← Product Code [JKA](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA) · K181754

# Blood Collection Accessory (BUCA) (K181754)

_Nsp Tech Pte, Ltd. · JKA · Jul 30, 2019 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA/K181754

## Device Facts

- **Applicant:** Nsp Tech Pte, Ltd.
- **Product Code:** [JKA](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA.md)
- **Decision Date:** Jul 30, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1675
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Blood Collection Accessory (BUCA) is a single use, sterile and non invasive device used for connecting to venous access device for the collection of blood specimens.

## Device Story

BUCA is a single-use, sterile, non-invasive blood collection accessory. It consists of a polypropylene holder with a male luer feature, a stainless-steel needle, and a rubber sleeve. The device connects to venous access devices to facilitate blood specimen collection into evacuated tubes. The holder encases the needle to prevent needle-stick injuries and protects personnel from direct blood contact. The user inserts a sample collection tube into the holder; the needle pierces the tube stopper to allow blood flow. After collection, the tube is removed. The device is intended for clinical use by healthcare personnel. It provides a safe, standardized interface for blood transfer, reducing exposure risks and ensuring secure connection to venous access points.

## Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemocompatibility) per ISO 10993-1 and performance testing (flow rate, leakage, insertion/evacuation force, holding strength of needle/sleeve, and deformation resistance).

## Technological Characteristics

Materials: Polypropylene (holder), Acrylonitrile butadiene styrene (adapter), stainless steel (needle), rubber (sleeve). Sterilization: Ethylene Oxide (EtO) per ISO 11135:2014 (SAL 10^-6). Connectivity: Mechanical luer connection. Standards: ISO 80369-7:2016, ISO 80369-20:2015, ISO 10993 series.

## Regulatory Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

## Predicate Devices

- Vacutainer® Brand Multiple Sample Luer Adapter ([K991088](/device/K991088.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

July 30, 2019

NSP Tech Pte Ltd HC Tan 10 Admiralty Street, Northlink Building, #02-06 757695 Singapore

Re: K181754

Trade/Device Name: Blood Collection Accessory (BUCA) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: April 16, 2019 Received: June 28, 2019

Dear HC Tan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181754

Device Name Blood Collection Accessory (BUCA)

Indications for Use (Describe)

The Blood Collection Accessory (BUCA) is a single use, sterile and non invasive device used for connecting to venous access device for the collection of blood specimens.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 1. SUBMITTER:

### Applicant Name:

NSP Tech Pte Ltd 10 Admiralty Street, Northlink Building, #02-06 Singapore 757695

### Contact Person:

Name: HC Tan Phone: (65)98180450 Fax: (65)67476533 Email: hockchoon.tan@nsptech.com.sg

| Establishment registration number | : 3008337059    |
|-----------------------------------|-----------------|
| Date prepared                     | : July 29, 2019 |

#### DEVICE 2.

| Trade Name :       | Blood Collection Accessory (BUCA) |
|--------------------|-----------------------------------|
| Common Name :      | Blood Specimen Collection Device  |
| Regulatory Class : | Class II                          |
| Regulation Number: | 21 CFR 862.1675                   |
| Regulation Name:   | Blood Specimen Collection Device  |
| Product Code :     | JKA                               |
| Panel :            | Clinical Chemistry                |

#### PREDICATE DEVICE 3.

Vacutainer® Brand Multiple Sample Luer Adapter

| 510(k) Number:     | K991088        |
|--------------------|----------------|
| Regulatory Class:  | Class II       |
| Regulation Number: | 21 CR 862.1675 |
| Product Code:      | JKA            |

#### DEVICE DESCRIPTION: 4.

The Blood Collection Accessory (BUCA) is a single use device comprising of a holder made from polypropylene plastic with a male luer feature. A stainless-steel needle covered with a rubber sleeve is attached to the inside of the holder.

