← Product Code [JKA](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA) · K220258

# 20G Open-System Ally Device Kit (AV100000), 22G Open-System Ally Device Kit (AV110000) (K220258)

_Avia Vascular · JKA · Jul 14, 2022 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA/K220258

## Device Facts

- **Applicant:** Avia Vascular
- **Product Code:** [JKA](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA.md)
- **Decision Date:** Jul 14, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1675
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Ally device is attached to a PIV catheter system for blood sampling into a vacuum tube or syringe.

## Device Story

The Ally device is a sterile, single-use needleless blood draw system designed to attach to a peripheral intravenous (PIV) catheter. It functions by extending a conduit tube through the internal lumen of the PIV catheter, with the distal tip protruding beyond the catheter tip to facilitate blood aspiration. The device enables clinicians to collect venous blood samples directly into vacuum tubes or syringes without requiring additional venipuncture. By utilizing the existing PIV access, the device aims to reduce the need for repeated needle sticks, thereby improving patient comfort and procedural efficiency in clinical settings.

## Clinical Evidence

Bench testing only. Performance testing included flow rate, tensile strength, leak decay, and particulate testing (USP <788>). Biocompatibility testing was conducted per ISO 10993-1 for limited (<24 hour) blood-contacting devices, including cytotoxicity, sensitization, irritation, pyrogenicity, acute systemic toxicity, and hemocompatibility (hemolysis, complement activation, PTT, and platelet/leucocyte counts). All predetermined acceptance criteria were met.

## Technological Characteristics

Materials: Polyimide (distal tube), Polyurethane (proximal tube). Dimensions: 20G and 22G configurations. Connectivity: Attaches to PIV catheter systems. Sterilization: Ethylene Oxide. Testing standards: ISO 10993-1, ISO 10993-4, ISO 10993-5, ISO 10993-10, ISO 10993-11, ASTM F-756-17, USP <788>.

## Regulatory Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

## Predicate Devices

- PIVO™ ([K190604](/device/K190604.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 14, 2022

AVIA Vascular % Jonathan Holmes Senior Manager, Regulatory Affairs MedVenture Health 299 S Main Street, Suite 2300 Salt Lake City, Utah 84111

### Re: K220258

Trade/Device Name: 20G Open-System Ally Device Kit (AV10000), 22G Open-System Ally Device Kit (AV110000) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: June 14, 2022 Received: June 15, 2022

Dear Jonathan Holmes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Wolloscheck, Ph.D. For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220258

#### Device Name

20G Open-System Ally Device Kit (AV100000); 22G Open-System Ally Device Kit (AV110000)

Indications for Use (Describe)

The Ally device is attached to a PIV catheter system for blood sampling into a vacuum tube or syringe.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo and address of Avia Vascular. The logo is a stylized, intertwined shape in shades of purple and pink. Below the logo is the text "AVIA VASCULAR", followed by the address "6030 W. Harold Gatty Dr., Salt Lake City, UT 84116".

# K220258 510(k) SUMMARY

### Submitter:

Kevin Cook President & CEO Avia Vascular 6030 W. Harold Gatty Dr. Salt Lake City, UT 84116 Phone: (801) 824-5375 Email: kcook(@avia-vascular.com

-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

#### Correspondent:

Jonathan Holmes Senior Manager, Regulatory Affairs 299 S. Main St. Suite 2300 Salt Lake City, UT 84111 Phone: (801) 503-9065 Email: jholmes@medventurehealth.com

# DATE PREPARED: July 14, 2022

### SUBJECT DEVICE:

| Trade Name:           | 20G Open-System Ally Device Kit (AV100000);              |
|-----------------------|----------------------------------------------------------|
|                       | 22G Open-System Ally Device Kit (AV110000)               |
| Common Name:          | Blood Collection Tubes, Vials, Systems, Serum Separators |
| Regulation Name:      | Blood Specimen Collection Device                         |
| Classification Panel: | Clinical Chemistry                                       |
| Regulatory Class:     | Class II                                                 |
| Product Code:         | JKA                                                      |
| Regulation Number:    | 21 CFR 862.1675                                          |

### PREDICATE DEVICE:

| Proprietary Name: | PIVOTTM                                                  |
|-------------------|----------------------------------------------------------|
| Common Name:      | Blood Collection Tubes, Vials, Systems, Serum Separators |
| Regulation Name:  | Blood Specimen Collection Device                         |

Avia Vascular

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Avia Vascular logo. The logo consists of a stylized letter "A" in a gradient of purple and pink. Below the logo, the words "AVIA VASCULAR" are written in a simple sans-serif font.

