Company / Institution name: | Richard Wolf Medical Instruments Corp. | FDA establishment regulation number: | 14 184 79 | | Division name (if applicable): | N.A. | Phone number (include area code): | (847) 913-1113 | | Street address: | 353 Corporate Woods Parkway | FAX number (include area code): | (847) 913-0924 | | City: | Vernon Hills | State/Province: | Illinois | | Country: | USA | ZIP/Postal Code: | 60061 | | Contact name: | Mr. Robert L. Casarsa | | | | Contact title: | Quality Assurance Manager | | | | Product Information: | | | | | Trade name: | RIWO System Trays | | | | Model number: | See section 3, "Submitted Devices" | | | | Common name: | Trays for Sterilization, Storage and Transport | | | | Classification Name: | Accessories, cleaning, for endoscope | | | | Information on devices to which substantial equivalence is claimed: | | | | | 510(k) Number | Trade or proprietary or model name | Manufacturer | | | 1 pre-enact. | 1 Instrument Container 8582.06, 8584.06 | 1 Richard Wolf | | | 2 | 2 Micropac, Endopak, Scopepak, Multipak | 2 Riley | | | 3 | 3 Scope trays, Silicone Nipple Mat System, Standard Sterilization Containers | 3 Transmedica | | | 4 | 4 C.A.S.E. System: Containers and Baskets | 4 Genesis | | | 5 | 5 Instrument Protection and Sterilizing System | 5 Micromedics | | #### 1.0 Description The devices are assorted open or closed trays for sterilization, storage and transport. ### 2.0 Intended Use The RIWO SYSTEM TRA Ys 3820x.xxx have been designed - to be equipped as required - · for steam and EtO sterilization - · for sterile storage - · for transport {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the text "Pg 2 of 2" in a handwritten style. The text is written in black ink on a white background. The letters are connected and have a cursive appearance. #### 3.0 Technological Characteristics The RIWO System Trays are blue in color. The assorted trays vary in size and begin with product number 3820x.xxx. Other characteristics are a base, latched cover and silicone peg mat construction. These trays hold the device to be sterilized and the tray must be wrapped for sterilization and sterile storage. They can be used in steam or ethylene oxide sterilization. #### 4.0 Substantial Equivalence These devices are substantially equivalent to existing pre-enactment devices and 510(k) devices sold by Richard Wolf and 510(k) devices sold by Riley, Transmedica, Genesis, and Micromedics. #### 5.0 Performance Data The steam sterilization tests performed by Richard Wolf show that the steam sterilization has no influence on the functional performance of the submitted devices when using the pre-vac method. The EO sterilization tests performed, shows that the ethylene oxide sterilization has no influence on the functional performance of the submitted devices. #### 6.0 Clinical Tests No clinical tests performed. #### 7.0 Conclusions Drawn These devices are designed and tested to guarantee the safety and effectiveness when used according to the instruction manual. By: Robert L. Casasa Robert L. Casarsa Quality Assurance Manager Date: July 18, 2001 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 19 2001 Mr. Robert L. Casarsa Quality Assurance Manager Richard Wolf Medical Instruments Corp. 353 Corporate Woods Parkway VERNON HILLS IL 60061 Re: K012309 Trade/Device Name: RIWO System Trays for Sterilization and Transport Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulation Number: 21 CFR §880.6850 Regulation Name: Sterilization wrap Regulatory Class: II Product Code: 78 FEB and KCT Dated: July 18, 2001 Received: July 23, 2001 Dear Mr. Casarsa: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your bector of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to the enactment date of the Medical Device Amendments, or to commerce price to that 2011-17-15, in accordance with the provisions of the Federal Food, Drug, devices that have occh rocussion in quire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The Tou may, dicrorely mains of the Act include requirements for annual registration, listing of general controll profitions practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that I Dri has intact and regulations administered by other Federal agencies. You must or any I oderal buttatoo the ect's requirements, including, but not limited to: registration and listing Compry with an the Prequently Part 801); good manufacturing practice requirements as set (21 OF IC Part 807), accems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific ac-ripliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on of reference to pictures under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): // 0 / 2 3 0 9 Trays for Sterilization Device Name:__________________________________________________________________________________________________________________________________________________________________ Intended Use: The RIWO System Trays 38201.xxx, 38202.xxx, 38301.xxx have been designed to hold devices during sterile storage, transport, and while undergoing steam and ethylene oxide sterilization. | Material Composition: | PSU granulate UdelP-1700, PPSU granulate Radel R 5000, silicone, medical grade stainless steel, polyester foil | | |-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Physical Properties: | The instrument container ranges in height from 60mm to 100mm. The tray system consists of:<br>perforated instrument container with peg mat inlay spring catches perforated lid w/ fixed peg mat coding plate for container & lid marker pen | | | Specifications: | The inner dimension of the trays range from a length of 300mm to 705 mm and widths of 120mm to 200mm.<br><br>The outer dimensions of the trays range from a length of 366mm to 766mm; widths of 150mm to 266mm; and heights range from 47mm to 103mm. | | | Design: | The RIWO System Trays are designed as an accessory for housing and holding endoscopes and instruments. The design provides spaces to position the devices without placing them on top of each other; thus, reducing the potential for damage during sterile storage, transport, and sterilization. They are designed as an aid to sterilization. The trays with the included endoscope/instruments must be wrapped in sterile packaging material (in accordance with DIN | | 58953). The wrapping, not part of this submission, maintains sterility and barrier integrity. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) ![Signature](signature.png) | (Division Sign-Off) | | |--------------------------------------|---------| | Division of Reproductive, Abdominal, | | | and Radiological Devices | | | 510(k) Number | K012309 | Prescription Use Over-The Counter_
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...