The Blood Collection Accessory (BUCA) has a Holder that encases the Needle preventing any needle stick issue with user. The Holder of the Blood Collection Accessory (BUCA) also house the sample collection tube and protects the user or healthcare personnel from direct

{4}------------------------------------------------

contact with blood specimen. The Adapter connected to the blood collection device for the blood transfer. The sample collection tube is inserted from the far end of the holder where the opening is located. This transfer is conducted until the required amount of blood is completed. The sample collection tube is then evacuated from the Blood Collection Accessory (BUCA) by pulling it away from the latter.

#### INDICATION FOR USE STATEMENT 5.

The Blood Collection Accessory (BUCA) is a single use, sterile and non invasive device used for connecting to venous access device for the collection of blood specimens.

#### SUBSTANTIAL EQUIVALENCE COMPARISON 6.

### TABLE OF COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

|                          | NSP Tech Pte Ltd<br>Blood Collection<br>Accessory (BUCA)                                                                                                                                          | Vacutainer® Brand<br>Multiple Sample Luer<br>Adapter                                                                                                                                                                                                                                                                                                                                                                                                          | Discussion/Comment                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510 (k) Number           | K181754                                                                                                                                                                                           | K991088                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification           | Class II                                                                                                                                                                                          | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                      | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code             | JKA                                                                                                                                                                                               | JKA                                                                                                                                                                                                                                                                                                                                                                                                                                                           | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Indication For Use       | The Blood Collection<br>Accessory (BUCA) is<br>a single use, sterile<br>and non invasive<br>device used for<br>connecting to<br>venous access device<br>for the collection of<br>blood specimens. | The Vacutainer Brand<br>Luer Adapter is a sterile,<br>non-invasive device used<br>to connect venous<br>access devices such as<br>needles, blood collection<br>sets, and infusion sets to<br>blood collection tubes.<br>They are also used in<br>connection with non-<br>needle devices for<br>collection of blood from<br>catheters. The<br>Vacutainer Brand Luer<br>Adapter is sold by itself<br>and as a component of<br>other Vacutainer Brand<br>devices. | Similar- The difference is<br>in the name of the<br>device, this does not<br>raise any questions of<br>safety or effectiveness.<br>The predicate specifies<br>the devices that it<br>connects with and the<br>subject device does not.                                                                                                                                                                                                                                 |
| Manufacturer             | NSP Tech Pte Ltd                                                                                                                                                                                  | Becton Dickinson                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product<br>Configuration | Holder –<br>Polypropylene<br>Adapter - Acrylonitrile<br>butadiene<br>styrene<br>Sleeve – Not made<br>with natural<br>rubber latex<br>Needle - Stainless<br>Steel                                  | Holder – Polypropylene<br>Adapter - Polycarbonate<br>Sleeve - Not made<br>with natural<br>rubber latex<br>Needle - Stainless Steel                                                                                                                                                                                                                                                                                                                            | The only major<br>difference between the<br>subject device and the<br>predicate device is in the<br>material used for the<br>part, Adapter. The<br>subject device, Blood<br>Collection Accessory<br>(BUCA) uses the<br>polymer, Acrylonitrile<br>butadiene styrene (ABS)<br>for the Adapter part<br>while the predicate<br>device uses<br>Polycarbonate (PC)<br>polymer.<br>Biocompatibility Testing<br>and Performance<br>Testing was conducted<br>to demonstrate SE. |
| Sterilization            | Ethylene Oxide (EtO)<br>Gas sterilization                                                                                                                                                         | Ethylene Oxide (EtO) Gas<br>sterilization same                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Sterility                | Meet the Sterility<br>Assurance Level of 10-<br>6 as per the<br>requirement of ISO<br>11135:2014                                                                                                  | Not known                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Non-Clinical Performance<br>Testing was conducted to<br>demonstrate SE.                                                                                                                                                                                                                                                                                                                                                                                                |
| Packaging                | Sealed Blister<br>Packaging                                                                                                                                                                       | Sealed Blister Packaging                                                                                                                                                                                                                                                                                                                                                                                                                                      | Non-Clinical<br>Performance Testing<br>was conducted to<br>demonstrate SE.                                                                                                                                                                                                                                                                                                                                                                                             |
| Use                      | Single-Use only                                                                                                                                                                                   | Single-Use only                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Bio compatibility        | ISO 10993 series                                                                                                                                                                                  | ISO 10993 series                                                                                                                                                                                                                                                                                                                                                                                                                                              | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Shelf Life               | 1 year                                                                                                                                                                                            | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Non-Clinical Performance<br>Testing was conducted to<br>demonstrate SE.                                                                                                                                                                                                                                                                                                                                                                                                |

