The CBC-5D Plus Retics is a tri-level hematology control intended for use in clinical laboratories to monitor the precision of hematology analyzers for the parameters listed in the expected values table.
Device Story
R&D 5D Retic Hematology Control is an in vitro diagnostic reagent; composed of stabilized human cells in plasma-like fluid with preservatives. Used as a quality control material to monitor reticulocyte counting methods (RET%, RET#, IRF, MRV) on Coulter® hematology analyzers. Sampled identically to patient specimens. Used in clinical laboratory settings by laboratory personnel. Provides verification of analyzer performance; ensures accuracy of patient test results.
Clinical Evidence
Bench testing only. Verification and validation activities performed per design control procedures (21 CFR 820.30) and ISO 14971 risk management. Results demonstrated that predetermined acceptance criteria were met.
Technological Characteristics
Tri-level, assayed hematology control. Composition: stabilized human cells in plasma-like fluid with preservatives. Form factor: liquid reagent. Intended for use on Coulter® hematology analyzers. No electronic or software components.
Indications for Use
Indicated for use in clinical laboratories as a tri-level hematology control to monitor the precision of hematology analyzers for reticulocyte parameters.
Regulatory Classification
Identification
A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).
Special Controls
*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
R&D Systems CBC-5D Plus Retics Hematology Control (K072096)
Related Devices
K091433 — R&D SYSTEMS TETIC-I PLUS HEMOTOLOGY CONTROL · R&D Systems, Inc. · Oct 28, 2009
K072096 — CBC-5D PLUS RETICS HEMATOLOGY CONTROL, MODELS: 5DR03, 5DR04 · R&D Systems, Inc. · Aug 20, 2007
K100050 — R & D SYSTEMS XERET HEMATOLOGY CONTROL · R&D Systems, Inc. · Jan 29, 2010
K993496 — RETIC-QUINOX · Hematronix, Inc. · Nov 30, 1999
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER k130962
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II device requiring 510(k). The following items are present and acceptable:
1. The name and 510(k) number of the Submitters previously cleared device: CBC-5D Plus Retics, k072096.
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, and package labeling.
3. A description of the device MODIFICATION(S), including device description and package insert in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was to change the tri-level Hematology control to reticulocyte parameters only, instead of the combining with complete blood count parameters in the original version.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and stability.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis according to R&D Systems Risk Analysis and Management Procedure compliant with ISO 14971. The risk analysis for the predicate was performed using failure mode and effects analysis (FMEA) method (page 37 of 59.)
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied (page 37 of 59).
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met (page 53 of 59), and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review (page 53 of 59).
6. A Truthful and Accurate Statement (page 3 of 59), a 510(k) Summary (page 35 of 59) and the Indications for Use Enclosure (page 59 of 59).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.