510(k) Number:

K190604

### PRODUCT DESCRIPTION:

The Ally device by Avia Vascular is intended to be attached to a peripheral intravenous (PIV) catheter system for use as a needleless blood draw device into a vacuum tube or syringe. Once attached to a PIV catheter system, the Ally device functions by extending a conduit tube through the internal lumen of the PIV catheter so that the distal tip of the Ally conduit extends beyond the tip of the PIV catheter, allowing for sampling of blood. The Ally device is a sterile, single use device.

#### INTENDED USE/INDICATION FOR USE:

Intended Use: Venous blood drawing.

Indications for Use: The Ally device is attached to a PIV catheter system for blood sampling into a vacuum tube or syringe.

# TECHNOLOGICAL COMPARISION TO PREDICATE DEVICE:

| Characteristic                                   | Ally Blood Collection<br>Device<br>(K220258)<br>(Subject Device)                                                   | PIVO™<br>(K190604)<br>(Predicate Device)                                                                                                           | Analysis of<br>Differences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use                           | The Ally device is<br>attached to a PIV<br>catheter system for<br>blood sampling into a<br>vacuum tube or syringe. | The PIVO™ device<br>attaches to a peripheral<br>IV catheter system for<br>use as a direct blood draw<br>device into a vacuum<br>tube or a syringe. | Substantially<br>Equivalent<br>The indication for<br>use is the same. The<br>difference in<br>phrasing does not<br>alter the meaning of<br>the indication for use<br>or negatively impact<br>the clarity.                                                                                                                                                                                                                                                                                                                                  |
| Device<br>Configurations<br>and Colors           | • 20G – Pink<br>• 22G – Blue                                                                                       | • 20G – Pink<br>• 22G – Blue<br>• 24G – Yellow                                                                                                     | Substantially<br>Equivalent<br>The subject device<br>utilizes the same<br>primary color as the<br>predicate for<br>equivalent gauge<br>sizes.                                                                                                                                                                                                                                                                                                                                                                                              |
| Distal Tube<br>Material                          | Polyimide                                                                                                          | Polyimide                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Characteristic                                   | Ally Blood Collection<br>Device<br>(K220258)<br>(Subject Device)                                                   | PIVOTM<br>(K190604)<br>(Predicate Device)                                                                                                          | Analysis of<br>Differences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Proximal Tube<br>Material                        | Polyurethane                                                                                                       | Pebax                                                                                                                                              | Different<br>Performance and<br>biocompatibility<br>testing pertaining to<br>the proximal tubing<br>(leak, tensile,<br>hemolysis) show the<br>component to be<br>equivalent to the<br>predicate with<br>respect to applicable<br>requirements. The<br>subject devices were<br>determined to be<br>biocompatible for<br>their intended use<br>per ISO 10993-1<br>testing as noted in<br>the summary below.<br>Material difference<br>does not raise new<br>questions for safety<br>or effectiveness with<br>respect to the<br>intended use. |
| Inner Tubing<br>Length                           | • 20G and 22G Open-<br>System - 6.253"                                                                             | • 20G – 5.85"<br>• 22G – 5.85"<br>• 24G – 141.2mm                                                                                                  | Different<br>Distal tube lengths<br>of the subject device<br>are designed such<br>that the tubing can<br>be extended<br>throughout a PIV<br>catheter system in<br>the same manner as<br>the predicate.<br>Appropriateness was<br>demonstrated<br>through testing<br>(dimensional,<br>hemolysis, flow<br>testing, usability).<br>Differences in length<br>do not raise new<br>questions for safety<br>or effectiveness of the                                                                                                               |
| Characteristic                                   | Ally Blood Collection<br>Device<br>(K220258)<br>(Subject Device)                                                   | PIVOTM<br>(K190604)<br>(Predicate Device)                                                                                                          | Analysis of<br>Differences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                  |                                                                                                                    |                                                                                                                                                    | or effectiveness with<br>respect to the<br>intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Outer Diameter<br>(OD) of Distal<br>Inner Tubing | • 20G Open-System –<br>0.0270”<br>• 22G Open-System –<br>0.0210”                                                   | • 20G = 0.709mm max<br>• 22G = 0.543mm max<br>• 24G = 0.400mm max                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Packaging                                        | Tyvek Pouch                                                                                                        | Tyvek Pouch                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Sterilization                                    | Ethylene Oxide                                                                                                     | Gamma                                                                                                                                              | Different<br>The method of<br>sterilization for the<br>subject device was<br>found to be valid for<br>a sterility claim and<br>appropriate through<br>packaging,<br>biocompatibility,<br>and sterilization<br>evaluations. Both<br>subject device and<br>predicate devices are<br>classified as sterile.                                                                                                                                                                                                                                   |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Avia Vascular. The logo consists of a stylized, intertwined ribbon-like shape in shades of purple and pink. Below the logo, the words "AVIA VASCULAR" are written in a simple, sans-serif font.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with a stylized, abstract design. The logo features a shape that resembles two intertwined or overlapping curves, creating a sense of connection or flow. The color gradient transitions from a deep purple at the top to a lighter pink at the bottom, adding depth and visual interest to the design. The logo is simple, modern, and potentially represents a company or organization with a focus on innovation or collaboration.