{5}------------------------------------------------

# 7. TECHNOLOGICAL CHARACTERISTICS COMPARISON

The Blood Collection Accessory (BUCA) device has similarities with the predicate device in areas such as Intended Use Product Configuration, and performance.. The only difference between the subject device and the predicate device is in the use of material for the part, Adapter. The Blood Collection Accessory (BUCA) uses the polymer material, Acrylonitrile butadiene styrene (ABS) while the Page 3 of 5

{6}------------------------------------------------

predicate device uses the polymer material, Polycarbonate (PC). Biocompatibility tests and performance tests based on international standards demonstrated the device performs as intended. The indications for use of the predicate device specifies the blood collection devices it connects to. The Subject device does not specify the blood collection devices, this does not raise different questions of safety or effectiveness.

#### 8. Performance Data/Non-Clinical Testing:

The Blood Collection Accessory (BUCA) was tested for validation and verification of functions based on risk analysis and the results passed predetermined acceptance criteria.

## Biocompatibility

Biocompatibility was evaluated per the FDA Guidance titled Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" - Guidance for Industry and Food and Drug Administration Staff issued in June 2016 with the following endpoints:

Cytotoxicity

Sensitization

Irritation

Acute Systemic Toxicity

Pyrogenicity

Hemocompatibility (Coagulation, Platelets and Hemolysis)

### PerformanceData

Performance testing was carried out on the Blood Collection Accessory (BUCA) using applicable international standards as reference and predicate device as benchmark for comparison.

- . Luer Lok Adapter
For international standard related test, the Luer Lok adapter is tested with reference to the ISO 80369-7:2016 and ISO 80369-20:2015.

- Performance Testing
As for the benchmarking tests related to the predicate device, blood collection performances such as the flow rate and quality of holding strength of sleeve to adapter and needle to adapter are assessed and compared against that of the predicate device.. The test methods employed are as summarized below:

- o Flow Rate The time taken to fill up a 3.0 ml blood collection Test tube with liquid of viscosity similar to blood is recorded.

{7}------------------------------------------------

- o Leakage With the device held in place by a fixture, the maximum numbers of round of insertion and Testevacuation of blood collection tube the device can withstand before leakage start occurring at the device is recorded.
- The device is held in place by a fixture and a force Evacuation o Force Test taken to insert or evacuate a blood collection tube into the holder of the device is recorded.
- Using fixture to hold the adapter in place, the o Holding Strength of needle is pull in a direction away from the adapter. The force taken to dislodge the needle Needle to Adapter Test from the adapter is recorded.
- With the adapter held in place by a fixture, the Holding o sleeve is pull in a direction away from the adapter Strength of and the force taken to dislodge the sleeve from the Sleeve to adapter is recorded. Adapter Test -
- Deformation The device is held in place by a fixture and a 5 kg o Test force is applied on the device for a period of 5 minutes. The permanent change in outer diameter of the holder of the device before and after compression by the applied force is measured and recorded. This test assesses the quality of the holder.

### Sterilization

Sterilization is done in accordance to ISO 11135-1:2014.

The Sterility Assurance Level (SAL) is 10-6.

#### SUBSTANTIAL EQUIVALENCE CONCLUSION 9.

The result of non-clinical testing demonstrates that the Blood Collection Accessory (BUCA) is substantially to the predicate device, Vacutainer® Brand Multiple Sample Luer Adapter and has been demonstrated to meet its predetermined specifications.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA/K181754](https://fda.innolitics.com/submissions/HO/subpart-b%E2%80%94clinical-chemistry-test-systems/JKA/K181754)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