# AVIA VASCULAR

### PERFORMANCE TESTING

The Ally Blood Collection Device was thoroughly tested and verifies that it performs as designed and is suitable for its intended use.

Performance Testing included the following:

- Flow Conduit Inner Diameter
- Flow Conduit Outer Diameter
- · Flow Conduit Effective Length and Flow Conduit Exposure Length
- Open-System Introducer Length
- Extension Tubing Inner Diameter
- · Flow Conduit-to-Extension Tubing Assembly Tensile
- Extension Leg-to-Stop Ring Assembly Tensile
- · Extension Tubing-to-Proximal Luer Assembly Tensile
- Guide Cannula-to-Introducer Assembly Tensile
- Introducer to Housing Retention

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Avia Vascular. The logo consists of a stylized, intertwined ribbon-like shape in shades of purple and pink. Below the logo, the words "AVIA VASCULAR" are written in a simple, sans-serif font.

- · Sheathing Cannula Tensile Testing
- · Kink Diameter
- · Aspiration Flow Rate
- · Leak Decay
- · Housing Seal Leak Decay Testing
- · Particulate testing per USP <788>

Biocompatibility per ISO 10993-1 for an external communicating device, limited (<24 hour) blood contacting device.

- · Cytotoxicity MEM Elution ISO 10993-5
- · Sensitization ISO 10993-10
- · Irritation ISO 10993-10
- · Material Mediated Pyrogenicity ISO 10993-11
- · Acute Systemic Toxicity ISO 10993-11
- · Hemolysis (direct and indirect) ISO 10993-4
- · Mechanically Induced Hemolysis ASTM F-756-17
- · Complement Activation Assay C3a and SC5b9 Methods-ISO 10993-4
- · Partial Thromboplastin Time (PTT) ISO 10993-4
- · Platelet and Leucocyte Count (PLC) ISO 10993-4

The subject Ally device met all predetermined acceptance criteria and raised no new concerns regarding safety or effectiveness.

### CONCLUSION:

The subject Ally has been demonstrated to be substantially equivalent in design, materials, sterilization, principles of operation, performance, and indications for use to the predicate PIVO™ device (K190604). Any differences do not raise new or different questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA/K220258](https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA/K220258)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
